Published in: Journal of Artificial Organs 2/2012

01-06-2012 | Original Article

Experimental evaluation of flow and dialysis performance of hollow-fiber dialyzers with different packing densities

Authors: Ayaka Hirano, Shoko Kida, Ken-ichiro Yamamoto, Kiyotaka Sakai

Published in: Journal of Artificial Organs | Issue 2/2012

Login to get access

Abstract

The dialyzer housing structure should be designed in such a way that high dialysis performance is achieved. To achieve high dialysis performance, the flow of the dialysis fluid and blood should be uniform, without channeling and dead spaces. The objective of this study was to evaluate the effect of fiber packing density on the flow of dialysis fluid and blood, and on the dialysis performance of a hollow-fiber dialyzer at defined flow rates for blood (Q B = 200 mL/min), dialysis fluid (Q D = 500 mL/min), and filtrate (Q F = 0 mL/min). We measured Q D, Q B, and solute clearance for 3 test dialyzers with dialyzer housing different diameters. To evaluate the flow of dialysis fluid and blood, we measured the residence time of the dialysis fluid and blood in the test dialyzers by use of the pulse–response method. We also measured the clearances of urea, creatinine, vitamin B12, and lysozyme to evaluate the dialysis performance of the test dialyzers. At packing densities ranging from 48 to 67%, higher packing densities and lower housing diameters of the dialyzer resulted in higher dialysis performance because the dialysis fluid and blood entered the hollow-fiber bundle smoothly and, hence, increased contact area between the dialysis fluid and the blood led to better dialysis performance.
Literature
1.
go back to reference Hirano A, Yamamoto K, Matsuda M, Ogawa T, Yakushiji T, Miyasaka T, Sakai K. Evaluation of dialyzer housing structure and hollow-fiber dialysis membranes to achieve high dialysis performance. Ther Aphaer Dial. 2011;15:66–74.CrossRef Hirano A, Yamamoto K, Matsuda M, Ogawa T, Yakushiji T, Miyasaka T, Sakai K. Evaluation of dialyzer housing structure and hollow-fiber dialysis membranes to achieve high dialysis performance. Ther Aphaer Dial. 2011;15:66–74.CrossRef
2.
go back to reference Kim JC, Kim JH, Kim H-C, Kang E, Ronco C, Kim HC. Analysis of blood and dialysis-fluid flow in a hemodialyzer by perfusion computed tomography. WC2009, IFMBE Proceedings 2009; 25/VII; 831–834. Kim JC, Kim JH, Kim H-C, Kang E, Ronco C, Kim HC. Analysis of blood and dialysis-fluid flow in a hemodialyzer by perfusion computed tomography. WC2009, IFMBE Proceedings 2009; 25/VII; 831–834.
3.
go back to reference Ronco C, Brendolan A, Crepaldi C, Rodighiero M, Scabardi M. Blood and dialysis-fluid flow distributions in hollow-fiber hemodialyzers analyzed by computerized helical scanning technique. J Am Soc Nephrol. 2002;13:S53–61.PubMed Ronco C, Brendolan A, Crepaldi C, Rodighiero M, Scabardi M. Blood and dialysis-fluid flow distributions in hollow-fiber hemodialyzers analyzed by computerized helical scanning technique. J Am Soc Nephrol. 2002;13:S53–61.PubMed
5.
go back to reference Fukuda M, Miyazaki M, Uezumi S, Yoshida M. Design and assessment of the new APS dialyzer. J Artif Organs. 2006;9:192–8.PubMedCrossRef Fukuda M, Miyazaki M, Uezumi S, Yoshida M. Design and assessment of the new APS dialyzer. J Artif Organs. 2006;9:192–8.PubMedCrossRef
6.
go back to reference Fijimura T, Uchi Y, Fukuda M, Miyazaki M, Uezumi S, Hiyoshi T. Development of a dialyzer with enhanced internal filtration to increase the clearance of low molecular weight proteins. J Artif Organs. 2004;7:149–54.CrossRef Fijimura T, Uchi Y, Fukuda M, Miyazaki M, Uezumi S, Hiyoshi T. Development of a dialyzer with enhanced internal filtration to increase the clearance of low molecular weight proteins. J Artif Organs. 2004;7:149–54.CrossRef
7.
go back to reference Sato Y, Mineshima M, Ishimori I, Kaneko I, Akiba T, Terasawa S. Effect of hollow fiber length on solute removal and quantification of internal filtration rate by Doppler ultrasound. Int J Artif Organs. 2003;26:129–34.PubMed Sato Y, Mineshima M, Ishimori I, Kaneko I, Akiba T, Terasawa S. Effect of hollow fiber length on solute removal and quantification of internal filtration rate by Doppler ultrasound. Int J Artif Organs. 2003;26:129–34.PubMed
8.
go back to reference Leypoldt JK, Cheung AK, Chirananthavat T, Gilson JF, Kamerath CD, Deeter RB. Hollow fiber shape alters solute clearances in high flux hemodialyzers. ASAIO J. 2003;49:81–7.PubMedCrossRef Leypoldt JK, Cheung AK, Chirananthavat T, Gilson JF, Kamerath CD, Deeter RB. Hollow fiber shape alters solute clearances in high flux hemodialyzers. ASAIO J. 2003;49:81–7.PubMedCrossRef
