Published in: Journal of Interventional Cardiac Electrophysiology 3/2013

01-04-2013

Extraction of defibrillator leads recalled for cable externalization and failure

Authors: Divyang Patel, Evan Adelstein, Jan Nemec, G. Stuart Mendenhall, Raveen Bazaz, Sandeep Jain, Samir Saba

Published in: Journal of Interventional Cardiac Electrophysiology | Issue 3/2013

Login to get access

Abstract

Background

The Riata implantable cardioverter–defibrillator lead was recalled by the Food and Drug Administration because of an increased rate of failure associated with cable externalization. Because of its mechanical separation, the Riata lead may be challenging to extract. We therefore examined the experience with Riata lead extractions at our institution.

Methods

All patients implanted with the Riata lead underwent detailed review of their electronic medical records and operative notes. Procedural complications were ascertained by reviewing the medical records up to 30 days after Riata extraction.

Results

From a total of 627 patients implanted with the Riata lead at our institution, 20 patients (age at time of extraction, 57 ± 11 years; 85 % men; lead dwell time, 48 ± 27 months) underwent lead extraction. Extraction was successful in 19 of 20 (95 %) patients and required the use of laser-powered sheaths in 18 (90 %) patients. Over a 30-day follow-up period, 1 of 20 (5 %) patients had a minor procedure-related complication consisting of a new pericardial effusion that did not require drainage.

