Published in: International Urogynecology Journal 8/2022

14-06-2021 | Original Article

Evaluation of the histological and biomechanical properties of poly-4-hydroxybutyrate scaffold for pelvic organ prolapse, compared with polypropylene mesh in a rabbit model

Authors: Danielle O’Shaughnessy, Daniel Grande, Diana El-Neemany, Sujata Sajjan, Nirmala Pillalamarri, Dara Shalom, Harvey Winkler

Published in: International Urogynecology Journal | Issue 8/2022

Login to get access

Abstract

Introduction and hypothesis

Poly-4-hydroxybutyrate (P4HB) is a biopolymer produced by Escherichia coli K12 bacteria. P4HB is fully resorbed in vivo by 18–24 months post-implantation. The aim of this study is to evaluate P4HB in the rabbit abdomen and vagina to determine that the biomechanical and histological properties are similar to the standard polypropylene mesh. Our hypothesis is that the histological and biomechanical properties of a fully absorbable graft will be similar to a lightweight polypropylene (PP) mesh when implanted in rabbit vagina and abdomen.

Methods

Sixteen (n = 16) female New Zealand White (retired breeder) rabbits were equally divided between two time points (3 and 9 months). A total of 17 rabbits were used owing to one death secondary to suspected cardiomyopathy. P4HB scaffold and PP mesh were subcutaneously and peri-vaginally implanted into the rabbit abdomen and vagina respectively. All rabbits had both posterior and anterior vaginal implants, and half of the rabbits had four abdominal implants in addition to the vaginal implants. The abdominal implants were 4.5 cm long × 1.5 cm wide whereas the vaginal implants were 1.5 cm long × 0.5 cm wide. At 3 and 9 months, gross necropsy was performed and samples were obtained, sectioned, stained and evaluated via histological analysis. Specimens were assessed for host inflammatory response, neovascularization, elastin content, and collagen deposition/maturation. Specimens were also biomechanically evaluated via uniaxial tensile test to determine the stiffness, ultimate tensile strength and load at ultimate tensile strength of the device/tissue composite.

Results

No abdominal mesh exposures were noted. A comparable number of asymptomatic partial vaginal exposures were observed at 3 months (P4HB: n = 3; PP: n = 2) and 9 months (P4HB: n = 3; PP: n = 2) respectively. Histological analysis of specimens showed comparable results in the P4HB and PP groups at 3 and 9 months post-implantation. Although no acute inflammation was seen, chronic inflammation was demonstrated in all specimens. Elastic fibers were present in the 3-month vaginal PP and P4HB specimens, but were not seen again. There was an increase in type I/III collagen noted over time. Biomechanical evaluation of the vaginal mesh tissue complex showed ultimate tensile strength was not significantly different between P4HB and PP groups at 3 (P = 0.625) and 9 months (P = 0.250) respectively.

