Evaluation of pulse oximeter at the nasal septum during general anesthesia: comparison with finger oximeter

Authors: Yena Oh, Duk Kyung Kim, Dae Kyun Ryu, Ji Won Choi

Published in: Journal of Anesthesia | Issue 3/2024

Login to get access

Abstract

Purpose

Though the finger is generally recommended for pulse oxygen saturation (SpO2) monitoring site, its reliability may be compromised in conditions of poor peripheral perfusion. Therefore, we compared the performance of nasal septum SpO2 monitoring with finger SpO2 monitoring relative to simultaneous arterial oxygen saturation (SaO2) monitoring in generally anesthetized patients.

Methods

In 23 adult patients, comparisons of SpO2 measured at the nasal septum and finger with simultaneous SaO2 were made at four time points during the 90 min study period. A pulse oximetry monitoring failure was defined as a > 10 s continuous failure of in an adequate SpO2 data acquisition. Core temperature as well as finger-tip and nasal septum temperatures were simultaneously measured at 10 min intervals.

Results

A total of 92 sets of SpO2 and SaO2 measurements were obtained in 23 patients. The bias and precision for SpO2 measured at the nasal septum were − 0.8 ± 1.3 (95% confidence interval: − 1.1 to − 0.6), which was similar to those for SpO2 measured at the finger (− 0.6 ± 1.4; 95% confidence interval: − 0.9 to − 0.4) (p = 0.154). Finger-tip temperatures were consistently lower than other two temperatures at all time points (p < 0.05), reaching 33.5 ± 2.3 °C at 90 min after induction of anesthesia. While pulse oximetry monitoring failure did not occur for nasal septum probe, two cases of failure occurred for finger probe.

