01-06-2018 | Short Communication

Evaluation of implant screw loosening by resonance frequency analysis with triaxial piezoelectric pick-up: in vitro model and in vivo animal study

Authors: Makiko Miyashita, Toru Ogawa, Hideki Naito, Aya Shibamoto, Angelina Shuhua Wang, Kenta Shobara, Keiichi Sasaki

Published in: Clinical Oral Investigations | Issue 5/2018

Login to get access

Abstract

Objective

The aim of this study was to evaluate implant screw loosening using resonance frequency (RF) analysis with triaxial piezoelectric pick-up in vitro and in vivo.

Methods

For the in vitro experiment, a titanium implant was inserted into a mandibular model. The abutment screws were tightened to 10 N torque and loosened from 0 to 90° for RF measurement using 13 different screw conditions. In the in vivo experiment, three titanium implants were installed in each tibia of a New Zealand white rabbit, and the RF values were recorded after 8 weeks. For the RF analyses, a small 3D accelerometer was mounted rigidly onto the implant abutment, and impulsive vibration was directly applied to the abutment to vibrate the implant in a direction perpendicular to the tibia and implant (x-axis). The y-axis was defined as parallel to the tibia. The RF values of the x- and y-axis directions (RF-X and RF-Y) were used for evaluation.

Results

The RF values significantly decreased according to the degree of screw loosening (p < 0.05, ANOVA). In the in vitro model, RF-Y with x- and y-axis vibrations (RF-Yx and RF-Yy) significantly differed from the initial value at 10 and 15°, respectively (p < 0.05, Dunnett’s test). In the in vivo experiment, the RF-Yy significantly differed from the initial value between 5 and 20° (p < 0.05).

Conclusion

The results suggest that RF analysis with triaxial piezoelectric pick-up can be used to detect implant screw loosening.

Clinical relevance

RF analysis with the triaxial piezoelectric pick-up can be used to detect screw loosening after mounting the superstructure.
Literature
1.
go back to reference Tey VH, Philips R, Tan K (2017) Five-year retrospective study on success, survival and incidence of complications of single crowns supported by dental implants. Clin Oral Implants Res 5:620–625CrossRef Tey VH, Philips R, Tan K (2017) Five-year retrospective study on success, survival and incidence of complications of single crowns supported by dental implants. Clin Oral Implants Res 5:620–625CrossRef
2.
go back to reference Kourtis S, Damanaki M, Kaitatzidou S, Kaitatzidou A, Roussou V (2017) Loosening of the fixing screw in single implant crowns: predisposing factors, prevention and treatment options. J Esthet Restor Dent 29(4):233–246CrossRefPubMed Kourtis S, Damanaki M, Kaitatzidou S, Kaitatzidou A, Roussou V (2017) Loosening of the fixing screw in single implant crowns: predisposing factors, prevention and treatment options. J Esthet Restor Dent 29(4):233–246CrossRefPubMed
3.
go back to reference Papaspyridakos P, Chen CJ, Chuang SK, Weber HP, Gallucci GO (2012) A systematic review of biologic and technical complications with fixed implant rehabilitations for edentulous patients. Int J Oral Maxillofac Implants 1:102–110 Papaspyridakos P, Chen CJ, Chuang SK, Weber HP, Gallucci GO (2012) A systematic review of biologic and technical complications with fixed implant rehabilitations for edentulous patients. Int J Oral Maxillofac Implants 1:102–110
4.
go back to reference Jung RE, Zembic A, Pjetursson BE, Zwahlen M, Thoma DS (2012) Systematic review of the survival rate and the incidence of biological, technical, and aesthetic complications of single crowns on implants reported in longitudinal studies with a mean follow-up of 5 years. Clin Oral Implants Res 6:2–21CrossRef Jung RE, Zembic A, Pjetursson BE, Zwahlen M, Thoma DS (2012) Systematic review of the survival rate and the incidence of biological, technical, and aesthetic complications of single crowns on implants reported in longitudinal studies with a mean follow-up of 5 years. Clin Oral Implants Res 6:2–21CrossRef
5.
go back to reference Ogawa T, Dhaliwal S, Naert I, Mine A, Kronstrom M, Sasaki K, Duyck J (2010) Effect of tilted and short distal implants on axial forces and bending moments in implants supporting fixed dental prostheses: an in vitro study. Int J Prosthodont 23:566–573PubMed Ogawa T, Dhaliwal S, Naert I, Mine A, Kronstrom M, Sasaki K, Duyck J (2010) Effect of tilted and short distal implants on axial forces and bending moments in implants supporting fixed dental prostheses: an in vitro study. Int J Prosthodont 23:566–573PubMed
6.
go back to reference Meredith N, Alleyne D, Cawley P (1996) Quantitative determination of the stability of the implant-tissue interface using resonance frequency analysis. Clin Oral Implants Res 3:261–267CrossRef Meredith N, Alleyne D, Cawley P (1996) Quantitative determination of the stability of the implant-tissue interface using resonance frequency analysis. Clin Oral Implants Res 3:261–267CrossRef
7.
go back to reference Glauser R, Sennerby L, Meredith N, Ree A, Lundgren A, Gottlow J, Hammerle CH (2004) Resonance frequency analysis of implants subjected to immediate or early functional occlusal loading. Successful vs. failing implants. Clin Oral Implants Res 4:428–434CrossRef Glauser R, Sennerby L, Meredith N, Ree A, Lundgren A, Gottlow J, Hammerle CH (2004) Resonance frequency analysis of implants subjected to immediate or early functional occlusal loading. Successful vs. failing implants. Clin Oral Implants Res 4:428–434CrossRef
8.
go back to reference Kitagawa T, Tanimoto Y, Odaki M, Nemoto K, Aida M (2005) Influence of implant/abutment joint designs on abutment screw loosening in a dental implant system. J Biomed Mater Res B Appl Biomater 75:457–463CrossRefPubMed Kitagawa T, Tanimoto Y, Odaki M, Nemoto K, Aida M (2005) Influence of implant/abutment joint designs on abutment screw loosening in a dental implant system. J Biomed Mater Res B Appl Biomater 75:457–463CrossRefPubMed
Metadata
Title
Evaluation of implant screw loosening by resonance frequency analysis with triaxial piezoelectric pick-up: in vitro model and in vivo animal study
Authors
Makiko Miyashita
Toru Ogawa
Hideki Naito
Aya Shibamoto
Angelina Shuhua Wang
Kenta Shobara
Keiichi Sasaki
Publication date
01-06-2018
Publisher
Springer Berlin Heidelberg
Published in
Clinical Oral Investigations / Issue 5/2018
Print ISSN: 1432-6981
Electronic ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-018-2467-1

Other articles of this Issue 5/2018

Clinical Oral Investigations 5/2018 Go to the issue