Fast calibration of electromagnetically tracked oblique-viewing rigid endoscopes

Authors: Xinyang Liu, Christina E. Rice, Raj Shekhar

Published in: International Journal of Computer Assisted Radiology and Surgery | Issue 10/2017

Login to get access

Abstract

Purpose

The oblique-viewing (i.e., angled) rigid endoscope is a commonly used tool in conventional endoscopic surgeries. The relative rotation between its two moveable parts, the telescope and the camera head, creates a rotation offset between the actual and the projection of an object in the camera image. A calibration method tailored to compensate such offset is needed.

Methods

We developed a fast calibration method for oblique-viewing rigid endoscopes suitable for clinical use. In contrast to prior approaches based on optical tracking, we used electromagnetic (EM) tracking as the external tracking hardware to improve compactness and practicality. Two EM sensors were mounted on the telescope and the camera head, respectively, with considerations to minimize EM tracking errors. Single-image calibration was incorporated into the method, and a sterilizable plate, laser-marked with the calibration pattern, was also developed. Furthermore, we proposed a general algorithm to estimate the rotation center in the camera image. Formulas for updating the camera matrix in terms of clockwise and counterclockwise rotations were also developed.

Results

The proposed calibration method was validated using a conventional \(30{^{\circ }}\), 5-mm laparoscope. Freehand calibrations were performed using the proposed method, and the calibration time averaged 2 min and 8 s. The calibration accuracy was evaluated in a simulated clinical setting with several surgical tools present in the magnetic field of EM tracking. The root-mean-square re-projection error averaged 4.9 pixel (range 2.4–8.5 pixel, with image resolution of \(1280 \times 720)\) for rotation angles ranged from \(-40.3{^{\circ }}\) to \(174.7{^{\circ }}\).

