Effects of zero reference position on bladder pressure measurements: an observational study

Authors: Caridad De Dios Soler Morejón, Tomás Ariel Lombardo, Teddy Osmin Tamargo Barbeito, Barquín García Sandra

Published in: Annals of Intensive Care | Special Issue 1/2012

Login to get access

Abstract

Background

Although the World Society for Abdominal Compartment Syndrome in its guidelines recommends midaxillary line (MAL) as zero reference level in intra-abdominal pressure (IAP) measurements in aiming at standardizing the technique, evidence supporting this suggestion is scarce. The aim of this study is to study if the zero reference position influences bladder pressure measurements as estimate for IAP.

Methods

The IAP of 100 surgical patients was measured during the first 24 h of admission to the surgical intensive care unit of General Calixto Garcia Hospital in Havana (Cuba) following laparotomy. The period was January 2009 to January 2010. The IAP was measured twice with a six-hour interval using the transurethral technique with a priming volume of 25 ml. IAP was first measured with the zero reference level placed at MAL (IAPMAL), followed by a second measurement at the level of the symphysis pubis (SP) after 3 minutes (IAPSP). Correlations were made between IAP and body mass index (BMI), type of surgery, gender, and age.

Results

Mean IAPMAL was 8.5 ± 2.8 mmHg vs. IAPSP 6.5 ± 2.8 mmHg (p < 0.0001). The bias between measurements was 2.0 ± 1.5, 95% confidence interval of 1.4 to 3.0, upper limit of 4.9, lower limit of -0.9, and a percentage error of 35.1%. IAPMAL was consistently higher than IAPSP regardless of the type of surgery. The BMI correlated with IAP values regardless of the zero reference level (R2 = 0.4 and 0.3 with IAPMAL and IAPSP respectively, p < 0.0001).

