Published in: European Journal of Applied Physiology 3/2008

01-06-2008 | Original Article

Estimation of the distribution of intramuscular current during electrical stimulation of the quadriceps muscle

Authors: Jerrold Petrofsky, Michelle Prowse, Melanie Bain, Elaine Ebilane, Hye Jin Suh, Jennifer Batt, Daryl Lawson, Viviana Hernandez, Armia Abdo, Tien-ning Yang, Enrique Mendoza, Kelly Collins, Michael Laymon

Published in: European Journal of Applied Physiology | Issue 3/2008

Login to get access

Abstract

Electrical stimulation is commonly used for strengthening muscle but little evidence exists as to the optimal electrode size, waveform, or frequency to apply. Three male and three female subjects (22–40 years old) were examined during electrical stimulation of the quadriceps muscle. Two self adhesive electrode sizes were examined, 2 cm × 2 cm and 2 cm × 4 cm. Electrical stimulation was applied with square and sine waveforms, currents of 5, 10 and 15 mA, and pulse widths of 100–500 μs above the quadriceps muscle. Frequencies of stimulation were 20, 30, and 50 Hz. Current on the skin above the quadriceps muscle was measured with surface electrodes at five positions and at three positions with needle electrodes in the same muscle. Altering pulse width in the range of 100–500 μs, the frequency over a range of 20–50 Hz, or current from 5 to 15 mA had no effect on current dispersion either in the skin or within muscle. In contrast, the distance separating the electrodes caused large changes in current dispersion on the skin or into muscle. The most significant finding in the present investigation was that, while on the surface of the skin current dispersion was not different between sine and square wave stimulation, significantly more current was transferred deep in the muscle with sine versus square wave stimulation. The use of sine wave stimulation with electrode separation distances of less then 15 cm is recommended for electrical stimulation with a sine wave to achieve deep muscle stimulation.
Literature
go back to reference Abram SE, Asiddao CB, Reynolds AC (1980) Increased skin temperature during transcutaneous nerve stimulation. Anesth Analg 59:22–25PubMed Abram SE, Asiddao CB, Reynolds AC (1980) Increased skin temperature during transcutaneous nerve stimulation. Anesth Analg 59:22–25PubMed
go back to reference Alon G (1985) High voltage stimulation; effects of electrode size on basic excitatory responses. Phys Ther 65:890–895PubMed Alon G (1985) High voltage stimulation; effects of electrode size on basic excitatory responses. Phys Ther 65:890–895PubMed
go back to reference Ashley EA, Laskin JJ, Olenik LM, Burnham R, Steadward RD, Cumming DC, Wheeler GD (1993) Evidence of autonomic dysreflexia during functional electrical stimulation in individuals with spinal cord injuries. Paraplegia 31:593–605PubMed Ashley EA, Laskin JJ, Olenik LM, Burnham R, Steadward RD, Cumming DC, Wheeler GD (1993) Evidence of autonomic dysreflexia during functional electrical stimulation in individuals with spinal cord injuries. Paraplegia 31:593–605PubMed
go back to reference Baker LL, Bowman BR, McNeal DR (1988) Effects of waveforms on comfort during neuromuscular electrical stimulation. Clin Orthop 233:75–85PubMed Baker LL, Bowman BR, McNeal DR (1988) Effects of waveforms on comfort during neuromuscular electrical stimulation. Clin Orthop 233:75–85PubMed
go back to reference Bennie SD, Petrofsky JS, Nisperos J, Tsurudome M, Laymon M (2002) Toward the optimal waveform for electrical stimulation of human muscle. Eur J Appl Physiol 88:13–19PubMedCrossRef Bennie SD, Petrofsky JS, Nisperos J, Tsurudome M, Laymon M (2002) Toward the optimal waveform for electrical stimulation of human muscle. Eur J Appl Physiol 88:13–19PubMedCrossRef
