Published in: Pediatric Radiology 11/2010

01-11-2010 | Original Article

Eye-lens bismuth shielding in paediatric head CT: artefact evaluation and reduction

Authors: Maria Raissaki, Kostas Perisinakis, John Damilakis, Nicholas Gourtsoyiannis

Published in: Pediatric Radiology | Issue 11/2010

Login to get access

Abstract

Background

CT scans of the brain, sinuses and petrous bones performed as the initial imaging test for a variety of indications have the potential to expose the eye-lens, considered among the most radiosensitive human tissues, to a radiation dose. There are several studies in adults discussing the reduction of orbital dose resulting from the use of commercially available bismuth-impregnated latex shields during CT examinations of the head.

Objective

To evaluate bismuth shielding-induced artefacts and to provide suggestions for optimal eye-lens shielding in paediatric head CT.

Materials and methods

A bismuth shield was placed over the eyelids of 60 consecutive children undergoing head CT. Images were assessed for the presence and severity of artefacts with regard to eye-shield distance and shield wrinkling. An anthropomorphic paediatric phantom and thermoluminescence dosimeters (TLDs) were used to study the effect of eye lens-to-shield distance on shielding efficiency.

Results

Shields were tolerated by 56/60 children. Artefacts were absent in 45% of scans. Artefacts on orbits, not affecting and affecting orbit evaluation were noted in 39% and 14% of scans, respectively. Diagnostically insignificant artefacts on intracranial structures were noted in 1 case (2%) with shield misplacement. Mean eye-lens-to-shield distance was 8.8 mm in scans without artefacts, and 4.3 mm and 2.2 mm in scans with unimportant and diagnostically important artefacts, respectively. Artefacts occurred in 8 out of 9 cases with shield wrinkling. Dose reduction remained unchanged for different shield-to-eye distances.

