Skip to main content
Top
Published in: Clinical Orthopaedics and Related Research® 10/2016

01-10-2016 | Symposium: 2015 International Hip Society Proceedings

Early Subsidence Predicts Failure of a Cemented Femoral Stem With Minor Design Changes

Authors: Per-Erik Johanson, MD, Martin Antonsson, MStat MScCSE, Bita Shareghi, BSc, Johan Kärrholm, MD, PhD

Published in: Clinical Orthopaedics and Related Research® | Issue 10/2016

Login to get access

Abstract

Background

Radiostereometry (RSA) measurements of early micromotion can predict later failure in hip and knee prostheses. In hip implants, RSA has been particularly helpful in the evaluation of composite-beam stem designs. The Spectron EF Primary stem (Smith & Nephew, London, UK) has shown inferior performance compared with its predecessors in both clinical studies and registry reports. Early RSA studies have shown somewhat greater subsidence for the Spectron EF Primary stem compared with the earlier Spectron EF, but still within boundaries considered to be safe.

Questions/purposes

Our primary research question was whether stem subsidence and rotation for this stem design measured with RSA at 2 years can predict later stem failure. A secondary question was whether high femoral stem offset and small stem sizes, both features specific to the Spectron EF Primary stem compared with its predecessors, are associated with stem failure rate.

Methods

Two hundred forty-seven hips (209 patients with median age 63 years [range, 29–80 years], 65% female, and 77% primary osteoarthritis) with a valid RSA examination at 2 years were selected from four different RSA studies (totaling 279 hips in 236 patients) in our department. The studies were primarily aimed at evaluating cup fixation, bone cement, and polyethylene types. All study patients received a cemented Spectron EF Primary stem. The selected hips had complete followup until stem failure, death, or the end of the followup period. Stem failure was defined as revision of a loose femoral stem or radiological failure with significant osteolysis in Gruen zones 2 to 6. Cox regression analyses were performed to evaluate if stem subsidence and rotation after 2 years, adjusted for age, sex, stem size, standard/high stem offset, and conventional/highly crosslinked polyethylene, could predict later clinical aseptic failure of the stem. We identified 32 stem failures (27 revisions, five radiological failures) at 14 years median followup (range, 3–18 years). Ten-year stem survival was 94% (95% confidence interval [CI], 90%–96%).

Results

Stem subsidence at 2 years (adjusted hazard ratio [HR], 6.0; 95% CI, 2.5–15; p < 0.001) and retrotorsion of the stem (adjusted HR, 1.7; 95% CI, 1.1–2.5; p = 0.018) were associated with later stem failure. Further risk factors were male sex (subsidence analysis HR, 6.9; p > 0.001), use of the two smallest stem sizes (HRsize 1, 8.0; p > 0.001, HRsize 2, 1 [reference], HRsize 3+, 0.06; p = 0.035), and the high offset option (HR, 3.1; p = 0.005).

Conclusions

Stem subsidence and retrotorsion at 2 years can predict later failure in the Spectron EF Primary stem, consistent with earlier findings on composite-beam cemented stems. Small stem size and high-offset stems comprise the main group of underperforming stems. We recommend that premarket small-scale RSA studies be performed after any design change to a THA femoral component, because even seemingly minor design changes may unexpectedly result in inferior performance.