Conclusions

P4HB scaffold may represent a fully absorbable alternative to permanent mesh for pelvic organ prolapse (POP) repair.
Literature
1.
go back to reference Campbell P, Jha S, Cutner A. Vaginal mesh in prolapse surgery. Obstet Gynaecol. 2018;20(1):49–56.CrossRef Campbell P, Jha S, Cutner A. Vaginal mesh in prolapse surgery. Obstet Gynaecol. 2018;20(1):49–56.CrossRef
2.
go back to reference Maher C, Feiner B, Baessler K, Christmann-Schmid C, Haya N, Marjoribanks J. Transvaginal mesh or grafts compared with native tissue repair for vaginal prolapse. Cochrane Database Syst Rev. 2016;2(2):CD012079.PubMed Maher C, Feiner B, Baessler K, Christmann-Schmid C, Haya N, Marjoribanks J. Transvaginal mesh or grafts compared with native tissue repair for vaginal prolapse. Cochrane Database Syst Rev. 2016;2(2):CD012079.PubMed
3.
go back to reference Martin DP, Williams SF. Medical applications of poly-4-hydroxybutyrate: a strong flexible absorbable biomaterial. Biochem Eng J. 2003;16(2):97–105.CrossRef Martin DP, Williams SF. Medical applications of poly-4-hydroxybutyrate: a strong flexible absorbable biomaterial. Biochem Eng J. 2003;16(2):97–105.CrossRef
4.
go back to reference Roth JS, Anthone GJ, Selzer DJ, et al. Prospective evaluation of poly-4-hydroxybutyrate mesh in CDC class I/high-risk ventral and incisional hernia repair: 18-month follow-up. Surg Endosc. 2018;32(4):1929–36.CrossRef Roth JS, Anthone GJ, Selzer DJ, et al. Prospective evaluation of poly-4-hydroxybutyrate mesh in CDC class I/high-risk ventral and incisional hernia repair: 18-month follow-up. Surg Endosc. 2018;32(4):1929–36.CrossRef
5.
go back to reference Adams WP Jr, Baxter R, Glicksman C, Mast BA, Tantillo M, Van Natta BW. The use of poly-4-hydroxybutyrate (P4HB) scaffold in the ptotic breast: a multicenter clinical study. Aesthet Surg J. 2018;38(5):502–18.CrossRef Adams WP Jr, Baxter R, Glicksman C, Mast BA, Tantillo M, Van Natta BW. The use of poly-4-hydroxybutyrate (P4HB) scaffold in the ptotic breast: a multicenter clinical study. Aesthet Surg J. 2018;38(5):502–18.CrossRef
6.
go back to reference Deeken CR, Matthews BD. Characterization of the mechanical strength, resorption properties, and histologic characteristics of a fully absorbable material (poly-4-hydroxybutyrate—PHASIX mesh) in a porcine model of hernia repair. ISRN Surg. 2013;2013:238067.CrossRef Deeken CR, Matthews BD. Characterization of the mechanical strength, resorption properties, and histologic characteristics of a fully absorbable material (poly-4-hydroxybutyrate—PHASIX mesh) in a porcine model of hernia repair. ISRN Surg. 2013;2013:238067.CrossRef
7.
go back to reference Abramov Y, Golden B, Sullivan M, et al. Histologic characterization of vaginal vs. abdominal surgical wound healing in a rabbit model. Wound Repair Regen. 2007;15(1):80–6.CrossRef Abramov Y, Golden B, Sullivan M, et al. Histologic characterization of vaginal vs. abdominal surgical wound healing in a rabbit model. Wound Repair Regen. 2007;15(1):80–6.CrossRef
8.
go back to reference Huffaker RK, Muir TW, Rao A, Baumann SS, Kuehl TJ, Pierce LM. Histologic response of porcine collagen-coated and uncoated polypropylene grafts in a rabbit vagina model. Am J Obstet Gynecol. 2008;198(5):582.e1–7.CrossRef Huffaker RK, Muir TW, Rao A, Baumann SS, Kuehl TJ, Pierce LM. Histologic response of porcine collagen-coated and uncoated polypropylene grafts in a rabbit vagina model. Am J Obstet Gynecol. 2008;198(5):582.e1–7.CrossRef
9.
go back to reference Ziats NP, Miller KM, Anderson JM. In vitro and in vivo interactions of cells with biomaterials. Biomaterials. 1988;9(1):5–13.CrossRef Ziats NP, Miller KM, Anderson JM. In vitro and in vivo interactions of cells with biomaterials. Biomaterials. 1988;9(1):5–13.CrossRef
10.
go back to reference Patel H, Ostergard DR, Sternschuss G. Polypropylene mesh and the host response. Int Urogynecol J. 2012;23(6):669–79.CrossRef Patel H, Ostergard DR, Sternschuss G. Polypropylene mesh and the host response. Int Urogynecol J. 2012;23(6):669–79.CrossRef
11.
