Published in: European Radiology 1/2018

Open Access 01-01-2018 | Cardiac

Four-dimensional flow MRI of stented versus stentless aortic valve bioprostheses

Authors: Floortje van Kesteren, Laurens W. Wollersheim, Jan Baan Jr., Aart. J. Nederveen, Abdullah Kaya, S. Matthijs Boekholdt, Bas A. de Mol, Pim van Ooij, R. Nils Planken

Published in: European Radiology | Issue 1/2018

Login to get access

Abstract

Objectives

To evaluate aortic velocity, wall shear stress (WSS) and viscous energy loss (EL) of stented and stentless bioprostheses using 4D flow MRI 1 year after surgical aortic valve replacement.

Methods

For this cross-sectional study 28 patients with stented (n = 14) or stentless (n = 14) bioprosthesis underwent non-contrast-enhanced 4D-flow MRI at 1.5 T. Analyses included a comparison of velocity, WSS and EL in the ascending aorta during peak systole for both spatially averaged values and a comparison of local differences using per-voxel analysis.

Results

No significant differences were found in peak and mean velocity (stented vs. stentless: 2.45 m/s vs. 2.11 m/s; p = 0.09 and 0.60 m/s vs. 0.62 m/s; p = 0.89), WSS (0.60 Pa vs. 0.59 Pa; p = 0.55) and EL (10.17 mW vs. 7.82 mW; p = 0.10). Per-voxel analysis revealed significantly higher central lumen velocity, and lower outer lumen velocity, WSS and EL for stentless versus stented prostheses.

Conclusion

One year after aortic valve implantation with stented and stentless bioprostheses, velocity, WSS and EL were comparable when assessed for averaged values in the ascending aorta. However, the flow profile described with local analysis for stentless prosthesis is potentially favourable with a significantly higher central velocity profile and lower values for outer lumen velocity, WSS and EL.

