Published in: Intensive Care Medicine 11/2003

01-11-2003 | Technical Note

Evaluation of a new invasive continuous cardiac output monitoring system: the truCCOMS system

Authors: Stéphane Thierry, Dominique Thebert, Elsa Brocas, Fereshte Razzaghi, Andry Van de Louw, Daniel Loisance, Jean Louis Teboul

Published in: Intensive Care Medicine | Issue 11/2003

Login to get access

Abstract

Objective

To compare measurements of cardiac output using a new pulmonary artery catheter with those obtained using two " gold standard " methods: the periaortic transit time ultrasonic flow probe and the conventional pulmonary artery thermodilution.

Design

Prospective clinical trial.

Setting

Cardiac surgery operating room and surgical ICU in a university hospital.

Material and methods

In the operating room, a new pulmonary artery catheter (truCCOMS system) was inserted in eight patients. A periaortic flow probe was inserted in four of them. Measurements of cardiac output obtained with the truCCOMS catheter and with the flow probe were compared at different phases of the surgical procedure. In the intensive care unit, the cardiac output displayed by the truCCOMS monitor was compared with the value obtained after bolus injection performed subsequently.

Results

In the operating room (70 measurements), the coefficient of correlation between cardiac output measured by the flow probe and the truCCOMS system was r 2 = 0.79, the bias was +0.11 l/min with a precision of 0.47 l/min, and limits of agreement −0.83 to +1.05 l/min. In the intensive care unit (108 measurements), the coefficient of correlation between cardiac output measured by thermodilution and the truCCOMS system was r 2 = 0.56, the bias was −0.07 l/min, the precision was 0.66 l/min, and the limits of agreement were −1.39 to +1.25 l/min.

