Published in: Archives of Orthopaedic and Trauma Surgery 12/2013

01-12-2013 | Hip Arthroplasty

Early loosening of a press-fit cup with ceramic-on-ceramic articulation: our early results

Authors: Daniël Haverkamp, Stijn Westerbos, Martin M. Campo, Ritsert H. Boonstra, G. H. Rob Albers, Harm M. van der Vis

Published in: Archives of Orthopaedic and Trauma Surgery | Issue 12/2013

Login to get access

Abstract

Introduction

In this study, we present the short-term results of the Selexys TH+ cup with the Ceramys inlay which is a press-fit cup with a ceramic-on-ceramic articulation. (Mathys, Bettlach, Switzerland). We compared the results with a retrospective-matched control group with a Delta PF cup (Lima, Udine, Italy), which is also a press-fit cup with a ceramic-on-ceramic articulation.

Materials and methods

257 elective hip arthroplasties with the Selexys TH+ cup in 250 patients placed in 2009 and 2010 were analyzed and compared with a control group retrospective analysis of the uncemented Delta PF cup (Lima, Udine, Italy) placed in 2007 and 2008 in 208 patients (222 hips). Surgical technique and surgeons were identical in both groups.

Results

During a follow-up period of 3–21 months, 19 aseptic loosenings (7.4 %) were found for the Selexys TH+ cup. The survival plotted by a Kaplan–Meier curve shows a 1-year survival of 87.4 %. The Lima Delta PF cup showed a 1-year survival of 99.5 %. Failure analysis showed no clear explanation for this early loosening.

