Factors affecting hemoglobin measurement

Author: Lauren Berkow

Published in: Journal of Clinical Monitoring and Computing | Issue 5/2013

Login to get access

Abstract

A review of the literature shows that current “standard” laboratory measurements for hemoglobin are subject to numerous factors that affect both accuracy and reliability. In addition, total hemoglobin concentration measurements are subject to numerous factors that affect the “true” hemoglobin value. This article discusses both the physiologic factors that influence hemoglobin levels and the technical aspects and variability among the different measurement methodologies currently available.
Footnotes
1
Beer-Lambert law is the linear relationship between absorbance and concentration of an absorbing species and is typically written as:
A = a(λ) × b × c
where A is the measured absorbance, a(λ) is a wavelength-dependent absorptivity coefficient, b is the path length, and c is the analyte concentration.
 
Literature
1.
go back to reference Hünefeld FL. Die Chemismus in der thierischen Organization. Leipzig: Brockhaus FA; 1840. Hünefeld FL. Die Chemismus in der thierischen Organization. Leipzig: Brockhaus FA; 1840.
2.
go back to reference Perutz MF. Structure of hemoglobin. Brookhaven Symp Biol. 1960;13:165–83.PubMed Perutz MF. Structure of hemoglobin. Brookhaven Symp Biol. 1960;13:165–83.PubMed
3.
go back to reference Hoppe-Seyler F. Über das Verhalten des Blutfarbstoffes im Spektrum des Sonnenlichtes. Arch Pathol Anat Physiol. 1862;23:446. Hoppe-Seyler F. Über das Verhalten des Blutfarbstoffes im Spektrum des Sonnenlichtes. Arch Pathol Anat Physiol. 1862;23:446.
4.
go back to reference Lichtman MA, Kaushansky K, Kipps TJ, Prchal JT, Levi MM. Williams hematology. 8th ed. New York: McGraw Hill Medical; 2011. Lichtman MA, Kaushansky K, Kipps TJ, Prchal JT, Levi MM. Williams hematology. 8th ed. New York: McGraw Hill Medical; 2011.
5.
go back to reference Medical laboratories—Reduction of error through risk management and continual improvement: International Organization for Standardization: ISO/TS 22367; 2008. Medical laboratories—Reduction of error through risk management and continual improvement: International Organization for Standardization: ISO/TS 22367; 2008.
6.
go back to reference Lippi G, Guidi GC, Mattiuzzi C, Plebani M. Preanalytical variability: the dark side of the moon in laboratory testing. Clin Chem Lab Med. 2006;44:358–65.PubMed Lippi G, Guidi GC, Mattiuzzi C, Plebani M. Preanalytical variability: the dark side of the moon in laboratory testing. Clin Chem Lab Med. 2006;44:358–65.PubMed
7.
go back to reference Daae LN, Halvorsen S, Mathisen PM, Mironska K. A comparison between haematological parameters in ‘capillary’ and venous blood from healthy adults. Scand J Clin Lab Invest. 1988;48:723–6.PubMed Daae LN, Halvorsen S, Mathisen PM, Mironska K. A comparison between haematological parameters in ‘capillary’ and venous blood from healthy adults. Scand J Clin Lab Invest. 1988;48:723–6.PubMed
8.
go back to reference Neufeld L, Garcia-Guerra A, Sanchez-Francia D, Newton-Sanchez O, Ramirez-Villalobos MD, Rivera-Dommarco J. Hemoglobin measured by Hemocue and a reference method in venous and capillary blood: a validation study. Salud Publica Mex. 2002;44:219–27.PubMedCrossRef Neufeld L, Garcia-Guerra A, Sanchez-Francia D, Newton-Sanchez O, Ramirez-Villalobos MD, Rivera-Dommarco J. Hemoglobin measured by Hemocue and a reference method in venous and capillary blood: a validation study. Salud Publica Mex. 2002;44:219–27.PubMedCrossRef
9.
