Effect of computer assistance on observer performance of approximal caries diagnosis using intraoral digital radiography

Authors: Kazuyuki Araki, Yukiko Matsuda, Kenji Seki, Tomohiro Okano

Published in: Clinical Oral Investigations | Issue 3/2010

Login to get access

Abstract

Logicon Caries Detector™ (LDDC) is the only commercially available computer-assisted diagnostic system for caries diagnosis. The object of this study is to elucidate the efficacy of LDDC when used by inexperienced dentists. Fifty extracted teeth were imaged using an RVG6000. Seven dentists who had just passed the Japanese National Dental Board Examination observed those images without LDDC (woLDDC) and assessed the probability that caries lesions were present, then re-assessed the same teeth using LDDC (wLDDC). The areas under the receiver operating characteristic curves (Az) were compared. No statistically significant difference was found between woLDDC Az values and wLDDC Az values when caries lesions of all depths were considered. When positive cases were restricted to caries lesions in the inner half of the enamel or to dentine caries lesions, however, wLDDC Az values were significantly larger than woLDDC (p = 0.043 and 0.018, respectively).
Literature
1.
go back to reference Schneiderman A, Elbaum M, Shultz T, Keem S, Greenebaum M, Driller J (1997) Assessment of dental caries with Digital Imaging Fiber-Optic TransIllumination (DIFOTI): in vitro study. Caries Res 31:103–110CrossRefPubMed Schneiderman A, Elbaum M, Shultz T, Keem S, Greenebaum M, Driller J (1997) Assessment of dental caries with Digital Imaging Fiber-Optic TransIllumination (DIFOTI): in vitro study. Caries Res 31:103–110CrossRefPubMed
2.
go back to reference Matalon S, Feuerstein O, Calderon S, Mittleman A, Kaffe I (2007) Detection of cavitated carious lesions in approximal tooth surfaces by ultrasonic caries detector. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 103:109–113CrossRefPubMed Matalon S, Feuerstein O, Calderon S, Mittleman A, Kaffe I (2007) Detection of cavitated carious lesions in approximal tooth surfaces by ultrasonic caries detector. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 103:109–113CrossRefPubMed
3.
go back to reference Peker I, Toraman Alkurt M, Altunkaynak B (2007) Film tomography compared with film and digital bitewing radiography for proximal caries detection. Dento-Maxillo-Facial Radiol 36:495–499 Peker I, Toraman Alkurt M, Altunkaynak B (2007) Film tomography compared with film and digital bitewing radiography for proximal caries detection. Dento-Maxillo-Facial Radiol 36:495–499
4.
go back to reference Kidd EA, Pitts NB (1990) A reappraisal of the value of the bitewing radiograph in the diagnosis of posterior approximal caries. Br Dent J 169:195–200PubMed Kidd EA, Pitts NB (1990) A reappraisal of the value of the bitewing radiograph in the diagnosis of posterior approximal caries. Br Dent J 169:195–200PubMed
5.
go back to reference Pitts NB, Rimmer PA (1992) An in vivo comparison of radiographic and directly assessed clinical caries status of posterior approximal surfaces in primary and permanent teeth. Caries Res 26:146–152CrossRefPubMed Pitts NB, Rimmer PA (1992) An in vivo comparison of radiographic and directly assessed clinical caries status of posterior approximal surfaces in primary and permanent teeth. Caries Res 26:146–152CrossRefPubMed
6.
go back to reference Pitts NB, Renson CE (1985) Reproducibility of computer-aided image-analysis-derived estimates of the depth and area of radiolucencies in approximal enamel. J Dent Res 64:1221–1224PubMed Pitts NB, Renson CE (1985) Reproducibility of computer-aided image-analysis-derived estimates of the depth and area of radiolucencies in approximal enamel. J Dent Res 64:1221–1224PubMed
7.
go back to reference Pitts NB, Renson CE (1986) Image analysis of bitewing radiographs: a histologically validated comparison with visual assessments of radiolucency depth in enamel. Br Dent J 160:205–209CrossRefPubMed Pitts NB, Renson CE (1986) Image analysis of bitewing radiographs: a histologically validated comparison with visual assessments of radiolucency depth in enamel. Br Dent J 160:205–209CrossRefPubMed
8.