Conclusions

Considering the higher stability to hypothermia with a similar accuracy, nasal septum pulse oximetry may be an attractive alternative to finger pulse oximetry.
Trail registration This study was registered with Clinical Research Information Service (CRIS: https://​cris.​nih.​go.​kr/​cris/​en/​; ref: KCT0008352).
Literature
1.
go back to reference Falconer RJ, Robinson BJ. Comparison of pulse oximeters: accuracy at low arterial pressure in volunteers. Br J Anaesth. 1990;65:552–7.CrossRefPubMed Falconer RJ, Robinson BJ. Comparison of pulse oximeters: accuracy at low arterial pressure in volunteers. Br J Anaesth. 1990;65:552–7.CrossRefPubMed
2.
go back to reference Lawson D, Norley I, Korbon G, Loeb R, Ellis J. Blood flow limits and pulse oximeter signal detection. Anesthesiology. 1987;67:599–603.CrossRefPubMed Lawson D, Norley I, Korbon G, Loeb R, Ellis J. Blood flow limits and pulse oximeter signal detection. Anesthesiology. 1987;67:599–603.CrossRefPubMed
3.
go back to reference Palve H, Vuori A. Accuracy of three pulse oximeters at low cardiac index and peripheral temperature. Crit Care Med. 1991;19:560–2.CrossRefPubMed Palve H, Vuori A. Accuracy of three pulse oximeters at low cardiac index and peripheral temperature. Crit Care Med. 1991;19:560–2.CrossRefPubMed
4.
go back to reference Ezri T, Lurie S, Konichezky S, Soroker D. Pulse oximetry from the nasal septum. J Clin Anesth. 1991;3:447–50.CrossRefPubMed Ezri T, Lurie S, Konichezky S, Soroker D. Pulse oximetry from the nasal septum. J Clin Anesth. 1991;3:447–50.CrossRefPubMed
5.
go back to reference Clayton DG, Webb RK, Ralston AC, Duthie D, Runciman WB. Pulse oximeter probes. A comparison between finger, nose, ear and forehead probes under conditions of poor perfusion. Anaesthesia. 1991;46:260–5.CrossRefPubMed Clayton DG, Webb RK, Ralston AC, Duthie D, Runciman WB. Pulse oximeter probes. A comparison between finger, nose, ear and forehead probes under conditions of poor perfusion. Anaesthesia. 1991;46:260–5.CrossRefPubMed
6.
go back to reference Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.CrossRefPubMed Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.CrossRefPubMed
7.
go back to reference Ralston AC, Webb RK, Runciman WB. Potential errors in pulse oximetry. I. Pulse oximeter evaluation. Anaesthesia. 1991;46:202–6.CrossRefPubMed Ralston AC, Webb RK, Runciman WB. Potential errors in pulse oximetry. I. Pulse oximeter evaluation. Anaesthesia. 1991;46:202–6.CrossRefPubMed
8.
go back to reference Evans ML, Geddes LA. An assessment of blood vessel vasoactivity using photoplethysmography. Med Instrum. 1988;22:29–32.PubMed Evans ML, Geddes LA. An assessment of blood vessel vasoactivity using photoplethysmography. Med Instrum. 1988;22:29–32.PubMed
9.
go back to reference Clayton D, Webb R, Ralston A, Duthie D, Runciman W. Pulse oximeter probes- a comparison between finger, nose, ear and forehead probes under conditions of poor perfusion. Anaesthesia. 1991;46:260–5.CrossRefPubMed Clayton D, Webb R, Ralston A, Duthie D, Runciman W. Pulse oximeter probes- a comparison between finger, nose, ear and forehead probes under conditions of poor perfusion. Anaesthesia. 1991;46:260–5.CrossRefPubMed
10.
go back to reference Budidha K, Kyriacou PA. In vivo investigation of ear canal pulse oximetry during hypothermia. J Clin Monit Comput. 2018;32:97–107.CrossRefPubMed Budidha K, Kyriacou PA. In vivo investigation of ear canal pulse oximetry during hypothermia. J Clin Monit Comput. 2018;32:97–107.CrossRefPubMed
11.
go back to reference Coté CJ, Daniels AL, Connoll M, Szyfelbein SK, Wickens CD. Tongue oximetry in children with extensive thermal injury: comparison with peripheral oximetry. Can J Anaesth. 1992;39:454–7.CrossRefPubMed Coté CJ, Daniels AL, Connoll M, Szyfelbein SK, Wickens CD. Tongue oximetry in children with extensive thermal injury: comparison with peripheral oximetry. Can J Anaesth. 1992;39:454–7.CrossRefPubMed
12.
go back to reference Pälve H, Vuori A. Pulse oximetry during low cardiac output and hypothermia states immediately after open heart surgery. Crit Care Med. 1989;17:66–9.CrossRefPubMed Pälve H, Vuori A. Pulse oximetry during low cardiac output and hypothermia states immediately after open heart surgery. Crit Care Med. 1989;17:66–9.CrossRefPubMed
13.
go back to reference Lim HS, Kim BR, Kim DC, Lee SK, Ko SH. A comparison of the temperature difference according to the placement of a nasopharyngeal temperature probe. Korean J Anesthesiol. 2016;69:357–61.CrossRefPubMedPubMedCentral Lim HS, Kim BR, Kim DC, Lee SK, Ko SH. A comparison of the temperature difference according to the placement of a nasopharyngeal temperature probe. Korean J Anesthesiol. 2016;69:357–61.CrossRefPubMedPubMedCentral
14.
15.
go back to reference Shelley KH, Tamai D, Jablonka D, Gesquiere M, Stout RG, Silverman DG. The effect of venous pulsation on the forehead pulse oximeter waveform as a possible source of error in Spo2 calculation. Anesth Analg. 2005;100:743–7.CrossRefPubMed Shelley KH, Tamai D, Jablonka D, Gesquiere M, Stout RG, Silverman DG. The effect of venous pulsation on the forehead pulse oximeter waveform as a possible source of error in Spo2 calculation. Anesth Analg. 2005;100:743–7.CrossRefPubMed
16.
go back to reference Berge KH, Lanier WL, Scanlon PD. Ischemic digital skin necrosis: a complication of the reusable Nelcor pulse oximeter probe. Anesth Analg. 1988;67:712–3.CrossRefPubMed Berge KH, Lanier WL, Scanlon PD. Ischemic digital skin necrosis: a complication of the reusable Nelcor pulse oximeter probe. Anesth Analg. 1988;67:712–3.CrossRefPubMed
17.
go back to reference Goodell TT. An in vitro quantification of pressures exerted by earlobe pulse oximeter probes following reports of device-related pressure ulcers in ICU patients. Ostomy Wound Manage. 2012;58:30–4.PubMed Goodell TT. An in vitro quantification of pressures exerted by earlobe pulse oximeter probes following reports of device-related pressure ulcers in ICU patients. Ostomy Wound Manage. 2012;58:30–4.PubMed
18.
go back to reference Severinghaus JW, Naifeh KH. Accuracy of response of six pulse oximeters to profound hypoxia. Anesthesiology. 1987;67:551–8.CrossRefPubMed Severinghaus JW, Naifeh KH. Accuracy of response of six pulse oximeters to profound hypoxia. Anesthesiology. 1987;67:551–8.CrossRefPubMed
19.
go back to reference Desgranges FP, Desebbe O, Ghazouani A, Gilbert K, Keller G, Chiari P, Robin J, Bastien O, Lehot JJ, Cannesson M. Influence of the site of measurement on the ability of plethysmographic variability index to predict fluid responsiveness. Br J Anaesth. 2011;107:329–35.CrossRefPubMed Desgranges FP, Desebbe O, Ghazouani A, Gilbert K, Keller G, Chiari P, Robin J, Bastien O, Lehot JJ, Cannesson M. Influence of the site of measurement on the ability of plethysmographic variability index to predict fluid responsiveness. Br J Anaesth. 2011;107:329–35.CrossRefPubMed
Metadata
Title
Evaluation of pulse oximeter at the nasal septum during general anesthesia: comparison with finger oximeter
Authors
Yena Oh
Duk Kyung Kim
Dae Kyun Ryu
Ji Won Choi
Publication date
19-03-2024
Publisher
Springer Nature Singapore
Published in
Journal of Anesthesia / Issue 3/2024
Print ISSN: 0913-8668
Electronic ISSN: 1438-8359
DOI
https://doi.org/10.1007/s00540-024-03317-5

Other articles of this Issue 3/2024

Journal of Anesthesia 3/2024 Go to the issue