Conclusions

We developed a method for fast and accurate calibration of oblique-viewing rigid endoscopes. The method was also designed to be performed in the operating room and will therefore support clinical translation of many emerging endoscopic computer-assisted surgical systems.
Appendix
Available only for authorised users
Literature
1.
go back to reference Tsai RY (1987) A versatile camera calibration technique for high-accuracy 3-D machine vision methodology using off-the-shelf TV cameras and lenses. IEEE J Robot Automat 3:323–344CrossRef Tsai RY (1987) A versatile camera calibration technique for high-accuracy 3-D machine vision methodology using off-the-shelf TV cameras and lenses. IEEE J Robot Automat 3:323–344CrossRef
2.
go back to reference Heikkila J, Silven O (1997) A four-step camera calibration procedure with implicit image correction. In: Proceedings of IEEE computer society conference computer vision pattern recognition, pp 1106–1112 Heikkila J, Silven O (1997) A four-step camera calibration procedure with implicit image correction. In: Proceedings of IEEE computer society conference computer vision pattern recognition, pp 1106–1112
3.
go back to reference Zhang Z (1999) Flexible camera calibration by viewing a plane from unknown orientations. In: Proceedings of international conference on computer vision, pp 666–673 Zhang Z (1999) Flexible camera calibration by viewing a plane from unknown orientations. In: Proceedings of international conference on computer vision, pp 666–673
5.
go back to reference Shiu Y, Ahmad S (1989) Calibration of wrist-mounted robotic sensors by solving homogeneous transform equations of the form ax = xb. IEEE Trans Bobot Autom 5(1):16–29CrossRef Shiu Y, Ahmad S (1989) Calibration of wrist-mounted robotic sensors by solving homogeneous transform equations of the form ax = xb. IEEE Trans Bobot Autom 5(1):16–29CrossRef
6.
go back to reference Feuerstein M, Mussack T, Heining SM, Navab N (2008) Intraoperative laparoscope augmentation for port placement and resection planning in minimally invasive liver resection. IEEE Trans Med Imaging 27(3):355–369CrossRefPubMed Feuerstein M, Mussack T, Heining SM, Navab N (2008) Intraoperative laparoscope augmentation for port placement and resection planning in minimally invasive liver resection. IEEE Trans Med Imaging 27(3):355–369CrossRefPubMed
7.
go back to reference Shekhar R, Dandekar O, Bhat V, Philip M, Lei P, Godinez C, Sutton E, George I, Kavic S, Mezrich R, Park A (2010) Live augmented reality: a new visualization method for laparoscopic surgery using continuous volumetric computed tomography. Surg Endosc 24(8):1976–1985CrossRefPubMed Shekhar R, Dandekar O, Bhat V, Philip M, Lei P, Godinez C, Sutton E, George I, Kavic S, Mezrich R, Park A (2010) Live augmented reality: a new visualization method for laparoscopic surgery using continuous volumetric computed tomography. Surg Endosc 24(8):1976–1985CrossRefPubMed
8.
go back to reference Kang X, Azizian M, Wilson E, Wu K, Martin AD, Kane TD, Peters CA, Cleary K, Shekhar R (2014) Stereoscopic augmented reality for laparoscopic surgery. Surg Endosc 28(7):2227–2235CrossRefPubMed Kang X, Azizian M, Wilson E, Wu K, Martin AD, Kane TD, Peters CA, Cleary K, Shekhar R (2014) Stereoscopic augmented reality for laparoscopic surgery. Surg Endosc 28(7):2227–2235CrossRefPubMed
9.
go back to reference Cheung CL, Wedlake C, Moore J, Pautler SE, Peters TM (2010) Fused video and ultrasound images for minimally invasive partial nephrectomy: a phantom study. Proc Med Image Comput Comput Assist Interv 13(Pt 3):408–415 Cheung CL, Wedlake C, Moore J, Pautler SE, Peters TM (2010) Fused video and ultrasound images for minimally invasive partial nephrectomy: a phantom study. Proc Med Image Comput Comput Assist Interv 13(Pt 3):408–415
10.
go back to reference Liu X, Kang S, Plishker W, Zaki G, Kane TD, Shekhar R (2016) Laparoscopic stereoscopic augmented reality: toward a clinically viable electromagnetic tracking solution. J Med Imaging 3(4):045001CrossRef Liu X, Kang S, Plishker W, Zaki G, Kane TD, Shekhar R (2016) Laparoscopic stereoscopic augmented reality: toward a clinically viable electromagnetic tracking solution. J Med Imaging 3(4):045001CrossRef
11.
go back to reference Yamaguchi T, Nakamoto M, Sato Y, Konishi K, Hashizume M, Sugano N, Yoshikawa H, Tamura S (2004) Development of a camera model and calibration procedure for oblique-viewing endoscopes. Comput Aided Surg 9(5):203–214PubMed Yamaguchi T, Nakamoto M, Sato Y, Konishi K, Hashizume M, Sugano N, Yoshikawa H, Tamura S (2004) Development of a camera model and calibration procedure for oblique-viewing endoscopes. Comput Aided Surg 9(5):203–214PubMed
12.
go back to reference Wu C, Jaramaz B, Narasimhan SG (2010) A full geometric and photometric calibration method for oblique-viewing endoscopes. Comput Aided Surg 15(1–3):19–31CrossRefPubMed Wu C, Jaramaz B, Narasimhan SG (2010) A full geometric and photometric calibration method for oblique-viewing endoscopes. Comput Aided Surg 15(1–3):19–31CrossRefPubMed