go back to reference Brown BH, Karatzas T, Nakielny R, Clarke RG (1988) Determination of upper arm muscle and fat areas using electrical impedance measurements. Clin Phys Physiol Meas 9(1):47–55PubMedCrossRef Brown BH, Karatzas T, Nakielny R, Clarke RG (1988) Determination of upper arm muscle and fat areas using electrical impedance measurements. Clin Phys Physiol Meas 9(1):47–55PubMedCrossRef
go back to reference Delitto A, Rose SJ (1986) Comparative comfort of three waveforms used in electrically elicitating quadriceps femoris muscle contractions. Phys Ther 66:1704–1707PubMed Delitto A, Rose SJ (1986) Comparative comfort of three waveforms used in electrically elicitating quadriceps femoris muscle contractions. Phys Ther 66:1704–1707PubMed
go back to reference Delitto A, Strube MJ, Shulman AD, Minor SD (1992) A study of discomfort with electrical stimulation. Phys Ther 72:410–424PubMed Delitto A, Strube MJ, Shulman AD, Minor SD (1992) A study of discomfort with electrical stimulation. Phys Ther 72:410–424PubMed
go back to reference Eriksson E, Haggmark T (1979) Comparison of isometric muscle training and electrical stimulation supplementing isometric muscle training in the recovery after major knee ligament surgery. Am J Sports Med 7:169–171PubMedCrossRef Eriksson E, Haggmark T (1979) Comparison of isometric muscle training and electrical stimulation supplementing isometric muscle training in the recovery after major knee ligament surgery. Am J Sports Med 7:169–171PubMedCrossRef
go back to reference Forrester B, Petrofsky JS (2004) Effect of electrode size and shape on electrical stimulation. Eur J Appl Physiol 4:346–354 Forrester B, Petrofsky JS (2004) Effect of electrode size and shape on electrical stimulation. Eur J Appl Physiol 4:346–354
go back to reference Greco EC Jr, Clark JW Jr (1977) The field from an isolated nerve in a volume conductor. IEEE Trans Biomed Eng 24(1):18–23PubMedCrossRef Greco EC Jr, Clark JW Jr (1977) The field from an isolated nerve in a volume conductor. IEEE Trans Biomed Eng 24(1):18–23PubMedCrossRef
go back to reference Hewett P (2006) Electric current, chap 23. In: Conceptual physics. Addison Wesely, St Petersburg Hewett P (2006) Electric current, chap 23. In: Conceptual physics. Addison Wesely, St Petersburg
go back to reference Iwasaki T, Shiba N, Matsuse H, Nago T, Umezu Y, Tagawa Y, Nagata K, Basford JR (2006) Improvement in knee extension strength through training by means of combined electrical stimulation and voluntary muscle contraction. Tohoku J Exp Med 209(1):33–40PubMedCrossRef Iwasaki T, Shiba N, Matsuse H, Nago T, Umezu Y, Tagawa Y, Nagata K, Basford JR (2006) Improvement in knee extension strength through training by means of combined electrical stimulation and voluntary muscle contraction. Tohoku J Exp Med 209(1):33–40PubMedCrossRef
go back to reference Leon LJ, Roberge FA (1993) A model study of extracellular stimulation of cardiac cells. IEEE Trans Biomed Eng 40(12):1307–1319PubMedCrossRef Leon LJ, Roberge FA (1993) A model study of extracellular stimulation of cardiac cells. IEEE Trans Biomed Eng 40(12):1307–1319PubMedCrossRef
go back to reference Matthews JM, Wheeler GD, Burnham RS, Malone LA, Steadward RD (1997) The effects of surface anesthesia on the autonomic dysreflexia response during functional electrical stimulation. Spinal Cord 35:647–651PubMedCrossRef Matthews JM, Wheeler GD, Burnham RS, Malone LA, Steadward RD (1997) The effects of surface anesthesia on the autonomic dysreflexia response during functional electrical stimulation. Spinal Cord 35:647–651PubMedCrossRef