go back to reference Mokken FC, van der Waart FJ, Henny CP, Goedhart PT, Gelb AW. Differences in peripheral arterial and venous hemorheologic parameters. Ann Hematol. 1996;73:135–7.PubMedCrossRef Mokken FC, van der Waart FJ, Henny CP, Goedhart PT, Gelb AW. Differences in peripheral arterial and venous hemorheologic parameters. Ann Hematol. 1996;73:135–7.PubMedCrossRef
10.
go back to reference Yang ZW, Yang SH, Chen L, Qu J, Zhu J, Tang Z. Comparison of blood counts in venous, fingertip and arterial blood and their measurement variation. Clin Lab Haematol. 2001;23:155–9.PubMedCrossRef Yang ZW, Yang SH, Chen L, Qu J, Zhu J, Tang Z. Comparison of blood counts in venous, fingertip and arterial blood and their measurement variation. Clin Lab Haematol. 2001;23:155–9.PubMedCrossRef
11.
go back to reference Junge B, Hoffmeister H, Feddersen HM, Rocker L. Standardisation of obtaining blood samples: influence of tourniquet application on 33 constituents of blood and serum (author’s transl). Dtsch Med Wochenschr. 1978;103:260–5.PubMedCrossRef Junge B, Hoffmeister H, Feddersen HM, Rocker L. Standardisation of obtaining blood samples: influence of tourniquet application on 33 constituents of blood and serum (author’s transl). Dtsch Med Wochenschr. 1978;103:260–5.PubMedCrossRef
12.
go back to reference Lippi G, Salvagno GL, Solero GP, Guidi GC. The influence of the tourniquet time on hematological testing for antidoping purposes. Int J Sports Med. 2006;27:359–62.PubMedCrossRef Lippi G, Salvagno GL, Solero GP, Guidi GC. The influence of the tourniquet time on hematological testing for antidoping purposes. Int J Sports Med. 2006;27:359–62.PubMedCrossRef
13.
go back to reference Lippi G, Salvagno GL, Montagnana M, Franchini M, Guidi GC. Venous stasis and routine hematologic testing. Clin Lab Haematol. 2006;28:332–7.PubMedCrossRef Lippi G, Salvagno GL, Montagnana M, Franchini M, Guidi GC. Venous stasis and routine hematologic testing. Clin Lab Haematol. 2006;28:332–7.PubMedCrossRef
14.
go back to reference Cengiz M, Ulker P, Meiselman HJ, Baskurt OK. Influence of tourniquet application on venous blood sampling for serum chemistry, hematological parameters, leukocyte activation and erythrocyte mechanical properties. Clin Chem Lab Med. 2009;47:769–76.PubMedCrossRef Cengiz M, Ulker P, Meiselman HJ, Baskurt OK. Influence of tourniquet application on venous blood sampling for serum chemistry, hematological parameters, leukocyte activation and erythrocyte mechanical properties. Clin Chem Lab Med. 2009;47:769–76.PubMedCrossRef
15.
go back to reference Lundvall J, Bjerkhoel P. Failure of hemoconcentration during standing to reveal plasma volume decline induced in the erect posture. J Appl Physiol. 1994;77:2155–62.PubMed Lundvall J, Bjerkhoel P. Failure of hemoconcentration during standing to reveal plasma volume decline induced in the erect posture. J Appl Physiol. 1994;77:2155–62.PubMed
16.
go back to reference Ahlgrim C, Pottgiesser T, Robinson N, Sottas PE, Ruecker G, Schumacher YO. Are 10 min of seating enough to guarantee stable haemoglobin and haematocrit readings for the athlete’s biological passport? Int J Lab Hematol. 2012;32:506–11.CrossRef Ahlgrim C, Pottgiesser T, Robinson N, Sottas PE, Ruecker G, Schumacher YO. Are 10 min of seating enough to guarantee stable haemoglobin and haematocrit readings for the athlete’s biological passport? Int J Lab Hematol. 2012;32:506–11.CrossRef
17.
