Abstract

Background

Two sizes of endoscopic gastrointestinal anastomosis (Endo-GIA) staples are currently available for vascular control. The purpose of this study was to compare the Endo-GIA 30–2.0 mm with the Endo-GIA 30–2.5 mm vascular staple.

Methods

The renal blood vessels, aorta, and vena cava were divided laparoscopically using the Endo-GIA 30–2.0 and 30–2.5 mm in four pigs, respectively. The arterial and venous segments were harvested en bloc for ex vivo burst pressure studies.

Results

Both 2.0-mm and 2.5-mm staples achieved vessel sealing at supraphysiologic mean burst pressures (≥310 mmHg). There were no statistical differences between mean burst pressures of vessels stapled by 2.0-mm compared to 2.5-mm staples. Both staples sealed arteries up to 17 mm and veins up to 22 mm in diameter.

Conclusion

Endo-GIA 30–2.0 mm and 30–2.5 mm vascular staples are equivalent in sealing vessels at supraphysiologic mean burst pressures.
Literature
1.
go back to reference Chan, D, Bishoff, JT, Ratner, L, Kavoussi, LR, Jarrett, TW 2000Endovascular gastrointestinal stapler device malfunction during laparoscopic nephrectomy: early recognition and management.J Urol164319321CrossRef Chan, D, Bishoff, JT, Ratner, L, Kavoussi, LR, Jarrett, TW 2000Endovascular gastrointestinal stapler device malfunction during laparoscopic nephrectomy: early recognition and management.J Urol164319321CrossRef
2.
go back to reference Craig, SR, Walker, WS 1995Potential complications of vascular stapling in thoracoscopic pulmonary resection.Ann Thorac Surg59736738CrossRef Craig, SR, Walker, WS 1995Potential complications of vascular stapling in thoracoscopic pulmonary resection.Ann Thorac Surg59736738CrossRef
3.
go back to reference Hanash, KA, Peracha, AM, Al-Zahrani, HM, Merdad, TA, Hafeez Kardar, A, Aslam, M, Mohamed, GH 2000Radical cystectomy: minimizing operative blood loss with a “stapling technique.”Urology56488491CrossRef Hanash, KA, Peracha, AM, Al-Zahrani, HM, Merdad, TA, Hafeez Kardar, A, Aslam, M, Mohamed, GH 2000Radical cystectomy: minimizing operative blood loss with a “stapling technique.”Urology56488491CrossRef
4.
go back to reference Hsu, TH, Su, LM, Ratner, LE, Kavoussi, LR 2002Renovascular complications of laparoscopic donor nephrectomy.Urology60811815CrossRef Hsu, TH, Su, LM, Ratner, LE, Kavoussi, LR 2002Renovascular complications of laparoscopic donor nephrectomy.Urology60811815CrossRef
5.
go back to reference Kennedy, JS, Stranahan, PL, Taylor, KD, Chandler, JG 1998High-burst-strength, feedback-controlled bipolar vessel sealing.Surg Endosc12878878CrossRef Kennedy, JS, Stranahan, PL, Taylor, KD, Chandler, JG 1998High-burst-strength, feedback-controlled bipolar vessel sealing.Surg Endosc12878878CrossRef
6.
go back to reference Kerbl, K, Chandhoke, PS, Clayman, RV, McDougall, E, Stone, AM, Figenshau, RS 1993Ligation of the renal pedicle during laparoscopic nephrectomy: a comparison of staples, clips, and sutures.J Laparoendosc Surg3912CrossRef Kerbl, K, Chandhoke, PS, Clayman, RV, McDougall, E, Stone, AM, Figenshau, RS 1993Ligation of the renal pedicle during laparoscopic nephrectomy: a comparison of staples, clips, and sutures.J Laparoendosc Surg3912CrossRef
7.
go back to reference Landman, J, Kerbl, , K, , Rehman, J, Andreoni, C, Humphrey, PA, Collyer, W, Olweny, E, Sundaram, C, Clayman, RV 2003Evaluation of vessel sealing system, bipolar electrosurgery, harmonic scalpel, titanium clips, endoscopic gastrointestinal anastomosis vascular staples and sutures for arterial and venous ligation in a porcine model.J Urol169697700CrossRef Landman, J, Kerbl, , K, , Rehman, J, Andreoni, C, Humphrey, PA, Collyer, W, Olweny, E, Sundaram, C, Clayman, RV 2003Evaluation of vessel sealing system, bipolar electrosurgery, harmonic scalpel, titanium clips, endoscopic gastrointestinal anastomosis vascular staples and sutures for arterial and venous ligation in a porcine model.J Urol169697700CrossRef
8.
go back to reference Patsner, B 1998Radical abdominal hysterectomy using the Endo-GIA stapler: a report of 150 cases and literature review.Eur J Gynaecol Oncol19215219PubMed Patsner, B 1998Radical abdominal hysterectomy using the Endo-GIA stapler: a report of 150 cases and literature review.Eur J Gynaecol Oncol19215219PubMed
9.
go back to reference Sharp, HT, Dorsey, JH, Holtz, PM, Melick, CF 1996Electrocoagulation versus the Endo GIA in LAVH.J Am Assoc Gynecol Laparosc3S45S46CrossRef Sharp, HT, Dorsey, JH, Holtz, PM, Melick, CF 1996Electrocoagulation versus the Endo GIA in LAVH.J Am Assoc Gynecol Laparosc3S45S46CrossRef
Metadata
Title
Effect of Endo-GIA vascular staple size on laparoscopic vessel sealing in a porcine model
Authors
A. El-Hakim
Y. Cai
R. Marcovich
P. Pinto
B. R. Lee
Publication date
01-06-2004
Publisher
Springer US
Published in
Surgical Endoscopy / Issue 6/2004
Print ISSN: 0930-2794
Electronic ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-003-8546-6

Other articles of this Issue 6/2004

Surgical Endoscopy 6/2004 Go to the issue

Editorial Notes

News and notices