Level of Evidence

Level III, therapeutic study.
Literature
1.
go back to reference Baldursson H, Egund N, Hansson LI, Selvik G. Instability and wear of total hip prostheses determined with roentgen stereophotogrammetry. Arch Orthop Trauma Surg. 1979;95:257–263.CrossRefPubMed Baldursson H, Egund N, Hansson LI, Selvik G. Instability and wear of total hip prostheses determined with roentgen stereophotogrammetry. Arch Orthop Trauma Surg. 1979;95:257–263.CrossRefPubMed
2.
go back to reference Biau DJ, Latouche A, Porcher R. Competing events influence estimated survival probability: when is Kaplan-Meier analysis appropriate? Clin Orthop Relat Res. 2007;462:229–233.CrossRefPubMed Biau DJ, Latouche A, Porcher R. Competing events influence estimated survival probability: when is Kaplan-Meier analysis appropriate? Clin Orthop Relat Res. 2007;462:229–233.CrossRefPubMed
3.
go back to reference Crowe JF, Sculco TP, Kahn B. Revision total hip arthroplasty: hospital cost and reimbursement analysis. Clin Orthop Relat Res. 2003;413:175–182.CrossRefPubMed Crowe JF, Sculco TP, Kahn B. Revision total hip arthroplasty: hospital cost and reimbursement analysis. Clin Orthop Relat Res. 2003;413:175–182.CrossRefPubMed
4.
go back to reference de Vries LM, van der Weegen W, Pilot P, Stolarczyk PA, Sijbesma T, Hoffman EL. The predictive value of radiostereometric analysis for stem survival in total hip arthroplasty. A systematic review. Hip Int. 2014;24:215–222.CrossRefPubMed de Vries LM, van der Weegen W, Pilot P, Stolarczyk PA, Sijbesma T, Hoffman EL. The predictive value of radiostereometric analysis for stem survival in total hip arthroplasty. A systematic review. Hip Int. 2014;24:215–222.CrossRefPubMed
5.
go back to reference Espehaug B, Furnes O, Engesaeter LB, Havelin LI. 18 years of results with cemented primary hip prostheses in the Norwegian Arthroplasty Register: concerns about some newer implants. Acta Orthop. 2009;80:402–412.CrossRefPubMedPubMedCentral Espehaug B, Furnes O, Engesaeter LB, Havelin LI. 18 years of results with cemented primary hip prostheses in the Norwegian Arthroplasty Register: concerns about some newer implants. Acta Orthop. 2009;80:402–412.CrossRefPubMedPubMedCentral
6.
go back to reference Garellick G, Malchau H, Herberts P. The Charnley versus the Spectron hip prosthesis: clinical evaluation of a randomized, prospective study of 2 different hip implants. J Arthroplasty. 1999;14:407–413.CrossRefPubMed Garellick G, Malchau H, Herberts P. The Charnley versus the Spectron hip prosthesis: clinical evaluation of a randomized, prospective study of 2 different hip implants. J Arthroplasty. 1999;14:407–413.CrossRefPubMed
7.
go back to reference Garellick G, Malchau H, Regner H, Herberts P. The Charnley versus the Spectron hip prosthesis: radiographic evaluation of a randomized, prospective study of 2 different hip implants. J Arthroplasty. 1999;14:414–425.CrossRefPubMed Garellick G, Malchau H, Regner H, Herberts P. The Charnley versus the Spectron hip prosthesis: radiographic evaluation of a randomized, prospective study of 2 different hip implants. J Arthroplasty. 1999;14:414–425.CrossRefPubMed
8.
go back to reference Gooley TA, Leisenring W, Crowley J, Storer BE. Estimation of failure probabilities in the presence of competing risks: new representations of old estimators. Stat Med. 1999;18:695–706.CrossRefPubMed Gooley TA, Leisenring W, Crowley J, Storer BE. Estimation of failure probabilities in the presence of competing risks: new representations of old estimators. Stat Med. 1999;18:695–706.CrossRefPubMed
9.
go back to reference Gruen TA, McNeice GM, Amstutz HC. ‘Modes of failure’ of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res. 1979;141:17–27.PubMed Gruen TA, McNeice GM, Amstutz HC. ‘Modes of failure’ of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res. 1979;141:17–27.PubMed
10.