go back to reference Goldner F, Jacob G, Raj SR, Robertson D. The importance of recognizing postural pseudoanemia. Compr Ther. 2006;32:51–5.PubMedCrossRef Goldner F, Jacob G, Raj SR, Robertson D. The importance of recognizing postural pseudoanemia. Compr Ther. 2006;32:51–5.PubMedCrossRef
18.
go back to reference Pocock SJ, Ashby D, Shaper AG, Walker M, Broughton PM. Diurnal variations in serum biochemical and haematological measurements. J Clin Pathol. 1989;42:172–9.PubMedCrossRef Pocock SJ, Ashby D, Shaper AG, Walker M, Broughton PM. Diurnal variations in serum biochemical and haematological measurements. J Clin Pathol. 1989;42:172–9.PubMedCrossRef
19.
go back to reference Schumacher YO, Wenning M, Robinson N, Sottas PE, Ruecker G, Pottgiesser T. Diurnal and exercise-related variability of haemoglobin and reticulocytes in athletes. Int J Sports Med. 2010;31:225–30.PubMedCrossRef Schumacher YO, Wenning M, Robinson N, Sottas PE, Ruecker G, Pottgiesser T. Diurnal and exercise-related variability of haemoglobin and reticulocytes in athletes. Int J Sports Med. 2010;31:225–30.PubMedCrossRef
20.
go back to reference Looker AC, Sempos CT, Liu KA, Johnson CL, Gunter EW. Within-person variance in biochemical indicators of iron status: effects on prevalence estimates. Am J Clin Nutr. 1990;52:541–7.PubMed Looker AC, Sempos CT, Liu KA, Johnson CL, Gunter EW. Within-person variance in biochemical indicators of iron status: effects on prevalence estimates. Am J Clin Nutr. 1990;52:541–7.PubMed
21.
go back to reference Morris SS, Ruel MT, Cohen RJ, Dewey KG, de la Briere B, Hassan MN. Precision, accuracy, and reliability of hemoglobin assessment with use of capillary blood. Am J Clin Nutr. 1999;69:1243–8.PubMed Morris SS, Ruel MT, Cohen RJ, Dewey KG, de la Briere B, Hassan MN. Precision, accuracy, and reliability of hemoglobin assessment with use of capillary blood. Am J Clin Nutr. 1999;69:1243–8.PubMed
22.
go back to reference Boulton FE, Nightingale MJ, Reynolds W. Improved strategy for screening prospective blood donors for anaemia. Transfus Med. 1994;4:221–5.PubMedCrossRef Boulton FE, Nightingale MJ, Reynolds W. Improved strategy for screening prospective blood donors for anaemia. Transfus Med. 1994;4:221–5.PubMedCrossRef
23.
go back to reference Ebert RV, Stead EA Jr. Demonstration that the cell plasma ratio of blood contained in minute vessels is lower than that of venous blood. J Clin Invest. 1941;20:317–21.PubMedCrossRef Ebert RV, Stead EA Jr. Demonstration that the cell plasma ratio of blood contained in minute vessels is lower than that of venous blood. J Clin Invest. 1941;20:317–21.PubMedCrossRef
24.
go back to reference Recommendations for reference method for haemoglobinometry in human blood (ICSH standard 1986) and specifications for international haemiglobincyanide reference preparation (3rd edition). International Committee for Standardization in Haematology; Expert Panel on Haemoglobinometry. Clin Lab Haematol. 1987;9(1):73–79. Recommendations for reference method for haemoglobinometry in human blood (ICSH standard 1986) and specifications for international haemiglobincyanide reference preparation (3rd edition). International Committee for Standardization in Haematology; Expert Panel on Haemoglobinometry. Clin Lab Haematol. 1987;9(1):73–79.
