01-11-2012 | Symposium: Papers Presented at the 2011 Meeting of the International Hip Society

Does Standing Affect Acetabular Component Inclination and Version After THA?

Authors: Gregory G. Polkowski, MD, Ryan M. Nunley, MD, Erin L. Ruh, MS, Brandon M. Williams, DC, Robert L. Barrack, MD

Published in: Clinical Orthopaedics and Related Research® | Issue 11/2012

Login to get access

Abstract

Background

Avoiding complications after hip arthroplasty with hard-on-hard bearings, especially metal-on-metal, correlates with the position of the acetabular component. Supine imaging with conventional radiography has traditionally been utilized to assess component inclination (abduction), as well as anteversion, after THA and surface replacement arthroplasty (SRA). However, most adverse events with hard bearings (excessive wear and squeaking) have occurred with loading. Standing imaging, therefore, should provide more appropriate measurements.

Questions/purposes

We determined whether standing changed standard measurements of acetabular component position using a novel biplanar imaging system compared to traditional supine imaging.

Methods

We performed simultaneous biplanar standing imaging of the lower extremity with a novel imaging system using low radiation collimated beam on 46 patients who underwent THA (23) or SRA (23). Patients who had previously undergone THA had standard CT scans performed. For patients who underwent SRA, we compared acetabular inclination in the supine versus double-limb and single-limb standing.

Results

Standing anteversion differed from supine anteversion by greater than 5° for 12 of 23 patients who underwent THA (range, 5°–16°). For patients who underwent SRA, 13 of 23 patients exhibited a difference of greater than 3° in inclination between supine and double-limb standing images, and six of 23 patients exhibited a difference of greater than 3° in inclination between supine and single-limb standing images.