go back to reference Harris WH, McCarthy JC Jr, O’Neill DA. Femoral component loosening using contemporary techniques of femoral cement fixation. J Bone Joint Surg Am. 1982;64:1063–1067.PubMed Harris WH, McCarthy JC Jr, O’Neill DA. Femoral component loosening using contemporary techniques of femoral cement fixation. J Bone Joint Surg Am. 1982;64:1063–1067.PubMed
11.
go back to reference Hauptfleisch J, Glyn-Jones S, Beard DJ, Gill HS, Murray DW. The premature failure of the Charnley Elite-Plus stem: a confirmation of RSA predictions. J Bone Joint Surg Br. 2006;88:179–183.CrossRefPubMed Hauptfleisch J, Glyn-Jones S, Beard DJ, Gill HS, Murray DW. The premature failure of the Charnley Elite-Plus stem: a confirmation of RSA predictions. J Bone Joint Surg Br. 2006;88:179–183.CrossRefPubMed
12.
go back to reference Kadar T, Hallan G, Aamodt A, Indrekvam K, Badawy M, Havelin LI, Stokke T, Haugan K, Espehaug B, Furnes O. A randomized study on migration of the Spectron EF and the Charnley flanged 40 cemented femoral components using radiostereometric analysis at 2 years. Acta Orthop. 2011;82:538–544.CrossRefPubMedPubMedCentral Kadar T, Hallan G, Aamodt A, Indrekvam K, Badawy M, Havelin LI, Stokke T, Haugan K, Espehaug B, Furnes O. A randomized study on migration of the Spectron EF and the Charnley flanged 40 cemented femoral components using radiostereometric analysis at 2 years. Acta Orthop. 2011;82:538–544.CrossRefPubMedPubMedCentral
13.
go back to reference Karrholm J, Borssen B, Lowenhielm G, Snorrason F. Does early micromotion of femoral stem prostheses matter? 4-7-year stereoradiographic follow-up of 84 cemented prostheses. J Bone Joint Surg Br. 1994;76:912–917.PubMed Karrholm J, Borssen B, Lowenhielm G, Snorrason F. Does early micromotion of femoral stem prostheses matter? 4-7-year stereoradiographic follow-up of 84 cemented prostheses. J Bone Joint Surg Br. 1994;76:912–917.PubMed
14.
go back to reference Klerken T, Mohaddes M, Nemes S, Karrholm J. High early migration of the revised acetabular component is a predictor of late cup loosening: 312 cup revisions followed with radiostereometric analysis for 2-20 years. Hip Int. 2015;25:471–476.CrossRefPubMed Klerken T, Mohaddes M, Nemes S, Karrholm J. High early migration of the revised acetabular component is a predictor of late cup loosening: 312 cup revisions followed with radiostereometric analysis for 2-20 years. Hip Int. 2015;25:471–476.CrossRefPubMed
15.
go back to reference Malchau H, Bragdon CR, Muratoglu OK. The stepwise introduction of innovation into orthopedic surgery: the next level of dilemmas. J Arthroplasty. 2011;26:825–831.CrossRefPubMed Malchau H, Bragdon CR, Muratoglu OK. The stepwise introduction of innovation into orthopedic surgery: the next level of dilemmas. J Arthroplasty. 2011;26:825–831.CrossRefPubMed
16.
go back to reference Middleton RG, Howie DW, Costi K, Sharpe P. Effects of design changes on cemented tapered femoral stem fixation. Clin Orthop Relat Res. 1998;355:47–56.CrossRefPubMed Middleton RG, Howie DW, Costi K, Sharpe P. Effects of design changes on cemented tapered femoral stem fixation. Clin Orthop Relat Res. 1998;355:47–56.CrossRefPubMed
17.
go back to reference Olofsson K, Digas G, Karrholm J. Influence of design variations on early migration of a cemented stem in THA. Clin Orthop Relat Res. 2006;448:67–72.CrossRefPubMed Olofsson K, Digas G, Karrholm J. Influence of design variations on early migration of a cemented stem in THA. Clin Orthop Relat Res. 2006;448:67–72.CrossRefPubMed
18.
go back to reference Pijls BG, Nieuwenhuijse MJ, Fiocco M, Plevier JW, Middeldorp S, Nelissen RG, Valstar ER. Early proximal migration of cups is associated with late revision in THA: a systematic review and meta-analysis of 26 RSA studies and 49 survivalstudies. Acta Orthop. 2012;83:583–591.CrossRefPubMedPubMedCentral Pijls BG, Nieuwenhuijse MJ, Fiocco M, Plevier JW, Middeldorp S, Nelissen RG, Valstar ER. Early proximal migration of cups is associated with late revision in THA: a systematic review and meta-analysis of 26 RSA studies and 49 survivalstudies. Acta Orthop. 2012;83:583–591.CrossRefPubMedPubMedCentral