25.
go back to reference Zwart A, van Assendelft OW, Bull BS, England JM, Lewis SM, Zijlstra WG. Recommendations for reference method for haemoglobinometry in human blood (ICSH standard 1995) and specifications for international haemiglobinocyanide standard (4th edition). J Clin Pathol. 1996;49:271–4.PubMedCrossRef Zwart A, van Assendelft OW, Bull BS, England JM, Lewis SM, Zijlstra WG. Recommendations for reference method for haemoglobinometry in human blood (ICSH standard 1995) and specifications for international haemiglobinocyanide standard (4th edition). J Clin Pathol. 1996;49:271–4.PubMedCrossRef
26.
go back to reference Gehring H, Duembgen L, Peterlein M, Hagelberg S, Dibbelt L. Hemoximetry as the “gold standard?” Error assessment based on differences among identical blood gas analyzer devices of five manufacturers. Anesth Analg. 2007;105:S24–30.PubMedCrossRef Gehring H, Duembgen L, Peterlein M, Hagelberg S, Dibbelt L. Hemoximetry as the “gold standard?” Error assessment based on differences among identical blood gas analyzer devices of five manufacturers. Anesth Analg. 2007;105:S24–30.PubMedCrossRef
27.
go back to reference Graham MD. The Coulter principal: foundation of an industry. J Lab Autom. 2003;8(6):72–81.CrossRef Graham MD. The Coulter principal: foundation of an industry. J Lab Autom. 2003;8(6):72–81.CrossRef
28.
go back to reference Pinkerton PH, Spence I, Ogilvie JC, Ronald WA, Marchant P, Ray PK. An assessment of the Coulter counter model S. J Clin Pathol. 1970;23(1):68–76.PubMedCrossRef Pinkerton PH, Spence I, Ogilvie JC, Ronald WA, Marchant P, Ray PK. An assessment of the Coulter counter model S. J Clin Pathol. 1970;23(1):68–76.PubMedCrossRef
29.
go back to reference Hematocrit determination in the i-STAT system and comparison to other methods. iSTAT Technical Bulletin. Abbot Point of Care Inc. 2011. Hematocrit determination in the i-STAT system and comparison to other methods. iSTAT Technical Bulletin. Abbot Point of Care Inc. 2011.
30.
go back to reference Gehring H, Hornberger C, Dibbelt L, Rothsiqkeit A, Gerlach K, Schumacher J, et al. Accuracy of point-of-care-testing (POCT) for determining hemoglobin concentrations. Acta Anaesthesiol Scand. 2002;46:980–6.PubMedCrossRef Gehring H, Hornberger C, Dibbelt L, Rothsiqkeit A, Gerlach K, Schumacher J, et al. Accuracy of point-of-care-testing (POCT) for determining hemoglobin concentrations. Acta Anaesthesiol Scand. 2002;46:980–6.PubMedCrossRef
31.
go back to reference Brunelle JA, Degtiarov AM, Moran RF, Race LA. Simultaneous measurement of total hemoglobin and its derivatives in blood using CO-Oximeters: analytical principles; their application in selecting analytical wavelengths and reference methods; a comparison of the results of the choices made. Scand J Clin Lab Invest Suppl. 1996;224:47–69.PubMedCrossRef Brunelle JA, Degtiarov AM, Moran RF, Race LA. Simultaneous measurement of total hemoglobin and its derivatives in blood using CO-Oximeters: analytical principles; their application in selecting analytical wavelengths and reference methods; a comparison of the results of the choices made. Scand J Clin Lab Invest Suppl. 1996;224:47–69.PubMedCrossRef
32.
go back to reference Holbek CC. New developments in the measurement of CO-Oximetry. Anesth Analg. 2002;94(Suppl 1):S89–92.PubMed Holbek CC. New developments in the measurement of CO-Oximetry. Anesth Analg. 2002;94(Suppl 1):S89–92.PubMed
33.
go back to reference Patel KP, Hay GW, Cheteri MK, Holt DW. Hemoglobin test result variability and cost analysis of eight different analyzers during open heart surgery. J Extra Corpor Technol. 2007;39:10–7.PubMed Patel KP, Hay GW, Cheteri MK, Holt DW. Hemoglobin test result variability and cost analysis of eight different analyzers during open heart surgery. J Extra Corpor Technol. 2007;39:10–7.PubMed
34.
go back to reference Hematocrit/HCT and calculated hemoglobin/Hb. Abbott Point of Care, Abbott Park; 2011. Hematocrit/HCT and calculated hemoglobin/Hb. Abbott Point of Care, Abbott Park; 2011.
