Published in: Langenbeck's Archives of Surgery 3/2009

01-05-2009 | Original Article

Electro-physiological parameters of hepatic radiofrequency ablation—a comparison of an in vitro versus an in vivo porcine liver model

Authors: R. Wahba, C. Bangard, R. Kleinert, S. Rösgen, J.-H. Fischer, K.-J. Lackner, A. H. Hölscher, D. L. Stippel

Published in: Langenbeck's Archives of Surgery | Issue 3/2009

Login to get access

Abstract

Background

Radiofrequency ablation (RFA) is an inherent part of curative treatment within a multimodal therapy concept of malignant liver tumors. The biggest problem is the high rate of local recurrences in tumors with a diameter of more than 3 cm because of the high variability and poor reproducibility of the zone of ablation. No imaging modality facilitates monitoring during neither intraoperativ nor percutaneous RFA. This experimental study describes and compares an in vitro and in vivo porcine model by its electro-physiological parameters with the aim of monitoring RFA procedures.

Materials and methods

RFA was performed in a perfused in vitro porcine (one RFA per liver) and in vivo porcine model (24 animals) with three different RFA systems (Rita XL 5 cm, Rita XLi 7 cm, LeVeen 5 cm). In the in vivo model, percutaneous placement of the RFA device was guided by native and contrast-enhanced CT scan. The electro-physical parameters during RFA were online (in real time) recorded by a dedicated software. After the RFA, the livers were explanted, sliced, and measured according to the consensus technique.

Results

The delivered energy was in vivo versus in vitro: Rita XL 238 ± 135 kJ versus 135 ± 53 kJ (p = 0.247); Rita XLi 711 ± 180 kJ versus 159 ± 54 (p = 0.016) and with LeVeen 212 ± 71 kJ (in vivo). The LeVeen system was inconsistent in the in vitro model. This correlates to an energy consumption per ml of necrosis in vivo versus in vitro Rita XL of 8 ± 3 kJ/ml versus 6.4 ± 3.9 kJ/ml (p = 0.537), Rita XLi of 10 ± 6 kJ/ml versus 1.8 ± 0.2 kJ/ml (p = 0.016), and LeVeen of 14.0 ± 12 kJ/ml (in vivo). The volume of ablation was in vivo versus in vitro Rita XL 30 ± 10 ml versus 26 ± 17 ml (p = 0.329), Rita XLi 90 ± 58 ml versus 88 ± 21 ml (p = 0.905), and LeVeen 22 ± 11 ml versus 50 ± 12 ml (p = 0.04). The impedance during RFA were in vivo versus in vitro Rita XL 39 ± 4 Ω versus 50 ± 14 Ω (p < 0.247), Rita XLi 33 ± 5 Ω versus 61 ± 16 Ω (p = 0.016) and LeVeen 31 ± 2 Ω (in vivo).

