Published in: Surgical Endoscopy 7/2021

01-07-2021 | Endoscopy | 2020 SAGES Oral

Electromagnetic interference on cardiac pacemakers and implantable cardioverter defibrillators during endoscopy as reported to the US Federal Drug Administration

Authors: Jason M. Samuels, Douglas M. Overbey, Krzysztof J. Wikiel, Teresa S. Jones, Thomas N. Robinson, Edward L. Jones

Published in: Surgical Endoscopy | Issue 7/2021

Login to get access

Abstract

Introduction

More than 3 million patients have a cardiac implanted electronic device (CIED) such as a pacemaker or implanted cardioverter-defibrillator in the USA. These devices are susceptible to electromagnetic interference (EMI) leading to malfunction and injury. Radiofrequency energy, the most common modality for obtaining hemostasis during endoscopy, is the most common source of EMI. Few studies have evaluated the effect of endoscopic radiofrequency energy on CIEDs. We aim to characterize CIED dysfunction related to endoscopic procedures. We hypothesize that EMI from endoscopic energy can result in patient injury.

Methods

We queried the Manufacturer and User Facility Device Experience (MAUDE) database for CIED dysfunction related to electrosurgical devices over a 10-year period (2009–2019). CIED dysfunction events were identified using seven problem codes (malfunction, electromagnetic interference, ambient noise, pacing problem, over-sensing, inappropriate shock, defibrillation). These were cross-referenced for the terms “cautery, electrocautery, endoscopy, esophagus, colonoscopy, colon, esophagoscopy, and esophagogastroduodenoscopy.” Reports were individually reviewed to confirm and characterize CIED malfunction due to an endoscopic procedure.

Results

A search for CIED dysfunction resulted in 43,759 reports. Three hundred and eleven reports (0.7%) were associated with electrocautery, and of these, 45 reports (14.5%) included endoscopy. Ten reports involving endoscopy (22%) specified upper (3, 7%) or lower (7, 16%) endoscopy while the remainder were non-specific. Twenty-six of reports involving endoscopy (58%) suffered injury because of CIED dysfunction: Of these, 17 (65%) received inappropriate shocks, 5 (19%) had pacing inhibition with bradycardia or asystole, 3 (12%) had CIED damage requiring explant and replacement, and 1 (4%) patient suffered ventricular tachycardia requiring hospital admission.

