01-12-2024 | Electroencephalography | Original Article

Boltless nylon-suture technique for stereotactic electroencephalography as a safe, effective alternative when the anchor bolt is inappropriate

Authors: Manabu Mutoh, Satoshi Maesawa, Daisuke Nakatsubo, Tomotaka Ishizaki, Takafumi Tanei, Jun Torii, Yoshiki Ito, Miki Hashida, Ryuta Saito

Published in: Acta Neurochirurgica | Issue 1/2024

Login to get access

Abstract

Background

The use of anchor bolts to secure electrodes to the skull can be difficult in some clinical situations. Herein, we present the boltless technique to secure electrodes to the scalp using nylon sutures to overcome the problems associated with anchor bolts. We investigated the safety, accuracy errors, and patient-related and operative factors affecting errors in the boltless technique.

Methods

This single-institution retrospective series analyzed 103 electrodes placed in 12 patients. The target-point localization error (TPLE), entry-point localization error (EPLE), radial error (RE), and depth error (DE) of the electrodes were calculated.

Results

The median of the mean operative time per electrode was 9.3 min. The median TPLE, EPLE, RE, and absolute DE value were 4.1 mm, 1.6 mm, 2.7 mm, and 1.9 mm, respectively. Positive correlations were observed between the preoperative scalp thickness, mean operative time per electrode, EPLE, RE, and the absolute value of DE versus TPLE (r = .228, p = .02; r = .678, p = .015; r = .228, p = .02; r = .445, p < .01; r = .630, p < .01, respectively), and electrode approach angle versus EPLE (r = .213, p = .031). Multivariate analysis revealed that the absolute value of DE had the strongest influence on the TPLE, followed by RE and preoperative scalp thickness, respectively (β = .938, .544, .060, respectively, p < .001). No complications related to SEEG insertion and monitoring were encountered.

