Double left ventricular pacing following accidental malpositioning of the right ventricular electrode during implantation of a cardiac resynchronization therapy device

Authors: Ruediger Dissmann, Udo Wolthoff, Markus Zabel

Published in: Journal of Cardiothoracic Surgery | Issue 1/2013

Login to get access

Abstract

Accidental malpositioning of a right ventricular (RV) electrode has not been previously reported in the context of cardiac resynchronization therapy (CRT).
The case of a 75-year old male patient with dilative cardiomyopathy, left ventricular (LV) ejection fraction 23%, New York Heart Association functional heart failure status stage III, left bundle branch block (LBBB) with QRS width of 136 ms, and misplacement of the RV lead to the LV apex during implantation of a CRT defibrillator is described.
Following unremarkable implantation, routine interrogation of the CRT device on the first day after the implantation revealed uneventful technical findings. The 12-lead surface electrocardiogram (ECG) showed biventricular stimulation featuring a narrow QRS complex with incomplete right bundle branch block (RBBB) and R>S in V1. The biplane postoperative chest X-ray was graded normal. On routine follow-up one month later, a transthoracic echocardiogram revealed an increased ejection fraction of 51% but the RV lead was placed in the LV apex. An additional transesophageal echocardiogram exhibited an Eustachian valve guiding the lead via the patent foramen ovale through the mitral valve into the LV apex. Operative revision was scheduled and the active fixation lead was uneventful removed from the LV. A new electrode was inserted and placed in the RV apex.
Accidental malplacement of the RV electrode to the LV may be difficult to diagnose in the context of CRT patients as a stimulated biventricular ECG with incomplete RBBB appearance is expected in this situation. Careful analysis of lateral radiographic views during the operation is important to ensure correct lead positioning. As timely revision is the preferred procedure, early routine transthoracic echocardiography may be considered for detection of malplacement.
Appendix
Available only for authorised users
Literature
1.
go back to reference Dissmann R: Entrapment of pacemaker lead by a large netlike Eustachian valve within the right atrium. Clin Res Cardiol. 2006, 95: 241-243. 10.1007/s00392-006-0368-9.CrossRefPubMed Dissmann R: Entrapment of pacemaker lead by a large netlike Eustachian valve within the right atrium. Clin Res Cardiol. 2006, 95: 241-243. 10.1007/s00392-006-0368-9.CrossRefPubMed
2.
go back to reference Oginosawa Y, Abe H, Nakashima Y: Prevalence of venous anatomic variants and occlusion among patients undergoing implantation of transvenous leads. PACE. 2005, 25: 425-428.CrossRef Oginosawa Y, Abe H, Nakashima Y: Prevalence of venous anatomic variants and occlusion among patients undergoing implantation of transvenous leads. PACE. 2005, 25: 425-428.CrossRef
3.
go back to reference Engström A, Holmberg B, Månsson A, Carlsson J: Inadvertent malposition of a transvenous pacing lead in the left ventricle. Herzschrittmacherther Elektrophysiol. 2006, 17: 221-224. 10.1007/s00399-006-0537-6.CrossRefPubMed Engström A, Holmberg B, Månsson A, Carlsson J: Inadvertent malposition of a transvenous pacing lead in the left ventricle. Herzschrittmacherther Elektrophysiol. 2006, 17: 221-224. 10.1007/s00399-006-0537-6.CrossRefPubMed
4.
go back to reference Rodriquez Y, Baltodano P, Tower A, Martinez C, Carrillo R: Management of symptomatic inadvertently placed endocardial leads in the left ventricle. Pace. 2011, 34: 1192-1200. 10.1111/j.1540-8159.2011.03146.x.CrossRef Rodriquez Y, Baltodano P, Tower A, Martinez C, Carrillo R: Management of symptomatic inadvertently placed endocardial leads in the left ventricle. Pace. 2011, 34: 1192-1200. 10.1111/j.1540-8159.2011.03146.x.CrossRef
5.
go back to reference McManus DD, Mattei M-L, Rose K, Rashkin J, Rosenthal LS: Inadvertent lead placement in the left ventricle: a case report and brief review. Indian Pacing Electrophysiol J. 2009, 9: 224-228.PubMedPubMedCentral McManus DD, Mattei M-L, Rose K, Rashkin J, Rosenthal LS: Inadvertent lead placement in the left ventricle: a case report and brief review. Indian Pacing Electrophysiol J. 2009, 9: 224-228.PubMedPubMedCentral
6.