Conclusion

The use of energy during endoscopy can cause dysfunction of CIEDs. This most commonly results in inappropriate defibrillation, symptomatic bradycardia, and asystole. Patients with CIEDs undergoing endoscopy should undergo pre- and post-procedure device interrogation and re-programming to avoid patient injury.
Literature
1.
go back to reference Morris ML, Tucker RD, Baron TH, Song LM (2009) Electrosurgery in gastrointestinal endoscopy: principles to practice. Am J Gastroenterol 104:1563–1574CrossRef Morris ML, Tucker RD, Baron TH, Song LM (2009) Electrosurgery in gastrointestinal endoscopy: principles to practice. Am J Gastroenterol 104:1563–1574CrossRef
2.
go back to reference Zhan C, Baine WB, Sedrakyan A, Steiner C (2008) Cardiac device implantation in the United States from 1997 through 2004: a population-based analysis. J Gen Intern Med 23(Suppl 1):13–19CrossRef Zhan C, Baine WB, Sedrakyan A, Steiner C (2008) Cardiac device implantation in the United States from 1997 through 2004: a population-based analysis. J Gen Intern Med 23(Suppl 1):13–19CrossRef
3.
go back to reference Mond HG, Proclemer A (2011) The 11th world survey of cardiac pacing and implantable cardioverter-defibrillators: calendar year 2009—a World Society of Arrhythmia's project. Pacing Clin Electrophysiol 34:1013–1027CrossRef Mond HG, Proclemer A (2011) The 11th world survey of cardiac pacing and implantable cardioverter-defibrillators: calendar year 2009—a World Society of Arrhythmia's project. Pacing Clin Electrophysiol 34:1013–1027CrossRef
4.
go back to reference Misiri J, Kusumoto F, Goldschlager N (2012) Electromagnetic interference and implanted cardiac devices: the nonmedical environment (part I). Clin Cardiol 35:276–280CrossRef Misiri J, Kusumoto F, Goldschlager N (2012) Electromagnetic interference and implanted cardiac devices: the nonmedical environment (part I). Clin Cardiol 35:276–280CrossRef
5.
go back to reference Misiri J, Kusumoto F, Goldschlager N (2012) Electromagnetic interference and implanted cardiac devices: the medical environment (part II). Clin Cardiol 35:321–328CrossRef Misiri J, Kusumoto F, Goldschlager N (2012) Electromagnetic interference and implanted cardiac devices: the medical environment (part II). Clin Cardiol 35:321–328CrossRef
6.
go back to reference Yerra L, Reddy PC (2007) Effects of electromagnetic interference on implanted cardiac devices and their management. Cardiol Rev 15:304–309CrossRef Yerra L, Reddy PC (2007) Effects of electromagnetic interference on implanted cardiac devices and their management. Cardiol Rev 15:304–309CrossRef
7.
go back to reference Govekar HR, Robinson TN, Varosy PD, Girard G, Montero PN, Dunn CL, Jones EL, Stiegmann GV (2012) Effect of monopolar radiofrequency energy on pacemaker function. Surg Endosc 26:2784–2788CrossRef Govekar HR, Robinson TN, Varosy PD, Girard G, Montero PN, Dunn CL, Jones EL, Stiegmann GV (2012) Effect of monopolar radiofrequency energy on pacemaker function. Surg Endosc 26:2784–2788CrossRef
8.
go back to reference Mangar D, Atlas GM, Kane PB (1991) Electrocautery-induced pacemaker malfunction during surgery. Can J Anaesth 38:616–618CrossRef Mangar D, Atlas GM, Kane PB (1991) Electrocautery-induced pacemaker malfunction during surgery. Can J Anaesth 38:616–618CrossRef
9.
go back to reference Casavant D, Haffajee C, Stevens S, Pacetti P (1998) Aborted implantable cardioverter defibrillator shock during facial electrosurgery. Pacing Clin Electrophysiol 21:1325–1326CrossRef Casavant D, Haffajee C, Stevens S, Pacetti P (1998) Aborted implantable cardioverter defibrillator shock during facial electrosurgery. Pacing Clin Electrophysiol 21:1325–1326CrossRef
10.
go back to reference Nercessian OA, Wu H, Nazarian D, Mahmud F (1998) Intraoperative pacemaker dysfunction caused by the use of electrocautery during a total hip arthroplasty. J Arthroplasty 13:599–602CrossRef Nercessian OA, Wu H, Nazarian D, Mahmud F (1998) Intraoperative pacemaker dysfunction caused by the use of electrocautery during a total hip arthroplasty. J Arthroplasty 13:599–602CrossRef
11.
go back to reference Goel AK, Korotkin S, Walsh D, Bess M, Frawley S (2009) Monomorphic ventricular tachycardia caused by electrocautery during pacemaker generator change in a patient with normal left ventricular function. Pacing Clin Electrophysiol 32:957–958CrossRef Goel AK, Korotkin S, Walsh D, Bess M, Frawley S (2009) Monomorphic ventricular tachycardia caused by electrocautery during pacemaker generator change in a patient with normal left ventricular function. Pacing Clin Electrophysiol 32:957–958CrossRef
12.