Conclusions

Standing changed the acetabular inclination and version in a substantial percentage of patients undergoing hip arthroplasty.
Literature
1.
go back to reference Babisch JW, Layher F, Amiot LP. The rationale for tilt-adjusted acetabular cup navigation. J Bone Joint Surg Am. 2008;90:357–365.PubMedCrossRef Babisch JW, Layher F, Amiot LP. The rationale for tilt-adjusted acetabular cup navigation. J Bone Joint Surg Am. 2008;90:357–365.PubMedCrossRef
2.
go back to reference Beckmann J, Luring C, Tingart M, Anders S, Grifka J, Kock FX. Cup positioning in THA: current status and pitfalls: a systematic evaluation of the literature. Arch Orthop Trauma Surg. 2009;129:863–872.PubMedCrossRef Beckmann J, Luring C, Tingart M, Anders S, Grifka J, Kock FX. Cup positioning in THA: current status and pitfalls: a systematic evaluation of the literature. Arch Orthop Trauma Surg. 2009;129:863–872.PubMedCrossRef
3.
go back to reference Bernasek T, Fisher D, Dalury D, Levering M, Dimitris K. Is metal-on-metal squeaking related to acetabular angle of inclination? Clin Orthop Relat Res. 2011;469:2577–2582.PubMedCrossRef Bernasek T, Fisher D, Dalury D, Levering M, Dimitris K. Is metal-on-metal squeaking related to acetabular angle of inclination? Clin Orthop Relat Res. 2011;469:2577–2582.PubMedCrossRef
4.
go back to reference Blondel B, Parratte S, Tropiano P, Pauly V, Aubaniac JM, Argenson JN. Pelvic tilt measurement before and after total hip arthroplasty. Orthop Traumatol Surg Res. 2009;95:568–572.PubMedCrossRef Blondel B, Parratte S, Tropiano P, Pauly V, Aubaniac JM, Argenson JN. Pelvic tilt measurement before and after total hip arthroplasty. Orthop Traumatol Surg Res. 2009;95:568–572.PubMedCrossRef
5.
go back to reference Callanan MC, Jarrett B, Bragdon CR, Zurakowski D, Rubash HE, Freiberg AA, Malchau H. The John Charnley Award. Risk factors for cup malpositioning: quality improvement through a joint registry at a tertiary hospital. Clin Orthop Relat Res. 2011;469:319–329.PubMedCrossRef Callanan MC, Jarrett B, Bragdon CR, Zurakowski D, Rubash HE, Freiberg AA, Malchau H. The John Charnley Award. Risk factors for cup malpositioning: quality improvement through a joint registry at a tertiary hospital. Clin Orthop Relat Res. 2011;469:319–329.PubMedCrossRef
6.
go back to reference Chaibi Y, Cresson T, Aubert B, Hausselle J, Neyret P, Hauger O, de Guise JA, Skalli W. Fast 3D reconstruction of the lower limb using a parametric model and statistical inferences and clinical measurements calculation from biplanar X-rays. Comput Methods Biomech Biomed Engin. 2012;15:457–466.PubMedCrossRef Chaibi Y, Cresson T, Aubert B, Hausselle J, Neyret P, Hauger O, de Guise JA, Skalli W. Fast 3D reconstruction of the lower limb using a parametric model and statistical inferences and clinical measurements calculation from biplanar X-rays. Comput Methods Biomech Biomed Engin. 2012;15:457–466.PubMedCrossRef
7.
go back to reference Dandachli W, Nakhla A, Iranpour F, Kannan V, Cobb JP. Can the acetabular position be derived from a pelvic frame of reference? Clin Orthop Relat Res. 2009;467:886–893.PubMedCrossRef Dandachli W, Nakhla A, Iranpour F, Kannan V, Cobb JP. Can the acetabular position be derived from a pelvic frame of reference? Clin Orthop Relat Res. 2009;467:886–893.PubMedCrossRef
8.
go back to reference Deschenes S, Charron G, Beaudoin G, Labelle H, Dubois J, Miron MC, Parent S. Diagnostic imaging of spinal deformities: reducing patients’ radiation dose with a new slot-scanning X-ray imager. Spine (Phila Pa 1976). 2010;35:989–994.CrossRef Deschenes S, Charron G, Beaudoin G, Labelle H, Dubois J, Miron MC, Parent S. Diagnostic imaging of spinal deformities: reducing patients’ radiation dose with a new slot-scanning X-ray imager. Spine (Phila Pa 1976). 2010;35:989–994.CrossRef
9.
go back to reference Frank ED, Long BW, Smith BJ, Merrill V. Merrill’s Atlas of Radiographic Positioning and Procedures. 11th ed. St Louis, MO: Mosby/Elsevier; 2007. Frank ED, Long BW, Smith BJ, Merrill V. Merrill’s Atlas of Radiographic Positioning and Procedures. 11th ed. St Louis, MO: Mosby/Elsevier; 2007.
10.
go back to reference Hart AJ, Skinner JA, Henckel J, Sampson B, Gordon F. Insufficient acetabular version increases blood metal ion levels after metal-on-metal hip resurfacing. Clin Orthop Relat Res. 2011;469:2590–2597.PubMedCrossRef Hart AJ, Skinner JA, Henckel J, Sampson B, Gordon F. Insufficient acetabular version increases blood metal ion levels after metal-on-metal hip resurfacing. Clin Orthop Relat Res. 2011;469:2590–2597.PubMedCrossRef
11.
go back to reference Jolles BM, Zangger P, Leyvraz PF. Factors predisposing to dislocation after primary total hip arthroplasty: a multivariate analysis. J Arthroplasty. 2002;17:282–288.PubMedCrossRef Jolles BM, Zangger P, Leyvraz PF. Factors predisposing to dislocation after primary total hip arthroplasty: a multivariate analysis. J Arthroplasty. 2002;17:282–288.PubMedCrossRef
12.
go back to reference Journe A, Belicourt C, Sautet A, Dousounian L. [New method for measuring the anteversion and inclination of an acetabular implant using the EOS stereoradiographic system: evaluation of the inter- and intra-observer reproducibility] [in French]. Rev Chir Orthop Traumatol. 2010;96(suppl):80S. Journe A, Belicourt C, Sautet A, Dousounian L. [New method for measuring the anteversion and inclination of an acetabular implant using the EOS stereoradiographic system: evaluation of the inter- and intra-observer reproducibility] [in French]. Rev Chir Orthop Traumatol. 2010;96(suppl):80S.
13.