go back to reference Vanhercke D, Heytens W, Verloove H: Eight years of left ventricular pacing due to inadvertent malposition of a transvenous lead in the left ventricle. Eur J Echocard. 2008, 9: 825-827. 10.1093/ejechocard/jen187.CrossRef Vanhercke D, Heytens W, Verloove H: Eight years of left ventricular pacing due to inadvertent malposition of a transvenous lead in the left ventricle. Eur J Echocard. 2008, 9: 825-827. 10.1093/ejechocard/jen187.CrossRef
7.
go back to reference Van Gelder BM, Bracke FA, Oto A, Yildirir A, Haas PC, Seger JJ, Stainback RF, Botman KJ, Meijer A: Diagnosis and management in inadvertently placed pacing and ICD leads in the left ventricle: A multicenter experience and review of the literature. PACE. 2000, 23: 877-883. 10.1111/j.1540-8159.2000.tb00858.x.CrossRefPubMed Van Gelder BM, Bracke FA, Oto A, Yildirir A, Haas PC, Seger JJ, Stainback RF, Botman KJ, Meijer A: Diagnosis and management in inadvertently placed pacing and ICD leads in the left ventricle: A multicenter experience and review of the literature. PACE. 2000, 23: 877-883. 10.1111/j.1540-8159.2000.tb00858.x.CrossRefPubMed
8.
go back to reference Schön N: Inadvertently placed pacing lead: a case report. J Kardiol. 2007, 14: 228-230. Schön N: Inadvertently placed pacing lead: a case report. J Kardiol. 2007, 14: 228-230.
9.
go back to reference Edwards NC, Varma M, Pitcher DW: Routine chest radiography after permanent pacemaker implantation: is it necessary. J Postgrad Med. 2005, 51: 92-97.PubMed Edwards NC, Varma M, Pitcher DW: Routine chest radiography after permanent pacemaker implantation: is it necessary. J Postgrad Med. 2005, 51: 92-97.PubMed
10.
go back to reference Bauersfeld UK, Thakur RK, Ghani M, Yee R, Klein GJ: Malposition of transvenous pacing lead in the left ventricle: radiographic findings. AJR. 1994, 162: 290-292. 10.2214/ajr.162.2.8310911.CrossRefPubMed Bauersfeld UK, Thakur RK, Ghani M, Yee R, Klein GJ: Malposition of transvenous pacing lead in the left ventricle: radiographic findings. AJR. 1994, 162: 290-292. 10.2214/ajr.162.2.8310911.CrossRefPubMed
11.
go back to reference Yang YN, Yin WH, Young MS: Safe right bundle branch block during permanent right ventricular right ventricular pacing. J Electrocardiol. 2003, 36: 67-71. 10.1054/jelc.2003.50002.CrossRefPubMed Yang YN, Yin WH, Young MS: Safe right bundle branch block during permanent right ventricular right ventricular pacing. J Electrocardiol. 2003, 36: 67-71. 10.1054/jelc.2003.50002.CrossRefPubMed
13.
go back to reference Ross WB, Mohiuddin SM, Pagano T, Hughes D: Malposition of transvenous cardiac electrode associated with amaurosis fugax. PACE. 1983, 6: 119-123. 10.1111/j.1540-8159.1983.tb06589.x.CrossRefPubMed Ross WB, Mohiuddin SM, Pagano T, Hughes D: Malposition of transvenous cardiac electrode associated with amaurosis fugax. PACE. 1983, 6: 119-123. 10.1111/j.1540-8159.1983.tb06589.x.CrossRefPubMed
14.
go back to reference Seki H, Fukui T, Shimokawa T, Manabe S, Watanabe Y, Chino K, Takanashi S: Malpositioning of a pacemaker lead to the left ventricle accompanied by posterior mitral leaflet injury. Interact Cardiovasc Thorac Surg. 2009, 8: 235-237.CrossRefPubMed Seki H, Fukui T, Shimokawa T, Manabe S, Watanabe Y, Chino K, Takanashi S: Malpositioning of a pacemaker lead to the left ventricle accompanied by posterior mitral leaflet injury. Interact Cardiovasc Thorac Surg. 2009, 8: 235-237.CrossRefPubMed
15.
go back to reference Shmuely H, Erdman S, Strasberg B, Rosenfeld JB: Seven years of left ventricular pacing due to malposition of pacing electrode. PACE. 1992, 15: 369-372. 10.1111/j.1540-8159.1992.tb05129.x.CrossRefPubMed Shmuely H, Erdman S, Strasberg B, Rosenfeld JB: Seven years of left ventricular pacing due to malposition of pacing electrode. PACE. 1992, 15: 369-372. 10.1111/j.1540-8159.1992.tb05129.x.CrossRefPubMed
Metadata
Title
Double left ventricular pacing following accidental malpositioning of the right ventricular electrode during implantation of a cardiac resynchronization therapy device
Authors
Ruediger Dissmann
Udo Wolthoff
Markus Zabel
Publication date
01-12-2013
Publisher
BioMed Central
Published in
Journal of Cardiothoracic Surgery / Issue 1/2013
Electronic ISSN: 1749-8090
DOI
https://doi.org/10.1186/1749-8090-8-162

Other articles of this Issue 1/2013

Journal of Cardiothoracic Surgery 1/2013 Go to the issue