Published in: Neuroradiology 7/2016

01-07-2016 | Interventional Neuroradiology

Dose comparison of classical 2-plane DSA and 3D rotational angiography for the assessment of intracranial aneurysms

Authors: N. Guberina, U. Lechel, M. Forsting, C. Mönninghoff, U. Dietrich, A. Ringelstein

Published in: Neuroradiology | Issue 7/2016

Login to get access

Abstract

Introduction

The purpose of this experimental phantom study was to compare radiation doses imparted to patients undergoing classical two-plane digital subtraction angiography (2-plane DSA) and 3D rotational angiography in interventional neuroradiology.

Methods

Thermoluminescence dosimeter (TLD) measurements were performed at an anthropomorphic phantom using a digital interventional angiography system. Two-plane DSA included posterior/anterior (PA) and lateral (LAT) projections (frame-rate, 7.6 frames (PA) and 9.8 frames (LAT) for a scan time of approximately 8 s; image intensifier 27 cm (PA) and 25 cm (LAT)). For 3D rotational angiography, 122 images were acquired from one single image run with the imaging system rotating 240° around the phantom’s head (image intensifier 37 cm).

Results

Effective dose was 0.4 mSv for 2-plane DSA compared to 0.1 mSv for 3D rotational angiography. Organ doses were approximately two to five times higher for classical 2-plane technique compared to the 3D rotational angiography, respectively: brain (11.4 vs. 2.4 mSv), eye lens (4.5 vs.1 mSv), salivary glands (7 vs. 1,7 mSv), oral mucosa (2.7 vs.0.9 mSv), thyroid (0.5 vs. 0.2 mSv), thymus (0.2 vs. 0.05 mSv), bone marrow within imaged region (1 vs. 0.2 mSv), oesophagus (0.07 vs. 0.03 mSv), endotracheal system (2.6 vs. 0.7 mSv) and skeletal components in the imaged region (0.7 vs. 0.2 mSv).