Conclusion

The Selexys TH+ cup combined with the Ceramys ceramic-on-ceramic inlay coupling show an unacceptable high early revision rate. Therefore, we advice against using this combination.
Literature
1.
go back to reference Haverkamp D (2009) The latest information on total hip arthroplasty bearing surfaces. Minerva Ortopedica e Traumatologica 60(3):233–240 Haverkamp D (2009) The latest information on total hip arthroplasty bearing surfaces. Minerva Ortopedica e Traumatologica 60(3):233–240
2.
go back to reference Tsukada S, Wakui M, Matsueda M (2013) Uncemented third-generation ceramic-on-ceramic total hip arthroplasty using metal acetabular shell with direct taper locking liner. Arch Orthop Trauma Surg 133(6):861–868. doi:10.1007/s00402-013-1724-7 PubMedCrossRef Tsukada S, Wakui M, Matsueda M (2013) Uncemented third-generation ceramic-on-ceramic total hip arthroplasty using metal acetabular shell with direct taper locking liner. Arch Orthop Trauma Surg 133(6):861–868. doi:10.​1007/​s00402-013-1724-7 PubMedCrossRef
3.
go back to reference Benazzo F, Cuzzocrea F, Stroppa S, Ravasi F, Dalla Pria P (2007) Modular stems in DDH. Hip Int 17(Suppl 5):S138–S141PubMed Benazzo F, Cuzzocrea F, Stroppa S, Ravasi F, Dalla Pria P (2007) Modular stems in DDH. Hip Int 17(Suppl 5):S138–S141PubMed
4.
go back to reference Massin P, Schmidt L, Engh CA (1989) Evaluation of cementless acetabular component migration. An experimental study. J Arthroplast 4(3):245–251CrossRef Massin P, Schmidt L, Engh CA (1989) Evaluation of cementless acetabular component migration. An experimental study. J Arthroplast 4(3):245–251CrossRef
5.
go back to reference Malchau H, Kärrholm J, Wang YX, Herberts P (1995) Accuracy of migration analysis in hip arthroplasty. Digitized and conventional radiography, compared to radiostereometry in 51 patients. Acta Orthop Scand 66(5):418–424PubMedCrossRef Malchau H, Kärrholm J, Wang YX, Herberts P (1995) Accuracy of migration analysis in hip arthroplasty. Digitized and conventional radiography, compared to radiostereometry in 51 patients. Acta Orthop Scand 66(5):418–424PubMedCrossRef
6.
go back to reference Udomkiat P, Wan Z, Dorr LD (2001) Comparison of preoperative radiographs and intraoperative findings of fixation of hemispheric porous-coated sockets. J Bone Jt Surg Am 83:1865–1870 Udomkiat P, Wan Z, Dorr LD (2001) Comparison of preoperative radiographs and intraoperative findings of fixation of hemispheric porous-coated sockets. J Bone Jt Surg Am 83:1865–1870
7.
go back to reference Ilchmann T, Pannhorst C, Abächerli C, Clauss M (2010) Early failure of a spherical pressfit cup. Annual meeting of the Swiss Society of Orthopaedics and Traumatology. St Gallen, Switzerland Ilchmann T, Pannhorst C, Abächerli C, Clauss M (2010) Early failure of a spherical pressfit cup. Annual meeting of the Swiss Society of Orthopaedics and Traumatology. St Gallen, Switzerland
8.
go back to reference Jovanovic M, Warzecha C, Muschik M, Hoffmann F (2009) Early results and differences in clinical outcome of 32 and 36 mm ceramic on ceramic articulation. EFORT, Vienna Jovanovic M, Warzecha C, Muschik M, Hoffmann F (2009) Early results and differences in clinical outcome of 32 and 36 mm ceramic on ceramic articulation. EFORT, Vienna
9.
go back to reference Hoffmann F, Jovanovic M, Muschik M (2010) Head size in relation to noise occurrence in ceramic on ceramic bearings. EFORT, Madrid Hoffmann F, Jovanovic M, Muschik M (2010) Head size in relation to noise occurrence in ceramic on ceramic bearings. EFORT, Madrid
10.
go back to reference Hothan A, Huber G, Weiss C, Hoffmann N, Morlock M (2011) Deformation characteristics and eigenfrequencies of press-fit acetabular cups. Clin Biomech (Bristol, Avon) 26(1):46–51CrossRef Hothan A, Huber G, Weiss C, Hoffmann N, Morlock M (2011) Deformation characteristics and eigenfrequencies of press-fit acetabular cups. Clin Biomech (Bristol, Avon) 26(1):46–51CrossRef
11.
go back to reference Hasegawa M, Sudo A, Hirata H, Uchida A (2003) Ceramic acetabular liner fracture in total hip arthroplasty with a ceramic sandwich cup. J Arthroplast 18(5):658–661CrossRef Hasegawa M, Sudo A, Hirata H, Uchida A (2003) Ceramic acetabular liner fracture in total hip arthroplasty with a ceramic sandwich cup. J Arthroplast 18(5):658–661CrossRef
12.
go back to reference Zhou ZK, Li MG, Borlin N, Wood DJ, Nivbrant B (2006) No increased migration in cups with ceramic-on-ceramic bearing: an RSA study. Clin Orthop Relat Res 448:39–45PubMedCrossRef Zhou ZK, Li MG, Borlin N, Wood DJ, Nivbrant B (2006) No increased migration in cups with ceramic-on-ceramic bearing: an RSA study. Clin Orthop Relat Res 448:39–45PubMedCrossRef