go back to reference Pierce LM, Rao A, Baumann SS, Glassberg JE, Kuehl TJ, Muir TW. Long-term histologic response to synthetic and biologic graft materials implanted in the vagina and abdomen of a rabbit model. Am J Obstet Gynecol. 2009;200(5):546.e1–8.CrossRef Pierce LM, Rao A, Baumann SS, Glassberg JE, Kuehl TJ, Muir TW. Long-term histologic response to synthetic and biologic graft materials implanted in the vagina and abdomen of a rabbit model. Am J Obstet Gynecol. 2009;200(5):546.e1–8.CrossRef
12.
go back to reference Stadelmann WK, Digenis AG, Tobin GR. Physiology and healing dynamics of chronic cutaneous wounds. Am J Surg. 1998;176(2):26S–38S.CrossRef Stadelmann WK, Digenis AG, Tobin GR. Physiology and healing dynamics of chronic cutaneous wounds. Am J Surg. 1998;176(2):26S–38S.CrossRef
13.
go back to reference Junge K, Klinge U, Rosch R, et al. Decreased collagen type I/III ratio in patients with recurring hernia after implantation of alloplastic prostheses. Langenbecks Arch Surg. 2004;389(1):17–22.CrossRef Junge K, Klinge U, Rosch R, et al. Decreased collagen type I/III ratio in patients with recurring hernia after implantation of alloplastic prostheses. Langenbecks Arch Surg. 2004;389(1):17–22.CrossRef
14.
go back to reference Nolfi AL, Brown BN, Liang R, Palcsey SL, Bonidie MJ, Abramowitch SD, et al. Host response to synthetic mesh in women with mesh complications. Am J Obstet Gynecol. 2016;215(2):206.e1–8CrossRef Nolfi AL, Brown BN, Liang R, Palcsey SL, Bonidie MJ, Abramowitch SD, et al. Host response to synthetic mesh in women with mesh complications. Am J Obstet Gynecol. 2016;215(2):206.e1–8CrossRef
15.
go back to reference Manodoro S, Endo M, Uvin P, et al. Graft-related complications and biaxial tensiometry following experimental vaginal implantation of flat mesh of variable dimensions. BJOG Gynaecol. 2013;120(2):244–50.CrossRef Manodoro S, Endo M, Uvin P, et al. Graft-related complications and biaxial tensiometry following experimental vaginal implantation of flat mesh of variable dimensions. BJOG Gynaecol. 2013;120(2):244–50.CrossRef
16.
go back to reference Feiner B, Maher C. Vaginal mesh contraction: definition, clinical presentation, and management. Obstet Gynecol. 2010;115(2):325–30.CrossRef Feiner B, Maher C. Vaginal mesh contraction: definition, clinical presentation, and management. Obstet Gynecol. 2010;115(2):325–30.CrossRef
17.
go back to reference Ulrich D, Edwards SL, Alexander DL, et al. Changes in pelvic organ prolapse mesh mechanical properties following implantation in rats. Am J Obstet Gynecol. 2016;214(2):260.e1–8.CrossRef Ulrich D, Edwards SL, Alexander DL, et al. Changes in pelvic organ prolapse mesh mechanical properties following implantation in rats. Am J Obstet Gynecol. 2016;214(2):260.e1–8.CrossRef
18.
go back to reference Pierce LM, Grunlan MA, Hou Y, Baumann SS, Kuehl TJ, Muir TW. Biomechanical properties of synthetic and biologic graft materials following long-term implantation in the rabbit abdomen and vagina. Am J Obstet Gynecol. 2009;200(5):549.e1–8.CrossRef Pierce LM, Grunlan MA, Hou Y, Baumann SS, Kuehl TJ, Muir TW. Biomechanical properties of synthetic and biologic graft materials following long-term implantation in the rabbit abdomen and vagina. Am J Obstet Gynecol. 2009;200(5):549.e1–8.CrossRef
19.
go back to reference Liang R, Abramowitch S, Knight K, et al. Vaginal degeneration following implantation of synthetic mesh with increased stiffness. BJOG. 2013;120(2):233–43.CrossRef Liang R, Abramowitch S, Knight K, et al. Vaginal degeneration following implantation of synthetic mesh with increased stiffness. BJOG. 2013;120(2):233–43.CrossRef
Metadata
Title
Evaluation of the histological and biomechanical properties of poly-4-hydroxybutyrate scaffold for pelvic organ prolapse, compared with polypropylene mesh in a rabbit model
Authors
Danielle O’Shaughnessy
Daniel Grande
Diana El-Neemany
Sujata Sajjan
Nirmala Pillalamarri
Dara Shalom
Harvey Winkler
Publication date
14-06-2021
Publisher
Springer International Publishing
Published in
International Urogynecology Journal / Issue 8/2022
Print ISSN: 0937-3462
Electronic ISSN: 1433-3023
DOI
https://doi.org/10.1007/s00192-021-04851-6

Other articles of this Issue 8/2022

International Urogynecology Journal 8/2022 Go to the issue