Evaluation of different POCT devices for glucose measurement in a clinical neonatal setting

Authors: Matthias Roth-Kleiner, Corinne Stadelmann Diaw, Jocelyne Urfer, Christiane Ruffieux, Dominique Werner

Published in: European Journal of Pediatrics | Issue 11/2010

Login to get access

Abstract

Hypoglycaemia is a major cause of neonatal morbidity and may induce long-term developmental sequelae. Clinical signs of hypoglycaemia in neonatal infants are unspecific or even absent, and therefore, precise and accurate methods for the assessment of glycaemia are needed. Glycaemia measurement in newborns has some particularities like a very low limit of normal glucose concentration compared to adults and a large range of normal haematocrit values. Many bedside point-of-care testing (POCT) systems are available, but literature about their accuracy in newborn infants is scarce and not very convincing. In this retrospective study, we identified over a 1-year study period 1,324 paired glycaemia results, one obtained at bedside with one of three different POCT systems (Elite™ XL, Ascensia™ Contour™ and ABL 735) and the other in the central laboratory of the hospital with the hexokinase reference method. All three POCT systems tended to overestimate glycaemia values, and none of them fulfilled the ISO 15197 accuracy criteria. The Elite XL appeared to be more appropriate than Contour to detect hypoglycaemia, however with a low specificity. Contour additionally showed an important inaccuracy with increasing haematocrit. The bench analyzer ABL 735 was the most accurate of the three tested POCT systems. Both of the tested handheld glucometers have important drawbacks in their use as screening tools for hypoglycaemia in newborn infants. ABL 735 could be a valuable alternative, but the blood volume needed is more than 15 times higher than for handheld glucometers. Before daily use in the newborn population, careful clinical evaluation of each new POCT system for glucose measurement is of utmost importance.
Literature
2.
3.
go back to reference Chan AY, Swaminathan R, Cockram CS (1989) Effectiveness of sodium fluoride as a preservative of glucose in blood. Clin Chem 35:315–317PubMed Chan AY, Swaminathan R, Cockram CS (1989) Effectiveness of sodium fluoride as a preservative of glucose in blood. Clin Chem 35:315–317PubMed
4.
go back to reference Cornblath M, Hawdon JM, Williams AF, Aynsley-Green A, Ward-Platt MP, Schwartz R, Kalhan SC (2000) Controversies regarding definition of neonatal hypoglycemia: suggested operational thresholds. Pediatrics 105:1141–1145CrossRefPubMed Cornblath M, Hawdon JM, Williams AF, Aynsley-Green A, Ward-Platt MP, Schwartz R, Kalhan SC (2000) Controversies regarding definition of neonatal hypoglycemia: suggested operational thresholds. Pediatrics 105:1141–1145CrossRefPubMed
5.
go back to reference Dietzen DJ, Lee HR, Dunne NA, Viggiani MT (2007) Performance of the new Bayer CONTOUR blood glucose monitoring system in neonates. Bayer HealthCare, March 2007, pp 1–5 Dietzen DJ, Lee HR, Dunne NA, Viggiani MT (2007) Performance of the new Bayer CONTOUR blood glucose monitoring system in neonates. Bayer HealthCare, March 2007, pp 1–5
6.
go back to reference Duvanel CB, Fawer CL, Cotting J, Hohlfeld P, Matthieu JM (1999) Long-term effects of neonatal hypoglycemia on brain growth and psychomotor development in small-for-gestational-age preterm infants. J Pediatr 134:492–498CrossRefPubMed Duvanel CB, Fawer CL, Cotting J, Hohlfeld P, Matthieu JM (1999) Long-term effects of neonatal hypoglycemia on brain growth and psychomotor development in small-for-gestational-age preterm infants. J Pediatr 134:492–498CrossRefPubMed
7.
go back to reference Germagnoli L, Bonini P, DuBois J, Bierens de Haan J, Tartarotti C (2009) Suitability assessment of a new bedside interference-free glucose system for use in critical care when compared with current technology. Point of Care 8:96–100 Germagnoli L, Bonini P, DuBois J, Bierens de Haan J, Tartarotti C (2009) Suitability assessment of a new bedside interference-free glucose system for use in critical care when compared with current technology. Point of Care 8:96–100