13.
go back to reference Aldrian S, Nau T, Gillesberger F, Petras N, Ehall R (2009) Medium-term analysis of modern ceramic-on-ceramic bearing in THA. Hip Int 19(1):36–40PubMed Aldrian S, Nau T, Gillesberger F, Petras N, Ehall R (2009) Medium-term analysis of modern ceramic-on-ceramic bearing in THA. Hip Int 19(1):36–40PubMed
14.
go back to reference Capello WN, D’Antonio JA, Feinberg JR, Manley MT, Naughton M (2008) Ceramic-on-ceramic total hip arthroplasty: update. J Arthroplast 23(Suppl 7):39–43CrossRef Capello WN, D’Antonio JA, Feinberg JR, Manley MT, Naughton M (2008) Ceramic-on-ceramic total hip arthroplasty: update. J Arthroplast 23(Suppl 7):39–43CrossRef
15.
go back to reference D’Antonio J, Capello W, Manley M, Naughton M, Sutton K (2005) Alumina ceramic bearings for total hip arthroplasty: five-year results of a prospective randomized study. Clin Orthop Relat Res 436:164–171PubMedCrossRef D’Antonio J, Capello W, Manley M, Naughton M, Sutton K (2005) Alumina ceramic bearings for total hip arthroplasty: five-year results of a prospective randomized study. Clin Orthop Relat Res 436:164–171PubMedCrossRef
17.
go back to reference Shon WY, Baldini T, Peterson MG, Wright TM, Salvati EA (2005) Impingement in total hip arthroplasty a study of retrieved acetabular components. J Arthroplast 20(4):427–435CrossRef Shon WY, Baldini T, Peterson MG, Wright TM, Salvati EA (2005) Impingement in total hip arthroplasty a study of retrieved acetabular components. J Arthroplast 20(4):427–435CrossRef
18.
go back to reference Elkins JM, O’Brien MK, Stroud NJ, Pedersen DR, Callaghan JJ, Brown TD (2010) Hard-on-hard total hip impingement causes extreme contact stress concentrations. Clin Orthop Relat Res 469(2):454–463. doi:10.1007/s11999-010-1632-x CrossRef Elkins JM, O’Brien MK, Stroud NJ, Pedersen DR, Callaghan JJ, Brown TD (2010) Hard-on-hard total hip impingement causes extreme contact stress concentrations. Clin Orthop Relat Res 469(2):454–463. doi:10.​1007/​s11999-010-1632-x CrossRef
19.
go back to reference Brockett C, Williams S, Jin Z, Isaac G, Fisher J (2007) Friction of total hip replacements with different bearings and loading conditions. J Biomed Mater Res B Appl Biomater 81(2):508–515PubMed Brockett C, Williams S, Jin Z, Isaac G, Fisher J (2007) Friction of total hip replacements with different bearings and loading conditions. J Biomed Mater Res B Appl Biomater 81(2):508–515PubMed
20.
go back to reference Flanagan S, Jones E, Birkinshaw C (2010) In vitro friction and lubrication of large bearing hip prostheses. Proc Inst Mech Eng H 224(7):853–864PubMedCrossRef Flanagan S, Jones E, Birkinshaw C (2010) In vitro friction and lubrication of large bearing hip prostheses. Proc Inst Mech Eng H 224(7):853–864PubMedCrossRef
21.
go back to reference Schreiner U, Simnacher M, Scheller G, Scharf HP (2007) The influence of different surface treatments on the primary stability of cementless acetabular cups: an in vitro study. Biomed Tech (Berl) 52(3):243–247CrossRef Schreiner U, Simnacher M, Scheller G, Scharf HP (2007) The influence of different surface treatments on the primary stability of cementless acetabular cups: an in vitro study. Biomed Tech (Berl) 52(3):243–247CrossRef
22.
go back to reference Klika AK, Murray TG, Darwiche H, Barsoum WK (2007) Options for acetabular fixation surfaces. J Long Term Eff Med Implant 17(3):187–192CrossRef Klika AK, Murray TG, Darwiche H, Barsoum WK (2007) Options for acetabular fixation surfaces. J Long Term Eff Med Implant 17(3):187–192CrossRef
23.
go back to reference Mahoney OM, Dimon JH III (1990) Unsatisfactory results with a ceramic total hip prosthesis. J Bone Jt Surg Am 72(5):663–671 Mahoney OM, Dimon JH III (1990) Unsatisfactory results with a ceramic total hip prosthesis. J Bone Jt Surg Am 72(5):663–671
Metadata
Title
Early loosening of a press-fit cup with ceramic-on-ceramic articulation: our early results
Authors
Daniël Haverkamp
Stijn Westerbos
Martin M. Campo
Ritsert H. Boonstra
G. H. Rob Albers
Harm M. van der Vis
Publication date
01-12-2013
Publisher
Springer Berlin Heidelberg
Published in
Archives of Orthopaedic and Trauma Surgery / Issue 12/2013
Print ISSN: 0936-8051
Electronic ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-013-1861-z

Other articles of this Issue 12/2013

Archives of Orthopaedic and Trauma Surgery 12/2013 Go to the issue