19.
go back to reference Pijls BG, Valstar ER, Nouta KA, Plevier JW, Fiocco M, Middeldorp S, Nelissen RG. Early migration of tibial components is associated with late revision: a systematic review and meta-analysis of 21,000 knee arthroplasties. Acta Orthop. 2012;83:614–624.CrossRefPubMedPubMedCentral Pijls BG, Valstar ER, Nouta KA, Plevier JW, Fiocco M, Middeldorp S, Nelissen RG. Early migration of tibial components is associated with late revision: a systematic review and meta-analysis of 21,000 knee arthroplasties. Acta Orthop. 2012;83:614–624.CrossRefPubMedPubMedCentral
20.
go back to reference Ranstam J, Karrholm J, Pulkkinen P, Makela K, Espehaug B, Pedersen AB, Mehnert F, Furnes O, group Ns. Statistical analysis of arthroplasty data. II. Guidelines. Acta Orthop. 2011;82:258–267. Ranstam J, Karrholm J, Pulkkinen P, Makela K, Espehaug B, Pedersen AB, Mehnert F, Furnes O, group Ns. Statistical analysis of arthroplasty data. II. Guidelines. Acta Orthop. 2011;82:258–267.
21.
go back to reference Ryd L, Albrektsson BE, Carlsson L, Dansgard F, Herberts P, Lindstrand A, Regner L, Toksvig-Larsen S. Roentgen stereophotogrammetric analysis as a predictor of mechanical loosening of knee prostheses. J Bone Joint Surg Br. 1995;77:377–383.PubMed Ryd L, Albrektsson BE, Carlsson L, Dansgard F, Herberts P, Lindstrand A, Regner L, Toksvig-Larsen S. Roentgen stereophotogrammetric analysis as a predictor of mechanical loosening of knee prostheses. J Bone Joint Surg Br. 1995;77:377–383.PubMed
22.
go back to reference Swedish Hip Arthroplasty Register. Annual Report 2010. Gothenburg, Sweden: Swedish Hip Arthroplasty Register; 2010. Swedish Hip Arthroplasty Register. Annual Report 2010. Gothenburg, Sweden: Swedish Hip Arthroplasty Register; 2010.
23.
go back to reference Thanner J, Freij-Larsson C, Karrholm J, Malchau H, Wesslen B. Evaluation of Boneloc. Chemical and mechanical properties, and a randomized clinical study of 30 total hip arthroplasties. Acta Orthop Scand. 1995;66:207–214.CrossRefPubMed Thanner J, Freij-Larsson C, Karrholm J, Malchau H, Wesslen B. Evaluation of Boneloc. Chemical and mechanical properties, and a randomized clinical study of 30 total hip arthroplasties. Acta Orthop Scand. 1995;66:207–214.CrossRefPubMed
25.
go back to reference Valstar ER, Gill R, Ryd L, Flivik G, Borlin N, Karrholm J. Guidelines for standardization of radiostereometry (RSA) of implants. Acta Orthop. 2005;76:563–572.CrossRefPubMed Valstar ER, Gill R, Ryd L, Flivik G, Borlin N, Karrholm J. Guidelines for standardization of radiostereometry (RSA) of implants. Acta Orthop. 2005;76:563–572.CrossRefPubMed
26.
go back to reference van der Voort P, Pijls BG, Nieuwenhuijse MJ, Jasper J, Fiocco M, Plevier JW, Middeldorp S, Valstar ER, Nelissen RG. Early subsidence of shape-closed hip arthroplasty stems is associated with late revision. A systematic review and meta-analysis of 24 RSA studies and 56 survival studies. Acta Orthop. 2015;86:575–585.CrossRefPubMedPubMedCentral van der Voort P, Pijls BG, Nieuwenhuijse MJ, Jasper J, Fiocco M, Plevier JW, Middeldorp S, Valstar ER, Nelissen RG. Early subsidence of shape-closed hip arthroplasty stems is associated with late revision. A systematic review and meta-analysis of 24 RSA studies and 56 survival studies. Acta Orthop. 2015;86:575–585.CrossRefPubMedPubMedCentral
Metadata
Title
Early Subsidence Predicts Failure of a Cemented Femoral Stem With Minor Design Changes
Authors
Per-Erik Johanson, MD
Martin Antonsson, MStat MScCSE
Bita Shareghi, BSc
Johan Kärrholm, MD, PhD
Publication date
01-10-2016
Publisher
Springer US
Published in
Clinical Orthopaedics and Related Research® / Issue 10/2016
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-016-4884-2

Other articles of this Issue 10/2016

Clinical Orthopaedics and Related Research® 10/2016 Go to the issue