Conclusion

The boltless technique using our unique planning and technical method is a safe, effective, and low-cost alternative in cases where anchor bolts are contraindicated.
Literature
1.
go back to reference Abel JT, Muthiah N, Hect JL, Gonzalez-Martinez J, Salehi A, Smyth MD, Smith KJ (2022) Cost-effectiveness of invasive monitoring strategies in epilepsy surgery. J Neurosurg 1:1–7CrossRef Abel JT, Muthiah N, Hect JL, Gonzalez-Martinez J, Salehi A, Smyth MD, Smith KJ (2022) Cost-effectiveness of invasive monitoring strategies in epilepsy surgery. J Neurosurg 1:1–7CrossRef
2.
go back to reference Arnulfo G, Narizzano M, Cardinale F, Fato MM, Palva MJ (2015) Automatic segmentation of deep intracerebral electrodes in computed tomography scans. BMC Bioinformatics 16:99CrossRefPubMedPubMedCentral Arnulfo G, Narizzano M, Cardinale F, Fato MM, Palva MJ (2015) Automatic segmentation of deep intracerebral electrodes in computed tomography scans. BMC Bioinformatics 16:99CrossRefPubMedPubMedCentral
3.
go back to reference Beghi E, Giussani G, Nichols E et al (2019) Global, regional, and national burden of epilepsy, 1990–2016: a systematic analysis for the Global Burden of Disease Study 2016. Lancet Neurol 18:357–375CrossRef Beghi E, Giussani G, Nichols E et al (2019) Global, regional, and national burden of epilepsy, 1990–2016: a systematic analysis for the Global Burden of Disease Study 2016. Lancet Neurol 18:357–375CrossRef
4.
go back to reference Bourdillon P, Châtillon EC, Moles A, Rheims S, Catenoix H, Montavont A, Ostrowsky-Coste K, Boulogne S, Isnard J, Guénot M (2019) Effective accuracy of stereoelectroencephalography: robotic 3D versus Talairach orthogonal approaches. J Neurosurg 131:1938–1946CrossRef Bourdillon P, Châtillon EC, Moles A, Rheims S, Catenoix H, Montavont A, Ostrowsky-Coste K, Boulogne S, Isnard J, Guénot M (2019) Effective accuracy of stereoelectroencephalography: robotic 3D versus Talairach orthogonal approaches. J Neurosurg 131:1938–1946CrossRef
5.
go back to reference Cardinale F, Cossu M, Castana L, Casaceli G, Schiariti MP, Miserocchi A, Fuschillo D, Moscato A, Caborni C, Arnulfo G, Lo Russo G (2013) Stereoelectroencephalography: surgical methodology, safety, and stereotactic application accuracy in 500 procedures. Neurosurgery 72:353–366CrossRefPubMed Cardinale F, Cossu M, Castana L, Casaceli G, Schiariti MP, Miserocchi A, Fuschillo D, Moscato A, Caborni C, Arnulfo G, Lo Russo G (2013) Stereoelectroencephalography: surgical methodology, safety, and stereotactic application accuracy in 500 procedures. Neurosurgery 72:353–366CrossRefPubMed
6.
go back to reference Fiest MK, Sauro MK, Wiebe S, Patten SB, Kwon CS, Dykeman J, Pringsheim T, Lorenzetti DL, Jetté N (2017) Prevalence and incidence of epilepsy: a systematic review and meta-analysis of international studies. Neurology 88:296–303CrossRefPubMedPubMedCentral Fiest MK, Sauro MK, Wiebe S, Patten SB, Kwon CS, Dykeman J, Pringsheim T, Lorenzetti DL, Jetté N (2017) Prevalence and incidence of epilepsy: a systematic review and meta-analysis of international studies. Neurology 88:296–303CrossRefPubMedPubMedCentral
7.
go back to reference Garcia-Lorenzo B, Pino-Sedeño DT, Rocamora R, López EJ, Serrano-Aguilar P, Trujillo-Martín MM (2019) Stereoelectroencephalography for refractory epileptic patients considered for surgery: systematic review, meta-analysis, and economic evaluation. Neurosurgery 84:326–338CrossRefPubMed Garcia-Lorenzo B, Pino-Sedeño DT, Rocamora R, López EJ, Serrano-Aguilar P, Trujillo-Martín MM (2019) Stereoelectroencephalography for refractory epileptic patients considered for surgery: systematic review, meta-analysis, and economic evaluation. Neurosurgery 84:326–338CrossRefPubMed
8.
go back to reference Ho LA, Muftuoglu Y, Pendharkar VA, Sussman ES, Porter BE, Halpern CH, Grant GA (2018) Robot-guided pediatric stereoelectroencephalography: single-institution experience. J Neurosurg Pediatr 22:489–496CrossRef Ho LA, Muftuoglu Y, Pendharkar VA, Sussman ES, Porter BE, Halpern CH, Grant GA (2018) Robot-guided pediatric stereoelectroencephalography: single-institution experience. J Neurosurg Pediatr 22:489–496CrossRef
9.