8.
go back to reference Girouard J, Forest JC, Masse J, Leroux M, Bradburn NC, Noblet TC, Joynes JO, Baum J (2000) Multicenter evaluation of the glucometer Elite XL meter, an instrument specifically designed for use with neonates. Diab Care 23:1149–1153CrossRef Girouard J, Forest JC, Masse J, Leroux M, Bradburn NC, Noblet TC, Joynes JO, Baum J (2000) Multicenter evaluation of the glucometer Elite XL meter, an instrument specifically designed for use with neonates. Diab Care 23:1149–1153CrossRef
10.
go back to reference Holtrop PC, Madison KA, Kiechle FL, Karcher RE, Batton DG (1990) A comparison of chromogen test strip (Chemstrip bG) and serum glucose values in newborns. Am J Dis Child 144:183–185PubMed Holtrop PC, Madison KA, Kiechle FL, Karcher RE, Batton DG (1990) A comparison of chromogen test strip (Chemstrip bG) and serum glucose values in newborns. Am J Dis Child 144:183–185PubMed
11.
go back to reference Holtzinger C, Szelag E, DuBois J, Shirey T, Presti S (2008) Evaluation of a new POCT bedside glucose meter and strip with hematocrit and interference corrections. Point of Care 7:1–6 Holtzinger C, Szelag E, DuBois J, Shirey T, Presti S (2008) Evaluation of a new POCT bedside glucose meter and strip with hematocrit and interference corrections. Point of Care 7:1–6
12.
go back to reference Hussain K, Sharief N (2000) The inaccuracy of venous and capillary blood glucose measurement using reagent strips in the newborn period and the effect of haematocrit. Early Hum Dev 57:111–121CrossRefPubMed Hussain K, Sharief N (2000) The inaccuracy of venous and capillary blood glucose measurement using reagent strips in the newborn period and the effect of haematocrit. Early Hum Dev 57:111–121CrossRefPubMed
13.
go back to reference International Organization for Standardization (2003) In vitro diagnostic test systems—requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. ISO 15197 International Organization for Standardization (2003) In vitro diagnostic test systems—requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. ISO 15197
14.
go back to reference Jopling J, Henry E, Wiedmeier SE, Christensen RD (2009) Reference ranges for hematocrit and blood hemoglobin concentration during the neonatal period: data from a multihospital health care system. Pediatrics 123:e333–e337. doi:10.1542/peds.2008-2654 PubMed Jopling J, Henry E, Wiedmeier SE, Christensen RD (2009) Reference ranges for hematocrit and blood hemoglobin concentration during the neonatal period: data from a multihospital health care system. Pediatrics 123:e333–e337. doi:10.​1542/​peds.​2008-2654 PubMed
16.
go back to reference Leibundgut KE, Bucher HU, Mieth D, Duc G (1989) Clinical assessment of a new glucose reflectance meter (Glucoscot II) for the detection of neonatal hypoglycemias. Monatsschr Kinderheilkd 137:330–332PubMed Leibundgut KE, Bucher HU, Mieth D, Duc G (1989) Clinical assessment of a new glucose reflectance meter (Glucoscot II) for the detection of neonatal hypoglycemias. Monatsschr Kinderheilkd 137:330–332PubMed
17.
go back to reference Leonard M, Chessall M, Manning D (1997) The use of a Hemocue blood glucose analyser in a neonatal unit. Ann Clin Biochem 34(Pt 3):287–290PubMed Leonard M, Chessall M, Manning D (1997) The use of a Hemocue blood glucose analyser in a neonatal unit. Ann Clin Biochem 34(Pt 3):287–290PubMed
18.
go back to reference Louie RF, Tang Z, Sutton DV, Lee JH, Kost GJ (2000) Point-of-care glucose testing: effects of critical care variables, influence of reference instruments, and a modular glucose meter design. Arch Pathol Lab Med 124:257–266PubMed Louie RF, Tang Z, Sutton DV, Lee JH, Kost GJ (2000) Point-of-care glucose testing: effects of critical care variables, influence of reference instruments, and a modular glucose meter design. Arch Pathol Lab Med 124:257–266PubMed
19.