Conclusion

Three-dimensional rotational angiography clearly reduces radiation doses compared to the classical 2-plane technique. Replacement of additional 2-plane DSA projections with 3D rotational angiography will lead to a remarkable decrease in patient radiation dose, without loss of image quality. Thus, we recommend routine application of 3D rotational angiography, in particular for diagnostic assessment of aneurysm morphology.
Literature
1.
go back to reference Ringelstein A, Mueller O, Mönninghoff C, Hahnemann ML, Sure U, Forsting M, Schlamann M (2014) 3D rotational angiography after non-traumatic SAH. Röfo 186(7):675–679PubMed Ringelstein A, Mueller O, Mönninghoff C, Hahnemann ML, Sure U, Forsting M, Schlamann M (2014) 3D rotational angiography after non-traumatic SAH. Röfo 186(7):675–679PubMed
2.
go back to reference Abe T, Hirohata M, Tanaka N et al (2002) Clinical benefits of rotational 3D angiography in endovascular treatment of ruptured cerebral aneurysms. AJNR 23:686–688PubMed Abe T, Hirohata M, Tanaka N et al (2002) Clinical benefits of rotational 3D angiography in endovascular treatment of ruptured cerebral aneurysms. AJNR 23:686–688PubMed
3.
go back to reference Hirai T, Korogi Y, Suginohara K et al (2003) Clinical usefulness of unsubtracted 3D digital angiography compared with rotational digital angiography in the pretreatment evaluation of intracranial aneurysms. AJNR 24:1067–1074PubMed Hirai T, Korogi Y, Suginohara K et al (2003) Clinical usefulness of unsubtracted 3D digital angiography compared with rotational digital angiography in the pretreatment evaluation of intracranial aneurysms. AJNR 24:1067–1074PubMed
4.
go back to reference Alexander MD, Oliff MC, Olorunsola OG et al (2010) Patient radiation exposure during diagnostic and therapeutic interventional neuroradiology procedures. J Neurointerv Surg 2:6–10CrossRefPubMed Alexander MD, Oliff MC, Olorunsola OG et al (2010) Patient radiation exposure during diagnostic and therapeutic interventional neuroradiology procedures. J Neurointerv Surg 2:6–10CrossRefPubMed
5.
go back to reference Marshall NW, Noble J, Faulkner K (1995) Patient and staff dosimetry in neuroradiological procedures. Br J Radiol 68:495–501CrossRefPubMed Marshall NW, Noble J, Faulkner K (1995) Patient and staff dosimetry in neuroradiological procedures. Br J Radiol 68:495–501CrossRefPubMed
6.
go back to reference Gosch D, Kurze W, Deckert F, Schulz T, Patz A, Kahn T (2006) Radiation exposure with 3D DSA of the skull. Röfo 178(9):880–885PubMed Gosch D, Kurze W, Deckert F, Schulz T, Patz A, Kahn T (2006) Radiation exposure with 3D DSA of the skull. Röfo 178(9):880–885PubMed
7.
go back to reference Bridcut RR, Murphy E, Workman A et al (2007) Patient dose from 3D rotational neurovascular studies. Br J Radiol 80:362–366CrossRefPubMed Bridcut RR, Murphy E, Workman A et al (2007) Patient dose from 3D rotational neurovascular studies. Br J Radiol 80:362–366CrossRefPubMed
8.
go back to reference Struffert T, Hauer M, Banckwitz R, Köhler C, Royalty K, Doerfler A (2014) Effective dose to patient measurements in flat-detector and multislice computed tomography: a comparison of applications in neuroradiology. Eur Radiol 24(6):1257–1265CrossRefPubMed Struffert T, Hauer M, Banckwitz R, Köhler C, Royalty K, Doerfler A (2014) Effective dose to patient measurements in flat-detector and multislice computed tomography: a comparison of applications in neuroradiology. Eur Radiol 24(6):1257–1265CrossRefPubMed
9.
go back to reference Lechel U, Becker C, Langenfeld-Jäger G, Brix G (2009) Dose reduction by automatic exposure control in multi-slice computed tomography − comparison between measurement and calculation. Eur Radiology 19:1027–1034CrossRef Lechel U, Becker C, Langenfeld-Jäger G, Brix G (2009) Dose reduction by automatic exposure control in multi-slice computed tomography − comparison between measurement and calculation. Eur Radiology 19:1027–1034CrossRef
10.
go back to reference ICRP Publication 103 (2007) Recommendations of the international commission on radiological protection. Annals ICRP Elsevier Sci, Oxford 37:2–4 ICRP Publication 103 (2007) Recommendations of the international commission on radiological protection. Annals ICRP Elsevier Sci, Oxford 37:2–4
11.