Conclusions

The zero reference level has an important impact on IAP measurement in surgical patients after laparotomy and can potentially lead to over or underestimation. Further anthropometric studies are needed with regard to the relative MAL and SP zero reference position in relation to the theoretical ideal reference level at midpoint of the abdomen. Until better evidence is available, MAL remains the recommended zero reference position due to its best anatomical localization at iliac crest.
Literature
1.
go back to reference Malbrain ML: Different techniques to measure intra-abdominal pressure (IAP): time for a critical re-appraisal. Intensive Care Med 2004, 30: 357–371. 10.1007/s00134-003-2107-2PubMedCrossRef Malbrain ML: Different techniques to measure intra-abdominal pressure (IAP): time for a critical re-appraisal. Intensive Care Med 2004, 30: 357–371. 10.1007/s00134-003-2107-2PubMedCrossRef
3.
go back to reference Waele JJ, Cheatham ML, Malbrain MLNG, Kirkpatrick A, Sugrue M, Balogh Z, Ivatury R, De Keulenaer B, Kimball EJ: Recommendations for research from the International Conference of Experts on Intra-abdominal Hypertension and Abdominal Compartment Syndrome. Acta Clinica Belgica 2009,64(3):203–209.PubMedCrossRef Waele JJ, Cheatham ML, Malbrain MLNG, Kirkpatrick A, Sugrue M, Balogh Z, Ivatury R, De Keulenaer B, Kimball EJ: Recommendations for research from the International Conference of Experts on Intra-abdominal Hypertension and Abdominal Compartment Syndrome. Acta Clinica Belgica 2009,64(3):203–209.PubMedCrossRef
4.
go back to reference Malbrain M, Jones F: Intra-abdominal pressure measurement techniques. In Abdominal compartment syndrome. Edited by: Ivatury R, Cheatham M, Malbrain M, Sugrue M. Georgetown: Landes Bioscience; 2006:19–68. Malbrain M, Jones F: Intra-abdominal pressure measurement techniques. In Abdominal compartment syndrome. Edited by: Ivatury R, Cheatham M, Malbrain M, Sugrue M. Georgetown: Landes Bioscience; 2006:19–68.
5.
go back to reference Malbrain ML, Cheatham M, Kirkpatrick A, Sugrue M, Parr M, de Waele J, Balogh Z, Leppäniemi A, Olvera C, Ivatury R, D'Amours S, Wendon J, Hillman K, Johansson K, Kolkman K, Wilmer A: Results from the International Conference of Experts on Intra-abdominal Hypertension and Abdominal Compartment Syndrome. I Definitions. Intensive Care Med 2006, 32: 1722–1732. 10.1007/s00134-006-0349-5PubMedCrossRef Malbrain ML, Cheatham M, Kirkpatrick A, Sugrue M, Parr M, de Waele J, Balogh Z, Leppäniemi A, Olvera C, Ivatury R, D'Amours S, Wendon J, Hillman K, Johansson K, Kolkman K, Wilmer A: Results from the International Conference of Experts on Intra-abdominal Hypertension and Abdominal Compartment Syndrome. I Definitions. Intensive Care Med 2006, 32: 1722–1732. 10.1007/s00134-006-0349-5PubMedCrossRef
6.
go back to reference Cheatham M, Malbrain M, Kirkpatrick A, Sugrue M, Parr M, de Waele J, Balogh Z, Leppäniemi A, Olvera C, Ivatury R, D'Amours S, Wendon J, Hillman K, Wilmer A: Results from the International Conference of Experts on Intra-abdominal Hypertension and Abdominal Compartment Syndrome. II Recommendations. Intensive Care Med 2007,33(6):951–962. 10.1007/s00134-007-0592-4PubMedCrossRef Cheatham M, Malbrain M, Kirkpatrick A, Sugrue M, Parr M, de Waele J, Balogh Z, Leppäniemi A, Olvera C, Ivatury R, D'Amours S, Wendon J, Hillman K, Wilmer A: Results from the International Conference of Experts on Intra-abdominal Hypertension and Abdominal Compartment Syndrome. II Recommendations. Intensive Care Med 2007,33(6):951–962. 10.1007/s00134-007-0592-4PubMedCrossRef
9.
go back to reference Kron IL, Harman PK, Nolan SP: The measurement of intraabdominal pressure as a criterion for abdominal re-exploration. Ann Surg 1984,199(1):28–30. 10.1097/00000658-198401000-00005PubMedCentralPubMedCrossRef Kron IL, Harman PK, Nolan SP: The measurement of intraabdominal pressure as a criterion for abdominal re-exploration. Ann Surg 1984,199(1):28–30. 10.1097/00000658-198401000-00005PubMedCentralPubMedCrossRef
10.
go back to reference Balogh Z, De Waele JJ, Malbrain ML: Continuous intra-abdominal pressure monitoring. Acta Clin Belg Suppl 2007,62(1):26–32.CrossRef Balogh Z, De Waele JJ, Malbrain ML: Continuous intra-abdominal pressure monitoring. Acta Clin Belg Suppl 2007,62(1):26–32.CrossRef
11.
go back to reference Cheatham ML, Safcsak K: Intra-abdominal pressure: a revised method for measurement. J A M Coll Surg 1998,186(5):594–595. 10.1016/S1072-7515(98)00122-7CrossRef Cheatham ML, Safcsak K: Intra-abdominal pressure: a revised method for measurement. J A M Coll Surg 1998,186(5):594–595. 10.1016/S1072-7515(98)00122-7CrossRef
12.
go back to reference De Keulenaer BL, De Waele JJ, Powell B, Malbrain ML: What is normal intra-abdominal pressure and how is it affected by positioning, body mass and positive end-expiratory pressure? Intensive Care Med 2009,35(6):969–976. 10.1007/s00134-009-1445-0PubMedCrossRef De Keulenaer BL, De Waele JJ, Powell B, Malbrain ML: What is normal intra-abdominal pressure and how is it affected by positioning, body mass and positive end-expiratory pressure? Intensive Care Med 2009,35(6):969–976. 10.1007/s00134-009-1445-0PubMedCrossRef
13.
go back to reference De Waele J, De Laet I, De Keulenaer B, Widder S, Kirkpatrick A, Cresswell A, Malbrain M, Bodnar Z, Mejia-Mantilla J, Reis R, Parr M, Schulze R, Compano S, Cheatham M: The effect of different reference transducer positions on intra-abdominal pressure measurement: a multicenter analysis. Intensive Care Med 2008, 34: 1299–1303. 10.1007/s00134-008-1098-4PubMedCrossRef De Waele J, De Laet I, De Keulenaer B, Widder S, Kirkpatrick A, Cresswell A, Malbrain M, Bodnar Z, Mejia-Mantilla J, Reis R, Parr M, Schulze R, Compano S, Cheatham M: The effect of different reference transducer positions on intra-abdominal pressure measurement: a multicenter analysis. Intensive Care Med 2008, 34: 1299–1303. 10.1007/s00134-008-1098-4PubMedCrossRef