Conclusions

Extraction of the Riata lead seems to be successful and safe and frequently requires the use of powered sheaths.
Literature
1.
go back to reference Cannom, D. S., & Prystowky, E. N. (2004). Evolution of the implantable cardioverter defibrillator. Journal of Cardiovascular Electrophysiology, 15, 375–385.PubMedCrossRef Cannom, D. S., & Prystowky, E. N. (2004). Evolution of the implantable cardioverter defibrillator. Journal of Cardiovascular Electrophysiology, 15, 375–385.PubMedCrossRef
3.
go back to reference Israel, C. W., Duray, G. Z., Schmitt, J., & Hohnloser, S. H. (2008). Implantable cardioverter–defibrillator lead disintegration at the level of the tricuspid valve. Heart Rhythm, 5, 1224–1225.PubMedCrossRef Israel, C. W., Duray, G. Z., Schmitt, J., & Hohnloser, S. H. (2008). Implantable cardioverter–defibrillator lead disintegration at the level of the tricuspid valve. Heart Rhythm, 5, 1224–1225.PubMedCrossRef
4.
go back to reference Jalal, Z., Derval, N., Ploux, S., & Bordachar, P. (2010). Unusual failure of a multilumen, small-diameter implantable cardioverter–defibrillator lead. Heart Rhythm, 7, 1166–1167.PubMedCrossRef Jalal, Z., Derval, N., Ploux, S., & Bordachar, P. (2010). Unusual failure of a multilumen, small-diameter implantable cardioverter–defibrillator lead. Heart Rhythm, 7, 1166–1167.PubMedCrossRef
5.
go back to reference Hauser, R. G., Abdelhadi, R., McGriff, D., & Retel, L. K. (2012). Deaths caused by the failure of Riata and Riata ST implantable cardioverter–defibrillator leads. Heart Rhythm, 9, 1227–1235.PubMedCrossRef Hauser, R. G., Abdelhadi, R., McGriff, D., & Retel, L. K. (2012). Deaths caused by the failure of Riata and Riata ST implantable cardioverter–defibrillator leads. Heart Rhythm, 9, 1227–1235.PubMedCrossRef
6.
go back to reference Parvathaneni, S. V., Ellis, C. R., & Rottmann, J. N. (2012). High prevalence of insulation failure with externalized cables in St. Jude Medical Riata family ICD leads: Fluoroscopic grading scale and correlation to extracted leads. Heart Rhythm, 9, 1218–1224. Parvathaneni, S. V., Ellis, C. R., & Rottmann, J. N. (2012). High prevalence of insulation failure with externalized cables in St. Jude Medical Riata family ICD leads: Fluoroscopic grading scale and correlation to extracted leads. Heart Rhythm, 9, 1218–1224.
8.
go back to reference Wilkoff, B. L., Love, C. J., Byrd, C. L., Bongiorni, M. G., Carrillo, R. G., Crossley, G. H., 3rd, et al. (2009). Transvenous lead extraction: Heart Rhythm Society’s expert consensus on facilities, training, indications, and patient management. Heart Rhythm, 6, 1085–1104.PubMedCrossRef Wilkoff, B. L., Love, C. J., Byrd, C. L., Bongiorni, M. G., Carrillo, R. G., Crossley, G. H., 3rd, et al. (2009). Transvenous lead extraction: Heart Rhythm Society’s expert consensus on facilities, training, indications, and patient management. Heart Rhythm, 6, 1085–1104.PubMedCrossRef
9.
go back to reference Kennergren, C., Bjurman, C., Wiklund, R., & Gabel, J. (2009). A single-centre experience of over one thousand lead extractions. Europace, 11, 612–617.PubMedCrossRef Kennergren, C., Bjurman, C., Wiklund, R., & Gabel, J. (2009). A single-centre experience of over one thousand lead extractions. Europace, 11, 612–617.PubMedCrossRef
10.
go back to reference Jones, S. O., Eckart, R. E., Albert, C. M., & Epstein, L. M. (2008). Large, single-center, single-operator experience with transvenous lead extraction: Outcomes and changing indications. Heart Rhythm, 5, 520–525.PubMedCrossRef Jones, S. O., Eckart, R. E., Albert, C. M., & Epstein, L. M. (2008). Large, single-center, single-operator experience with transvenous lead extraction: Outcomes and changing indications. Heart Rhythm, 5, 520–525.PubMedCrossRef
11.
go back to reference Maytin, M., Love, C. J., Fischer, A., Carrillo, R. G., Garisto, J. D., Bongiorni, M. G., et al. (2010). Multicenter experience with extraction of the Sprint Fidelis implantable cardioverter–defibrillator lead. Journal of the American College of Cardiology, 56, 646–650.PubMedCrossRef Maytin, M., Love, C. J., Fischer, A., Carrillo, R. G., Garisto, J. D., Bongiorni, M. G., et al. (2010). Multicenter experience with extraction of the Sprint Fidelis implantable cardioverter–defibrillator lead. Journal of the American College of Cardiology, 56, 646–650.PubMedCrossRef
12.
