Feasibility of data-driven cardiac respiratory motion correction of myocardial perfusion CZT SPECT: A pilot study

Authors: Doumit Daou, MD, PhD, Rémy Sabbah, MD, Carlos Coaguila, MD, Hatem Boulahdour, MD, PhD

Published in: Journal of Nuclear Cardiology | Issue 5/2017

Login to get access

Abstract

Background

We developed a data-driven respiratory motion (RM) correction method (REGAT program) for multiple-pinhole detector CZT SPECT. We verified its clinical feasibility with myocardial perfusion imaging (MPI) and studied its impact on image characteristics.

Methods

This retrospective study included 18 patients having stress/rest 99mTc-Tetrofosmin MPI SPECT. List mode was acquired on CZT SPECT and processed with REGAT. REGAT generates reconstructed RM-gated volumes that are summed either without realignment (NR-SPECT) or after realignment (R-SPECT). For both stress and rest, we calculated the maximal RM in the 3 axis, and image characteristics of both R-SPECT and NR-SPECT: minimum left ventricular (LV) cavity counts (LV-Min), maximum LV myocardial counts (LV-Max), LV contrast, and FWHM of both anterior (FWHM-ant) and inferior (FWHM-inf) LV myocardial walls.

Results

At both stress and rest, cranio-caudal motion was the dominant axial movement and REGAT had a positive impact on image characteristics as reflected by variations between R-SPECT and NR-SPECT in LV-Min, LV-Max, FWHM-ant, FWHM-inf, and contrast. These latter were well correlated to the amplitude of cranio-caudal motion at both stress and rest.