Conclusion

The volume of ablation showed analogue data in vivo and in vitro. The delivered energy and energy consumption was in vivo up to five times (Rita XLi) higher than in vitro and the impedance in vivo was always lower than in vitro. These differences observed between in vivo and in vitro were more pronounced than previously described. Thus the use of an in vitro model for research of the RFA technique must be challenged. The large deployment of the Rita XLi was a problem for percutaneous positioning of the device without direct contact to liver surface or major vessels in 80-kg pigs and to a lesser extent in in vitro liver originating from 130- to 140-kg pigs. Modern RFA systems which generate large volume of tissue necrosis can therefore only be adequately tested in a porcine model with a liver weight of at least 1.5–2 kg. Alternatively, a bovine liver model (with a liver weight up to 10 kg) should be developed in the future.
Literature
2.
go back to reference Stippel DL, Brochhagen HG, Arenja M, Hunkemoller J, Holscher AH, Beckurts KT (2004) Variability of size and shape of necrosis induced by radiofrequency ablation in human livers: a volumetric evaluation. Ann Surg Oncol 11(4):420–425. doi:10.1245/ASO.2004.09.012 PubMedCrossRef Stippel DL, Brochhagen HG, Arenja M, Hunkemoller J, Holscher AH, Beckurts KT (2004) Variability of size and shape of necrosis induced by radiofrequency ablation in human livers: a volumetric evaluation. Ann Surg Oncol 11(4):420–425. doi:10.​1245/​ASO.​2004.​09.​012 PubMedCrossRef
3.
4.
go back to reference Lepetit-Coiffe M, Quesson B, Seror O, Dumont E, Le Bail B, Moonen CT, Trillaud H (2006) Real-time monitoring of radiofrequency ablation of rabbit liver by respiratory-gated quantitative temperature MRI. J Magn Reson Imaging 24(1):152–159. doi:10.1002/jmri.20605 PubMedCrossRef Lepetit-Coiffe M, Quesson B, Seror O, Dumont E, Le Bail B, Moonen CT, Trillaud H (2006) Real-time monitoring of radiofrequency ablation of rabbit liver by respiratory-gated quantitative temperature MRI. J Magn Reson Imaging 24(1):152–159. doi:10.​1002/​jmri.​20605 PubMedCrossRef
5.
go back to reference Seror O, Lepetit-Coiffe M, Quesson B, Trillaud H, Moonen CT (2006) Quantitative magnetic resonance temperature mapping for real-time monitoring of radiofrequency ablation of the liver: an ex vivo study. Eur Radiol 16(10):2265–2274. doi:10.1007/s00330-006-0210-9 PubMedCrossRef Seror O, Lepetit-Coiffe M, Quesson B, Trillaud H, Moonen CT (2006) Quantitative magnetic resonance temperature mapping for real-time monitoring of radiofrequency ablation of the liver: an ex vivo study. Eur Radiol 16(10):2265–2274. doi:10.​1007/​s00330-006-0210-9 PubMedCrossRef
6.
go back to reference Stippel DL, Bangard C, Prenzel K, Yavuzyasar S, Fischer JH, Holscher AH (2008) Which parameters are needed for targeting a multitined radiofrequency device—an approach to a simple algorithm. Langenbecks Arch Surg. doi:10.1007/s00423-088-0306-6 Stippel DL, Bangard C, Prenzel K, Yavuzyasar S, Fischer JH, Holscher AH (2008) Which parameters are needed for targeting a multitined radiofrequency device—an approach to a simple algorithm. Langenbecks Arch Surg. doi:10.​1007/​s00423-088-0306-6
7.
go back to reference Solazzo SA, Liu Z, Lobo SM, Ahmed M, Hines-Peralta AU, Lenkinski RE, Goldberg SN (2005) Radiofrequency ablation: importance of background tissue electrical conductivity—an agar phantom and computer modeling study. Radiology 236(2):495–502. doi:10.1148/radiol.2362040965 PubMedCrossRef Solazzo SA, Liu Z, Lobo SM, Ahmed M, Hines-Peralta AU, Lenkinski RE, Goldberg SN (2005) Radiofrequency ablation: importance of background tissue electrical conductivity—an agar phantom and computer modeling study. Radiology 236(2):495–502. doi:10.​1148/​radiol.​2362040965 PubMedCrossRef
8.
go back to reference Mulier S, Ni Y, Frich L, Burdio F, Denys AL, De Wispelaere JF, Dupas B, Habib N, Hoey M, Jansen MC et al (2007) Experimental and clinical radiofrequency ablation: proposal for standardized description of coagulation size and geometry. Ann Surg Oncol 14(4):1381–1396. doi:10.1245/s10434-006-9033-9 PubMedCrossRef Mulier S, Ni Y, Frich L, Burdio F, Denys AL, De Wispelaere JF, Dupas B, Habib N, Hoey M, Jansen MC et al (2007) Experimental and clinical radiofrequency ablation: proposal for standardized description of coagulation size and geometry. Ann Surg Oncol 14(4):1381–1396. doi:10.​1245/​s10434-006-9033-9 PubMedCrossRef
