Duration of ultrasonic activation causing secondary fractures during the removal of the separated instruments with different tapers

Authors: Hakan Arslan, Ezgi Doğanay Yıldız, Gizem Taş, Nuray Akbıyık, Hüseyin Sinan Topçuoğlu

Published in: Clinical Oral Investigations | Issue 1/2020

Login to get access

Abstract

Objectives

The aim of the present study was to determine the effect of taper (.08, .06, and .04) of separated K3XF instruments on duration taken for the secondary fracture formation during ultrasonic activation.

Materials and methods

Ten 25/.08 K3XF (SybronEndo, Orange, CA, USA), ten 25/.06 K3XF, and ten 25/.04 K3XF instruments were used for the study. The apical 5 mm of the instruments was cut to simulate the fragments in root canals. Fragments of the instruments were sandwiched between two straight dentin blocks. An ultrasonic tip was used to cause a secondary fracture of the fragment. The time needed for the secondary fracture was recorded for each instrument. The data were statistically analyzed using the Kruskal-Wallis H test (alpha = 0.05).

Results

Secondary fractures occurred in all instruments. In the .08 taper group, secondary fractures took longer than in the case of the .06 and the .04 taper groups (P < 0.05). There were no significant differences between the .06 and the .04 taper groups in terms of the time required for the occurrence of a secondary fracture (P > 0.05).

Conclusions

In the .08 taper group, secondary fracture took longer time than in the case of the .06 and the .04 taper groups due to its larger cross-sectional area involved.