go back to reference Pitts NB, Renson CE (1986) Further development of a computer-aided image analysis method of quantifying radiolucencies in approximal enamel. Caries Res 20:361–370CrossRefPubMed Pitts NB, Renson CE (1986) Further development of a computer-aided image analysis method of quantifying radiolucencies in approximal enamel. Caries Res 20:361–370CrossRefPubMed
9.
go back to reference Pitts NB (1987) Detection of approximal radiolucencies in enamel: a preliminary comparison between experienced clinicians and an image analysis method. J Dent 15:191–197CrossRefPubMed Pitts NB (1987) Detection of approximal radiolucencies in enamel: a preliminary comparison between experienced clinicians and an image analysis method. J Dent 15:191–197CrossRefPubMed
10.
go back to reference Heaven TJ, Firestone AR, Feagin FF (1992) Computer-based image analysis of natural approximal caries on radiographic films. J Dent Res 71:846–849 Spec NoPubMed Heaven TJ, Firestone AR, Feagin FF (1992) Computer-based image analysis of natural approximal caries on radiographic films. J Dent Res 71:846–849 Spec NoPubMed
11.
go back to reference Zubery Y, Dove SB, Ebersole J (1993) An in vitro study of the characteristics of a computer-aided radiographic evaluation (CARE) system for longitudinal assessment of density changes. J Periodontal Res 28:233–240CrossRefPubMed Zubery Y, Dove SB, Ebersole J (1993) An in vitro study of the characteristics of a computer-aided radiographic evaluation (CARE) system for longitudinal assessment of density changes. J Periodontal Res 28:233–240CrossRefPubMed
12.
go back to reference Sund T, Møystad A (2006) Sliding window adaptive histogram equalization of intraoral radiographs: effect on image quality. Dento-Maxillo-Facial Radiol 35:133–138 Sund T, Møystad A (2006) Sliding window adaptive histogram equalization of intraoral radiographs: effect on image quality. Dento-Maxillo-Facial Radiol 35:133–138
13.
go back to reference Maggio JJ, Hausmann EM, Allen K, Potts TV (1990) A model for dentinal caries progression by digital subtraction radiography. J Prosthet Dent 64:727–732CrossRefPubMed Maggio JJ, Hausmann EM, Allen K, Potts TV (1990) A model for dentinal caries progression by digital subtraction radiography. J Prosthet Dent 64:727–732CrossRefPubMed
14.
go back to reference Heaven TJ, Weems RA, Firestone AR (1994) The use of a computer-based image analysis program for the diagnosis of approximal caries from bitewing radiographs. Caries Res 28:55–58CrossRefPubMed Heaven TJ, Weems RA, Firestone AR (1994) The use of a computer-based image analysis program for the diagnosis of approximal caries from bitewing radiographs. Caries Res 28:55–58CrossRefPubMed
15.
go back to reference Duncan RC, Heaven T, Weems RA, Firestone AR, Greer DF, Patel JR (1995) Using computers to diagnose and plan treatment of approximal caries. Detected in radiographs. J Am Dent Assoc 126:873–882PubMed Duncan RC, Heaven T, Weems RA, Firestone AR, Greer DF, Patel JR (1995) Using computers to diagnose and plan treatment of approximal caries. Detected in radiographs. J Am Dent Assoc 126:873–882PubMed
16.
go back to reference Gakenheimer DC (2002) The efficacy of a computerized caries detector in intraoral digital radiography. J Am Dent Assoc 133:883–890PubMed Gakenheimer DC (2002) The efficacy of a computerized caries detector in intraoral digital radiography. J Am Dent Assoc 133:883–890PubMed
17.
go back to reference Schmidlin PR, Tepper SA, Scriba H, Lutz F (2002) In vitro assessment of incipient approximal carious lesions using computer-assisted densitometric image analysis. J Dent 30:305–311CrossRefPubMed Schmidlin PR, Tepper SA, Scriba H, Lutz F (2002) In vitro assessment of incipient approximal carious lesions using computer-assisted densitometric image analysis. J Dent 30:305–311CrossRefPubMed
18.
go back to reference Li S, Fevens T, Krzyzak A, Jin C, Li S (2005) Toward automatic computer aided dental X-ray analysis using level set method. Med Image Comput Comput Assist Interv 8:670–678CrossRefPubMed Li S, Fevens T, Krzyzak A, Jin C, Li S (2005) Toward automatic computer aided dental X-ray analysis using level set method. Med Image Comput Comput Assist Interv 8:670–678CrossRefPubMed
19.
go back to reference Kositbowornchai S, Siriteptawee S, Plermkamon S, Bureerat S, Chetchotsak D (2006) An artificial neural network for detection of simulated dental caries. Int J CARS 1:91–96CrossRef Kositbowornchai S, Siriteptawee S, Plermkamon S, Bureerat S, Chetchotsak D (2006) An artificial neural network for detection of simulated dental caries. Int J CARS 1:91–96CrossRef