9.
go back to reference Ng KK, Lam CM, Poon RT, Shek TW, Yu WC, To JY, Wo YH, Lau CP, Tang TC, Ho DW et al (2005) Porcine liver: morphologic characteristics and cell viability at experimental radiofrequency ablation with internally cooled electrodes. Radiology 235(2):478–486. doi:10.1148/radiol.2352040425 PubMedCrossRef Ng KK, Lam CM, Poon RT, Shek TW, Yu WC, To JY, Wo YH, Lau CP, Tang TC, Ho DW et al (2005) Porcine liver: morphologic characteristics and cell viability at experimental radiofrequency ablation with internally cooled electrodes. Radiology 235(2):478–486. doi:10.​1148/​radiol.​2352040425 PubMedCrossRef
10.
go back to reference Grundmann RT, Hermanek P, Merkel S, Germer CT, Hauss J, Henne-Bruns D, Herfarth K, Hopt UT, Junginger T, Klar E et al (2008) Diagnosis and treatment of colorectal liver metastases—workflow. Zentralbl Chir 133(3):267–284. doi:10.1055/s-2008-1076796 PubMedCrossRef Grundmann RT, Hermanek P, Merkel S, Germer CT, Hauss J, Henne-Bruns D, Herfarth K, Hopt UT, Junginger T, Klar E et al (2008) Diagnosis and treatment of colorectal liver metastases—workflow. Zentralbl Chir 133(3):267–284. doi:10.​1055/​s-2008-1076796 PubMedCrossRef
11.
go back to reference Itoh T, Orba Y, Takei H, Ishida Y, Saitoh M, Nakamura H, Meguro T, Horita S, Fujita M, Nagashima K (2002) Immunohistochemical detection of hepatocellular carcinoma in the setting of ongoing necrosis after radiofrequency ablation. Mod Pathol 15(2):110–115. doi:10.1038/modpathol.3880501 PubMedCrossRef Itoh T, Orba Y, Takei H, Ishida Y, Saitoh M, Nakamura H, Meguro T, Horita S, Fujita M, Nagashima K (2002) Immunohistochemical detection of hepatocellular carcinoma in the setting of ongoing necrosis after radiofrequency ablation. Mod Pathol 15(2):110–115. doi:10.​1038/​modpathol.​3880501 PubMedCrossRef
12.
go back to reference Brieger J, Pereira PL, Trubenbach J, Schenk M, Krober SM, Schmidt D, Aube C, Claussen CD, Schick F (2003) In vivo efficiency of four commercial monopolar radiofrequency ablation systems: a comparative experimental study in pig liver. Invest Radiol 38(10):609–616. doi:10.1097/01.rli.0000077122.70885.94 PubMedCrossRef Brieger J, Pereira PL, Trubenbach J, Schenk M, Krober SM, Schmidt D, Aube C, Claussen CD, Schick F (2003) In vivo efficiency of four commercial monopolar radiofrequency ablation systems: a comparative experimental study in pig liver. Invest Radiol 38(10):609–616. doi:10.​1097/​01.​rli.​0000077122.​70885.​94 PubMedCrossRef
13.
go back to reference Germer CT, Buhr HJ, Isbert C (2005) Nonoperative ablation for liver metastases. Possibilities and limitations as a curative treatment. Chirurg 76(6):552–554, 556–563. doi:10.1007/s00104-005-1047-5 Germer CT, Buhr HJ, Isbert C (2005) Nonoperative ablation for liver metastases. Possibilities and limitations as a curative treatment. Chirurg 76(6):552–554, 556–563. doi:10.1007/s00104-005-1047-5
14.
go back to reference Scott DJ, Fleming JB, Watumull LM, Lindberg G, Tesfay ST, Jones DB (2002) The effect of hepatic inflow occlusion on laparoscopic radiofrequency ablation using simulated tumors. Surg Endosc 16(9):1286–1291. doi:10.1007/s004640080167 PubMedCrossRef Scott DJ, Fleming JB, Watumull LM, Lindberg G, Tesfay ST, Jones DB (2002) The effect of hepatic inflow occlusion on laparoscopic radiofrequency ablation using simulated tumors. Surg Endosc 16(9):1286–1291. doi:10.​1007/​s004640080167 PubMedCrossRef
Metadata
Title
Electro-physiological parameters of hepatic radiofrequency ablation—a comparison of an in vitro versus an in vivo porcine liver model
Authors
R. Wahba
C. Bangard
R. Kleinert
S. Rösgen
J.-H. Fischer
K.-J. Lackner
A. H. Hölscher
D. L. Stippel
Publication date
01-05-2009
Publisher
Springer-Verlag
Published in
Langenbeck's Archives of Surgery / Issue 3/2009
Print ISSN: 1435-2443
Electronic ISSN: 1435-2451
DOI
https://doi.org/10.1007/s00423-009-0475-y

Other articles of this Issue 3/2009

Langenbeck's Archives of Surgery 3/2009 Go to the issue

Current Concepts in Endocrine Surgery

Neck lesions mimicking thyroid pathology