Early clinical experience utilizing scintillator with optical fiber (SOF) detector in clinical boron neutron capture therapy: its issues and solutions

Authors: Masayori Ishikawa, Tetsuya Yamamoto, Akira Matsumura, Junichi Hiratsuka, Shin-Ichi Miyatake, Itsuro Kato, Yoshinori Sakurai, Hiroaki Kumada, Shubhechha J. Shrestha, Koji Ono

Published in: Radiation Oncology | Issue 1/2016

Login to get access

Abstract

Background

Real-time measurement of thermal neutrons in the tumor region is essential for proper evaluation of the absorbed dose in boron neutron capture therapy (BNCT) treatment. The gold wire activation method has been routinely used to measure the neutron flux distribution in BNCT irradiation, but a real-time measurement using gold wire is not possible. To overcome this issue, the scintillator with optical fiber (SOF) detector has been developed. The purpose of this study is to demonstrate the feasibility of the SOF detector as a real-time thermal neutron monitor in clinical BNCT treatment and also to report issues in the use of SOF detectors in clinical practice and their solutions.

Material and methods

Clinical measurements using the SOF detector were carried out in 16 BNCT clinical trial patients from December 2002 until end of 2006 at the Japanese Atomic Energy Agency (JAEA) and Kyoto University Research Reactor Institute (KURRI).

Results

The SOF detector worked effectively as a real-time thermal neutron monitor. The neutron fluence obtained by the gold wire activation method was found to differ from that obtained by the SOF detector. The neutron fluence obtained by the SOF detector was in better agreement with the expected fluence than with gold wire activation. The estimation error for the SOF detector was small in comparison to the gold wire measurement. In addition, real-time monitoring suggested that the neutron flux distribution and intensity at the region of interest (ROI) may vary due to the reactor condition, patient motion and dislocation of the SOF detector.