20.
go back to reference Yoon DC, Wilensky GD, Neuhaus JA, Manukian N and Gakenheimer DC (1998) Quantitative dental caries detection system and method. US patent number5742700, Apr 21 Yoon DC, Wilensky GD, Neuhaus JA, Manukian N and Gakenheimer DC (1998) Quantitative dental caries detection system and method. US patent number5742700, Apr 21
21.
go back to reference Wenzel A, Hintze H, Kold LM, Kold S (2002) Accuracy of computer-automated caries detection in digital radiographs compared with human observers. Eur J Oral Sci 110:199–203CrossRefPubMed Wenzel A, Hintze H, Kold LM, Kold S (2002) Accuracy of computer-automated caries detection in digital radiographs compared with human observers. Eur J Oral Sci 110:199–203CrossRefPubMed
22.
go back to reference Navarro FL, Llena Puy MC, Godoy GF (2008) Diagnostic performance of radiovisiography in combination with a diagnosis assisting program versus conventional radiography and radiovisiography in basic mode and with magnification. Med Oral Patol Oral Cir Bucal 13:E261–E265 Navarro FL, Llena Puy MC, Godoy GF (2008) Diagnostic performance of radiovisiography in combination with a diagnosis assisting program versus conventional radiography and radiovisiography in basic mode and with magnification. Med Oral Patol Oral Cir Bucal 13:E261–E265
24.
go back to reference Wenzel A (2001) Computer-Automated Caries Detection in Digital Bitewings: Consistency of a Program and Its Influence on Observer Agreement. Oper Dent 35:12–20 Wenzel A (2001) Computer-Automated Caries Detection in Digital Bitewings: Consistency of a Program and Its Influence on Observer Agreement. Oper Dent 35:12–20
25.
go back to reference Harase Y, Araki K, Okano T (2006) Accuracy of extraoral tuned aperture computed tomography (TACT) for proximal caries detection. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 101:791–796CrossRefPubMed Harase Y, Araki K, Okano T (2006) Accuracy of extraoral tuned aperture computed tomography (TACT) for proximal caries detection. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 101:791–796CrossRefPubMed
26.
go back to reference Tsuchida R, Araki K, Okano T (2007) Evaluation of a limited cone-beam volumetric imaging system: comparison with film radiography in detecting incipient proximal caries. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 104:412–416CrossRefPubMed Tsuchida R, Araki K, Okano T (2007) Evaluation of a limited cone-beam volumetric imaging system: comparison with film radiography in detecting incipient proximal caries. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 104:412–416CrossRefPubMed
27.
go back to reference Hahn SK, Kim JW, Lee SH, Kim CC, Hahn SH, Jang KT (2004) Microcomputed tomographic assessment of chemomechanical caries removal. Caries Res 38:75–78CrossRefPubMed Hahn SK, Kim JW, Lee SH, Kim CC, Hahn SH, Jang KT (2004) Microcomputed tomographic assessment of chemomechanical caries removal. Caries Res 38:75–78CrossRefPubMed
28.
go back to reference Lee HS, Berg JH, Garcia-Godoy F, Jang KT (2008) Long-term evaluation of the remineralization of interproximal caries-like lesions adjacent to glass-ionomer restorations: a micro-CT study. Am J Dent 21:129–132PubMed Lee HS, Berg JH, Garcia-Godoy F, Jang KT (2008) Long-term evaluation of the remineralization of interproximal caries-like lesions adjacent to glass-ionomer restorations: a micro-CT study. Am J Dent 21:129–132PubMed
29.
go back to reference Khan EA, Tyndall DA, Ludlow JB, Caplan D (2005) Proximal caries detection: Sirona Sidexis versus Kodak Ektaspeed Plus. Gen Dent 53:43–48PubMed Khan EA, Tyndall DA, Ludlow JB, Caplan D (2005) Proximal caries detection: Sirona Sidexis versus Kodak Ektaspeed Plus. Gen Dent 53:43–48PubMed
30.
go back to reference Moystad A, Svanaes DB, Risnes S, Larheim TA, Grondahl HG (1996) Detection of approximal caries with a storage phosphor system. A comparison of enhanced digital images with dental X-ray film. Dento-Maxillo-Fac Radiol 25:202–206 Moystad A, Svanaes DB, Risnes S, Larheim TA, Grondahl HG (1996) Detection of approximal caries with a storage phosphor system. A comparison of enhanced digital images with dental X-ray film. Dento-Maxillo-Fac Radiol 25:202–206