go back to reference Ito S, Shibata H, Okahisa T, Okamura S, Wada S, Yano M, Saijyo T, Honda H, Hayashi H, Shimizu I et al (1994) Endoscopic therapy using monopolar and bipolar snare with a high-frequency current in patients with a pacemaker. Endoscopy 26:270CrossRef Ito S, Shibata H, Okahisa T, Okamura S, Wada S, Yano M, Saijyo T, Honda H, Hayashi H, Shimizu I et al (1994) Endoscopic therapy using monopolar and bipolar snare with a high-frequency current in patients with a pacemaker. Endoscopy 26:270CrossRef
13.
go back to reference Baeg MK, Kim SW, Ko SH, Lee YB, Hwang S, Lee BW, Choi HJ, Park JM, Lee IS, Oh YS, Choi MG (2016) Endoscopic electrosurgery in patients with cardiac implantable electronic devices. Clin Endosc 49:176–181CrossRef Baeg MK, Kim SW, Ko SH, Lee YB, Hwang S, Lee BW, Choi HJ, Park JM, Lee IS, Oh YS, Choi MG (2016) Endoscopic electrosurgery in patients with cardiac implantable electronic devices. Clin Endosc 49:176–181CrossRef
14.
go back to reference Cheng A, Nazarian S, Spragg DD, Bilchick K, Tandri H, Mark L, Halperin H, Calkins H, Berger RD, Henrikson CA (2008) Effects of surgical and endoscopic electrocautery on modern-day permanent pacemaker and implantable cardioverter-defibrillator systems. Pacing Clin Electrophysiol 31:344–350CrossRef Cheng A, Nazarian S, Spragg DD, Bilchick K, Tandri H, Mark L, Halperin H, Calkins H, Berger RD, Henrikson CA (2008) Effects of surgical and endoscopic electrocautery on modern-day permanent pacemaker and implantable cardioverter-defibrillator systems. Pacing Clin Electrophysiol 31:344–350CrossRef
15.
go back to reference Petersen BT, Hussain N, Marine JE, Trohman RG, Carpenter S, Chuttani R, Croffie J, Disario J, Chotiprasidhi P, Liu J, Somogyi L, Technology Assessment Committee (2007) Endoscopy in patients with implanted electronic devices. Gastrointest Endosc 65:561–568CrossRef Petersen BT, Hussain N, Marine JE, Trohman RG, Carpenter S, Chuttani R, Croffie J, Disario J, Chotiprasidhi P, Liu J, Somogyi L, Technology Assessment Committee (2007) Endoscopy in patients with implanted electronic devices. Gastrointest Endosc 65:561–568CrossRef
16.
go back to reference Chang KJ (2019) Endoscopic foregut surgery and interventions: the future is now. The state-of-the-art and my personal journey. World J Gastroenterol 25:1–41CrossRef Chang KJ (2019) Endoscopic foregut surgery and interventions: the future is now. The state-of-the-art and my personal journey. World J Gastroenterol 25:1–41CrossRef
17.
go back to reference Talukdar R, Inoue H, Nageshwar Reddy D (2015) Efficacy of peroral endoscopic myotomy (POEM) in the treatment of achalasia: a systematic review and meta-analysis. Surg Endosc 29:3030–3046CrossRef Talukdar R, Inoue H, Nageshwar Reddy D (2015) Efficacy of peroral endoscopic myotomy (POEM) in the treatment of achalasia: a systematic review and meta-analysis. Surg Endosc 29:3030–3046CrossRef
18.
go back to reference Kim MS, Holloway RH, Dent J, Utley DS (2003) Radiofrequency energy delivery to the gastric cardia inhibits triggering of transient lower esophageal sphincter relaxation and gastroesophageal reflux in dogs. Gastrointest Endosc 57:17–22CrossRef Kim MS, Holloway RH, Dent J, Utley DS (2003) Radiofrequency energy delivery to the gastric cardia inhibits triggering of transient lower esophageal sphincter relaxation and gastroesophageal reflux in dogs. Gastrointest Endosc 57:17–22CrossRef
19.
go back to reference Shaheen NJ, Sharma P, Overholt BF, Wolfsen HC, Sampliner RE, Wang KK, Galanko JA, Bronner MP, Goldblum JR, Bennett AE, Jobe BA, Eisen GM, Fennerty MB, Hunter JG, Fleischer DE, Sharma VK, Hawes RH, Hoffman BJ, Rothstein RI, Gordon SR, Mashimo H, Chang KJ, Muthusamy VR, Edmundowicz SA, Spechler SJ, Siddiqui AA, Souza RF, Infantolino A, Falk GW, Kimmey MB, Madanick RD, Chak A, Lightdale CJ (2009) Radiofrequency ablation in Barrett's esophagus with dysplasia. N Engl J Med 360:2277–2288CrossRef Shaheen NJ, Sharma P, Overholt BF, Wolfsen HC, Sampliner RE, Wang KK, Galanko JA, Bronner MP, Goldblum JR, Bennett AE, Jobe BA, Eisen GM, Fennerty MB, Hunter JG, Fleischer DE, Sharma VK, Hawes RH, Hoffman BJ, Rothstein RI, Gordon SR, Mashimo H, Chang KJ, Muthusamy VR, Edmundowicz SA, Spechler SJ, Siddiqui AA, Souza RF, Infantolino A, Falk GW, Kimmey MB, Madanick RD, Chak A, Lightdale CJ (2009) Radiofrequency ablation in Barrett's esophagus with dysplasia. N Engl J Med 360:2277–2288CrossRef
20.