Key Points

Stentless bioprostheses can be implanted instead of stented aortic valve bioprostheses.
Haemodynamic performance of valve prosthesis can be assessed using 4D flow MRI.
Averaged ascending aorta PSV, WSS and EL are comparable 1 year post-implantation.
Centreline velocity is highest, WSS and EL is lowest for stentless prosthesis.
Literature
1.
go back to reference Vahanian A, Alfieri O, Andreotti F et al (2012) Guidelines on the management of valvular heart disease (version 2012). Eur Heart J 33:2451–2496CrossRefPubMed Vahanian A, Alfieri O, Andreotti F et al (2012) Guidelines on the management of valvular heart disease (version 2012). Eur Heart J 33:2451–2496CrossRefPubMed
2.
go back to reference Nishimura RA, Otto CM, Bonow RO et al (2014) 2014 AHA/ACC guideline for the management of patients with valvular heart disease: executive summary: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines. J Am Coll Cardiol 63:2438–2488CrossRefPubMed Nishimura RA, Otto CM, Bonow RO et al (2014) 2014 AHA/ACC guideline for the management of patients with valvular heart disease: executive summary: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines. J Am Coll Cardiol 63:2438–2488CrossRefPubMed
3.
go back to reference Thourani VH, Suri RM, Gunter RL et al (2015) Contemporary real-world outcomes of surgical aortic valve replacement in 141,905 low-risk, intermediate-risk, and high-risk patients. Ann Thorac Surg 99:55–61CrossRefPubMed Thourani VH, Suri RM, Gunter RL et al (2015) Contemporary real-world outcomes of surgical aortic valve replacement in 141,905 low-risk, intermediate-risk, and high-risk patients. Ann Thorac Surg 99:55–61CrossRefPubMed
4.
go back to reference von Knobelsdorff-Brenkenhoff F, Trauzeddel RF, Barker AJ, Gruettner H, Markl M, Schulz-Menger J (2014) Blood flow characteristics in the ascending aorta after aortic valve replacement: a pilot study using 4D-flow MRI. Int J Cardiol 170:426–433CrossRef von Knobelsdorff-Brenkenhoff F, Trauzeddel RF, Barker AJ, Gruettner H, Markl M, Schulz-Menger J (2014) Blood flow characteristics in the ascending aorta after aortic valve replacement: a pilot study using 4D-flow MRI. Int J Cardiol 170:426–433CrossRef
5.
go back to reference Raja SG, Macarthur KJ, Pollock JC (2006) Impact of stentless aortic valves on left ventricular function and hypertrophy: current best available evidence. J Card Surg 21:313–319CrossRefPubMed Raja SG, Macarthur KJ, Pollock JC (2006) Impact of stentless aortic valves on left ventricular function and hypertrophy: current best available evidence. J Card Surg 21:313–319CrossRefPubMed
6.
go back to reference Dolan JM, Kolega J, Meng H (2013) High wall shear stress and spatial gradients in vascular pathology: a review. Ann Biomed Eng 41:1411–1427CrossRefPubMed Dolan JM, Kolega J, Meng H (2013) High wall shear stress and spatial gradients in vascular pathology: a review. Ann Biomed Eng 41:1411–1427CrossRefPubMed
7.
go back to reference Guzzardi DG, Barker AJ, van Ooij P et al (2015) Valve-related hemodynamics mediate human bicuspid aortopathy: insights from wall shear stress mapping. J Am Coll Cardiol 66:892–900CrossRefPubMedPubMedCentral Guzzardi DG, Barker AJ, van Ooij P et al (2015) Valve-related hemodynamics mediate human bicuspid aortopathy: insights from wall shear stress mapping. J Am Coll Cardiol 66:892–900CrossRefPubMedPubMedCentral
8.
go back to reference Barker AJ, van Ooij P, Bandi K et al (2014) Viscous energy loss in the presence of abnormal aortic flow. Magn Reson Med 72:620–628CrossRefPubMed Barker AJ, van Ooij P, Bandi K et al (2014) Viscous energy loss in the presence of abnormal aortic flow. Magn Reson Med 72:620–628CrossRefPubMed
9.
go back to reference Wigstrom L, Sjoqvist L, Wranne B (1996) Temporally resolved 3D phase-contrast imaging. Magn Reson Med 36:800–803CrossRefPubMed Wigstrom L, Sjoqvist L, Wranne B (1996) Temporally resolved 3D phase-contrast imaging. Magn Reson Med 36:800–803CrossRefPubMed
10.
go back to reference Markl M, Chan FP, Alley MT et al (2003) Time-resolved three-dimensional phase-contrast MRI. J Magn Reson Imaging 17:499–506CrossRefPubMed Markl M, Chan FP, Alley MT et al (2003) Time-resolved three-dimensional phase-contrast MRI. J Magn Reson Imaging 17:499–506CrossRefPubMed
11.
12.
go back to reference Bock J, Kreher W, Hennig J, Markl M (2007) Optimised preprocessing of time-resolved 2D and 3D phase contrast MRI data. Proc Intl Soc Magn Reson Med 15:3138 Bock J, Kreher W, Hennig J, Markl M (2007) Optimised preprocessing of time-resolved 2D and 3D phase contrast MRI data. Proc Intl Soc Magn Reson Med 15:3138
13.