go back to reference Iordanou CJ, Camara D, Ghatan S, Panov F (2019) Approach angle affects accuracy in robotic stereoelectroencephalography lead placement. World Neurosurg 128:e322–e228CrossRefPubMed Iordanou CJ, Camara D, Ghatan S, Panov F (2019) Approach angle affects accuracy in robotic stereoelectroencephalography lead placement. World Neurosurg 128:e322–e228CrossRefPubMed
10.
go back to reference Kalilani L, Sun X, Pelgrims B, Noack-Rink M, Villanueva V (2018) The epidemiology of drug-resistant epilepsy: a systematic review and meta-analysis. Epilepsia 59:2179–2193CrossRefPubMed Kalilani L, Sun X, Pelgrims B, Noack-Rink M, Villanueva V (2018) The epidemiology of drug-resistant epilepsy: a systematic review and meta-analysis. Epilepsia 59:2179–2193CrossRefPubMed
11.
go back to reference Karsonovich T, Alexander A, Graber S, O’Neill RB (2021) Placement of leads for stereotactic electroencephalography without the use of anchor bolts: technical note. J Neurosurg Pediatr 27:253–258CrossRef Karsonovich T, Alexander A, Graber S, O’Neill RB (2021) Placement of leads for stereotactic electroencephalography without the use of anchor bolts: technical note. J Neurosurg Pediatr 27:253–258CrossRef
12.
go back to reference Kim HL, Feng YA, Ho LA, Parker JJ, Kumar KK, Chen KS, Grant GA, Henderson JM, Halpern CH (2020) Robot-assisted versus manual navigated stereoelectroencephalography in adult medically-refractory epilepsy patients. Epilepsy Res 159:106253CrossRefPubMed Kim HL, Feng YA, Ho LA, Parker JJ, Kumar KK, Chen KS, Grant GA, Henderson JM, Halpern CH (2020) Robot-assisted versus manual navigated stereoelectroencephalography in adult medically-refractory epilepsy patients. Epilepsy Res 159:106253CrossRefPubMed
13.
go back to reference Lu C, Chen S, An Y, Meng F, Wang Y, Wei P, Fan X, Shan Y, Zhao G (2021) How can the accuracy of SEEG be increased?—an analysis of the accuracy of multilobe-spanning SEEG electrodes based on a frameless stereotactic robot-assisted system. Ann Palliat Med 10:3699–3705CrossRefPubMed Lu C, Chen S, An Y, Meng F, Wang Y, Wei P, Fan X, Shan Y, Zhao G (2021) How can the accuracy of SEEG be increased?—an analysis of the accuracy of multilobe-spanning SEEG electrodes based on a frameless stereotactic robot-assisted system. Ann Palliat Med 10:3699–3705CrossRefPubMed
14.
go back to reference Martinez GJ, Mullin J, Vadera S, Bulacio J, Hughes G, Jones S, Enatsu R, Najm I (2014) Stereotactic placement of depth electrodes in medically intractable epilepsy. J Neurosurg 120:639–644CrossRef Martinez GJ, Mullin J, Vadera S, Bulacio J, Hughes G, Jones S, Enatsu R, Najm I (2014) Stereotactic placement of depth electrodes in medically intractable epilepsy. J Neurosurg 120:639–644CrossRef
15.
go back to reference Matias MC, Kandregula S, Wu C, Sharan DA (2019) Accuracy of SEEG electrodes implanted in lateral position versus supine position. Clin Neurosurg 66:1CrossRef Matias MC, Kandregula S, Wu C, Sharan DA (2019) Accuracy of SEEG electrodes implanted in lateral position versus supine position. Clin Neurosurg 66:1CrossRef
16.
go back to reference Minotti L, Montavont A, Scholly J, Tyvaert L, Taussig D (2018) Indications and limits of stereoelectroencephalography (SEEG). Neurophysiol Clin 48:15–24CrossRefPubMed Minotti L, Montavont A, Scholly J, Tyvaert L, Taussig D (2018) Indications and limits of stereoelectroencephalography (SEEG). Neurophysiol Clin 48:15–24CrossRefPubMed
17.
go back to reference Mullin PJ, Shriver M, Alomar S, Najm I, Bulacio J, Chauvel P, Gonzalez-Martinez J (2016) Is SEEG safe? A systematic review and meta-analysis of stereo-electroencephalography–related complications. Epilepsia 57:386–401CrossRefPubMed Mullin PJ, Shriver M, Alomar S, Najm I, Bulacio J, Chauvel P, Gonzalez-Martinez J (2016) Is SEEG safe? A systematic review and meta-analysis of stereo-electroencephalography–related complications. Epilepsia 57:386–401CrossRefPubMed
18.
go back to reference Salehi A, Yang HP, Smyth DM (2022) Single-center cost comparison analysis of stereoelectroencephalography with subdural grid and strip implantation. J Neurosurg Pediatr 29:568–574CrossRefPubMed Salehi A, Yang HP, Smyth DM (2022) Single-center cost comparison analysis of stereoelectroencephalography with subdural grid and strip implantation. J Neurosurg Pediatr 29:568–574CrossRefPubMed
