Dual-lumen catheters for continuous venovenous hemofiltration: limits for blood delivery via femoral vein access and a potential alternative in an experimental setting in anesthetized pigs

Authors: Juliane K Unger, Klaus Pietzner, Roland C Francis, Juergen Birnbaum, Marc Michael Theisen, Arne-Joern Lemke, Stefan M Niehues

Published in: Critical Care | Issue 1/2007

Login to get access

Abstract

Introduction

Small intravascular volume, pathophysiological hemorheology, and/or low cardiac output [CO] are assumed to reduce available blood flow rates via common dual-lumen catheters (except for those with a right atrium catheter tip position) in the critically ill patient. We performed an experimental animal study to verify these assumptions.

Methods

Anesthetized, ventilated pigs (35 to 50 kg) were allocated to different hemorheological conditions based on the application of different volume substitutes (that is, colloids and crystalloids, n = 6 to 7 per volume substitute). In a second step, allocation to the final study group was performed after the determination of the highest values for access flow (Qa) via an axial dual-lumen catheter (11 French, 20 cm long, side holes) placed in the femoral vein. High Qa rates (>300 ml/minute) were allocated to the dual-lumen catheter group; low Qa rates were switched to a 'dual-vein approach' using an alternative catheter (8.5-French sheath) for separate blood delivery. Hemodynamics (CO and central venous pressure [CVP]) and blood composition (blood cell counts, plasma proteins, and colloid osmotic pressure) were measured. Catheter tip positions and vessel diameters were exemplified by computed tomography.

Results

Forty-four percent of the animals required an alternative vascular access due to only minimal Qa via the dual-lumen catheter. Neither hemorheologically relevant aspects nor CO and CVP correlated with the Qa achievable via the femoral vein access. Even though the catheter tip of the alternative catheter provided common iliac vein but not caval vein access, this catheter type enabled higher Qa than the dual-lumen catheter positioned in the caval vein.

