Dose reduction for fluoroscopically guided injections: phantom simulation and patient procedures

Authors: C. Y. Chang, F. J. Simeone, M. C. DeLorenzo, W. E. Palmer, M. A. Bredella, A. J. Huang

Published in: Skeletal Radiology | Issue 2/2018

Login to get access

Abstract

Objective

To demonstrate fluoroscopy dose reduction through both simulated injections on a phantom and patient injections.

Materials and methods

Our study was IRB-approved and HIPAA-compliant. Simulation on a phantom was used to estimate effective dose, entrance dose, and organ doses for hip joint injections without and with dose minimization technique (DMT). Additionally, 1,094 joint, bursae, and tendon sheath injections performed by three operators in the same fluoroscopy suite were evaluated both before and after application of DMT. Fluoroscopy time (FT), dose, and dose area product (DAP) of injections were compared using unpaired t-tests with P > 0.05 considered statistically significant.

Results

For the phantom simulation comparing injections without DMT and with DMT, the total DAP was 191.7 vs 18.7 μGy·m2, and the entrance dose was 10.2 vs 3.6 mGy, respectively. For both men and women, DMT reduces effective dose and organ doses.
For all injections, the FT (0.7 to 0.2 min), dose (5.6 to 1.9 mGy), and DAP (56.9 to 19.1 μGy·m2) for all three operators decreased with DMT and remained statistically significant when stratified by the two most common injections, glenohumeral and hip joint injections (P < 0.05).