go back to reference Cristy M, Eckerman KF (1987) Specific absorbed fractions of energy at various ages from internal photon sources. I. Methods. Oak Ridge National Laboratory, Oak Ridge, Tennessee; Report ORNL/TM-8381/V1 Cristy M, Eckerman KF (1987) Specific absorbed fractions of energy at various ages from internal photon sources. I. Methods. Oak Ridge National Laboratory, Oak Ridge, Tennessee; Report ORNL/TM-8381/V1
12.
go back to reference Conference of Radiation Control Program Directors, Inc (CRCPD) 205 Capital Avenue Frankfort, KY 40601; Reviewed/Republished: September 2008 www.crcpd.org Conference of Radiation Control Program Directors, Inc (CRCPD) 205 Capital Avenue Frankfort, KY 40601; Reviewed/Republished: September 2008 www.​crcpd.​org
13.
go back to reference Kiyosue H, Tanoue S, Okahara M et al (2002) Anatomic features predictive of complete aneurysm occlusion can be determined with three-dimensional digital subtraction angiography. AJNR 23:1206–1213PubMed Kiyosue H, Tanoue S, Okahara M et al (2002) Anatomic features predictive of complete aneurysm occlusion can be determined with three-dimensional digital subtraction angiography. AJNR 23:1206–1213PubMed
14.
go back to reference Anxionnat R, Bracard S, Ducrocq X et al (2001) Intracranial aneurysms: clinical value of 3D digital subtraction angiography in the therapeutic decision and endovascular treatment. Radiology 218:799–808CrossRefPubMed Anxionnat R, Bracard S, Ducrocq X et al (2001) Intracranial aneurysms: clinical value of 3D digital subtraction angiography in the therapeutic decision and endovascular treatment. Radiology 218:799–808CrossRefPubMed
15.
go back to reference Sugahara T, Korogi Y, Nakashima K et al (2002) Comparison of 2D and 3D digital subtraction angiography in evaluation of intracranial aneurysms. AJNR 23:1545–1552PubMed Sugahara T, Korogi Y, Nakashima K et al (2002) Comparison of 2D and 3D digital subtraction angiography in evaluation of intracranial aneurysms. AJNR 23:1545–1552PubMed
16.
go back to reference Van Rooij WJ, Sprengers ME, De Gast AN et al (2008) 3D rotational angiography: the new gold standard in the detection of additional intracranial aneurysms. AJNR 29:976–979CrossRefPubMed Van Rooij WJ, Sprengers ME, De Gast AN et al (2008) 3D rotational angiography: the new gold standard in the detection of additional intracranial aneurysms. AJNR 29:976–979CrossRefPubMed
17.
go back to reference Lescher S, Samaana T, Berkefeld J (2014) Evaluation of the pontine perforators of the basilar artery using digital subtraction angiography in high resolution and 3D rotation technique. AJNR 35(10):1942–1947CrossRefPubMed Lescher S, Samaana T, Berkefeld J (2014) Evaluation of the pontine perforators of the basilar artery using digital subtraction angiography in high resolution and 3D rotation technique. AJNR 35(10):1942–1947CrossRefPubMed
18.
go back to reference Van Rooij WJ, Pelusoa JPP, Sluzewskia M, Beute GN (2008) Additional value of 3D rotational angiography in angiographically negative aneurysmal subarachnoid hemorrhage: how negative is negative? AJNR 29:962–966CrossRefPubMed Van Rooij WJ, Pelusoa JPP, Sluzewskia M, Beute GN (2008) Additional value of 3D rotational angiography in angiographically negative aneurysmal subarachnoid hemorrhage: how negative is negative? AJNR 29:962–966CrossRefPubMed
19.
go back to reference Ishihara H, Kato S, Akimura T et al (2007) Angiogram-negative subarachnoid hemorrhage in the era of three dimensional rotational angiography. J Clin Neurosci 14:252–255CrossRefPubMed Ishihara H, Kato S, Akimura T et al (2007) Angiogram-negative subarachnoid hemorrhage in the era of three dimensional rotational angiography. J Clin Neurosci 14:252–255CrossRefPubMed
20.
go back to reference Brinjikji W, Cloft H, Lanzino G, Kallmes DF (2009) Comparison of 2D digital subtraction angiography and 3D rotational angiography in the evaluation of dome-to-neck ratio. AJNR 30:831–834CrossRefPubMed Brinjikji W, Cloft H, Lanzino G, Kallmes DF (2009) Comparison of 2D digital subtraction angiography and 3D rotational angiography in the evaluation of dome-to-neck ratio. AJNR 30:831–834CrossRefPubMed
Metadata
Title
Dose comparison of classical 2-plane DSA and 3D rotational angiography for the assessment of intracranial aneurysms
Authors
N. Guberina
U. Lechel
M. Forsting
C. Mönninghoff
U. Dietrich
A. Ringelstein
Publication date
01-07-2016
Publisher
Springer Berlin Heidelberg
Published in
Neuroradiology / Issue 7/2016
Print ISSN: 0028-3940
Electronic ISSN: 1432-1920
DOI
https://doi.org/10.1007/s00234-016-1671-4

Other articles of this Issue 7/2016

Neuroradiology 7/2016 Go to the issue