Conclusion

With respect to the femoral vein approach, none of the commonly assumed reasons for limited Qa via the arterial line of an axial dual-lumen catheter could be confirmed. The 8.5-French sheath, though not engineered for that purpose, performed quite well as an alternative catheter. Thus, in patients lacking right jugular vein access with tip positioning of large-French dual-lumen catheters in the right atrium, it would be of interest to obtain clinical data re-evaluating the 'dual-vein approach' with separate blood delivery via a tip-hole catheter in order to provide high-volume hemofiltration.
Appendix
Available only for authorised users
Literature
1.
go back to reference Baldwin I, Bellomo R, Koch B: Qb reductions during continuous renal replacement therapy and circuit life. Intensive Care Med 2004, 30: 2074-2079. 10.1007/s00134-004-2440-0CrossRefPubMed Baldwin I, Bellomo R, Koch B: Qb reductions during continuous renal replacement therapy and circuit life. Intensive Care Med 2004, 30: 2074-2079. 10.1007/s00134-004-2440-0CrossRefPubMed
2.
go back to reference Tal MG, Peixoto AJ, Crowley ST, Denbow N, Eliseo D, Pollak J: Comparison of side holes versus non side hole high flow hemodialysis catheters. Hemodial Int 2006, 10: 63-67. 10.1111/j.1542-4758.2006.01176.xCrossRefPubMed Tal MG, Peixoto AJ, Crowley ST, Denbow N, Eliseo D, Pollak J: Comparison of side holes versus non side hole high flow hemodialysis catheters. Hemodial Int 2006, 10: 63-67. 10.1111/j.1542-4758.2006.01176.xCrossRefPubMed
3.
go back to reference Ribot S, Siddiqi SW, Chen C: Right heart complications of dual-lumen tunneled venous catheters in hemodialysis patients. Am J Med Sci 2005, 330: 204-208. 10.1097/00000441-200510000-00011CrossRefPubMed Ribot S, Siddiqi SW, Chen C: Right heart complications of dual-lumen tunneled venous catheters in hemodialysis patients. Am J Med Sci 2005, 330: 204-208. 10.1097/00000441-200510000-00011CrossRefPubMed
4.
go back to reference Tokgoz H, Onaran M, Senocak , Polat F, Sert S: Temporary vascular access via the external iliac vein as a salvage procedure: a report of two cases. J Vasc Access 2005, 6: 200-202.PubMed Tokgoz H, Onaran M, Senocak , Polat F, Sert S: Temporary vascular access via the external iliac vein as a salvage procedure: a report of two cases. J Vasc Access 2005, 6: 200-202.PubMed
5.
go back to reference Unger JK, Janssen VR, Kashefi A, Haltern C, Klosterhalfen B, Fischer Y, Gressner AM, Rossaint R: Enhancing filtration rates by the use of blood flow around the capillaries of plasmafilters: an in vitro study. Int J Artif Organs 2001, 24: 821-831.PubMed Unger JK, Janssen VR, Kashefi A, Haltern C, Klosterhalfen B, Fischer Y, Gressner AM, Rossaint R: Enhancing filtration rates by the use of blood flow around the capillaries of plasmafilters: an in vitro study. Int J Artif Organs 2001, 24: 821-831.PubMed
6.
go back to reference Unger JK, Haltern C, Dohmen B, Gressner A, Grosse-Siestrup C, Groneberg DA, Rossaint R: Albumin and hydroxyethyl starch 130 kD/0.4 improve filter clearance and haemocompatibility in haemo- and plasmafiltration – an in vitro study. Nephrol Dial Transplant 2005, 20: 1922-1931. 10.1093/ndt/gfh913CrossRefPubMed Unger JK, Haltern C, Dohmen B, Gressner A, Grosse-Siestrup C, Groneberg DA, Rossaint R: Albumin and hydroxyethyl starch 130 kD/0.4 improve filter clearance and haemocompatibility in haemo- and plasmafiltration – an in vitro study. Nephrol Dial Transplant 2005, 20: 1922-1931. 10.1093/ndt/gfh913CrossRefPubMed
7.
go back to reference Unger JK, Lemke AJ, Grosse-Siestrup C: Thermography as potential real-time technique to assess changes in flow distribution in hemofiltration. Kidney Int 2006, 69: 520-525. 10.1038/sj.ki.5000143CrossRefPubMed Unger JK, Lemke AJ, Grosse-Siestrup C: Thermography as potential real-time technique to assess changes in flow distribution in hemofiltration. Kidney Int 2006, 69: 520-525. 10.1038/sj.ki.5000143CrossRefPubMed
8.
go back to reference Unger JK, Haltern C, Portz B, Dohmen B, Gressner A, Rossaint R: Relation of haemofilter type to venous catheter resistance is crucial for filtration performance and haemocompatibility in CVVH – an in vitro study. Nephrol Dial Transplant 2006, 21: 2191-2201. 10.1093/ndt/gfl068CrossRefPubMed Unger JK, Haltern C, Portz B, Dohmen B, Gressner A, Rossaint R: Relation of haemofilter type to venous catheter resistance is crucial for filtration performance and haemocompatibility in CVVH – an in vitro study. Nephrol Dial Transplant 2006, 21: 2191-2201. 10.1093/ndt/gfl068CrossRefPubMed