9.
go back to reference Yang M-C, Lin C-C. Influence of design of the hemodialyzer inlet chamber on red blood damage during hemodialysis. ASAIO J. 2001;47:92–6.PubMedCrossRef Yang M-C, Lin C-C. Influence of design of the hemodialyzer inlet chamber on red blood damage during hemodialysis. ASAIO J. 2001;47:92–6.PubMedCrossRef
10.
go back to reference Lu J, Lu W. Blood flow velocity and ultra-filtration velocity measured by CT imaging system inside a densely bundled hollow fiber dialyzer. Intern J Heat Mass Transf. 2010;53:1844–50.CrossRef Lu J, Lu W. Blood flow velocity and ultra-filtration velocity measured by CT imaging system inside a densely bundled hollow fiber dialyzer. Intern J Heat Mass Transf. 2010;53:1844–50.CrossRef
11.
go back to reference Yamashita AC, Fujita R, Tomisawa N, Jinbo Y, Yamamura M. Effect of packing density of hollow fibers on solute removal performances of dialyzers. Hemodial Int. 2009;13:S2–7.PubMedCrossRef Yamashita AC, Fujita R, Tomisawa N, Jinbo Y, Yamamura M. Effect of packing density of hollow fibers on solute removal performances of dialyzers. Hemodial Int. 2009;13:S2–7.PubMedCrossRef
12.
go back to reference Yamamoto K, Matsuda M, Hirano A, Takizawa N, Iwashima S, Yakushiji T, Fukuda M, Miyasaka T, Sakai K. Computational evaluation of dialysis fluid flow in dialyzers with variously designed housings. J Artif Organs. 2009;33:481–6.CrossRef Yamamoto K, Matsuda M, Hirano A, Takizawa N, Iwashima S, Yakushiji T, Fukuda M, Miyasaka T, Sakai K. Computational evaluation of dialysis fluid flow in dialyzers with variously designed housings. J Artif Organs. 2009;33:481–6.CrossRef
13.
go back to reference Poh CK, Hardy PA, Liao Z, Huang Z, Gao Dl. Effect of flow baffles on the dialysate flow distribution of hollow-fiber hemodialyzers a nonintrusive experimental study using MRI. J Biomech Eng. 2003;125:481–9.PubMedCrossRef Poh CK, Hardy PA, Liao Z, Huang Z, Gao Dl. Effect of flow baffles on the dialysate flow distribution of hollow-fiber hemodialyzers a nonintrusive experimental study using MRI. J Biomech Eng. 2003;125:481–9.PubMedCrossRef
14.
go back to reference Osuga T, Obata T, Ikehara H. Detection of small degree of nonuniformity in dialysate flow in hollow-fiber dialyzer using proton magnetic resonance imaging. Magn Reson Imaging. 2004;22:417–20.PubMedCrossRef Osuga T, Obata T, Ikehara H. Detection of small degree of nonuniformity in dialysate flow in hollow-fiber dialyzer using proton magnetic resonance imaging. Magn Reson Imaging. 2004;22:417–20.PubMedCrossRef
15.
go back to reference Takesawa S, Terasawa M, Sakagami M, Kobayashi T, Hidai H, Sakai K. Nondestructive evaluation by X-ray computed tomography of dialysate flow patterns in capillary dialyzers. ASAIO J. 1988;34:794–9. Takesawa S, Terasawa M, Sakagami M, Kobayashi T, Hidai H, Sakai K. Nondestructive evaluation by X-ray computed tomography of dialysate flow patterns in capillary dialyzers. ASAIO J. 1988;34:794–9.
16.
go back to reference Sakai Y, Wada S, Matsumoto H, Suyama T, Ohno O, Anno I. Nondestructive evaluation of blood flow in a dialyzer using X-ray computed tomography. Artif Organs. 2003;3:197–204. Sakai Y, Wada S, Matsumoto H, Suyama T, Ohno O, Anno I. Nondestructive evaluation of blood flow in a dialyzer using X-ray computed tomography. Artif Organs. 2003;3:197–204.
17.
go back to reference Yamamoto K, Matsukawa H, Yakushiji T, Fukuda M, Hiroshi T, Sakai K. Technical evaluation of dialysis-fluid flow in a newly designed dialyzer. ASAIO J. 2007;53:36–40.PubMedCrossRef Yamamoto K, Matsukawa H, Yakushiji T, Fukuda M, Hiroshi T, Sakai K. Technical evaluation of dialysis-fluid flow in a newly designed dialyzer. ASAIO J. 2007;53:36–40.PubMedCrossRef
18.
go back to reference Levenspiel O. Non-ideal flow. In: Chemical reaction engineering. 2nd ed. New York: Wiley; 1972. p. 253–325. Levenspiel O. Non-ideal flow. In: Chemical reaction engineering. 2nd ed. New York: Wiley; 1972. p. 253–325.
Metadata
Title
Experimental evaluation of flow and dialysis performance of hollow-fiber dialyzers with different packing densities
Authors
Ayaka Hirano
Shoko Kida
Ken-ichiro Yamamoto
Kiyotaka Sakai
Publication date
01-06-2012
Publisher
Springer Japan
Published in
Journal of Artificial Organs / Issue 2/2012
Print ISSN: 1434-7229
Electronic ISSN: 1619-0904
DOI
https://doi.org/10.1007/s10047-011-0620-6

Other articles of this Issue 2/2012

Journal of Artificial Organs 2/2012 Go to the issue