01-05-2009 | Computed Tomography

Dose reduction in computed tomography: the effect of eye and testicle shielding on radiation dose measured in patients with beryllium oxide-based optically stimulated luminescence dosimetry

Authors: Henrik Grobe, Marian Sommer, Arne Koch, Volker Hietschold, Jürgen Henniger, Nasreddin Abolmaali

Published in: European Radiology | Issue 5/2009

Login to get access

Abstract

The aim of this study was to assess the effect of eye and testicle shielding on radiation dose to the lens and the testes of patients undergoing CT examinations. Fifty-one male patients underwent CT twice with identical protocols initially without, the second time with protective garments. Doses to the testes and the lenses were recorded with beryllium oxide-based dosimeters. The dose to the testes and lenses from CT exposure was reduced by 96.2% ± 1.7% and 28.2% ± 18.5%, when testicle and eye shielding was used, respectively. The effect of the eye shielding on the eye lens dose was found to depend on the x-ray tube position when the eye is primarily exposed during the scan. The maximum eye lens dose reduction achieved was found to be 43.2% ± 6.5% corresponding to the anterior position of the tube. A significant correlation between the patient’s body mass index and dose exposure could not be found. Eye and testicle shields, apart from being inexpensive and easy to use, were proven to be effective in reducing eye lens and testicle radiation dose burden from CT exposures.
Literature
2.
go back to reference ICRP publication 103 (2007) 2007 recommendations of the International Commission on Radiological Protection. Elsevier, Amsterdam ICRP publication 103 (2007) 2007 recommendations of the International Commission on Radiological Protection. Elsevier, Amsterdam
3.
go back to reference Henk JM, Whitelocke RA et al (1993) Radiation dose to the lens and cataract formation. Int J Radiat Oncol Biol Phys 25(5):815–820PubMed Henk JM, Whitelocke RA et al (1993) Radiation dose to the lens and cataract formation. Int J Radiat Oncol Biol Phys 25(5):815–820PubMed
4.
go back to reference Chodick G, Bekiroglu N et al (2008) Risk of cataract after exposure to low doses of ionizing radiation: A 20-year prospective cohort study among US radiologic technologists. Am J Epidemiology 168(6):620–631CrossRef Chodick G, Bekiroglu N et al (2008) Risk of cataract after exposure to low doses of ionizing radiation: A 20-year prospective cohort study among US radiologic technologists. Am J Epidemiology 168(6):620–631CrossRef
5.
go back to reference Hietschold V, Koch A et al (2008) Computed tomography: influence of varying tube current on patient dose and correctness of effective dose calculations. Rofo 180(5):430–439PubMed Hietschold V, Koch A et al (2008) Computed tomography: influence of varying tube current on patient dose and correctness of effective dose calculations. Rofo 180(5):430–439PubMed
6.
go back to reference Greess H, Wolf H et al (2000) Dose reduction in computed tomography by attenuation-based on-line modulation of tube current: evaluation of six anatomical regions. Eur Radiol 10(2):391–394PubMedCrossRef Greess H, Wolf H et al (2000) Dose reduction in computed tomography by attenuation-based on-line modulation of tube current: evaluation of six anatomical regions. Eur Radiol 10(2):391–394PubMedCrossRef
7.
go back to reference Perisinakis K, Raissaki M et al (2005) Reduction of eye lens radiation dose by orbital bismuth shielding in pediatric patients undergoing CT of the head: a Monte Carlo study. Med Phys 32(4):1024–1030PubMedCrossRef Perisinakis K, Raissaki M et al (2005) Reduction of eye lens radiation dose by orbital bismuth shielding in pediatric patients undergoing CT of the head: a Monte Carlo study. Med Phys 32(4):1024–1030PubMedCrossRef
8.