31.
go back to reference Naitoh M, Yuasa H, Toyama M, Shiojima M, Nakamura M, Ushida M et al (1998) Observer agreement in the detection of proximal caries with direct digital intraoral radiography. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 85:107–112CrossRefPubMed Naitoh M, Yuasa H, Toyama M, Shiojima M, Nakamura M, Ushida M et al (1998) Observer agreement in the detection of proximal caries with direct digital intraoral radiography. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 85:107–112CrossRefPubMed
32.
go back to reference Syriopoulos K, Sanderink GC, Velders XL, van der Stelt PF (2000) Radiographic detection of approximal caries: a comparison of dental films and digital imaging systems. Dento-Maxillo-Fac Radiol 29:312–318 Syriopoulos K, Sanderink GC, Velders XL, van der Stelt PF (2000) Radiographic detection of approximal caries: a comparison of dental films and digital imaging systems. Dento-Maxillo-Fac Radiol 29:312–318
33.
go back to reference Tyndall DA, Ludlow JB, Platin E, Nair M (1998) A comparison of Kodak Ektaspeed Plus film and the Siemens Sidexis digital imaging system for caries detection using receiver operating characteristic analysis. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 85:113–118CrossRefPubMed Tyndall DA, Ludlow JB, Platin E, Nair M (1998) A comparison of Kodak Ektaspeed Plus film and the Siemens Sidexis digital imaging system for caries detection using receiver operating characteristic analysis. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 85:113–118CrossRefPubMed
34.
go back to reference Wenzel A (2004) Bitewing and digital bitewing radiography for detection of caries lesions. J Dent Res 83:C72–C75 Spec No CCrossRefPubMed Wenzel A (2004) Bitewing and digital bitewing radiography for detection of caries lesions. J Dent Res 83:C72–C75 Spec No CCrossRefPubMed
35.
go back to reference Pitts NB (2004) Are we ready to move from operative to non-operative/preventive treatment of dental caries in clinical practice? Caries Res 38:294–304CrossRefPubMed Pitts NB (2004) Are we ready to move from operative to non-operative/preventive treatment of dental caries in clinical practice? Caries Res 38:294–304CrossRefPubMed
36.
38.
go back to reference Faculty of GDP (1998) Selection criteria for dental radiography. Faculty of General Dental Practitioners The Royal college of Surgeons of England, London Faculty of GDP (1998) Selection criteria for dental radiography. Faculty of General Dental Practitioners The Royal college of Surgeons of England, London
39.
go back to reference Wenzel A, Haiter-Neto F, Gotfredsen E (2007) Risk factors for a false positive test outcome in diagnosis of caries in approximal surfaces: impact of radiographic modality and observer characteristics. Caries Res 41:170–176CrossRefPubMed Wenzel A, Haiter-Neto F, Gotfredsen E (2007) Risk factors for a false positive test outcome in diagnosis of caries in approximal surfaces: impact of radiographic modality and observer characteristics. Caries Res 41:170–176CrossRefPubMed
40.
go back to reference Mileman PA, van den Hout WB (2002) Comparing the accuracy of Dutch dentists and dental students in the radiographic diagnosis of dentinal caries. Dento-Maxillo-Facial Radiol 31:7–14 Mileman PA, van den Hout WB (2002) Comparing the accuracy of Dutch dentists and dental students in the radiographic diagnosis of dentinal caries. Dento-Maxillo-Facial Radiol 31:7–14
41.
go back to reference Schulze RK, Nackat D, D'hoedt B (2004) In vitro carious lesion detection on D-, E-, and F-speed radiographic films. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 97:529–534CrossRefPubMed Schulze RK, Nackat D, D'hoedt B (2004) In vitro carious lesion detection on D-, E-, and F-speed radiographic films. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 97:529–534CrossRefPubMed
Metadata
Title
Effect of computer assistance on observer performance of approximal caries diagnosis using intraoral digital radiography
Authors
Kazuyuki Araki
Yukiko Matsuda
Kenji Seki
Tomohiro Okano
Publication date
01-06-2010
Publisher
Springer-Verlag
Published in
Clinical Oral Investigations / Issue 3/2010
Print ISSN: 1432-6981
Electronic ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-009-0307-z

Other articles of this Issue 3/2010

Clinical Oral Investigations 3/2010 Go to the issue