Conclusions

Data-driven RM correction of MPI acquired with CZT SPECT is clinically feasible and easily applicable. It presents interesting impact on image characteristics.
Literature
1.
go back to reference Buechel RR, Pazhenkottil AP, Herzog BA, et al. Real-time breath-hold triggering of myocardial perfusion imaging with a novel cadmium-zinc-telluride detector gamma camera. Eur J Nucl Med Mol Imaging 2010;37:1903–8.CrossRefPubMed Buechel RR, Pazhenkottil AP, Herzog BA, et al. Real-time breath-hold triggering of myocardial perfusion imaging with a novel cadmium-zinc-telluride detector gamma camera. Eur J Nucl Med Mol Imaging 2010;37:1903–8.CrossRefPubMed
2.
go back to reference Cho K, Kumiata S, Okada S, Kumazaki T. Development of respiratory gated myocardial SPECT system. J Nucl Cardiol 1999;6:20–8.CrossRefPubMed Cho K, Kumiata S, Okada S, Kumazaki T. Development of respiratory gated myocardial SPECT system. J Nucl Cardiol 1999;6:20–8.CrossRefPubMed
3.
go back to reference Segars WP, Tsui BMW. Study of the efficacy of respiratory gating in myocardial SPECT using the new 4-D NCAT. IEEE Trans Nucl Sci 2000;47:1192–5.CrossRef Segars WP, Tsui BMW. Study of the efficacy of respiratory gating in myocardial SPECT using the new 4-D NCAT. IEEE Trans Nucl Sci 2000;47:1192–5.CrossRef
4.
go back to reference Livieratos L, Rajappan K, Stegger L, Schafers K, Bailey DL, Camici PG. Respiratory gating of cardiac PET data in list-mode acquisition. Eur J Nucl Med Mol Imaging 2006;33:584–8.CrossRefPubMed Livieratos L, Rajappan K, Stegger L, Schafers K, Bailey DL, Camici PG. Respiratory gating of cardiac PET data in list-mode acquisition. Eur J Nucl Med Mol Imaging 2006;33:584–8.CrossRefPubMed
5.
go back to reference Kovalski G, Israel O, Keidar Z, Frenkel A, Sachs J, Azhari H. Correction of heart motion due to respiration in clinical myocardial perfusion SPECT scans using respiratory gating. J Nucl Med 2007;48:630–6.CrossRefPubMed Kovalski G, Israel O, Keidar Z, Frenkel A, Sachs J, Azhari H. Correction of heart motion due to respiration in clinical myocardial perfusion SPECT scans using respiratory gating. J Nucl Med 2007;48:630–6.CrossRefPubMed
6.
go back to reference Martinez-Moller A, Zikic D, Botnar RM, et al. Dual cardiac-respiratory gated PET: Implementation and results from a feasibility study. Eur J Nucl Med Mol Imaging 2007;34:1447–54.CrossRefPubMed Martinez-Moller A, Zikic D, Botnar RM, et al. Dual cardiac-respiratory gated PET: Implementation and results from a feasibility study. Eur J Nucl Med Mol Imaging 2007;34:1447–54.CrossRefPubMed
7.
go back to reference Ko CL, Wu YW, Cheng MF, Yen RF, Wu WC, Tzen KY. Data-driven respiratory motion tracking and compensation in CZT cameras: A comprehensive analysis of phantom and human images. J Nucl Cardiol 2015;22:308–18.CrossRefPubMed Ko CL, Wu YW, Cheng MF, Yen RF, Wu WC, Tzen KY. Data-driven respiratory motion tracking and compensation in CZT cameras: A comprehensive analysis of phantom and human images. J Nucl Cardiol 2015;22:308–18.CrossRefPubMed
8.
go back to reference Bocher M, Blevis IM, Tsukerman L, Shrem Y, Kovalski G, Volokh L. A fast cardiac gamma camera with dynamic SPECT capabilities: Design, system validation and future potential. Eur J Nucl Med Mol Imaging 2010;37:1887–902.CrossRefPubMedPubMedCentral Bocher M, Blevis IM, Tsukerman L, Shrem Y, Kovalski G, Volokh L. A fast cardiac gamma camera with dynamic SPECT capabilities: Design, system validation and future potential. Eur J Nucl Med Mol Imaging 2010;37:1887–902.CrossRefPubMedPubMedCentral
9.
go back to reference Tawileh M. Development of an optimal method for respiratory gating in nuclear medicine. PhD thesis, University Paris XII, July 2009. Tawileh M. Development of an optimal method for respiratory gating in nuclear medicine. PhD thesis, University Paris XII, July 2009.
10.
go back to reference Daou D, Tawileh M, Khiar-Dechmi A, Coaguila C. Myocardial perfusion imaging with conventional gamma camera: To what level do data-driven, video and belt respiratory motion curves correlate? [abstract]. J Nucl Med 2015;56:1798P. Daou D, Tawileh M, Khiar-Dechmi A, Coaguila C. Myocardial perfusion imaging with conventional gamma camera: To what level do data-driven, video and belt respiratory motion curves correlate? [abstract]. J Nucl Med 2015;56:1798P.
11.