go back to reference Dauer LT, Casciotta KA et al (2007) Radiation dose reduction at a price: the effectiveness of a male gonadal shield during helical CT scans. BMC Med Imaging 16(7):5CrossRef Dauer LT, Casciotta KA et al (2007) Radiation dose reduction at a price: the effectiveness of a male gonadal shield during helical CT scans. BMC Med Imaging 16(7):5CrossRef
9.
go back to reference Geleijns J, Salvado Artells M et al (2006) Quantitative assessment of selective in-plane shielding of tissues in computed tomography through evaluation of absorbed dose and image quality. Eur Radiol 16(10):2334–2340PubMedCrossRef Geleijns J, Salvado Artells M et al (2006) Quantitative assessment of selective in-plane shielding of tissues in computed tomography through evaluation of absorbed dose and image quality. Eur Radiol 16(10):2334–2340PubMedCrossRef
10.
go back to reference Hohl C, Mahnken AH et al (2005) Radiation dose reduction to the male gonads during MDCT: the effectiveness of a lead shield. AJR Am J Roentgenol 184(1):128–130PubMed Hohl C, Mahnken AH et al (2005) Radiation dose reduction to the male gonads during MDCT: the effectiveness of a lead shield. AJR Am J Roentgenol 184(1):128–130PubMed
11.
go back to reference Hopper KD (2002) Orbital, thyroid, and breast superficial radiation shielding for patients undergoing diagnostic CT. Semin Ultrasound CT MR 23(5):423–427PubMedCrossRef Hopper KD (2002) Orbital, thyroid, and breast superficial radiation shielding for patients undergoing diagnostic CT. Semin Ultrasound CT MR 23(5):423–427PubMedCrossRef
12.
go back to reference Tzedakis A, Damilakis J et al (2005) The effect of z overscanning on patient effective dose from multidetector helical computed tomography examinations. Med Phys 32(6):1621–1629PubMedCrossRef Tzedakis A, Damilakis J et al (2005) The effect of z overscanning on patient effective dose from multidetector helical computed tomography examinations. Med Phys 32(6):1621–1629PubMedCrossRef
13.
go back to reference Sommer M, Henniger J (2006) Investigation of a BeO-based optically stimulated luminescence dosimeter. Radiat Prot Dosimetry 119(1-4):394–397PubMedCrossRef Sommer M, Henniger J (2006) Investigation of a BeO-based optically stimulated luminescence dosimeter. Radiat Prot Dosimetry 119(1-4):394–397PubMedCrossRef
14.
go back to reference Sommer M, Freudenberger R, Henniger J (2007) New aspects of a BeO-based optically stimulated luminescence dosimeter. Radiat Meas 42(4-5):617–620CrossRef Sommer M, Freudenberger R, Henniger J (2007) New aspects of a BeO-based optically stimulated luminescence dosimeter. Radiat Meas 42(4-5):617–620CrossRef
15.
go back to reference Sommer M, Jahn A et al (2008) Beryllium oxide as optically stimulated luminescence dosimeter. Radiat Meas 43(2-6):353–356CrossRef Sommer M, Jahn A et al (2008) Beryllium oxide as optically stimulated luminescence dosimeter. Radiat Meas 43(2-6):353–356CrossRef
16.
go back to reference Moeckel D et al (2007) Quantification of β + activity generated by hard photons by means of PET. Phys Med Biol 52(9):2515–2530CrossRef Moeckel D et al (2007) Quantification of β + activity generated by hard photons by means of PET. Phys Med Biol 52(9):2515–2530CrossRef
17.
go back to reference Straten van M, Venema HW et al (2004) Reproducibility of multi-slice spiral computed tomography scans: an experimental study. Med Phys 31(10):2785–2786CrossRef Straten van M, Venema HW et al (2004) Reproducibility of multi-slice spiral computed tomography scans: an experimental study. Med Phys 31(10):2785–2786CrossRef
Metadata
Title
Dose reduction in computed tomography: the effect of eye and testicle shielding on radiation dose measured in patients with beryllium oxide-based optically stimulated luminescence dosimetry
Authors
Henrik Grobe
Marian Sommer
Arne Koch
Volker Hietschold
Jürgen Henniger
Nasreddin Abolmaali
Publication date
01-05-2009
Publisher
Springer-Verlag
Published in
European Radiology / Issue 5/2009
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-008-1241-1

Other articles of this Issue 5/2009

European Radiology 5/2009 Go to the issue