go back to reference Rose MJ, Jarvis K, Chowdhary V et al (2016) Efficient method for volumetric assessment of peak blood flow velocity using 4D flow MRI. J Magn Reson Imaging 44:1673–1682CrossRefPubMedPubMedCentral Rose MJ, Jarvis K, Chowdhary V et al (2016) Efficient method for volumetric assessment of peak blood flow velocity using 4D flow MRI. J Magn Reson Imaging 44:1673–1682CrossRefPubMedPubMedCentral
14.
go back to reference Potters WV, van Ooij P, Marquering H, vanBavel E, Nederveen AJ (2015) Volumetric arterial wall shear stress calculation based on cine phase contrast MRI. J Magn Reson Imaging 41:505–516CrossRefPubMed Potters WV, van Ooij P, Marquering H, vanBavel E, Nederveen AJ (2015) Volumetric arterial wall shear stress calculation based on cine phase contrast MRI. J Magn Reson Imaging 41:505–516CrossRefPubMed
15.
go back to reference van Ooij P, Potters WV, Nederveen AJ et al (2015) A methodology to detect abnormal relative wall shear stress on the full surface of the thoracic aorta using four-dimensional flow MRI. Magn Reson Med 73:1216–1227CrossRefPubMed van Ooij P, Potters WV, Nederveen AJ et al (2015) A methodology to detect abnormal relative wall shear stress on the full surface of the thoracic aorta using four-dimensional flow MRI. Magn Reson Med 73:1216–1227CrossRefPubMed
16.
go back to reference Stevens SM, Reinier K, Chugh SS (2013) Increased left ventricular mass as a predictor of sudden cardiac death: is it time to put it to the test? Circ Arrhythm Electrophysiol 6:212–217CrossRefPubMedPubMedCentral Stevens SM, Reinier K, Chugh SS (2013) Increased left ventricular mass as a predictor of sudden cardiac death: is it time to put it to the test? Circ Arrhythm Electrophysiol 6:212–217CrossRefPubMedPubMedCentral
17.
go back to reference Bikkina M, Larson MG, Levy D (1993) Asymptomatic ventricular arrhythmias and mortality risk in subjects with left ventricular hypertrophy. J Am Coll Cardiol 22:1111–1116CrossRefPubMed Bikkina M, Larson MG, Levy D (1993) Asymptomatic ventricular arrhythmias and mortality risk in subjects with left ventricular hypertrophy. J Am Coll Cardiol 22:1111–1116CrossRefPubMed
18.
go back to reference Kweon J, Yang DH, Kim GB et al (2016) Four-dimensional flow MRI for evaluation of post-stenotic turbulent flow in a phantom: comparison with flowmeter and computational fluid dynamics. Eur Radiol 26:3588–3597CrossRefPubMed Kweon J, Yang DH, Kim GB et al (2016) Four-dimensional flow MRI for evaluation of post-stenotic turbulent flow in a phantom: comparison with flowmeter and computational fluid dynamics. Eur Radiol 26:3588–3597CrossRefPubMed
19.
go back to reference van Ooij P, Garcia J, Potters WV et al (2016) Age-related changes in aortic 3D blood flow velocities and wall shear stress: Implications for the identification of altered hemodynamics in patients with aortic valve disease. J Magn Reson Imaging 43:1239–1249CrossRefPubMed van Ooij P, Garcia J, Potters WV et al (2016) Age-related changes in aortic 3D blood flow velocities and wall shear stress: Implications for the identification of altered hemodynamics in patients with aortic valve disease. J Magn Reson Imaging 43:1239–1249CrossRefPubMed
20.
go back to reference Mensel B, Hesselbarth L, Wenzel M et al (2016) Thoracic and abdominal aortic diameters in a general population: MRI-based reference values and association with age and cardiovascular risk factors. Eur Radiol 26:969–978CrossRefPubMed Mensel B, Hesselbarth L, Wenzel M et al (2016) Thoracic and abdominal aortic diameters in a general population: MRI-based reference values and association with age and cardiovascular risk factors. Eur Radiol 26:969–978CrossRefPubMed
21.
go back to reference Payne DM, Koka HP, Karanicolas PJ et al (2008) Hemodynamic performance of stentless versus stented valves: a systematic review and meta-analysis. J Card Surg 23:556–564CrossRefPubMed Payne DM, Koka HP, Karanicolas PJ et al (2008) Hemodynamic performance of stentless versus stented valves: a systematic review and meta-analysis. J Card Surg 23:556–564CrossRefPubMed
22.
go back to reference Casas B, Lantz J, Dyverfeldt P, Ebbers T (2016) 4D Flow MRI-based pressure loss estimation in stenotic flows: evaluation using numerical simulations. Magn Reson Med 75:1808–1821CrossRefPubMed Casas B, Lantz J, Dyverfeldt P, Ebbers T (2016) 4D Flow MRI-based pressure loss estimation in stenotic flows: evaluation using numerical simulations. Magn Reson Med 75:1808–1821CrossRefPubMed
Metadata
Title
Four-dimensional flow MRI of stented versus stentless aortic valve bioprostheses
Authors
Floortje van Kesteren
Laurens W. Wollersheim
Jan Baan Jr.
Aart. J. Nederveen
Abdullah Kaya
S. Matthijs Boekholdt
Bas A. de Mol
Pim van Ooij
R. Nils Planken
Publication date
01-01-2018
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 1/2018
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-017-4953-2

Other articles of this Issue 1/2018

European Radiology 1/2018 Go to the issue