Published in: Acta Neurochirurgica 4/2014

01-04-2014 | Clinical Article - Neurosurgical Techniques

Flow-related noise in patients with ventriculoperitoneal shunt using gravitational adjustable valves

Authors: Florian Stockhammer, Christoph Miethke, Thoralf Knitter, Veit Rohde, Christian Sprung

Published in: Acta Neurochirurgica | Issue 4/2014

Login to get access

Abstract

Background

Noise disturbance arising from the valve is a rare event of ventriculoperitoneal shunts. We queried and investigated shunt patients for occurrence and evaluated the possible factors related to noise development.

Methods

Fifty ambulatory patients with implanted proGAV valve were investigated consecutively. Patients were asked for any noise arising from the shunt. In all cases, the valve was auscultated in sitting and upright position. The position of the gravitational unit (GU) was determined in respect to the Frankfurt horizontal plane (FHP) and in head reclination. Ten valves were perfused in vitro at different settings. One valve was opened for video documentation, and a frequency analysis of the noise was performed in nine valves.

Results

Eight percent (4/50) of the patients reported a noise arising from the valve only in upright position in combination with maximum head reclination, and immediately stopped when performing Vasalva’s maneuver. In three out of four of these patients, the noise was also audible for the investigator (FS) with a prepared stethoscope. Patients complaining about a noise had a larger GU deviation from vertical during head reclination (median: −80 vs −43°, p = 0.0007, t-test). A deviations threshold of less than −58.4° excluding audible noise by a negative predictive value of 1 (95 % confidence interval [CI] 0.9 to 1.0). In an experimental setting, the noise came from vibrations of the ball in the cone of the adjustable unit and was restricted to a flow of at least 220 ml/h. The noise frequencies tended to be higher at higher opening pressures.