14.
go back to reference Council for International Organizations of Medical Sciences (CIOMS): International Ethical Guidelines for Biomedical Research Involving Human Subjects: 2002 February/March, Geneva. CIOMS: WHO; 2002. Council for International Organizations of Medical Sciences (CIOMS): International Ethical Guidelines for Biomedical Research Involving Human Subjects: 2002 February/March, Geneva. CIOMS: WHO; 2002.
15.
go back to reference World Medical Association Declaration of Helsinki: Ethical Principles for Medical Research Involving Human Subject 59th WMA General Assembly: 2008 October; Seoul. World Medical Association; 2008. World Medical Association Declaration of Helsinki: Ethical Principles for Medical Research Involving Human Subject 59th WMA General Assembly: 2008 October; Seoul. World Medical Association; 2008.
16.
go back to reference Malbrain ML, Chiumello D, Pelosi P, Wilmer A, Brienza N, Malcangi V, Bihari D, Innes R, Cohen J, Singer P, Japiassu A, Kurtop E, De Keulenaer BL, Daelemans R, Del Turco M, Cosimini P, Ranieri M, Jacquet L, Laterre PF, Gattinoni L: Prevalence of intra-abdominal hypertension in critically ill patients: a multicentre epidemiological study. Intensive Care Med 2004, 30: 822–829. 10.1007/s00134-004-2169-9PubMedCrossRef Malbrain ML, Chiumello D, Pelosi P, Wilmer A, Brienza N, Malcangi V, Bihari D, Innes R, Cohen J, Singer P, Japiassu A, Kurtop E, De Keulenaer BL, Daelemans R, Del Turco M, Cosimini P, Ranieri M, Jacquet L, Laterre PF, Gattinoni L: Prevalence of intra-abdominal hypertension in critically ill patients: a multicentre epidemiological study. Intensive Care Med 2004, 30: 822–829. 10.1007/s00134-004-2169-9PubMedCrossRef
17.
go back to reference Malbrain M, Chiumello D, Pelosi P, Bihari D, Innes R, Ranieri M, Del Turco M, Wilmer A, Brienza N, Malcangi V, Cohen J, Japiassu A, De Keulenaer B, Daelemans R, Jacquet L, Laterre PF, Frank G, de Souza P, Cesana B, Gattinoni L: Incidence and prognosis of intraabdominal hypertension in a mixed population of critically ill patients: a multiple-center epidemiological study. Crit Care Med 2005, 33: 315–322. 10.1097/01.CCM.0000153408.09806.1BPubMedCrossRef Malbrain M, Chiumello D, Pelosi P, Bihari D, Innes R, Ranieri M, Del Turco M, Wilmer A, Brienza N, Malcangi V, Cohen J, Japiassu A, De Keulenaer B, Daelemans R, Jacquet L, Laterre PF, Frank G, de Souza P, Cesana B, Gattinoni L: Incidence and prognosis of intraabdominal hypertension in a mixed population of critically ill patients: a multiple-center epidemiological study. Crit Care Med 2005, 33: 315–322. 10.1097/01.CCM.0000153408.09806.1BPubMedCrossRef
19.
go back to reference Pita Fernández S, Pértega Díaz S, Rodríguez Maseda E: La fiabilidad de las mediciones clínicas: el análisis de concordancia para variables numéricas. Cad Aten Primaria 2003,10(4):290–96. Pita Fernández S, Pértega Díaz S, Rodríguez Maseda E: La fiabilidad de las mediciones clínicas: el análisis de concordancia para variables numéricas. Cad Aten Primaria 2003,10(4):290–96.
20.
go back to reference De Waele JJ, De Laet I, Malbrain ML: Rational intraabdominal pressure monitoring: how to do it? Acta Clin Belg Suppl 2007,62(1):16–25. [20]CrossRef De Waele JJ, De Laet I, Malbrain ML: Rational intraabdominal pressure monitoring: how to do it? Acta Clin Belg Suppl 2007,62(1):16–25. [20]CrossRef
21.
go back to reference Malbrain ML, De Laet I: Do we need to know body anthropomorphic data whilst measuring abdominal pressure? Intensive Care Med 2010,36(1):180–182. 10.1007/s00134-009-1685-zPubMedCrossRef Malbrain ML, De Laet I: Do we need to know body anthropomorphic data whilst measuring abdominal pressure? Intensive Care Med 2010,36(1):180–182. 10.1007/s00134-009-1685-zPubMedCrossRef
22.
go back to reference Sánchez NC, Tenofsky PL, Dort JM: What is normal intra-abdominal pressure? Am Surg 2001,67(3):243–248.PubMed Sánchez NC, Tenofsky PL, Dort JM: What is normal intra-abdominal pressure? Am Surg 2001,67(3):243–248.PubMed
23.
go back to reference Noblett KL, Jensen JK, Ostergard DR: The relationship of body mass index to intra-abdominal pressure as measured by multichannel cystometry. Int Urogynecol J Pelvis Flood Dysfunct 1997,8(6):323–326. 10.1007/BF02765589CrossRef Noblett KL, Jensen JK, Ostergard DR: The relationship of body mass index to intra-abdominal pressure as measured by multichannel cystometry. Int Urogynecol J Pelvis Flood Dysfunct 1997,8(6):323–326. 10.1007/BF02765589CrossRef
24.
go back to reference De Keulenaer BL, Cheatham ML, De Waele JJ, Kimball EJ, Powell B, Davis WA, Jenkins IR: Intra-abdominal pressure measurements in lateral decubitus versus supine position. Acta Clinica Belgica 2009,64(3):210–215.PubMedCrossRef De Keulenaer BL, Cheatham ML, De Waele JJ, Kimball EJ, Powell B, Davis WA, Jenkins IR: Intra-abdominal pressure measurements in lateral decubitus versus supine position. Acta Clinica Belgica 2009,64(3):210–215.PubMedCrossRef
Metadata
Title
Effects of zero reference position on bladder pressure measurements: an observational study
Authors
Caridad De Dios Soler Morejón
Tomás Ariel Lombardo
Teddy Osmin Tamargo Barbeito
Barquín García Sandra
Publication date
01-12-2012
Publisher
Springer Paris
Published in
Annals of Intensive Care / Issue Special Issue 1/2012
Electronic ISSN: 2110-5820
DOI
https://doi.org/10.1186/2110-5820-2-S1-S13

Other articles of this Special Issue 1/2012

Annals of Intensive Care 1/2012 Go to the issue