Conclusion

Bismuth shielding-related artefacts occurring in paediatric head CT are frequent, superficial and diagnostically insignificant when brain pathology is assessed. Shields should be placed 1 cm above the eyes when orbital pathology is addressed. Shield wrinkling should be avoided.
Literature
1.
go back to reference Brenner D, Elliston C, Hall E et al (2001) Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR 176:289–296PubMed Brenner D, Elliston C, Hall E et al (2001) Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR 176:289–296PubMed
2.
go back to reference International Commission on Radiological Protection (1990) Recommendations of the international commission on radiological protection. In: ICRP Publication 60. Ann ICRP 21:1–3, Pergamon, Oxford International Commission on Radiological Protection (1990) Recommendations of the international commission on radiological protection. In: ICRP Publication 60. Ann ICRP 21:1–3, Pergamon, Oxford
3.
go back to reference National Research Council (1990) Health effects exposure to low levels of ionizing radiation, BEIR V. National Academy, Washington National Research Council (1990) Health effects exposure to low levels of ionizing radiation, BEIR V. National Academy, Washington
4.
go back to reference Merriam GR Jr, Focht EF (1957) A clinical study of radiation cataracts and the relationship to dose. Am J Roentgenol Radium Ther Nucl Med 77:759–785PubMed Merriam GR Jr, Focht EF (1957) A clinical study of radiation cataracts and the relationship to dose. Am J Roentgenol Radium Ther Nucl Med 77:759–785PubMed
5.
go back to reference Minamoto A, Taniguchi H, Yoshitani N et al (2004) Cataracts in atomic bomb survivors. Int J Radiat Biol 80:339–345CrossRefPubMed Minamoto A, Taniguchi H, Yoshitani N et al (2004) Cataracts in atomic bomb survivors. Int J Radiat Biol 80:339–345CrossRefPubMed
6.
go back to reference Hall P, Granath F, Lundell M et al (1999) Lenticular opacities in individuals exposed to ionizing radiation in infancy. Radiat Res 152:190–195CrossRefPubMed Hall P, Granath F, Lundell M et al (1999) Lenticular opacities in individuals exposed to ionizing radiation in infancy. Radiat Res 152:190–195CrossRefPubMed
7.
go back to reference McLaughlin DJ, Mooney RB (2004) Dose reduction to radiosensitive tissues in CT. Do commercially available shields meet the users’ needs? Clin Radiol 59:446–450CrossRefPubMed McLaughlin DJ, Mooney RB (2004) Dose reduction to radiosensitive tissues in CT. Do commercially available shields meet the users’ needs? Clin Radiol 59:446–450CrossRefPubMed
8.
go back to reference Hopper KD, Neuman JD, King SH et al (2001) Radioprotection to the eye during CT scanning. AJNR 22:1194–1198PubMed Hopper KD, Neuman JD, King SH et al (2001) Radioprotection to the eye during CT scanning. AJNR 22:1194–1198PubMed
9.
go back to reference Colombo P, Pedroli G, Nicoloso M et al (2004) Evaluation of the efficacy of a bismuth shield during CT examinations. Radiol Med 108:560–568PubMed Colombo P, Pedroli G, Nicoloso M et al (2004) Evaluation of the efficacy of a bismuth shield during CT examinations. Radiol Med 108:560–568PubMed
10.
go back to reference Keil B, Wulff J, Schmitt R et al (2008) Protection of eye lens in computed tomography—dose evaluation on an anthropomorphic phantom using thermo-luminescent dosimeters and Monte-Carlo simulations. Rofo 180:1047–1053PubMed Keil B, Wulff J, Schmitt R et al (2008) Protection of eye lens in computed tomography—dose evaluation on an anthropomorphic phantom using thermo-luminescent dosimeters and Monte-Carlo simulations. Rofo 180:1047–1053PubMed
11.
go back to reference Kim S, Yoshizumi TT, Frush DP et al (2009) Dosimetric characterization of bismuth shields in CT: measurements and Monte Carlo simulations. Radiat Prot Dosim 133:105–110CrossRef Kim S, Yoshizumi TT, Frush DP et al (2009) Dosimetric characterization of bismuth shields in CT: measurements and Monte Carlo simulations. Radiat Prot Dosim 133:105–110CrossRef
12.
go back to reference Hein E, Rogalla P, Klingebiel R et al (2002) Low-dose CT of the paranasal sinuses with eye lens protection: effect on image quality and radiation dose. Eur Radiol 12:1693–1696CrossRefPubMed Hein E, Rogalla P, Klingebiel R et al (2002) Low-dose CT of the paranasal sinuses with eye lens protection: effect on image quality and radiation dose. Eur Radiol 12:1693–1696CrossRefPubMed
13.
go back to reference Perisinakis K, Raissaki M, Theocharopoulos N et al (2005) Reduction of eye lens radiation dose by orbital bismuth shielding in pediatric patients undergoing CT of the head: a Monte Carlo study. Med Phys 32:1024–1030CrossRefPubMed Perisinakis K, Raissaki M, Theocharopoulos N et al (2005) Reduction of eye lens radiation dose by orbital bismuth shielding in pediatric patients undergoing CT of the head: a Monte Carlo study. Med Phys 32:1024–1030CrossRefPubMed
14.
go back to reference Mukundan S Jr, Wang PI, Frush DP et al (2007) MOSFET dosimetry for radiation dose assessment of bismuth shielding of the eye in children. AJR 188:1648–1650CrossRefPubMed Mukundan S Jr, Wang PI, Frush DP et al (2007) MOSFET dosimetry for radiation dose assessment of bismuth shielding of the eye in children. AJR 188:1648–1650CrossRefPubMed
15.
go back to reference Hopper KD (2002) Orbital, thyroid and breast superficial radiation shielding for patients undergoing diagnostic CT. Semin Ultrasound CT MR 23:423–427CrossRefPubMed Hopper KD (2002) Orbital, thyroid and breast superficial radiation shielding for patients undergoing diagnostic CT. Semin Ultrasound CT MR 23:423–427CrossRefPubMed