go back to reference Lucas A, Morley R, Cole TJ (1988) Adverse neurodevelopmental outcome of moderate neonatal hypoglycaemia. BMJ 297:1304–1308CrossRefPubMed Lucas A, Morley R, Cole TJ (1988) Adverse neurodevelopmental outcome of moderate neonatal hypoglycaemia. BMJ 297:1304–1308CrossRefPubMed
20.
go back to reference Meloy L, Miller G, Chandrasekaran MH, Summitt C, Gutcher G (1999) Accuracy of glucose reflectance testing for detecting hypoglycemia in term newborns. Clin Pediatr (Phila) 38:717–724CrossRef Meloy L, Miller G, Chandrasekaran MH, Summitt C, Gutcher G (1999) Accuracy of glucose reflectance testing for detecting hypoglycemia in term newborns. Clin Pediatr (Phila) 38:717–724CrossRef
22.
go back to reference Movalia M, Ogino M (2006) Point of care glucose testing in neonatal hypoglycemia. Point of Care 5:95–99 Movalia M, Ogino M (2006) Point of care glucose testing in neonatal hypoglycemia. Point of Care 5:95–99
23.
go back to reference Papp M, Sharief N (2001) Comparison of two strip test methods of whole blood glucose measurement in the neonatal period. Acta Paediatr 90:1042–1046CrossRefPubMed Papp M, Sharief N (2001) Comparison of two strip test methods of whole blood glucose measurement in the neonatal period. Acta Paediatr 90:1042–1046CrossRefPubMed
24.
go back to reference Pelletier O, Arratoon C (1987) Precision of glucose measurements in control sera by isotope dilution/mass spectrometry: proposed definitive method compared with a reference method. Clin Chem 33:1397–1402PubMed Pelletier O, Arratoon C (1987) Precision of glucose measurements in control sera by isotope dilution/mass spectrometry: proposed definitive method compared with a reference method. Clin Chem 33:1397–1402PubMed
27.
go back to reference Sirkin A, Jalloh T, Lee L (2002) Selecting an accurate point-of-care testing system: clinical and technical issues and implications in neonatal blood glucose monitoring. J Spec Pediatr Nurs 7:104–112CrossRefPubMed Sirkin A, Jalloh T, Lee L (2002) Selecting an accurate point-of-care testing system: clinical and technical issues and implications in neonatal blood glucose monitoring. J Spec Pediatr Nurs 7:104–112CrossRefPubMed
28.
go back to reference St-Louis P, Ethier J (2002) An evaluation of three glucose meter systems and their performance in relation to criteria of acceptability for neonatal specimens. Clin Chim Acta 322:139–148CrossRefPubMed St-Louis P, Ethier J (2002) An evaluation of three glucose meter systems and their performance in relation to criteria of acceptability for neonatal specimens. Clin Chim Acta 322:139–148CrossRefPubMed
30.
go back to reference Tang Z, Lee JH, Louie RF, Kost GJ (2000) Effects of different hematocrit levels on glucose measurements with handheld meters for point-of-care testing. Arch Pathol Lab Med 124:1135–1140PubMed Tang Z, Lee JH, Louie RF, Kost GJ (2000) Effects of different hematocrit levels on glucose measurements with handheld meters for point-of-care testing. Arch Pathol Lab Med 124:1135–1140PubMed
31.
go back to reference Tang Z, Louie RF, Lee JH, Lee DM, Miller EE, Kost GJ (2001) Oxygen effects on glucose meter measurements with glucose dehydrogenase- and oxidase-based test strips for point-of-care testing. Crit Care Med 29:1062–1070CrossRefPubMed Tang Z, Louie RF, Lee JH, Lee DM, Miller EE, Kost GJ (2001) Oxygen effects on glucose meter measurements with glucose dehydrogenase- and oxidase-based test strips for point-of-care testing. Crit Care Med 29:1062–1070CrossRefPubMed
Metadata
Title
Evaluation of different POCT devices for glucose measurement in a clinical neonatal setting
Authors
Matthias Roth-Kleiner
Corinne Stadelmann Diaw
Jocelyne Urfer
Christiane Ruffieux
Dominique Werner
Publication date
01-11-2010
Publisher
Springer-Verlag
Published in
European Journal of Pediatrics / Issue 11/2010
Print ISSN: 0340-6199
Electronic ISSN: 1432-1076
DOI
https://doi.org/10.1007/s00431-010-1243-2

Other articles of this Issue 11/2010

European Journal of Pediatrics 11/2010 Go to the issue