13.
go back to reference De Buck S, Maes F, D’Hoore A, Suetens P (2007) Evaluation of a novel calibration technique for optically tracked oblique laparoscopes. Proc Med Image Comput Comput Assist Interv 10(Pt 1):467–474 De Buck S, Maes F, D’Hoore A, Suetens P (2007) Evaluation of a novel calibration technique for optically tracked oblique laparoscopes. Proc Med Image Comput Comput Assist Interv 10(Pt 1):467–474
14.
go back to reference Feuerstein M, Reichl T, Vogel J, Traub J, Navab N (2009) Magneto-optical tracking of flexible laparoscopic ultrasound: model-based online detection and correction of magnetic tracking errors. IEEE Trans Med Imaging 28(6):951–967CrossRefPubMed Feuerstein M, Reichl T, Vogel J, Traub J, Navab N (2009) Magneto-optical tracking of flexible laparoscopic ultrasound: model-based online detection and correction of magnetic tracking errors. IEEE Trans Med Imaging 28(6):951–967CrossRefPubMed
15.
go back to reference Melo R, Barreto JP, Falcão G (2012) A new solution for camera calibration and real-time image distortion correction in medical endoscopy-initial technical evaluation. IEEE Trans Biomed Eng 59(3):634–644CrossRefPubMed Melo R, Barreto JP, Falcão G (2012) A new solution for camera calibration and real-time image distortion correction in medical endoscopy-initial technical evaluation. IEEE Trans Biomed Eng 59(3):634–644CrossRefPubMed
16.
go back to reference Barreto JP, Roquette J, Sturm P, Fonseca F (2009) Automatic camera calibration applied to medical endoscopy. In: Proceedings of British machine vision conference, pp 1-10 Barreto JP, Roquette J, Sturm P, Fonseca F (2009) Automatic camera calibration applied to medical endoscopy. In: Proceedings of British machine vision conference, pp 1-10
17.
go back to reference Liu X, Plishker W, Zaki G, Kang S, Kane TD, Shekhar R (2016) On-demand calibration and evaluation for electromagnetically tracked laparoscope in augmented reality visualization. Int J Comput Assist Radiol Surg 11(6):1163–1171CrossRefPubMedPubMedCentral Liu X, Plishker W, Zaki G, Kang S, Kane TD, Shekhar R (2016) On-demand calibration and evaluation for electromagnetically tracked laparoscope in augmented reality visualization. Int J Comput Assist Radiol Surg 11(6):1163–1171CrossRefPubMedPubMedCentral
18.
go back to reference Franz AM, Haidegger T, Birkfellner W, Cleary K, Peters TM, Maier-Hein L (2014) Electromagnetic tracking in medicine-a review of technology, validation, and applications. IEEE Trans Med Imaging 33(8):1702–1725CrossRefPubMed Franz AM, Haidegger T, Birkfellner W, Cleary K, Peters TM, Maier-Hein L (2014) Electromagnetic tracking in medicine-a review of technology, validation, and applications. IEEE Trans Med Imaging 33(8):1702–1725CrossRefPubMed
19.
go back to reference Maier-Hein L, Franz AM, Birkfellner W, Hummel J, Gergel I, Wegner I, Meinzer HP (2012) Standardized assessment of new electromagnetic field generators in an interventional radiology setting. Med Phys 39(6):3424–3434CrossRefPubMed Maier-Hein L, Franz AM, Birkfellner W, Hummel J, Gergel I, Wegner I, Meinzer HP (2012) Standardized assessment of new electromagnetic field generators in an interventional radiology setting. Med Phys 39(6):3424–3434CrossRefPubMed
20.
go back to reference Liu X, Kang S, Wilson E, Peters CA, Kane TD, Shekhar R (2014) Evaluation of electromagnetic tracking for stereoscopic augmented reality laparoscopic visualization. In: Proceedings of MICCAI workshop on clinical image-based procedures: translational research in medical imaging, vol 8680, pp 84–91 Liu X, Kang S, Wilson E, Peters CA, Kane TD, Shekhar R (2014) Evaluation of electromagnetic tracking for stereoscopic augmented reality laparoscopic visualization. In: Proceedings of MICCAI workshop on clinical image-based procedures: translational research in medical imaging, vol 8680, pp 84–91
21.
go back to reference Nijkamp J, Schermers B, Schmitz S, de Jonge S, Kuhlmann K, van der Heijden F, Sonke J-J, Ruers T (2016) Comparing position and orientation accuracy of different electromagnetic sensors for tracking during interventions. Int J Comput Assist Radiol Surg 11(8):1487–1498CrossRefPubMed Nijkamp J, Schermers B, Schmitz S, de Jonge S, Kuhlmann K, van der Heijden F, Sonke J-J, Ruers T (2016) Comparing position and orientation accuracy of different electromagnetic sensors for tracking during interventions. Int J Comput Assist Radiol Surg 11(8):1487–1498CrossRefPubMed
Metadata
Title
Fast calibration of electromagnetically tracked oblique-viewing rigid endoscopes
Authors
Xinyang Liu
Christina E. Rice
Raj Shekhar
Publication date
01-10-2017
Publisher
Springer International Publishing
Published in
International Journal of Computer Assisted Radiology and Surgery / Issue 10/2017
Print ISSN: 1861-6410
Electronic ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-017-1623-4

Other articles of this Issue 10/2017

International Journal of Computer Assisted Radiology and Surgery 10/2017 Go to the issue