go back to reference Mesin L, Farina D (2005) A model for surface EMG generation in volume conductors with spherical inhomogeneities. IEEE Trans Biomed Eng 52(12):1984–1993PubMedCrossRef Mesin L, Farina D (2005) A model for surface EMG generation in volume conductors with spherical inhomogeneities. IEEE Trans Biomed Eng 52(12):1984–1993PubMedCrossRef
go back to reference Miranda PC, Correia L, Salvador R, Basser PJ (2007) Tissue heterogeneity as a mechanism for localized neural stimulation by applied electric fields. Phys Med Biol 52(18):5603–5617PubMedCrossRef Miranda PC, Correia L, Salvador R, Basser PJ (2007) Tissue heterogeneity as a mechanism for localized neural stimulation by applied electric fields. Phys Med Biol 52(18):5603–5617PubMedCrossRef
go back to reference Moore SR, Shurman J (1997) Combined neuromuscular electrical stimulation and transcutaneous electrical stimulation for treatment of chronic back pain; a double blind, repeated measures comparison. Arch Phys Med Rehabil 78(1):55–60PubMedCrossRef Moore SR, Shurman J (1997) Combined neuromuscular electrical stimulation and transcutaneous electrical stimulation for treatment of chronic back pain; a double blind, repeated measures comparison. Arch Phys Med Rehabil 78(1):55–60PubMedCrossRef
go back to reference Parker MG, Bennett MJ, Hieb MA, Hollar AC, Roe AA (2003) Strength response in human femoris muscle during 2 neuromuscular electrical stimulation programs. J Orthop Sports Phys Ther 33(12):719–726PubMed Parker MG, Bennett MJ, Hieb MA, Hollar AC, Roe AA (2003) Strength response in human femoris muscle during 2 neuromuscular electrical stimulation programs. J Orthop Sports Phys Ther 33(12):719–726PubMed
go back to reference Petrofsky JS, Smith J (1988) The 1987 Harry G. Armstrong lecture: computer aided rehabilitation. Aviat Space Environ Med 59:670–678PubMed Petrofsky JS, Smith J (1988) The 1987 Harry G. Armstrong lecture: computer aided rehabilitation. Aviat Space Environ Med 59:670–678PubMed
go back to reference Petrofsky JS, Schwab E (2007) A re-evaluation of modeling of current flow between electrodes: consideration of skin and muscle blood flow and wounds. J Med Eng Technol 31(1):62–74PubMedCrossRef Petrofsky JS, Schwab E (2007) A re-evaluation of modeling of current flow between electrodes: consideration of skin and muscle blood flow and wounds. J Med Eng Technol 31(1):62–74PubMedCrossRef
go back to reference Petrofsky JS, Schwab E, Lo T, Cuneo M, George J, Kim J, Almalty A (2005) Effects of electrical stimulation on skin blood flow in controls and in and around stage III and IV wounds in hairy and non hairy skin. Med Sci Monit 11:309–316 Petrofsky JS, Schwab E, Lo T, Cuneo M, George J, Kim J, Almalty A (2005) Effects of electrical stimulation on skin blood flow in controls and in and around stage III and IV wounds in hairy and non hairy skin. Med Sci Monit 11:309–316
go back to reference Petrofsky JS, Schwab E, Cuneo M, George J, Kim J, Almalty A, Lawson D, Johnson L, Regimio W (2006) Current distribution under electrodes in relation to stimulation current and skin blood flow: are modern electrodes really providing the current distribution during stimulation as we believe they are? J Med Eng Technol 30:368–381PubMedCrossRef Petrofsky JS, Schwab E, Cuneo M, George J, Kim J, Almalty A, Lawson D, Johnson L, Regimio W (2006) Current distribution under electrodes in relation to stimulation current and skin blood flow: are modern electrodes really providing the current distribution during stimulation as we believe they are? J Med Eng Technol 30:368–381PubMedCrossRef