Conclusions

Valve-related noise development may occur in patients with proGAV valves. This event could be prevented during shunt placement by avoiding posterior tilt of the gravitational unit, especially in patients with a good cervical mobility. The noise might indicate transient peak flows and was not associated with clinical or radiological signs of overdrainage.
Appendix
Available only for authorised users
Literature
1.
go back to reference Chong CC, van Gelder J, Sheridan M (2002) Clinical experience with the low pressure Novus valve in the treatment of adult hydrocephalus. J Clin Neurosci 9:539–543PubMedCrossRef Chong CC, van Gelder J, Sheridan M (2002) Clinical experience with the low pressure Novus valve in the treatment of adult hydrocephalus. J Clin Neurosci 9:539–543PubMedCrossRef
2.
go back to reference Cobb WS, Burns JM, Kercher KW, Matthews BD, James Norton H, Todd Heniford B (2005) Normal intraabdominal pressure in healthy adults. J Surg Res 129:231–235PubMedCrossRef Cobb WS, Burns JM, Kercher KW, Matthews BD, James Norton H, Todd Heniford B (2005) Normal intraabdominal pressure in healthy adults. J Surg Res 129:231–235PubMedCrossRef
3.
go back to reference Hara M, Kadowaki C, Konishi Y, Ogashiwa M, Numoto M, Takeuchi K (1983) A new method for measuring cerebrospinal fluid flow in shunts. J Neurosurg 58:557–561PubMedCrossRef Hara M, Kadowaki C, Konishi Y, Ogashiwa M, Numoto M, Takeuchi K (1983) A new method for measuring cerebrospinal fluid flow in shunts. J Neurosurg 58:557–561PubMedCrossRef
4.
go back to reference Hertel F, Zuchner M, Decker C, Schill S, Bosniak I, Bettag M (2008) The Miethke dual switch valve: experience in 169 adult patients with different kinds of hydrocephalus: an open field study. Minim Invasive Neurosurg 51:147–153PubMedCrossRef Hertel F, Zuchner M, Decker C, Schill S, Bosniak I, Bettag M (2008) The Miethke dual switch valve: experience in 169 adult patients with different kinds of hydrocephalus: an open field study. Minim Invasive Neurosurg 51:147–153PubMedCrossRef
5.
go back to reference Hidaka M, Matsumae M, Ito K, Tsugane R, Suzuki Y (2001) Dynamic measurement of the flow rate in cerebrospinal fluid shunts in hydrocephalic patients. Eur J Nucl Med 28:888–893PubMedCrossRef Hidaka M, Matsumae M, Ito K, Tsugane R, Suzuki Y (2001) Dynamic measurement of the flow rate in cerebrospinal fluid shunts in hydrocephalic patients. Eur J Nucl Med 28:888–893PubMedCrossRef
6.
go back to reference Kaestner S, Kruschat T, Nitzsche N, Deinsberger W (2009) Gravitational shunt units may cause under-drainage in bedridden patients. Acta Neurochir (Wien) 151:217–221, discussion 221CrossRef Kaestner S, Kruschat T, Nitzsche N, Deinsberger W (2009) Gravitational shunt units may cause under-drainage in bedridden patients. Acta Neurochir (Wien) 151:217–221, discussion 221CrossRef
7.
go back to reference Kolenda H, Schoener WF (1995) Shunt-valve noise as an unusual reason for shunt revision. Childs Nerv Syst 11:489–491PubMedCrossRef Kolenda H, Schoener WF (1995) Shunt-valve noise as an unusual reason for shunt revision. Childs Nerv Syst 11:489–491PubMedCrossRef
8.
go back to reference Lemcke J, Meier U (2010) Improved outcome in shunted iNPH with a combination of a Codman Hakim programmable valve and an Aesculap-Miethke ShuntAssistant. Cent Eur Neurosurg 71:113–116PubMedCrossRef Lemcke J, Meier U (2010) Improved outcome in shunted iNPH with a combination of a Codman Hakim programmable valve and an Aesculap-Miethke ShuntAssistant. Cent Eur Neurosurg 71:113–116PubMedCrossRef
9.
go back to reference Lemcke J, Meier U, Muller C, Fritsch M, Kiefer M, Eymann R, Kehler U, Langer N, Schuhmann MU, Speil A, Weber F, Remenez V, Rohde V, Ludwig HC, Stengel D (2012) On the method of a randomised comparison of programmable valves with and without gravitational units: the SVASONA study. Acta Neurochir Suppl 114:243–246PubMedCrossRef Lemcke J, Meier U, Muller C, Fritsch M, Kiefer M, Eymann R, Kehler U, Langer N, Schuhmann MU, Speil A, Weber F, Remenez V, Rohde V, Ludwig HC, Stengel D (2012) On the method of a randomised comparison of programmable valves with and without gravitational units: the SVASONA study. Acta Neurochir Suppl 114:243–246PubMedCrossRef
10.