go back to reference Crossley GH, Poole JE, Rozner MA, Asirvatham SJ, Cheng A, Chung MK, Ferguson TB, Gallagher JD, Gold MR, Hoyt RH, Irefin S, Kusumoto FM, Moorman LP, Thompson A (2011) The Heart Rhythm Society (HRS)/American Society of Anesthesiologists (ASA) Expert consensus statement on the perioperative management of patients with implantable defibrillators, pacemakers and arrhythmia monitors: facilities and patient management: this document was developed as a joint project with the American Society of Anesthesiologists (ASA), and in collaboration with the American Heart Association (AHA), and the Society of Thoracic Surgeons (STS). Heart Rhythm 8:1114–1154CrossRef Crossley GH, Poole JE, Rozner MA, Asirvatham SJ, Cheng A, Chung MK, Ferguson TB, Gallagher JD, Gold MR, Hoyt RH, Irefin S, Kusumoto FM, Moorman LP, Thompson A (2011) The Heart Rhythm Society (HRS)/American Society of Anesthesiologists (ASA) Expert consensus statement on the perioperative management of patients with implantable defibrillators, pacemakers and arrhythmia monitors: facilities and patient management: this document was developed as a joint project with the American Society of Anesthesiologists (ASA), and in collaboration with the American Heart Association (AHA), and the Society of Thoracic Surgeons (STS). Heart Rhythm 8:1114–1154CrossRef
21.
go back to reference Kusumoto FM, Schoenfeld MH, Barrett C, Edgerton JR, Ellenbogen KA, Gold MR, Goldschlager NF, Hamilton RM, Joglar JA, Kim RJ, Lee R, Marine JE, McLeod CJ, Oken KR, Patton KK, Pellegrini CN, Selzman KA, Thompson A, Varosy PD (2019) 2018 ACC/AHA/HRS Guideline on the evaluation and management of patients With bradycardia and cardiac conduction delay: a report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. Circulation 140:e382–e482PubMed Kusumoto FM, Schoenfeld MH, Barrett C, Edgerton JR, Ellenbogen KA, Gold MR, Goldschlager NF, Hamilton RM, Joglar JA, Kim RJ, Lee R, Marine JE, McLeod CJ, Oken KR, Patton KK, Pellegrini CN, Selzman KA, Thompson A, Varosy PD (2019) 2018 ACC/AHA/HRS Guideline on the evaluation and management of patients With bradycardia and cardiac conduction delay: a report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. Circulation 140:e382–e482PubMed
22.
go back to reference Al-Khatib SM, Stevenson WG, Ackerman MJ, Bryant WJ, Callans DJ, Curtis AB, Deal BJ, Dickfeld T, Field ME, Fonarow GC, Gillis AM, Granger CB, Hammill SC, Hlatky MA, Joglar JA, Kay GN, Matlock DD, Myerburg RJ, Page RL (2018) 2017 AHA/ACC/HRS guideline for management of patients with ventricular arrhythmias and the prevention of sudden cardiac death: executive summary: a report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. Heart Rhythm 15:e190–e252CrossRef Al-Khatib SM, Stevenson WG, Ackerman MJ, Bryant WJ, Callans DJ, Curtis AB, Deal BJ, Dickfeld T, Field ME, Fonarow GC, Gillis AM, Granger CB, Hammill SC, Hlatky MA, Joglar JA, Kay GN, Matlock DD, Myerburg RJ, Page RL (2018) 2017 AHA/ACC/HRS guideline for management of patients with ventricular arrhythmias and the prevention of sudden cardiac death: executive summary: a report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. Heart Rhythm 15:e190–e252CrossRef
23.
go back to reference (2020) Practice Advisory for the Perioperative Management of Patients with Cardiac Implantable Electronic Devices: Pacemakers and Implantable Cardioverter-Defibrillators 2020: An Updated Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Implantable Electronic Devices: Erratum Anesthesiology 132:225-252 (2020) Practice Advisory for the Perioperative Management of Patients with Cardiac Implantable Electronic Devices: Pacemakers and Implantable Cardioverter-Defibrillators 2020: An Updated Report by the American Society of Anesthesiologists Task Force on Perioperative Management of Patients with Cardiac Implantable Electronic Devices: Erratum Anesthesiology 132:225-252
24.
go back to reference Rasmussen MJ, Friedman PA, Hammill SC, Rea RF (2002) Unintentional deactivation of implantable cardioverter-defibrillators in health care settings. Mayo Clin Proc 77:855–859CrossRef Rasmussen MJ, Friedman PA, Hammill SC, Rea RF (2002) Unintentional deactivation of implantable cardioverter-defibrillators in health care settings. Mayo Clin Proc 77:855–859CrossRef
Metadata
Title
Electromagnetic interference on cardiac pacemakers and implantable cardioverter defibrillators during endoscopy as reported to the US Federal Drug Administration
Authors
Jason M. Samuels
Douglas M. Overbey
Krzysztof J. Wikiel
Teresa S. Jones
Thomas N. Robinson
Edward L. Jones
Publication date
01-07-2021
Publisher
Springer US
Published in
Surgical Endoscopy / Issue 7/2021
Print ISSN: 0930-2794
Electronic ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-020-07872-2

Other articles of this Issue 7/2021

Surgical Endoscopy 7/2021 Go to the issue