9.
go back to reference Unger JK, Horn NA, Kashefi A, Blumberg A, Klosterhalfen B, Rossaint R: The influence of hypoalbuminemia on maximal flow rates and transmembrane pressure during plasmapheresis – an in vitro study. Blood Purif 2001, 19: 408-416. 10.1159/000046972CrossRefPubMed Unger JK, Horn NA, Kashefi A, Blumberg A, Klosterhalfen B, Rossaint R: The influence of hypoalbuminemia on maximal flow rates and transmembrane pressure during plasmapheresis – an in vitro study. Blood Purif 2001, 19: 408-416. 10.1159/000046972CrossRefPubMed
10.
go back to reference Eger EI 2nd, Johnson BH, Weiskopf RB, Holmes MA, Yasuda N, Targ A, Rampil IJ: Minimum alveolar concentration of I-653 and isoflurane in pigs: definition of a supramaximal stimulus. Anesth Analg 1988, 67: 1174-1176. 10.1213/00000539-198812000-00012CrossRefPubMed Eger EI 2nd, Johnson BH, Weiskopf RB, Holmes MA, Yasuda N, Targ A, Rampil IJ: Minimum alveolar concentration of I-653 and isoflurane in pigs: definition of a supramaximal stimulus. Anesth Analg 1988, 67: 1174-1176. 10.1213/00000539-198812000-00012CrossRefPubMed
11.
go back to reference Journois D, Safran D, Castelain MH, Chanu D, Drévillon C, Barrier G: [Comparison of the anthithrombotic effects of heparin, enoxaparin and prostacyclin in continuous haemofiltration]. Ann Fr Anesth Réanim 1990, 9: 331-337.CrossRefPubMed Journois D, Safran D, Castelain MH, Chanu D, Drévillon C, Barrier G: [Comparison of the anthithrombotic effects of heparin, enoxaparin and prostacyclin in continuous haemofiltration]. Ann Fr Anesth Réanim 1990, 9: 331-337.CrossRefPubMed
12.
go back to reference Tan HK, Bridge N, Baldwin I, Bellomo R: An ex-vivo evaluation of vascular catheters for continuous hemofiltration. Ren Fail 2002, 24: 755-762. 10.1081/JDI-120015678CrossRefPubMed Tan HK, Bridge N, Baldwin I, Bellomo R: An ex-vivo evaluation of vascular catheters for continuous hemofiltration. Ren Fail 2002, 24: 755-762. 10.1081/JDI-120015678CrossRefPubMed
13.
go back to reference Gradman WS, Pozrikidis C: Analysis of options for migrating hemodialysis access-related ischemic steal phenomena. Ann Vasc Surg 2004, 18: 59-65. 10.1007/s10016-003-0103-1CrossRefPubMed Gradman WS, Pozrikidis C: Analysis of options for migrating hemodialysis access-related ischemic steal phenomena. Ann Vasc Surg 2004, 18: 59-65. 10.1007/s10016-003-0103-1CrossRefPubMed
14.
go back to reference Maeda N, Shiga T: Opposite effect of albumin on the erythrocyte aggregation induced by immunoglobulin G and fibrinogen. Biochem Biophys Acta 1986, 855: 127-135.CrossRefPubMed Maeda N, Shiga T: Opposite effect of albumin on the erythrocyte aggregation induced by immunoglobulin G and fibrinogen. Biochem Biophys Acta 1986, 855: 127-135.CrossRefPubMed
15.
go back to reference Stefanidis I, Heintz B, Frank D, Mertens PR, Kierdorf HP: Influence of hematocrit on hemostasis in continuous venovenous hemofiltration during acute renal failure. Kidney Int Suppl 1999, (72):S51-S55. 10.1046/j.1523-1755.56.s72.15.x Stefanidis I, Heintz B, Frank D, Mertens PR, Kierdorf HP: Influence of hematocrit on hemostasis in continuous venovenous hemofiltration during acute renal failure. Kidney Int Suppl 1999, (72):S51-S55. 10.1046/j.1523-1755.56.s72.15.x
16.
go back to reference Schnoor J, Schreck R, Baumert JH, Grosse-Siestrup C, Rossaint R, Unger JK: Influence of differences in body weight and volume management on experimental results in porcine models. Int J Artif Organs 2004, 27: 924-934.PubMed Schnoor J, Schreck R, Baumert JH, Grosse-Siestrup C, Rossaint R, Unger JK: Influence of differences in body weight and volume management on experimental results in porcine models. Int J Artif Organs 2004, 27: 924-934.PubMed
Metadata
Title
Dual-lumen catheters for continuous venovenous hemofiltration: limits for blood delivery via femoral vein access and a potential alternative in an experimental setting in anesthetized pigs
Authors
Juliane K Unger
Klaus Pietzner
Roland C Francis
Juergen Birnbaum
Marc Michael Theisen
Arne-Joern Lemke
Stefan M Niehues
Publication date
01-02-2007
Publisher
BioMed Central
Published in
Critical Care / Issue 1/2007
Electronic ISSN: 1364-8535
DOI
https://doi.org/10.1186/cc5693

Other articles of this Issue 1/2007

Critical Care 1/2007 Go to the issue