Conclusion

The truCCOMS system is a reliable method of continuous cardiac output measurement in cardiac surgery patients.
Literature
1.
go back to reference Shoemaker WC, Appel P, Kram HB, Waxman K, Lee TS (1988) Prospective trial of supranormal value of survivors as therapeutic goals in high-risk surgical patients. Chest 94:1176–1186PubMed Shoemaker WC, Appel P, Kram HB, Waxman K, Lee TS (1988) Prospective trial of supranormal value of survivors as therapeutic goals in high-risk surgical patients. Chest 94:1176–1186PubMed
2.
go back to reference Boyd O, Ground M, Bennett D (1993) A randomized clinical trial of the effect of deliberate perioperative increase of oxygen delivery on mortality in high risk surgical patients. JAMA 270:2699–2707PubMed Boyd O, Ground M, Bennett D (1993) A randomized clinical trial of the effect of deliberate perioperative increase of oxygen delivery on mortality in high risk surgical patients. JAMA 270:2699–2707PubMed
3.
go back to reference Wilson J, Woods I, Fawcett J, Whall R, Dibb W, Morris C, McManus E (1999) Reducing the risk of major elective surgery: randomised controlled trial of preoperative optimisation of oxygen delivery. BMJ 318:1099–1103PubMed Wilson J, Woods I, Fawcett J, Whall R, Dibb W, Morris C, McManus E (1999) Reducing the risk of major elective surgery: randomised controlled trial of preoperative optimisation of oxygen delivery. BMJ 318:1099–1103PubMed
4.
go back to reference Yelderman ML, Ramsay MA, Quinn MD (1992) Continuous thermodilution cardiac output measurement in intensive care unit patients. J Cardiothorac Vasc Anesth 6:270–274PubMed Yelderman ML, Ramsay MA, Quinn MD (1992) Continuous thermodilution cardiac output measurement in intensive care unit patients. J Cardiothorac Vasc Anesth 6:270–274PubMed
5.
go back to reference Haller M, Zöllner C, Briegel J, Forst H (1995) Evaluation of a new continuous thermodilution cardiac output monitor in critically ill patients: a prospective criterion standard study. Crit Care Med 23:860–866PubMed Haller M, Zöllner C, Briegel J, Forst H (1995) Evaluation of a new continuous thermodilution cardiac output monitor in critically ill patients: a prospective criterion standard study. Crit Care Med 23:860–866PubMed
6.
go back to reference Seed W, Wood N (1970) Development and evaluation of a hot-film velocity probe for cardiovascular studies. Cardiovasc Res 4:253–263 Seed W, Wood N (1970) Development and evaluation of a hot-film velocity probe for cardiovascular studies. Cardiovasc Res 4:253–263
7.
go back to reference Seed W (1972) A catheter-tip hot-film probe for the measurement of aortic blood velocity in man. J Physiol 223:5P–6P Seed W (1972) A catheter-tip hot-film probe for the measurement of aortic blood velocity in man. J Physiol 223:5P–6P
8.
go back to reference Dean DD, Jia CX, Cabreriza SE, D'Alessandro DA, Dickstein ML, Sardo MJ, Chalik N, Spotnitz HM (1996) Validation study of a new transit time ultrasonic flow probe for continuous great vessel measurements. ASAIO J 42:M671–676 Dean DD, Jia CX, Cabreriza SE, D'Alessandro DA, Dickstein ML, Sardo MJ, Chalik N, Spotnitz HM (1996) Validation study of a new transit time ultrasonic flow probe for continuous great vessel measurements. ASAIO J 42:M671–676
9.
go back to reference Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMed Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMed
10.
go back to reference Botero M, Klodell K, Blas M, Staples E, Lobato EB (2001) Evaluation of a new continuous thermodilution cardiac output system in cardiac surgical patients: clinical validation against aortic transit-time ultrasound. ASA Gainsville, USA Botero M, Klodell K, Blas M, Staples E, Lobato EB (2001) Evaluation of a new continuous thermodilution cardiac output system in cardiac surgical patients: clinical validation against aortic transit-time ultrasound. ASA Gainsville, USA
11.
go back to reference Boldt J, Menges T, Wollbrück M (1994) Is continuous cardiac output measurement using thermodilution reliable in the critically ill patient? Crit Care Med 22:1913–1918PubMed Boldt J, Menges T, Wollbrück M (1994) Is continuous cardiac output measurement using thermodilution reliable in the critically ill patient? Crit Care Med 22:1913–1918PubMed
12.
go back to reference Jacquet L, Hanique G, Glorieux D (1996) Analysis of the accuracy of continuous thermodilution cardiac output measurement. Intensive Care Med 22:1125–1129PubMed Jacquet L, Hanique G, Glorieux D (1996) Analysis of the accuracy of continuous thermodilution cardiac output measurement. Intensive Care Med 22:1125–1129PubMed
13.
go back to reference LeTulzo Y, Belgith M, Seguin P, Dall'Ava J, Monchi M, Thomas R, Dhainaut JF (1996) Reproducibility of thermodilution cardiac output determination in critically ill patients: comparison between bolus and continuous method. J Clin Monit 12:379–385PubMed LeTulzo Y, Belgith M, Seguin P, Dall'Ava J, Monchi M, Thomas R, Dhainaut JF (1996) Reproducibility of thermodilution cardiac output determination in critically ill patients: comparison between bolus and continuous method. J Clin Monit 12:379–385PubMed
14.
go back to reference Van Grondelle A, Ditchey RV, Groves BM, Wagner WW, Reeves JT (1983) Thermodilution method overestimates low cardiac output in humans. Am J Physiol 245:H690–692PubMed Van Grondelle A, Ditchey RV, Groves BM, Wagner WW, Reeves JT (1983) Thermodilution method overestimates low cardiac output in humans. Am J Physiol 245:H690–692PubMed
Metadata
Title
Evaluation of a new invasive continuous cardiac output monitoring system: the truCCOMS system
Authors
Stéphane Thierry
Dominique Thebert
Elsa Brocas
Fereshte Razzaghi
Andry Van de Louw
Daniel Loisance
Jean Louis Teboul
Publication date
01-11-2003
Publisher
Springer-Verlag
Published in
Intensive Care Medicine / Issue 11/2003
Print ISSN: 0342-4642
Electronic ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-003-1946-1

Other articles of this Issue 11/2003

Intensive Care Medicine 11/2003 Go to the issue