35.
go back to reference Hopfer SM, Nadeau FL, Sundra M, Makowski GS. Effect of protein on hemoglobin and hematocrit assays with a conductivity-based point-of-care testing device: comparison with optical methods. Ann Clin Lab Sci. 2004;34:75–82.PubMed Hopfer SM, Nadeau FL, Sundra M, Makowski GS. Effect of protein on hemoglobin and hematocrit assays with a conductivity-based point-of-care testing device: comparison with optical methods. Ann Clin Lab Sci. 2004;34:75–82.PubMed
36.
go back to reference Agarwal R, Heinz T. Bedside hemoglobinometry in hemodialysis patients: lessons from point-of-care testing. ASAIO J. 2001;47:240–3.PubMedCrossRef Agarwal R, Heinz T. Bedside hemoglobinometry in hemodialysis patients: lessons from point-of-care testing. ASAIO J. 2001;47:240–3.PubMedCrossRef
37.
go back to reference Bhaskaram P, Balakrishna N, Radhakrishna KV, Krishnaswamy K. Validation of hemoglobin estimation using Hemocue. Indian J Pediatr. 2003;70:25–8.PubMedCrossRef Bhaskaram P, Balakrishna N, Radhakrishna KV, Krishnaswamy K. Validation of hemoglobin estimation using Hemocue. Indian J Pediatr. 2003;70:25–8.PubMedCrossRef
38.
go back to reference 510(k) Premarket notification, K080238; Masimo rainbow set radical 7 pulse CO-Oximeter, Masimo rainbow set rad 87 pulse CO-Oximeter. In: Fa D, editor. Administration. Silver Spring, MD: Food and Drug Administration; 2008. 510(k) Premarket notification, K080238; Masimo rainbow set radical 7 pulse CO-Oximeter, Masimo rainbow set rad 87 pulse CO-Oximeter. In: Fa D, editor. Administration. Silver Spring, MD: Food and Drug Administration; 2008.
39.
go back to reference Miller RD, Ward TA, Shiboski SC, Cohen NH. A comparison of three methods of hemoglobin monitoring in patients undergoing spine surgery. Anesth Analg. 2011;112(4):858–63.PubMedCrossRef Miller RD, Ward TA, Shiboski SC, Cohen NH. A comparison of three methods of hemoglobin monitoring in patients undergoing spine surgery. Anesth Analg. 2011;112(4):858–63.PubMedCrossRef
40.
go back to reference Nguyen BV, Vincent JL, Nowak E, Coat M, Paleiron N, Gouny P, et al. The accuracy of noninvasive hemoglobin measurement by multiwavelength pulse oximetry after cardiac surgery. Anesth Analg. 2011;113:1052–7.PubMedCrossRef Nguyen BV, Vincent JL, Nowak E, Coat M, Paleiron N, Gouny P, et al. The accuracy of noninvasive hemoglobin measurement by multiwavelength pulse oximetry after cardiac surgery. Anesth Analg. 2011;113:1052–7.PubMedCrossRef
41.
go back to reference Macknet MR, Allard M, Applegate RL 2nd, Rook J. The accuracy of noninvasive and continuous total hemoglobin measurement by pulse CO-Oximetry in human subjects undergoing hemodilution. Anesth Analg. 2010;111:1424–6.PubMedCrossRef Macknet MR, Allard M, Applegate RL 2nd, Rook J. The accuracy of noninvasive and continuous total hemoglobin measurement by pulse CO-Oximetry in human subjects undergoing hemodilution. Anesth Analg. 2010;111:1424–6.PubMedCrossRef
42.