go back to reference Petrofsky JS, Lawson D, Suh HJ, Batt J (2007) Effects of a 2, 3, and 4 electrode stimulator design on current dispersion on the surface and into the limb during electrical stimulation in controls and patients with wounds. J Med Eng Technol (in press) Petrofsky JS, Lawson D, Suh HJ, Batt J (2007) Effects of a 2, 3, and 4 electrode stimulator design on current dispersion on the surface and into the limb during electrical stimulation in controls and patients with wounds. J Med Eng Technol (in press)
go back to reference Plonsey R (2000) Volume conductor theory. In: Biomedical engineering handbook, chap 9. CRC Press, Boca Raton Plonsey R (2000) Volume conductor theory. In: Biomedical engineering handbook, chap 9. CRC Press, Boca Raton
go back to reference Roth B (2000) The electrical conductivity of tissues. In: Handbook of bioengineering, chap 10. CRC Press, Boca Raton Roth B (2000) The electrical conductivity of tissues. In: Handbook of bioengineering, chap 10. CRC Press, Boca Raton
go back to reference Scheker LR, Chesher SP, Ramirez S (1999) Neuromuscular electrical stimulation and dynamic bracing as a treatment for upper-extremity spasticity in children with cerebral palsy. J Hand Surg 24(2):226–232 Scheker LR, Chesher SP, Ramirez S (1999) Neuromuscular electrical stimulation and dynamic bracing as a treatment for upper-extremity spasticity in children with cerebral palsy. J Hand Surg 24(2):226–232
go back to reference Selkowitz DM (1985) Improvement in isometric strength of the quadriceps femoris muscle after training with electrical stimulation. Phys Ther 65:186–196PubMed Selkowitz DM (1985) Improvement in isometric strength of the quadriceps femoris muscle after training with electrical stimulation. Phys Ther 65:186–196PubMed
go back to reference Thrasher A, Graham GM, Popovic MR (2005) Reducing muscle fatigue due to functional electrical stimulation using random modulation of stimulation parameters. Artif Organs 29(6):453–458PubMedCrossRef Thrasher A, Graham GM, Popovic MR (2005) Reducing muscle fatigue due to functional electrical stimulation using random modulation of stimulation parameters. Artif Organs 29(6):453–458PubMedCrossRef
go back to reference Ward AR, Robertson VJ (1998) Variation in torque production with frequency using medium frequency alternating current. Arch Phys Med Rehabil 79(11):1399–1404PubMedCrossRef Ward AR, Robertson VJ (1998) Variation in torque production with frequency using medium frequency alternating current. Arch Phys Med Rehabil 79(11):1399–1404PubMedCrossRef
go back to reference Zhu F, Leonard EF, Levin NW (2005) Body composition modeling in the calf using an equivalent circuit model of multi-frequency bioimpedance analysis. Physiol Meas 26(2):S133–S143PubMedCrossRef Zhu F, Leonard EF, Levin NW (2005) Body composition modeling in the calf using an equivalent circuit model of multi-frequency bioimpedance analysis. Physiol Meas 26(2):S133–S143PubMedCrossRef
Metadata
Title
Estimation of the distribution of intramuscular current during electrical stimulation of the quadriceps muscle
Authors
Jerrold Petrofsky
Michelle Prowse
Melanie Bain
Elaine Ebilane
Hye Jin Suh
Jennifer Batt
Daryl Lawson
Viviana Hernandez
Armia Abdo
Tien-ning Yang
Enrique Mendoza
Kelly Collins
Michael Laymon
Publication date
01-06-2008
Publisher
Springer-Verlag
Published in
European Journal of Applied Physiology / Issue 3/2008
Print ISSN: 1439-6319
Electronic ISSN: 1439-6327
DOI
https://doi.org/10.1007/s00421-008-0700-3

Other articles of this Issue 3/2008

European Journal of Applied Physiology 3/2008 Go to the issue