19.
go back to reference Talairach J, Bancaud J, Bonis A, Szikla G, Tournoux P (1962) Functional stereotaxic exploration of epilepsy. Confin Neurol 22:328–331CrossRefPubMed Talairach J, Bancaud J, Bonis A, Szikla G, Tournoux P (1962) Functional stereotaxic exploration of epilepsy. Confin Neurol 22:328–331CrossRefPubMed
20.
go back to reference Tandon N, Tong AB, Friedman RE, Johnson JA, Von Allmen G, Thomas MS, Hope OA, Kalamangalam GP, Slater JD, Thompson SA (2019) Analysis of morbidity and outcomes associated with use of subdural grids vs stereoelectroencephalography in patients with intractable epilepsy. JAMA Neurol 76:672–681CrossRefPubMedPubMedCentral Tandon N, Tong AB, Friedman RE, Johnson JA, Von Allmen G, Thomas MS, Hope OA, Kalamangalam GP, Slater JD, Thompson SA (2019) Analysis of morbidity and outcomes associated with use of subdural grids vs stereoelectroencephalography in patients with intractable epilepsy. JAMA Neurol 76:672–681CrossRefPubMedPubMedCentral
21.
go back to reference Tay SA, Menaker AS, Chan LJ, Mamelak AN (2022) Placement of stereotactic electroencephalography depth electrodes using the stealth autoguide robotic system: technical methods and initial results. Oper Neurosurg 22:E150–E157CrossRef Tay SA, Menaker AS, Chan LJ, Mamelak AN (2022) Placement of stereotactic electroencephalography depth electrodes using the stealth autoguide robotic system: technical methods and initial results. Oper Neurosurg 22:E150–E157CrossRef
22.
go back to reference Vakharia NV, Rodionov R, Miserocchi A, McEvoy AW, O’Keeffe A, Granados A, Shapoori S, Sparks R, Ourselin S, Duncan JS (2021) Comparison of robotic and manual implantation of intracerebral electrodes: a single centre, single blinded, randomized controlled trial. Sci Rep 11:17127CrossRefPubMedPubMedCentral Vakharia NV, Rodionov R, Miserocchi A, McEvoy AW, O’Keeffe A, Granados A, Shapoori S, Sparks R, Ourselin S, Duncan JS (2021) Comparison of robotic and manual implantation of intracerebral electrodes: a single centre, single blinded, randomized controlled trial. Sci Rep 11:17127CrossRefPubMedPubMedCentral
23.
go back to reference Wiebe S, Blume TW, Girvin PJ, Eliasziw M, Effectiveness and Efficiency of Surgery for Temporal Lobe Epilepsy Study Group (2001) A randomized, controlled trial of surgery for temporal-lobe epilepsy. N Engl J Med 345:311–318CrossRefPubMed Wiebe S, Blume TW, Girvin PJ, Eliasziw M, Effectiveness and Efficiency of Surgery for Temporal Lobe Epilepsy Study Group (2001) A randomized, controlled trial of surgery for temporal-lobe epilepsy. N Engl J Med 345:311–318CrossRefPubMed
25.
go back to reference Yu H, Pistol C, Franklin R, Barborica A (2018) Clinical accuracy of customized stereotactic fixtures for stereoelectroencephalography. World Neurosurg 109:82–88CrossRefPubMed Yu H, Pistol C, Franklin R, Barborica A (2018) Clinical accuracy of customized stereotactic fixtures for stereoelectroencephalography. World Neurosurg 109:82–88CrossRefPubMed
26.
go back to reference Zheng J, Liu LY, Zhang D, Cui XH, Sang LX, Xie T, Li WL (2021) Robot-assisted versus stereotactic frame-based stereoelectroencephalography in medically refractory epilepsy. Clin Neurophysiol 51:111–119CrossRef Zheng J, Liu LY, Zhang D, Cui XH, Sang LX, Xie T, Li WL (2021) Robot-assisted versus stereotactic frame-based stereoelectroencephalography in medically refractory epilepsy. Clin Neurophysiol 51:111–119CrossRef
27.
go back to reference Zumsteg D, Wieser GH (2000) Presurgical evaluation: current role of invasive EEG. Epilepsia 41:S55–S60CrossRefPubMed Zumsteg D, Wieser GH (2000) Presurgical evaluation: current role of invasive EEG. Epilepsia 41:S55–S60CrossRefPubMed
Metadata
Title
Boltless nylon-suture technique for stereotactic electroencephalography as a safe, effective alternative when the anchor bolt is inappropriate
Authors
Manabu Mutoh
Satoshi Maesawa
Daisuke Nakatsubo
Tomotaka Ishizaki
Takafumi Tanei
Jun Torii
Yoshiki Ito
Miki Hashida
Ryuta Saito
Publication date
01-12-2024
Publisher
Springer Vienna
Published in
Acta Neurochirurgica / Issue 1/2024
Print ISSN: 0001-6268
Electronic ISSN: 0942-0940
DOI
https://doi.org/10.1007/s00701-024-05889-3

Other articles of this Issue 1/2024

Acta Neurochirurgica 1/2024 Go to the issue