go back to reference Berkow L, Rotolo S, Mirski E. Continuous noninvasive hemoglobin monitoring during complex spine surgery. Anesth Analg. 2011;113:1396–402.PubMedCrossRef Berkow L, Rotolo S, Mirski E. Continuous noninvasive hemoglobin monitoring during complex spine surgery. Anesth Analg. 2011;113:1396–402.PubMedCrossRef
43.
go back to reference Frasca D, Dahyot-Fizelier C, Catherine K, Levrat Q, Debaene B, Mimoz O. Accuracy of a continuous noninvasive hemoglobin monitor in intensive care unit patients. Crit Care Med. 2011;39:2277–82.PubMedCrossRef Frasca D, Dahyot-Fizelier C, Catherine K, Levrat Q, Debaene B, Mimoz O. Accuracy of a continuous noninvasive hemoglobin monitor in intensive care unit patients. Crit Care Med. 2011;39:2277–82.PubMedCrossRef
44.
go back to reference O’Reilly M. Considerations for evaluating the accuracy of hemoglobin monitoring. Anesthesiology. 2012;117:429–30.PubMedCrossRef O’Reilly M. Considerations for evaluating the accuracy of hemoglobin monitoring. Anesthesiology. 2012;117:429–30.PubMedCrossRef
45.
go back to reference Coquin J, Dewitte A, Le Manach Y, Caujolle M, Joannes-Boyau O, Fleureau C, Janvier G, Ouattara A. Precision of noninvasive hemoglobin-level measurement by pulse CO-Oximetry in patients admitted to intensive care units for severe gastrointestinal bleeds. Crit Care Med. 2012;40:2576–82.PubMedCrossRef Coquin J, Dewitte A, Le Manach Y, Caujolle M, Joannes-Boyau O, Fleureau C, Janvier G, Ouattara A. Precision of noninvasive hemoglobin-level measurement by pulse CO-Oximetry in patients admitted to intensive care units for severe gastrointestinal bleeds. Crit Care Med. 2012;40:2576–82.PubMedCrossRef
46.
go back to reference Lamhaut L, Apriotesei R, Combes X, Lejay M, Carli P, Vivien B. Comparison of the accuracy of noninvasive hemoglobin monitoring by spectrophotometry (SpHb) and HemoCue(R) with automated laboratory hemoglobin measurement. Anesthesiology. 2011;115:548–54.PubMedCrossRef Lamhaut L, Apriotesei R, Combes X, Lejay M, Carli P, Vivien B. Comparison of the accuracy of noninvasive hemoglobin monitoring by spectrophotometry (SpHb) and HemoCue(R) with automated laboratory hemoglobin measurement. Anesthesiology. 2011;115:548–54.PubMedCrossRef
47.
go back to reference Karkouti K, Wijeysundera DN, Yau TM, McCluskey SA, van Rensburg A, Beattie WS. The influence of baseline hemoglobin concentration on tolerance of anemia in cardiac surgery. Transfusion. 2008;48:666–72.PubMedCrossRef Karkouti K, Wijeysundera DN, Yau TM, McCluskey SA, van Rensburg A, Beattie WS. The influence of baseline hemoglobin concentration on tolerance of anemia in cardiac surgery. Transfusion. 2008;48:666–72.PubMedCrossRef
48.
go back to reference Rivas Chirino L, da Silva Viana J, Tavares CA, Palmeiro A, Oliveira FJ. Use of a blood gas analyzer to measure blood hemoglobin during liver transplantation: a study of 935 paired samples. Transplant Proc. 2006;38:810–1.PubMedCrossRef Rivas Chirino L, da Silva Viana J, Tavares CA, Palmeiro A, Oliveira FJ. Use of a blood gas analyzer to measure blood hemoglobin during liver transplantation: a study of 935 paired samples. Transplant Proc. 2006;38:810–1.PubMedCrossRef
49.