go back to reference Lemcke J, Meier U, Muller C, Fritsch MJ, Kehler U, Langer N, Kiefer M, Eymann R, Schuhmann MU, Speil A, Weber F, Remenez V, Rohde V, Ludwig HC, Stengel D (2013) Safety and efficacy of gravitational shunt valves in patients with idiopathic normal pressure hydrocephalus: a pragmatic, randomised, open label, multicentre trial (SVASONA). J Neurol Neurosurg Psychiatry. doi:10.1136/jnnp-2012-303936 PubMedCentralPubMed Lemcke J, Meier U, Muller C, Fritsch MJ, Kehler U, Langer N, Kiefer M, Eymann R, Schuhmann MU, Speil A, Weber F, Remenez V, Rohde V, Ludwig HC, Stengel D (2013) Safety and efficacy of gravitational shunt valves in patients with idiopathic normal pressure hydrocephalus: a pragmatic, randomised, open label, multicentre trial (SVASONA). J Neurol Neurosurg Psychiatry. doi:10.​1136/​jnnp-2012-303936 PubMedCentralPubMed
12.
go back to reference Mavrocordatos P, Bissonnette B, Ravussin P (2000) Effects of neck position and head elevation on intracranial pressure in anaesthetized neurosurgical patients: preliminary results. J Neurosurg Anesthesiol 12:10–14PubMedCrossRef Mavrocordatos P, Bissonnette B, Ravussin P (2000) Effects of neck position and head elevation on intracranial pressure in anaesthetized neurosurgical patients: preliminary results. J Neurosurg Anesthesiol 12:10–14PubMedCrossRef
13.
go back to reference Meier U, Lemcke J (2006) First clinical experiences in patients with idiopathic normal-pressure hydrocephalus with the adjustable gravity valve manufactured by Aesculap (proGAV(Aesculap)). Acta Neurochir Suppl 96:368–372PubMedCrossRef Meier U, Lemcke J (2006) First clinical experiences in patients with idiopathic normal-pressure hydrocephalus with the adjustable gravity valve manufactured by Aesculap (proGAV(Aesculap)). Acta Neurochir Suppl 96:368–372PubMedCrossRef
14.
go back to reference Meier U, Lemcke J, Al-Zain F (2008) Course of disease in patients with idiopathic normal pressure hydrocephalus (iNPH): a follow-up study 3, 4 and 5 years following shunt implantation. Acta Neurochir Suppl 102:125–127PubMedCrossRef Meier U, Lemcke J, Al-Zain F (2008) Course of disease in patients with idiopathic normal pressure hydrocephalus (iNPH): a follow-up study 3, 4 and 5 years following shunt implantation. Acta Neurochir Suppl 102:125–127PubMedCrossRef
15.
go back to reference Moser E, Frenzel HU (1976) Qualitative function test of ventriculo-atrial drainage using Cordis-Hakim valves. Zentralbl Neurochir 37:191–195PubMed Moser E, Frenzel HU (1976) Qualitative function test of ventriculo-atrial drainage using Cordis-Hakim valves. Zentralbl Neurochir 37:191–195PubMed
16.
go back to reference Ordia JI, Mortara RW, Spatz EL (1987) Audible cerebrospinal fluid flow through a ventriculoperitoneal shunt. Report of two cases. J Neurosurg 67:460–462PubMedCrossRef Ordia JI, Mortara RW, Spatz EL (1987) Audible cerebrospinal fluid flow through a ventriculoperitoneal shunt. Report of two cases. J Neurosurg 67:460–462PubMedCrossRef
17.
go back to reference Park J, Kim GJ, Hwang SK (2007) Valve inclination influences the performance of gravity-assisted valve. Surg Neurol 68:14–18, discussion 18PubMedCrossRef Park J, Kim GJ, Hwang SK (2007) Valve inclination influences the performance of gravity-assisted valve. Surg Neurol 68:14–18, discussion 18PubMedCrossRef
18.
go back to reference Rossi GF, Maira G, Anile C (1987) Intracranial pressure behaviour and its relation to the outcome of surgical CSF shunting in normotensive hydrocephalus. Neurol Res 9:183–187PubMed Rossi GF, Maira G, Anile C (1987) Intracranial pressure behaviour and its relation to the outcome of surgical CSF shunting in normotensive hydrocephalus. Neurol Res 9:183–187PubMed
19.
go back to reference Sprung C, Schlosser HG, Lemcke J, Meier U, Messing-Junger M, Trost HA, Weber F, Schul C, Rohde V, Ludwig HC, Hopfner J, Sepehrnia A, Mirzayan MJ, Krauss JK (2010) The adjustable proGAV shunt: a prospective safety and reliability multicenter study. Neurosurgery 66:465–474PubMedCrossRef Sprung C, Schlosser HG, Lemcke J, Meier U, Messing-Junger M, Trost HA, Weber F, Schul C, Rohde V, Ludwig HC, Hopfner J, Sepehrnia A, Mirzayan MJ, Krauss JK (2010) The adjustable proGAV shunt: a prospective safety and reliability multicenter study. Neurosurgery 66:465–474PubMedCrossRef
Metadata
Title
Flow-related noise in patients with ventriculoperitoneal shunt using gravitational adjustable valves
Authors
Florian Stockhammer
Christoph Miethke
Thoralf Knitter
Veit Rohde
Christian Sprung
Publication date
01-04-2014
Publisher
Springer Vienna
Published in
Acta Neurochirurgica / Issue 4/2014
Print ISSN: 0001-6268
Electronic ISSN: 0942-0940
DOI
https://doi.org/10.1007/s00701-013-1876-9

Other articles of this Issue 4/2014

Acta Neurochirurgica 4/2014 Go to the issue