Conclusions

FT, effective dose, entrance dose, and DAP can be reduced with the use of simple easy-to-learn techniques, which will benefit both the patient and the radiologist.
Literature
1.
go back to reference Lambert RGW, Hutchings EJ, Grace MGA, Jhangri GS, Conner-Spady B, Maksymowych WP. Steroid injection for osteoarthritis of the hip: a randomized, double-blind, placebo-controlled trial. Arthritis Rheum. 2007;56:2278–87.CrossRefPubMed Lambert RGW, Hutchings EJ, Grace MGA, Jhangri GS, Conner-Spady B, Maksymowych WP. Steroid injection for osteoarthritis of the hip: a randomized, double-blind, placebo-controlled trial. Arthritis Rheum. 2007;56:2278–87.CrossRefPubMed
2.
3.
go back to reference Rastogi AK, Davis KW, Ross A, Rosas HG. Fundamentals of joint injection. AJR Am J Roentgenol. 2016;207:484–94.CrossRefPubMed Rastogi AK, Davis KW, Ross A, Rosas HG. Fundamentals of joint injection. AJR Am J Roentgenol. 2016;207:484–94.CrossRefPubMed
5.
go back to reference Goske MJ, Applegate KE, Boylan J, Butler PF, Callahan MJ, Coley BD, et al. Image gently(SM): a national education and communication campaign in radiology using the science of social marketing. J Am Coll Radiol JACR. 2008;5:1200–5.CrossRefPubMed Goske MJ, Applegate KE, Boylan J, Butler PF, Callahan MJ, Coley BD, et al. Image gently(SM): a national education and communication campaign in radiology using the science of social marketing. J Am Coll Radiol JACR. 2008;5:1200–5.CrossRefPubMed
6.
go back to reference Brink JA, Amis ES. Image wisely: a campaign to increase awareness about adult radiation protection. Radiology. 2010;257:601–2.CrossRefPubMed Brink JA, Amis ES. Image wisely: a campaign to increase awareness about adult radiation protection. Radiology. 2010;257:601–2.CrossRefPubMed
7.
go back to reference Ouellette HA, Munk PL. Response to letter to the editor. Skelet Radiol. 2016;45:1719.CrossRef Ouellette HA, Munk PL. Response to letter to the editor. Skelet Radiol. 2016;45:1719.CrossRef
8.
go back to reference Goldberg-Stein S, Goldschmiedt J. Letter to the editor. Skelet Radiol. 2016;45:279.CrossRef Goldberg-Stein S, Goldschmiedt J. Letter to the editor. Skelet Radiol. 2016;45:279.CrossRef
9.
go back to reference Messina C, Orlandi D, Sconfienza LM. Do we still need fluoroscopy to perform injections in the musculoskeletal system? Skelet Radiol. 2016;45:1717–8.CrossRef Messina C, Orlandi D, Sconfienza LM. Do we still need fluoroscopy to perform injections in the musculoskeletal system? Skelet Radiol. 2016;45:1717–8.CrossRef
10.
go back to reference Lopez-Ben R, Coumas J. Fluoroscopy versus sonography for image guidance in peripheral MSK intervention. Skelet Radiol. 2016;45:1721–2.CrossRef Lopez-Ben R, Coumas J. Fluoroscopy versus sonography for image guidance in peripheral MSK intervention. Skelet Radiol. 2016;45:1721–2.CrossRef
11.
go back to reference Saunders S, Longworth S. Injection techniques in orthopaedic and sports medicine: a practical manual for doctors and physiotherapists. Elsevier Health Sciences; 2006. Saunders S, Longworth S. Injection techniques in orthopaedic and sports medicine: a practical manual for doctors and physiotherapists. Elsevier Health Sciences; 2006.
12.
go back to reference ICRP. The 2007 recommendations of the International Commission on Radiological Protection. ICRP publication 103. Ann. ICRP. 2007;37:1–332. ICRP. The 2007 recommendations of the International Commission on Radiological Protection. ICRP publication 103. Ann. ICRP. 2007;37:1–332.
13.
go back to reference Hart D, Jones D, Wall B. Normalised organ doses for medical x-ray examinations calculated using Monte Carlo techniques. NRPB-SR262. 1994. Hart D, Jones D, Wall B. Normalised organ doses for medical x-ray examinations calculated using Monte Carlo techniques. NRPB-SR262. 1994.
14.
go back to reference Brenner DJ, Hall EJ. Computed tomography: an increasing source of radiation exposure. N Engl J Med. 2007;357:2277–84.CrossRefPubMed Brenner DJ, Hall EJ. Computed tomography: an increasing source of radiation exposure. N Engl J Med. 2007;357:2277–84.CrossRefPubMed
15.
go back to reference Kumar A, Dhir V, Sharma S, Sharma A, Singh S. Efficacy of methylprednisolone acetate versus triamcinolone acetonide intra-articular knee injection in patients with chronic inflammatory arthritis: a 24-week randomized controlled trial. Clin Ther. 2017;39:150–8.CrossRefPubMed Kumar A, Dhir V, Sharma S, Sharma A, Singh S. Efficacy of methylprednisolone acetate versus triamcinolone acetonide intra-articular knee injection in patients with chronic inflammatory arthritis: a 24-week randomized controlled trial. Clin Ther. 2017;39:150–8.CrossRefPubMed
16.
go back to reference Fazel R, Krumholz HM, Wang Y, Ross JS, Chen J, Ting HH, et al. Exposure to low-dose ionizing radiation from medical imaging procedures. N Engl J Med. 2009;361:849–57.CrossRefPubMedPubMedCentral Fazel R, Krumholz HM, Wang Y, Ross JS, Chen J, Ting HH, et al. Exposure to low-dose ionizing radiation from medical imaging procedures. N Engl J Med. 2009;361:849–57.CrossRefPubMedPubMedCentral
17.
go back to reference Nickoloff EL. AAPM/RSNA physics tutorial for residents: physics of flat-panel fluoroscopy systems: survey of modern fluoroscopy imaging: flat-panel detectors versus image intensifiers and more. Radiogr Rev Publ Radiol Soc N Am Inc. 2011;31:591–602. Nickoloff EL. AAPM/RSNA physics tutorial for residents: physics of flat-panel fluoroscopy systems: survey of modern fluoroscopy imaging: flat-panel detectors versus image intensifiers and more. Radiogr Rev Publ Radiol Soc N Am Inc. 2011;31:591–602.
18.
go back to reference Rauch P, Lin P-JP, Balter S, Fukuda A, Goode A, Hartwell G, et al. Functionality and operation of fluoroscopic automatic brightness control/automatic dose rate control logic in modern cardiovascular and interventional angiography systems: a report of task group 125 radiography/fluoroscopy subcommittee, imaging physics committee. Science Council Med Phys. 2012;39:2826–8.PubMed Rauch P, Lin P-JP, Balter S, Fukuda A, Goode A, Hartwell G, et al. Functionality and operation of fluoroscopic automatic brightness control/automatic dose rate control logic in modern cardiovascular and interventional angiography systems: a report of task group 125 radiography/fluoroscopy subcommittee, imaging physics committee. Science Council Med Phys. 2012;39:2826–8.PubMed
20.
go back to reference Singh J, Khan WS, Marwah S, Wells G, Tannous DK, Sharma HK. Do we need radiological guidance for intra-articular hip injections? Open Orthop J. 2014;8:114–7.CrossRefPubMedPubMedCentral Singh J, Khan WS, Marwah S, Wells G, Tannous DK, Sharma HK. Do we need radiological guidance for intra-articular hip injections? Open Orthop J. 2014;8:114–7.CrossRefPubMedPubMedCentral
21.
go back to reference Kurup H, Ward P. Do we need radiological guidance for hip joint injections? Acta Orthop Belg. 2010;76:205–7.PubMed Kurup H, Ward P. Do we need radiological guidance for hip joint injections? Acta Orthop Belg. 2010;76:205–7.PubMed
22.
go back to reference Diraçoğlu D, Alptekin K, Dikici F, Balci HI, Ozçakar L, Aksoy C. Evaluation of needle positioning during blind intra-articular hip injections for osteoarthritis: fluoroscopy versus arthrography. Arch Phys Med Rehabil. 2009;90:2112–5.CrossRefPubMed Diraçoğlu D, Alptekin K, Dikici F, Balci HI, Ozçakar L, Aksoy C. Evaluation of needle positioning during blind intra-articular hip injections for osteoarthritis: fluoroscopy versus arthrography. Arch Phys Med Rehabil. 2009;90:2112–5.CrossRefPubMed
23.
go back to reference Mei-Dan O, McConkey MO, Petersen B, McCarty E, Moreira B, Young DA. The anterior approach for a non-image-guided intra-articular hip injection. Arthrosc. J Arthrosc Relat Surg Off Publ Arthrosc Assoc N Am Int Arthrosc Assoc. 2013;29:1025–33.CrossRef Mei-Dan O, McConkey MO, Petersen B, McCarty E, Moreira B, Young DA. The anterior approach for a non-image-guided intra-articular hip injection. Arthrosc. J Arthrosc Relat Surg Off Publ Arthrosc Assoc N Am Int Arthrosc Assoc. 2013;29:1025–33.CrossRef
Metadata
Title
Dose reduction for fluoroscopically guided injections: phantom simulation and patient procedures
Authors
C. Y. Chang
F. J. Simeone
M. C. DeLorenzo
W. E. Palmer
M. A. Bredella
A. J. Huang
Publication date
01-02-2018
Publisher
Springer Berlin Heidelberg
Published in
Skeletal Radiology / Issue 2/2018
Print ISSN: 0364-2348
Electronic ISSN: 1432-2161
DOI
https://doi.org/10.1007/s00256-017-2788-1

Other articles of this Issue 2/2018

Skeletal Radiology 2/2018 Go to the issue

Browser's Notes

Browser’s notes