go back to reference Bourner G, Dhaliwal J, Sumner J. Performance evaluation of the latest fully automated hematology analyzers in a large, commercial laboratory setting: a 4-way, side-by-side study. Lab Hematol. 2005;11:285–97.PubMedCrossRef Bourner G, Dhaliwal J, Sumner J. Performance evaluation of the latest fully automated hematology analyzers in a large, commercial laboratory setting: a 4-way, side-by-side study. Lab Hematol. 2005;11:285–97.PubMedCrossRef
50.
go back to reference Coquin J, Bertarrex A, Dewitte A, Lefevre L, Joannes-Boyau O, Fleureau C, Winnock S, Leuillet S, Janvier G, Ouattara A. Accuracy of determining hemoglobin level using occlusion spectroscopy in patients with severe gastrointestinal bleeding. Anesthesiology. 2013;118:640–8.PubMedCrossRef Coquin J, Bertarrex A, Dewitte A, Lefevre L, Joannes-Boyau O, Fleureau C, Winnock S, Leuillet S, Janvier G, Ouattara A. Accuracy of determining hemoglobin level using occlusion spectroscopy in patients with severe gastrointestinal bleeding. Anesthesiology. 2013;118:640–8.PubMedCrossRef
51.
go back to reference Lacroix J, Tucci M. Noninvasive or invasive hemoglobin measurement? Crit Care Med. 2012;40:2715–6.PubMedCrossRef Lacroix J, Tucci M. Noninvasive or invasive hemoglobin measurement? Crit Care Med. 2012;40:2715–6.PubMedCrossRef
52.
go back to reference Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.PubMedCrossRef Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.PubMedCrossRef
53.
go back to reference Lippi G, Bassi A, Brocco G, Montagnana M, Salvagno GL, Guidi GC. Preanalytic error tracking in a laboratory medicine department: results of a 1-year experience. Clin Chem. 2006;52:1442–3.PubMedCrossRef Lippi G, Bassi A, Brocco G, Montagnana M, Salvagno GL, Guidi GC. Preanalytic error tracking in a laboratory medicine department: results of a 1-year experience. Clin Chem. 2006;52:1442–3.PubMedCrossRef
54.
go back to reference Dale JC, Novis DA. Outpatient phlebotomy success and reasons for specimen rejection. Arch Pathol Lab Med. 2002;126:416–9.PubMed Dale JC, Novis DA. Outpatient phlebotomy success and reasons for specimen rejection. Arch Pathol Lab Med. 2002;126:416–9.PubMed
55.
go back to reference Lippi G, Salvagno GL, Montagnana M, Franchini M, Guidi GC. Phlebotomy issues and quality improvement in results of laboratory testing. Clin Lab. 2006;52:217–30.PubMed Lippi G, Salvagno GL, Montagnana M, Franchini M, Guidi GC. Phlebotomy issues and quality improvement in results of laboratory testing. Clin Lab. 2006;52:217–30.PubMed
56.
go back to reference Lippi G, Guidi GC. Risk management in the preanalytical phase of laboratory testing. Clin Chem Lab Med. 2007;45:720–7.PubMed Lippi G, Guidi GC. Risk management in the preanalytical phase of laboratory testing. Clin Chem Lab Med. 2007;45:720–7.PubMed
57.
go back to reference Lippi G, Salvagno GL, Solero GP, Franchini M, Guidi GC. Stability of blood cell counts, hematologic parameters and reticulocytes indexes on the Advia A120 hematologic analyzer. J Lab Clin Med. 2005;146:333–40.PubMedCrossRef Lippi G, Salvagno GL, Solero GP, Franchini M, Guidi GC. Stability of blood cell counts, hematologic parameters and reticulocytes indexes on the Advia A120 hematologic analyzer. J Lab Clin Med. 2005;146:333–40.PubMedCrossRef
58.
go back to reference Plebani M. Appropriateness in programs for continuous quality improvement in clinical laboratories. Clin Chim Acta. 2003;333:131–9.PubMedCrossRef Plebani M. Appropriateness in programs for continuous quality improvement in clinical laboratories. Clin Chim Acta. 2003;333:131–9.PubMedCrossRef
59.