16.
go back to reference Geleijns J, Salvadó Artells M, Veldkamp WJ et al (2006) Quantitative assessment of selective in-plane shielding of tissues in computed tomography through evaluation of absorbed dose and image quality. Eur Radiol 16:2334–2340CrossRefPubMed Geleijns J, Salvadó Artells M, Veldkamp WJ et al (2006) Quantitative assessment of selective in-plane shielding of tissues in computed tomography through evaluation of absorbed dose and image quality. Eur Radiol 16:2334–2340CrossRefPubMed
17.
go back to reference Shah R, Gupta AK, Rehani MM et al (2005) Effect of reduction in tube current on reader confidence in paediatric computed tomography. Clin Radiol 60:224–231CrossRefPubMed Shah R, Gupta AK, Rehani MM et al (2005) Effect of reduction in tube current on reader confidence in paediatric computed tomography. Clin Radiol 60:224–231CrossRefPubMed
18.
go back to reference Fricke BL, Donnelly LF, Frush DP et al (2003) In-plane bismuth breast shields for pediatric CT: effects on radiation dose and image quality using experimental and clinical data. AJR 180:407–411PubMed Fricke BL, Donnelly LF, Frush DP et al (2003) In-plane bismuth breast shields for pediatric CT: effects on radiation dose and image quality using experimental and clinical data. AJR 180:407–411PubMed
19.
go back to reference Hohl C, Wildberger JE, Süss C et al (2006) Radiation dose reduction to breast and thyroid during MDCT: effectiveness of an in-plane bismuth shield. Acta Radiol 47:562–567CrossRefPubMed Hohl C, Wildberger JE, Süss C et al (2006) Radiation dose reduction to breast and thyroid during MDCT: effectiveness of an in-plane bismuth shield. Acta Radiol 47:562–567CrossRefPubMed
20.
go back to reference Yeoman LJ, Howarth L, Britten A et al (1992) Gantry angulation in brain CT: dosage implications, effect on posterior fossa artifacts, and current international practice. Radiology 184:113–116PubMed Yeoman LJ, Howarth L, Britten A et al (1992) Gantry angulation in brain CT: dosage implications, effect on posterior fossa artifacts, and current international practice. Radiology 184:113–116PubMed
21.
go back to reference Heaney DE, Norvill CA (2006) A comparison of reduction in CT dose through the use of gantry angulations or bismuth shields. Australas Phys Eng Sci Med 29:172–178CrossRefPubMed Heaney DE, Norvill CA (2006) A comparison of reduction in CT dose through the use of gantry angulations or bismuth shields. Australas Phys Eng Sci Med 29:172–178CrossRefPubMed
22.
go back to reference Tzedakis A, Damilakis J, Perisinakis K et al (2005) The effect of z overscanning on patient effective dose from multidetector helical computed tomography examinations. Med Phys 32:1621–1629CrossRefPubMed Tzedakis A, Damilakis J, Perisinakis K et al (2005) The effect of z overscanning on patient effective dose from multidetector helical computed tomography examinations. Med Phys 32:1621–1629CrossRefPubMed
23.
go back to reference Leswick DA, Hunt MM, Webster ST et al (2008) Thyroid shields versus z-axis automatic tube current modulation for dose reduction at neck CT. Radiology 249:572–580CrossRefPubMed Leswick DA, Hunt MM, Webster ST et al (2008) Thyroid shields versus z-axis automatic tube current modulation for dose reduction at neck CT. Radiology 249:572–580CrossRefPubMed
24.
go back to reference Smith AB, Dillon WP, Lau BC et al (2008) Radiation dose reduction strategy for CT protocols: successful implementation in neuroradiology section. Radiology 247:499–506CrossRefPubMed Smith AB, Dillon WP, Lau BC et al (2008) Radiation dose reduction strategy for CT protocols: successful implementation in neuroradiology section. Radiology 247:499–506CrossRefPubMed
25.
go back to reference Papadakis AE, Perisinakis K, Damilakis J (2007) Angular on-line tube current modulation in multidetector CT examinations of children and adults: the influence of different scanning parameters on dose reduction. Med Phys 34:2864–2874CrossRefPubMed Papadakis AE, Perisinakis K, Damilakis J (2007) Angular on-line tube current modulation in multidetector CT examinations of children and adults: the influence of different scanning parameters on dose reduction. Med Phys 34:2864–2874CrossRefPubMed
26.
go back to reference Papadakis AE, Perisinakis K, Damilakis J (2008) Automatic exposure control in pediatric and adult multidetector CT examinations: a phantom study on dose reduction and image quality. Med Phys 35:4567–4576CrossRefPubMed Papadakis AE, Perisinakis K, Damilakis J (2008) Automatic exposure control in pediatric and adult multidetector CT examinations: a phantom study on dose reduction and image quality. Med Phys 35:4567–4576CrossRefPubMed
27.
go back to reference Coursey C, Frush DP, Yoshizumi T et al (2008) Pediatric chest MDCT using tube current modulation: effect on radiation dose with breast shielding. AJR 190:W54–W61CrossRefPubMed Coursey C, Frush DP, Yoshizumi T et al (2008) Pediatric chest MDCT using tube current modulation: effect on radiation dose with breast shielding. AJR 190:W54–W61CrossRefPubMed
Metadata
Title
Eye-lens bismuth shielding in paediatric head CT: artefact evaluation and reduction
Authors
Maria Raissaki
Kostas Perisinakis
John Damilakis
Nicholas Gourtsoyiannis
Publication date
01-11-2010
Publisher
Springer-Verlag
Published in
Pediatric Radiology / Issue 11/2010
Print ISSN: 0301-0449
Electronic ISSN: 1432-1998
DOI
https://doi.org/10.1007/s00247-010-1715-6

Other articles of this Issue 11/2010

Pediatric Radiology 11/2010 Go to the issue