Abstract

Background

India has a large number of HIV infected patients being followed up at anti-retroviral therapy (ART) centers. The patients are regularly offered CD4 count estimation for deciding their eligibility for ART initiation as well as for monitoring response to ART, making CD4 count estimation a very critical test. Hence, quality control of CD4 testing is utmost important for ultimate success of ART program. As the commercial controls are very expensive, internal quality control (IQC), at present, is being done by duplicate analysis method using previous day samples in most of the laboratories. Hence the study was undertaken to review performance of duplicate analysis method for monitoring daily IQC.

Methods

Quality control (QC) data from 11 Indian laboratories using duplicate analysis and/or commercial controls for IQC of CD4 testing was collected for reviewing information on QC parameters such as precision, accuracy and trend monitoring. Precision was determined by r2 values and mean % variation for duplicate analysis and coefficient of variation (% CV) for commercial controls. Accuracy was monitored by rate of QC failures for both the types of control and trend monitoring was done by plotting LJ charts for commercial controls and by plotting daily % variation for duplicate analysis.

Results

The laboratories using duplicate analysis for IQC showed good precision with mean % variation ranging from 0.5 to 7.2. There was good match between r2 values and % CV of the laboratories performing both the types of QC methods. Rates of QC failures were 2.3 for duplicate analysis and 3 per laboratory-year for IMMUNO-TROL controls. Daily trend monitoring showed fluctuation of daily counts around mean in LJ charts and of percent variation around 0% in duplicate analysis method. Commercially available controls showed limitations such as altered specimen quality leading to difficulties in manual gating and issues with the establishment of laboratory range.

Conclusion

Duplicate analysis can serve as a cheaper alternative to commercially available controls for IQC of CD4 testing especially when supplemented with other QC measures for controlling variations caused by reagent, equipment, staff and environment in addition to the successful participation in External Quality Assurance programme.
Literature
2.
go back to reference Kinns H, Pitkin S, Housley D, Freedman DB (2013) Internal quality control: best practice. J Clin Pathol 66(12):1027–1032CrossRefPubMed Kinns H, Pitkin S, Housley D, Freedman DB (2013) Internal quality control: best practice. J Clin Pathol 66(12):1027–1032CrossRefPubMed
5.
go back to reference Westerman LE, Kohatsu L, Ortiz A, McClain B, Kaplan J, Spira T et al (2010) A quality management systems approach for CD4 testing in resource-poor settings. Am J Clin Pathol 134(4):556–567CrossRefPubMed Westerman LE, Kohatsu L, Ortiz A, McClain B, Kaplan J, Spira T et al (2010) A quality management systems approach for CD4 testing in resource-poor settings. Am J Clin Pathol 134(4):556–567CrossRefPubMed
7.
go back to reference Ramalingam VV, Mani M, Sundaresan VC, Karunaiya RJ, Sachithanandham J, Kannangai R (2012) Daily quality control in CD3+ and CD4+ T cell estimation by the FACSCount system at a tertiary care center in south India. Clin Vaccine Immunol 19(10):1693–1696PubMedCentralCrossRefPubMed Ramalingam VV, Mani M, Sundaresan VC, Karunaiya RJ, Sachithanandham J, Kannangai R (2012) Daily quality control in CD3+ and CD4+ T cell estimation by the FACSCount system at a tertiary care center in south India. Clin Vaccine Immunol 19(10):1693–1696PubMedCentralCrossRefPubMed
9.
go back to reference Ding T, Bergeron M, Seely P, Yang X, Diallo TO, Plews M et al (2014) Compatibility of stabilized whole blood products with CD4 technologies and their suitability for quality assessment programs. PLoS One 9(8):e103391PubMedCentralCrossRefPubMed Ding T, Bergeron M, Seely P, Yang X, Diallo TO, Plews M et al (2014) Compatibility of stabilized whole blood products with CD4 technologies and their suitability for quality assessment programs. PLoS One 9(8):e103391PubMedCentralCrossRefPubMed
11.
go back to reference Jahanmehr SA, Hamid RA, Sadrossadat M, Alizadeh S (2012) Peykan check, a simple continuous quality control method for hematology analyzers. Iran J Blood Cancer 4(2):67–73 Jahanmehr SA, Hamid RA, Sadrossadat M, Alizadeh S (2012) Peykan check, a simple continuous quality control method for hematology analyzers. Iran J Blood Cancer 4(2):67–73
Metadata
Title
Duplicate analysis method: a cheaper alternative to commercial IQC materials in limited resource settings for monitoring CD4 testing
Authors
Ashwini Shete
Dharmesh P Singh
Bharati Mahajan
Amol Kokare
Ramesh Paranjape
Madhuri Thakar
Publication date
01-12-2015
Publisher
BioMed Central
Published in
AIDS Research and Therapy / Issue 1/2015
Electronic ISSN: 1742-6405
DOI
https://doi.org/10.1186/s12981-015-0067-6

Other articles of this Issue 1/2015

AIDS Research and Therapy 1/2015 Go to the issue
Live Webinar | 27-06-2024 | 18:00 (CEST)

Keynote webinar | Spotlight on medication adherence

Live: Thursday 27th June 2024, 18:00-19:30 (CEST)

WHO estimates that half of all patients worldwide are non-adherent to their prescribed medication. The consequences of poor adherence can be catastrophic, on both the individual and population level.

Join our expert panel to discover why you need to understand the drivers of non-adherence in your patients, and how you can optimize medication adherence in your clinics to drastically improve patient outcomes.

Prof. Kevin Dolgin
Prof. Florian Limbourg
Prof. Anoop Chauhan
Developed by: Springer Medicine
Obesity Clinical Trial Summary

At a glance: The STEP trials

A round-up of the STEP phase 3 clinical trials evaluating semaglutide for weight loss in people with overweight or obesity.

Developed by: Springer Medicine