go back to reference Hayden SJ, Albert TJ, Watkins TR, Swenson ER. Anemia in critical illness: insights into etiology, consequences, and management. Am J Respir Crit Care Med. 2012;185(10):1049–57.PubMedCrossRef Hayden SJ, Albert TJ, Watkins TR, Swenson ER. Anemia in critical illness: insights into etiology, consequences, and management. Am J Respir Crit Care Med. 2012;185(10):1049–57.PubMedCrossRef
60.
go back to reference Shander A, Fink A, Javidroozi M, Erhard J, Farmer SL, Corwin H, et al. Appropriateness of allogeneic red blood cell transfusion: the international consensus conference on transfusion outcomes. Transfus Med Rev. 2011;25(232–246):e253. Shander A, Fink A, Javidroozi M, Erhard J, Farmer SL, Corwin H, et al. Appropriateness of allogeneic red blood cell transfusion: the international consensus conference on transfusion outcomes. Transfus Med Rev. 2011;25(232–246):e253.
61.
go back to reference Lippi G, Salvagno GL, Montagnana M, Brocco G, Guidi GC. Influence of short-term venous stasis on clinical chemistry testing. Clin Chem Lab Med. 2005;43:869–75.PubMed Lippi G, Salvagno GL, Montagnana M, Brocco G, Guidi GC. Influence of short-term venous stasis on clinical chemistry testing. Clin Chem Lab Med. 2005;43:869–75.PubMed
62.
go back to reference Nkrumah B, Nguah SB, Sarpong N, Dekker D, Idriss A, May J, et al. Hemoglobin estimation by the HemoCue® portable hemoglobin photometer in a resource poor setting. BMC Clin Pathol. 2011;11:5.PubMedCrossRef Nkrumah B, Nguah SB, Sarpong N, Dekker D, Idriss A, May J, et al. Hemoglobin estimation by the HemoCue® portable hemoglobin photometer in a resource poor setting. BMC Clin Pathol. 2011;11:5.PubMedCrossRef
63.
go back to reference Van de Louw A, Lasserre N, Drouhin F, Thierry S, Lecuyer L, Caen D, et al. Reliability of HemoCue in patients with gastrointestinal bleeding. Intensive Care Med. 2007;33:355–8.PubMedCrossRef Van de Louw A, Lasserre N, Drouhin F, Thierry S, Lecuyer L, Caen D, et al. Reliability of HemoCue in patients with gastrointestinal bleeding. Intensive Care Med. 2007;33:355–8.PubMedCrossRef
64.
go back to reference Mimoz O, Frasca D, Medard A, Soubiron L, Debaene B, Dahyot-Fizelier C. Reliability of the HemoCue® hemoglobinometer in critically ill patients: a prospective observational study. Minerva Anesthesiol. 2011;77:979–85. Mimoz O, Frasca D, Medard A, Soubiron L, Debaene B, Dahyot-Fizelier C. Reliability of the HemoCue® hemoglobinometer in critically ill patients: a prospective observational study. Minerva Anesthesiol. 2011;77:979–85.
65.
go back to reference Leino A, Kurvinen K. Interchangeability of blood gas, electrolyte and metabolite results measured with point-of-care, blood gas and core laboratory analyzers. Clin Chem Lab Med. 2011;49:1187–91.PubMedCrossRef Leino A, Kurvinen K. Interchangeability of blood gas, electrolyte and metabolite results measured with point-of-care, blood gas and core laboratory analyzers. Clin Chem Lab Med. 2011;49:1187–91.PubMedCrossRef
Metadata
Title
Factors affecting hemoglobin measurement
Author
Lauren Berkow
Publication date
01-10-2013
Publisher
Springer Netherlands
Published in
Journal of Clinical Monitoring and Computing / Issue 5/2013
Print ISSN: 1387-1307
Electronic ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-013-9456-3

Other articles of this Issue 5/2013

Journal of Clinical Monitoring and Computing 5/2013 Go to the issue