Exposure of medical staff to radiation during osteosynthesis of proximal femoral fracture: descriptive analysis and comparison of different devices
Authors:
Cédric Siedlecki, Rémi Gauthé, André Gillibert, Kevin Bellenger, Xavier Roussignol, Mourad Ould-Slimane
Published in:
Archives of Orthopaedic and Trauma Surgery
|
Issue 10/2017
Login to get access
Abstract
Background
The use of fluoroscopy is necessary during proximal femoral fracture (PFF) osteosynthesis. The frequency of these procedures justifies a description of radiation exposure and comparisons between different techniques and between the different surgical team members.
Methods
This observational prospective and comparative study includes a series of 68 patients with PFF receiving osteosynthesis. Radiation exposure was assessed for all members of the operating team. The radiation dose measurements for the different members of the surgical team during PFF osteosynthesis were compared. The factors affecting the radiation dose were investigated.
Results
The mean active dosimeter readings for each operation were 7.39 µSv for the primary surgeon, 3.93 µSv for the assistant surgeon, 1.92 µSv for the instrument nurse, 1.25 µSv for the circulating nurse, and 0.64 µSv for the anaesthesiologist, respectively. Doses decreased significantly between these different members of the medical team (all p < 0.001). The dose also varied with patient age and BMI, as well as with fluoroscopy time and operating time, but not with type of fracture or type of osteosynthesis.
Conclusion
Medical staff receives significantly different doses depending on their position in relation to the radiation source. Operating time and fluoroscopy time are the modifiable factors that affect the radiation dose. The radiation doses received by the different members of the medical teams involved in proximal femur osteosynthesis procedures all fall below the doses recommended by the International Commission on Radiation Units and Measurements.