go back to reference Kalteis TA, Handel M, Herbst B, Grifka J, Renkawitz T. In vitro investigation of the influence of pelvic tilt on acetabular cup alignment. J Arthroplasty. 2009;24:152–157.PubMedCrossRef Kalteis TA, Handel M, Herbst B, Grifka J, Renkawitz T. In vitro investigation of the influence of pelvic tilt on acetabular cup alignment. J Arthroplasty. 2009;24:152–157.PubMedCrossRef
14.
go back to reference Kosak R, Kralj-Iglic V, Iglic A, Daniel M. Polyethylene wear is related to patient-specific contact stress in THA. Clin Orthop Relat Res. 2011;469:3415–3422.PubMedCrossRef Kosak R, Kralj-Iglic V, Iglic A, Daniel M. Polyethylene wear is related to patient-specific contact stress in THA. Clin Orthop Relat Res. 2011;469:3415–3422.PubMedCrossRef
15.
go back to reference Lazennec JY, Boyer P, Gorin M, Catonne Y, Rousseau MA. Acetabular anteversion with CT in supine, simulated standing, and sitting positions in a THA patient population. Clin Orthop Relat Res. 2011;469:1103–1109.PubMedCrossRef Lazennec JY, Boyer P, Gorin M, Catonne Y, Rousseau MA. Acetabular anteversion with CT in supine, simulated standing, and sitting positions in a THA patient population. Clin Orthop Relat Res. 2011;469:1103–1109.PubMedCrossRef
16.
go back to reference Lazennec JY, Brusson A, Rousseau MA. Hip-spine relations and sagittal balance clinical consequences. Eur Spine J. 2011;20(suppl 5):686–698.PubMedCrossRef Lazennec JY, Brusson A, Rousseau MA. Hip-spine relations and sagittal balance clinical consequences. Eur Spine J. 2011;20(suppl 5):686–698.PubMedCrossRef
17.
go back to reference Lazennec JY, Rousseau MA, Rangel A, Gorin M, Belicourt C, Brusson A, Catonne Y. Pelvis and total hip arthroplasty acetabular component orientations in sitting and standing positions: measurements reproductibility with EOS imaging system versus conventional radiographies. Orthop Traumatol Surg Res. 2011;97:373–380.PubMedCrossRef Lazennec JY, Rousseau MA, Rangel A, Gorin M, Belicourt C, Brusson A, Catonne Y. Pelvis and total hip arthroplasty acetabular component orientations in sitting and standing positions: measurements reproductibility with EOS imaging system versus conventional radiographies. Orthop Traumatol Surg Res. 2011;97:373–380.PubMedCrossRef
18.
go back to reference Lembeck B, Mueller O, Reize P, Wuelker N. Pelvic tilt makes acetabular cup navigation inaccurate. Acta Orthop. 2005;76:517–523.PubMedCrossRef Lembeck B, Mueller O, Reize P, Wuelker N. Pelvic tilt makes acetabular cup navigation inaccurate. Acta Orthop. 2005;76:517–523.PubMedCrossRef
19.
go back to reference Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am. 1978;60:217–220.PubMed Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am. 1978;60:217–220.PubMed
20.
go back to reference Mall NA, Nunley RM, Zhu JJ, Maloney WJ, Barrack RL, Clohisy JC. The incidence of acetabular osteolysis in young patients with conventional versus highly crosslinked polyethylene. Clin Orthop Relat Res. 2011;469:372–381.PubMedCrossRef Mall NA, Nunley RM, Zhu JJ, Maloney WJ, Barrack RL, Clohisy JC. The incidence of acetabular osteolysis in young patients with conventional versus highly crosslinked polyethylene. Clin Orthop Relat Res. 2011;469:372–381.PubMedCrossRef
21.
go back to reference Murray DW. The definition and measurement of acetabular orientation. J Bone Joint Surg Br. 1993;75:228–232.PubMed Murray DW. The definition and measurement of acetabular orientation. J Bone Joint Surg Br. 1993;75:228–232.PubMed
22.
go back to reference Paterno SA, Lachiewicz PF, Kelley SS. The influence of patient-related factors and the position of the acetabular component on the rate of dislocation after total hip replacement. J Bone Joint Surg Am. 1997;79:1202–1210.PubMed Paterno SA, Lachiewicz PF, Kelley SS. The influence of patient-related factors and the position of the acetabular component on the rate of dislocation after total hip replacement. J Bone Joint Surg Am. 1997;79:1202–1210.PubMed
23.
go back to reference Sexton SA, Yeung E, Jackson MP, Rajaratnam S, Martell JM, Walter WL, Zicat BA, Walter WK. The role of patient factors and implant position in squeaking of ceramic-on-ceramic total hip replacements. J Bone Joint Surg Br. 2011;93:439–442.PubMed Sexton SA, Yeung E, Jackson MP, Rajaratnam S, Martell JM, Walter WL, Zicat BA, Walter WK. The role of patient factors and implant position in squeaking of ceramic-on-ceramic total hip replacements. J Bone Joint Surg Br. 2011;93:439–442.PubMed
24.
go back to reference Wan Z, Boutary M, Dorr LD. The influence of acetabular component position on wear in total hip arthroplasty. J Arthroplasty. 2008;23:51–56.PubMedCrossRef Wan Z, Boutary M, Dorr LD. The influence of acetabular component position on wear in total hip arthroplasty. J Arthroplasty. 2008;23:51–56.PubMedCrossRef
25.
go back to reference Zhu J, Wan Z, Dorr LD. Quantification of pelvic tilt in total hip arthroplasty. Clin Orthop Relat Res. 2010;468:571–575.PubMedCrossRef Zhu J, Wan Z, Dorr LD. Quantification of pelvic tilt in total hip arthroplasty. Clin Orthop Relat Res. 2010;468:571–575.PubMedCrossRef
Metadata
Title
Does Standing Affect Acetabular Component Inclination and Version After THA?
Authors
Gregory G. Polkowski, MD
Ryan M. Nunley, MD
Erin L. Ruh, MS
Brandon M. Williams, DC
Robert L. Barrack, MD
Publication date
01-11-2012
Publisher
Springer-Verlag
Published in
Clinical Orthopaedics and Related Research® / Issue 11/2012
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-012-2391-7

Other articles of this Issue 11/2012

Clinical Orthopaedics and Related Research® 11/2012 Go to the issue

Symposium: Papers Presented at the 2011 Meeting of the International Hip Society

Femoral Bone Is Preserved Using Cemented Polished Stems in Young Patients