go back to reference Tanaka H, Chikamori T, Tanaka N, et al. Diagnostic performance of a novel cadmium-zinc-telluride gamma camera system assessed using fractional flow reserve. Circ J 2014;78:2727–34.CrossRefPubMed Tanaka H, Chikamori T, Tanaka N, et al. Diagnostic performance of a novel cadmium-zinc-telluride gamma camera system assessed using fractional flow reserve. Circ J 2014;78:2727–34.CrossRefPubMed
12.
go back to reference Duvall WL, Sweeny JM, Croft LB, et al. Comparison of high efficiency CZT SPECT MPI to coronary angiography. J Nucl Cardiol 2011;18:595–604.CrossRefPubMed Duvall WL, Sweeny JM, Croft LB, et al. Comparison of high efficiency CZT SPECT MPI to coronary angiography. J Nucl Cardiol 2011;18:595–604.CrossRefPubMed
13.
go back to reference Duvall WL, Sweeny JM, Croft LB, Ginsberg E, Guma KA, Henzlova MJ. Reduced stress dose with rapid acquisition CZT SPECT MPI in a non-obese clinical population: Comparison to coronary angiography. J Nucl Cardiol 2012;19:19–27.CrossRefPubMed Duvall WL, Sweeny JM, Croft LB, Ginsberg E, Guma KA, Henzlova MJ. Reduced stress dose with rapid acquisition CZT SPECT MPI in a non-obese clinical population: Comparison to coronary angiography. J Nucl Cardiol 2012;19:19–27.CrossRefPubMed
14.
go back to reference Gimelli A, Bottai M, Genovesi D, Giorgetti A, Di Martino F, Marzullo P. High diagnostic accuracy of low-dose gated-SPECT with solid-state ultrafast detectors: Preliminary clinical results. Eur J Nucl Med Mol Imaging 2012;39:83–90.CrossRefPubMed Gimelli A, Bottai M, Genovesi D, Giorgetti A, Di Martino F, Marzullo P. High diagnostic accuracy of low-dose gated-SPECT with solid-state ultrafast detectors: Preliminary clinical results. Eur J Nucl Med Mol Imaging 2012;39:83–90.CrossRefPubMed
15.
go back to reference Buechel RR, Herzog BA, Husmann L, et al. Ultrafast nuclear myocardial perfusion imaging on a new gamma camera with semiconductor detector technique: First clinical validation. Eur J Nucl Med Mol Imaging 2010;37:773–8.CrossRefPubMed Buechel RR, Herzog BA, Husmann L, et al. Ultrafast nuclear myocardial perfusion imaging on a new gamma camera with semiconductor detector technique: First clinical validation. Eur J Nucl Med Mol Imaging 2010;37:773–8.CrossRefPubMed
16.
go back to reference Fiechter M, Ghadri JR, Kuest SM, et al. Nuclear myocardial perfusion imaging with a novel cadmium-zinc-telluride detector SPECT/CT device: First validation versus invasive coronary angiography. Eur J Nucl Med Mol Imaging 2011;38:2025–30.CrossRefPubMed Fiechter M, Ghadri JR, Kuest SM, et al. Nuclear myocardial perfusion imaging with a novel cadmium-zinc-telluride detector SPECT/CT device: First validation versus invasive coronary angiography. Eur J Nucl Med Mol Imaging 2011;38:2025–30.CrossRefPubMed
17.
go back to reference Mouden M, Ottervanger JP, Knollema S, et al. Myocardial perfusion imaging with a cadmium zinc telluride-based gamma camera versus invasive fractional flow reserve. Eur J Nucl Med Mol Imaging 2014;41:956–62.CrossRefPubMed Mouden M, Ottervanger JP, Knollema S, et al. Myocardial perfusion imaging with a cadmium zinc telluride-based gamma camera versus invasive fractional flow reserve. Eur J Nucl Med Mol Imaging 2014;41:956–62.CrossRefPubMed
18.
go back to reference Daou D. Respiratory motion handling is mandatory to accomplish the high-resolution PET destiny [editorial]. Eur J Nucl Med Mol Imaging 2008;35:1961–70.CrossRefPubMed Daou D. Respiratory motion handling is mandatory to accomplish the high-resolution PET destiny [editorial]. Eur J Nucl Med Mol Imaging 2008;35:1961–70.CrossRefPubMed
19.
go back to reference McLeish K, Hill DL, Atkinson D, Blackall JM, Razavi R. A study of the motion and deformation of the heart due to respiration. IEEE Trans Med Imaging 2002;21:1142–50.CrossRefPubMed McLeish K, Hill DL, Atkinson D, Blackall JM, Razavi R. A study of the motion and deformation of the heart due to respiration. IEEE Trans Med Imaging 2002;21:1142–50.CrossRefPubMed
Metadata
Title
Feasibility of data-driven cardiac respiratory motion correction of myocardial perfusion CZT SPECT: A pilot study
Authors
Doumit Daou, MD, PhD
Rémy Sabbah, MD
Carlos Coaguila, MD
Hatem Boulahdour, MD, PhD
Publication date
01-10-2017
Publisher
Springer US
Published in
Journal of Nuclear Cardiology / Issue 5/2017
Print ISSN: 1071-3581
Electronic ISSN: 1532-6551
DOI
https://doi.org/10.1007/s12350-016-0493-4

Other articles of this Issue 5/2017

Journal of Nuclear Cardiology 5/2017 Go to the issue

CME ARTICLE Review Article

Cardiac autonomic innervation

Technologist Corner

The EXXERT Study