go back to reference Hauser, R. G., Kallinen, L. M., Almquist, A. K., Gornick, C. C., & Katsiyiannis, W. T. (2007). Early failure of a small-diameter high-voltage implantable cardioverter–defibrillator lead. Heart Rhythm, 4, 892–896.PubMedCrossRef Hauser, R. G., Kallinen, L. M., Almquist, A. K., Gornick, C. C., & Katsiyiannis, W. T. (2007). Early failure of a small-diameter high-voltage implantable cardioverter–defibrillator lead. Heart Rhythm, 4, 892–896.PubMedCrossRef
13.
go back to reference Krahn, A. D., Champagne, J., Healey, J. S., Cameron, D., Simpson, C. S., Thibault, B., et al. (2008). Outcome of the Fidelis implantable cardioverter-defibrillator lead advisory: A report from the Canadian Heart Rhythm Society Device Advisory Committee. Heart Rhythm, 5, 639–642.PubMedCrossRef Krahn, A. D., Champagne, J., Healey, J. S., Cameron, D., Simpson, C. S., Thibault, B., et al. (2008). Outcome of the Fidelis implantable cardioverter-defibrillator lead advisory: A report from the Canadian Heart Rhythm Society Device Advisory Committee. Heart Rhythm, 5, 639–642.PubMedCrossRef
14.
go back to reference Gaca, J. G., Lima, B., Milano, A., Lin, S. S., Davis, R. D., Lowe, J. E., et al. (2009). Laser-assisted extraction of pacemaker and defibrillator leads: The role of the cardiac surgeon. The Annals of Thoracic Surgery, 87, 1446–1450.PubMedCrossRef Gaca, J. G., Lima, B., Milano, A., Lin, S. S., Davis, R. D., Lowe, J. E., et al. (2009). Laser-assisted extraction of pacemaker and defibrillator leads: The role of the cardiac surgeon. The Annals of Thoracic Surgery, 87, 1446–1450.PubMedCrossRef
15.
go back to reference Epstein, L. M., Byrd, C. L., Wilkoff, B. L., Love, C. J., Sellers, T. D., Hayes, D. L., et al. (1999). Initial experience with larger laser sheaths for the removal of transvenous pacemaker and implantable defibrillator leads. Circulation, 100, 516–525.PubMedCrossRef Epstein, L. M., Byrd, C. L., Wilkoff, B. L., Love, C. J., Sellers, T. D., Hayes, D. L., et al. (1999). Initial experience with larger laser sheaths for the removal of transvenous pacemaker and implantable defibrillator leads. Circulation, 100, 516–525.PubMedCrossRef
16.
go back to reference Wilkoff, B. L., Byrd, C. L., Love, C. J., Hayes, D. L., Sellers, T. D., Schaerf, R., et al. (1999). Pacemaker lead extraction with the laser sheath: Results of the pacing lead extraction with the excimer sheath (Plexes) trial. Journal of the American College of Cardiology, 33, 1671–1676.PubMedCrossRef Wilkoff, B. L., Byrd, C. L., Love, C. J., Hayes, D. L., Sellers, T. D., Schaerf, R., et al. (1999). Pacemaker lead extraction with the laser sheath: Results of the pacing lead extraction with the excimer sheath (Plexes) trial. Journal of the American College of Cardiology, 33, 1671–1676.PubMedCrossRef
17.
go back to reference Epstein, A. E., Kay, G. N., Plumb, V. J., Dailey, S. M., & Anderson, P. G. (1998). Gross and microscopic pathological changes associated with nonthoracotomy implantable defibrillator leads. Circulation, 98, 1517–1524.PubMedCrossRef Epstein, A. E., Kay, G. N., Plumb, V. J., Dailey, S. M., & Anderson, P. G. (1998). Gross and microscopic pathological changes associated with nonthoracotomy implantable defibrillator leads. Circulation, 98, 1517–1524.PubMedCrossRef
18.
go back to reference Kolodzinska, A., Kutarski, A., Koperski, L., Jarzyna, I., Małecka, B., & Opolski, G. (2012). Differences in encapsulating lead tissue in patients who underwent transvenous lead removal. Europace, 14, 994–1001.PubMedCrossRef Kolodzinska, A., Kutarski, A., Koperski, L., Jarzyna, I., Małecka, B., & Opolski, G. (2012). Differences in encapsulating lead tissue in patients who underwent transvenous lead removal. Europace, 14, 994–1001.PubMedCrossRef
19.
go back to reference Martin, D. T., Hook, B. G., Chaudhry, G. M., & Silver, J. (2011). Multicenter experience with extraction of the Sprint Fidelis implantable cardioverter–defibrillator lead. Journal of the American College of Cardiology, 57, 635–636.PubMedCrossRef Martin, D. T., Hook, B. G., Chaudhry, G. M., & Silver, J. (2011). Multicenter experience with extraction of the Sprint Fidelis implantable cardioverter–defibrillator lead. Journal of the American College of Cardiology, 57, 635–636.PubMedCrossRef
Metadata
Title
Extraction of defibrillator leads recalled for cable externalization and failure
Authors
Divyang Patel
Evan Adelstein
Jan Nemec
G. Stuart Mendenhall
Raveen Bazaz
Sandeep Jain
Samir Saba
Publication date
01-04-2013
Publisher
Springer US
Published in
Journal of Interventional Cardiac Electrophysiology / Issue 3/2013
Print ISSN: 1383-875X
Electronic ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-012-9751-9

Other articles of this Issue 3/2013

Journal of Interventional Cardiac Electrophysiology 3/2013 Go to the issue