Conclusion

Clinical measurements using the SOF detector to measure thermal neutron flux during BNCT confirmed that SOF detectors are effective as a real-time thermal neutron monitor. To minimize the estimation error due to the displacement of the SOF probe during treatment, a loop-type SOF probe was developed.
Literature
1.
go back to reference Farr LE, Sweet WH, Robertson J, Foster C, Locksley H, Sutherland D, et al. Neutron capture therapy with boron in the treatment of glioblastoma multiforme. Am J Roentgenol Radium Ther Nucl Med. 1954;71(2):279-93. Farr LE, Sweet WH, Robertson J, Foster C, Locksley H, Sutherland D, et al. Neutron capture therapy with boron in the treatment of glioblastoma multiforme. Am J Roentgenol Radium Ther Nucl Med. 1954;71(2):279-93.
2.
go back to reference Hatanaka H. A revised boron-neutron capture therapy for malignant brain tumors. J Neurol. 1975;209(2):81–94.CrossRefPubMed Hatanaka H. A revised boron-neutron capture therapy for malignant brain tumors. J Neurol. 1975;209(2):81–94.CrossRefPubMed
3.
go back to reference Mishima Y, Honda C, Ichihashi M, Obara H, Hiratsuka J, Fukuda H, et al. Treatment of malignant melanoma by single thermal neutron capture therapy with melanoma-seeking 10 B-compound. Lancet. 1989;334(8659):388–9.CrossRef Mishima Y, Honda C, Ichihashi M, Obara H, Hiratsuka J, Fukuda H, et al. Treatment of malignant melanoma by single thermal neutron capture therapy with melanoma-seeking 10 B-compound. Lancet. 1989;334(8659):388–9.CrossRef
4.
go back to reference Nakagawa Y, Hatanaka H. Boron neutron capture therapy: clinical brain tumor studies. J Neuro-Oncol. 1997;33(1–2):105–15.CrossRef Nakagawa Y, Hatanaka H. Boron neutron capture therapy: clinical brain tumor studies. J Neuro-Oncol. 1997;33(1–2):105–15.CrossRef
5.
go back to reference Barth RF, Soloway AH, Fairchild RG, Brugger RM. Boron neutron capture therapy for cancer. Realities and prospects. Cancer. 1992;70(12):2995–3007.CrossRefPubMed Barth RF, Soloway AH, Fairchild RG, Brugger RM. Boron neutron capture therapy for cancer. Realities and prospects. Cancer. 1992;70(12):2995–3007.CrossRefPubMed
6.
go back to reference Yamamoto T, Nakai K, Matsumura A. Boron neutron capture therapy for glioblastoma. Cancer Lett. 2008;262(2):143–52.CrossRefPubMed Yamamoto T, Nakai K, Matsumura A. Boron neutron capture therapy for glioblastoma. Cancer Lett. 2008;262(2):143–52.CrossRefPubMed
7.
go back to reference Ito Y, Katano G, Harano H, Matsumoto T, Uritani A, Kudo K, et al. Development of a tiny neutron probe with an optical fibre for BNCT. Radiat Prot Dosim. 2004;110(1–4):619–22.CrossRef Ito Y, Katano G, Harano H, Matsumoto T, Uritani A, Kudo K, et al. Development of a tiny neutron probe with an optical fibre for BNCT. Radiat Prot Dosim. 2004;110(1–4):619–22.CrossRef
8.
go back to reference Peurrung AJ. Recent developments in neutron detection. Nucl Instrum Methods Phys Res, Sect A. 2000;443(2):400–15.CrossRef Peurrung AJ. Recent developments in neutron detection. Nucl Instrum Methods Phys Res, Sect A. 2000;443(2):400–15.CrossRef
9.
go back to reference Ishikawa M, Ono K, Sakurai Y, Unesaki H, Uritani A, Bengua G, et al. Development of real-time thermal neutron monitor using boron-loaded plastic scintillator with optical fiber for boron neutron capture therapy. Appl Radiat Isot. 2004;61(5):775–9.CrossRefPubMed Ishikawa M, Ono K, Sakurai Y, Unesaki H, Uritani A, Bengua G, et al. Development of real-time thermal neutron monitor using boron-loaded plastic scintillator with optical fiber for boron neutron capture therapy. Appl Radiat Isot. 2004;61(5):775–9.CrossRefPubMed
10.
go back to reference Aschan C, Toivonen M, Savolainen S, Seppälä T, Auterinen I. Epithermal neutron beam dosimetry with thermoluminescence dosemeters for boron neutron capture therapy. Radiat Prot Dosim. 1999;81(1):47–56.CrossRef Aschan C, Toivonen M, Savolainen S, Seppälä T, Auterinen I. Epithermal neutron beam dosimetry with thermoluminescence dosemeters for boron neutron capture therapy. Radiat Prot Dosim. 1999;81(1):47–56.CrossRef
11.
go back to reference Gambarini G, Birattari C, Colombi C, Pirola L, Rosi G. Fricke gel dosimetry in boron neutron capture therapy. Radiat Prot Dosim. 2002;101(1–4):419–22.CrossRef Gambarini G, Birattari C, Colombi C, Pirola L, Rosi G. Fricke gel dosimetry in boron neutron capture therapy. Radiat Prot Dosim. 2002;101(1–4):419–22.CrossRef
12.
go back to reference Miller M, Mariani L, Gonçalves-Carralves M, Skumanic M, Thorp S. Implantable self-powered detector for on-line determination of neutron flux in patients during NCT treatment. Appl Radiat Isot. 2004;61(5):1033–7.CrossRefPubMed Miller M, Mariani L, Gonçalves-Carralves M, Skumanic M, Thorp S. Implantable self-powered detector for on-line determination of neutron flux in patients during NCT treatment. Appl Radiat Isot. 2004;61(5):1033–7.CrossRefPubMed
13.
go back to reference Ishikawa M, Kumada H, Yamamoto K, Kaneko J, Bengua G, Unesaki H, et al. Development of a wide-range paired scintillator with optical fiber neutron monitor for BNCT irradiation field study. Nucl Instrum Methods Phys Res, Sect A. 2005;551(2):448–57.CrossRef Ishikawa M, Kumada H, Yamamoto K, Kaneko J, Bengua G, Unesaki H, et al. Development of a wide-range paired scintillator with optical fiber neutron monitor for BNCT irradiation field study. Nucl Instrum Methods Phys Res, Sect A. 2005;551(2):448–57.CrossRef
14.
go back to reference Mouton J, Haug R, Bridier A, Dodinot B, Eschwege F. Influence of high-energy photon beam irradiation on pacemaker operation. Phys Med Biol. 2002;47(16):2879.CrossRefPubMed Mouton J, Haug R, Bridier A, Dodinot B, Eschwege F. Influence of high-energy photon beam irradiation on pacemaker operation. Phys Med Biol. 2002;47(16):2879.CrossRefPubMed
16.
go back to reference Fisher B, Abdurashitov J, Coakley K, Gavrin V, Gilliam D, Nico J, et al. Fast neutron detection with 6 Li-loaded liquid scintillator. Nucl Instrum Methods Phys Res, Sect A. 2011;646(1):126–34.CrossRef Fisher B, Abdurashitov J, Coakley K, Gavrin V, Gilliam D, Nico J, et al. Fast neutron detection with 6 Li-loaded liquid scintillator. Nucl Instrum Methods Phys Res, Sect A. 2011;646(1):126–34.CrossRef
Metadata
Title
Early clinical experience utilizing scintillator with optical fiber (SOF) detector in clinical boron neutron capture therapy: its issues and solutions
Authors
Masayori Ishikawa
Tetsuya Yamamoto
Akira Matsumura
Junichi Hiratsuka
Shin-Ichi Miyatake
Itsuro Kato
Yoshinori Sakurai
Hiroaki Kumada
Shubhechha J. Shrestha
Koji Ono
Publication date
01-12-2016
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2016
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/s13014-016-0680-0

Other articles of this Issue 1/2016

Radiation Oncology 1/2016 Go to the issue