Clinical relevance

Typically, when removing separated instruments, a much lower power setting is chosen. The purpose of this in vitro study was to determine which tapered files were more resilient to secondary fracture, thus allowing a higher power setting to be chosen. Thus, the results of the present study cannot be used in clinical practice. If the clinician knows the taper of the broken file, the clinician should be very careful with regard to secondary fractures when using ultrasonics to remove the separated smaller tapered instruments.
Literature
1.
go back to reference Melo M, Pereira E, Viana A, Fonseca A, Buono V, Bahia M (2008) Dimensional characterization and mechanical behaviour of K3 rotary instruments. Int Endod J 41:329–338CrossRef Melo M, Pereira E, Viana A, Fonseca A, Buono V, Bahia M (2008) Dimensional characterization and mechanical behaviour of K3 rotary instruments. Int Endod J 41:329–338CrossRef
2.
go back to reference Haikel Y, Serfaty R, Bateman G, Senger B, Allemann C (1999) Dynamic and cyclic fatigue of engine-driven rotary nickel-titanium endodontic instruments. J Endod 25:434–440CrossRef Haikel Y, Serfaty R, Bateman G, Senger B, Allemann C (1999) Dynamic and cyclic fatigue of engine-driven rotary nickel-titanium endodontic instruments. J Endod 25:434–440CrossRef
3.
go back to reference Pruett JP, Clement DJ, Carnes DL (1997) Cyclic fatigue testing of nickel-titanium endodontic instruments. J Endod 23:77–85CrossRef Pruett JP, Clement DJ, Carnes DL (1997) Cyclic fatigue testing of nickel-titanium endodontic instruments. J Endod 23:77–85CrossRef
4.
go back to reference Ullmann CJ, Peters OA (2005) Effect of cyclic fatigue on static fracture loads in ProTaper nickel-titanium rotary instruments. J Endod 31:183–186CrossRef Ullmann CJ, Peters OA (2005) Effect of cyclic fatigue on static fracture loads in ProTaper nickel-titanium rotary instruments. J Endod 31:183–186CrossRef
5.
go back to reference Parashos P, Messer HH (2006) Rotary NiTi instrument fracture and its consequences. J Endod 32:1031–1043CrossRef Parashos P, Messer HH (2006) Rotary NiTi instrument fracture and its consequences. J Endod 32:1031–1043CrossRef
6.
go back to reference Choi J, Oh S, Kim Y-C, Jee K-K, Kum K, Chang S (2016) Fracture resistance of K3 nickel-titanium files made from different thermal treatments. Bioinorg Chem Appl 2016:1–6CrossRef Choi J, Oh S, Kim Y-C, Jee K-K, Kum K, Chang S (2016) Fracture resistance of K3 nickel-titanium files made from different thermal treatments. Bioinorg Chem Appl 2016:1–6CrossRef
7.
go back to reference Shen Y, Zhou H, Campbell L, Wang Z, Wang R, Du T, Haapasalo M (2014) Fatigue and nanomechanical properties of K3 XF nickel-titanium instruments. Int Endod J 47:1160–1167CrossRef Shen Y, Zhou H, Campbell L, Wang Z, Wang R, Du T, Haapasalo M (2014) Fatigue and nanomechanical properties of K3 XF nickel-titanium instruments. Int Endod J 47:1160–1167CrossRef
8.
go back to reference Shen Y, Cheung GS-p, Bian Z, Peng B (2006) Comparison of defects in ProFile and ProTaper systems after clinical use. J Endod 32:61–65CrossRef Shen Y, Cheung GS-p, Bian Z, Peng B (2006) Comparison of defects in ProFile and ProTaper systems after clinical use. J Endod 32:61–65CrossRef
9.
go back to reference Cheung G, Peng B, Bian Z, Shen Y, Darvell B (2005) Defects in ProTaper S1 instruments after clinical use: fractographic examination. Int Endod J 38:802–809CrossRef Cheung G, Peng B, Bian Z, Shen Y, Darvell B (2005) Defects in ProTaper S1 instruments after clinical use: fractographic examination. Int Endod J 38:802–809CrossRef
10.
go back to reference Ha J-H, Kim SK, Cohenca N, Kim H-C (2013) Effect of R-phase heat treatment on torsional resistance and cyclic fatigue fracture. J Endod 39:389–393CrossRef Ha J-H, Kim SK, Cohenca N, Kim H-C (2013) Effect of R-phase heat treatment on torsional resistance and cyclic fatigue fracture. J Endod 39:389–393CrossRef
11.
go back to reference Gambarini G, Plotino G, Grande N, Al-Sudani D, De Luca M, Testarelli L (2011) Mechanical properties of nickel–titanium rotary instruments produced with a new manufacturing technique. Int Endod J 44:337–341CrossRef Gambarini G, Plotino G, Grande N, Al-Sudani D, De Luca M, Testarelli L (2011) Mechanical properties of nickel–titanium rotary instruments produced with a new manufacturing technique. Int Endod J 44:337–341CrossRef
16.
go back to reference Ruddle CJ (2002) Nonsurgical retreatment. In: Cohen S, Burns RC (eds) Pathways of the pulp, 8th edn. Mosby, St. Louis, pp 875–929 Ruddle CJ (2002) Nonsurgical retreatment. In: Cohen S, Burns RC (eds) Pathways of the pulp, 8th edn. Mosby, St. Louis, pp 875–929
17.
go back to reference Hulsmann M, Schinkel I (1999) Influence of several factors on the success or failure of removal of fractured instruments from the root canal. Endod Dent Traumatol 15:252–258CrossRef Hulsmann M, Schinkel I (1999) Influence of several factors on the success or failure of removal of fractured instruments from the root canal. Endod Dent Traumatol 15:252–258CrossRef
21.
go back to reference Gencoglu N, Helvacioglu D (2009) Comparison of the different techniques to remove fractured endodontic instruments from root canal systems. Eur J Dent 3:90–95CrossRef Gencoglu N, Helvacioglu D (2009) Comparison of the different techniques to remove fractured endodontic instruments from root canal systems. Eur J Dent 3:90–95CrossRef
Metadata
Title
Duration of ultrasonic activation causing secondary fractures during the removal of the separated instruments with different tapers
Authors
Hakan Arslan
Ezgi Doğanay Yıldız
Gizem Taş
Nuray Akbıyık
Hüseyin Sinan Topçuoğlu
Publication date
01-01-2020
Publisher
Springer Berlin Heidelberg
Published in
Clinical Oral Investigations / Issue 1/2020
Print ISSN: 1432-6981
Electronic ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-019-02936-7

Other articles of this Issue 1/2020

Clinical Oral Investigations 1/2020 Go to the issue