Do we need 3D tube current modulation information for accurate organ dosimetry in chest CT? Protocols dose comparisons

Authors: Xochitl Lopez-Rendon, Guozhi Zhang, Walter Coudyzer, Wim Develter, Hilde Bosmans, Federica Zanca

Published in: European Radiology | Issue 11/2017

Login to get access

Abstract

Objectives

To compare the lung and breast dose associated with three chest protocols: standard, organ-based tube current modulation (OBTCM) and fast-speed scanning; and to estimate the error associated with organ dose when modelling the longitudinal (z-) TCM versus the 3D-TCM in Monte Carlo simulations (MC) for these three protocols.

Method

Five adult and three paediatric cadavers with different BMI were scanned. The CTDIvol of the OBTCM and the fast-speed protocols were matched to the patient-specific CTDIvol of the standard protocol. Lung and breast doses were estimated using MC with both z- and 3D-TCM simulated and compared between protocols.

Results

The fast-speed scanning protocol delivered the highest doses. A slight reduction for breast dose (up to 5.1%) was observed for two of the three female cadavers with the OBTCM in comparison to the standard. For both adult and paediatric, the implementation of the z-TCM data only for organ dose estimation resulted in 10.0% accuracy for the standard and fast-speed protocols, while relative dose differences were up to 15.3% for the OBTCM protocol.

Conclusion

At identical CTDIvol values, the standard protocol delivered the lowest overall doses. Only for the OBTCM protocol is the 3D-TCM needed if an accurate (<10.0%) organ dosimetry is desired.

Key points

The z-TCM information is sufficient for accurate dosimetry for standard protocols.
The z-TCM information is sufficient for accurate dosimetry for fast-speed scanning protocols.
For organ-based TCM schemes, the 3D-TCM information is necessary for accurate dosimetry.
At identical CTDI vol , the fast-speed scanning protocol delivered the highest doses.
Lung dose was higher in XCare than standard protocol at identical CTDI vol .
Literature
1.
go back to reference Gunn ML, Kohr JR (2010) State of the art: technologies for computed tomography dose reduction. Emerg Radiol 17:209–218CrossRef Gunn ML, Kohr JR (2010) State of the art: technologies for computed tomography dose reduction. Emerg Radiol 17:209–218CrossRef
2.
go back to reference Van Straten M, Deak P, Shrimpton PC, Kalender WA (2009) The effect of angular and longitudinal tube current modulations on the estimation of organ and effective doses in x-ray computed tomography. Med Phys 36:4881–4889CrossRef Van Straten M, Deak P, Shrimpton PC, Kalender WA (2009) The effect of angular and longitudinal tube current modulations on the estimation of organ and effective doses in x-ray computed tomography. Med Phys 36:4881–4889CrossRef
3.
go back to reference Lungren MP, Yoshizumi TT, Brady SM et al (2012) Radiation dose estimations to the thorax using organ-based dose modulation. AJR Am J Roentgenol 199:W65–W73CrossRef Lungren MP, Yoshizumi TT, Brady SM et al (2012) Radiation dose estimations to the thorax using organ-based dose modulation. AJR Am J Roentgenol 199:W65–W73CrossRef
4.
go back to reference Vollmar SV, Kalender WA (2008) Reduction of dose to the female breast in thoracic CT: a comparison of standard-protocol, bismuth-shielded, partial and tube-current modulated CT examinations. Eur Radiol 18:674–1682CrossRef Vollmar SV, Kalender WA (2008) Reduction of dose to the female breast in thoracic CT: a comparison of standard-protocol, bismuth-shielded, partial and tube-current modulated CT examinations. Eur Radiol 18:674–1682CrossRef
5.
go back to reference Duan X, Wang J, Christner JA, Leng S, Grant KL, McCollough CH (2011) Dose reduction to anterior surfaces with organ-based-tube current modulation: evaluation of performance in a phantom study. AJR Am J Roentgenol 197:689–695CrossRef Duan X, Wang J, Christner JA, Leng S, Grant KL, McCollough CH (2011) Dose reduction to anterior surfaces with organ-based-tube current modulation: evaluation of performance in a phantom study. AJR Am J Roentgenol 197:689–695CrossRef
6.
go back to reference Seidenfuss A, Mayr A, Schmid M, Uder M, Lell MM (2014) Dose reduction of the female breasts in chest CT. AJR 202:W447–W452CrossRef Seidenfuss A, Mayr A, Schmid M, Uder M, Lell MM (2014) Dose reduction of the female breasts in chest CT. AJR 202:W447–W452CrossRef
7.
go back to reference Gandhi D, Crotty D, Stevens G, Schmidt TG (2015) Technical note: phantom study to evaluate the dose and image quality effects of a computed tomography organ-based tube current modulation technique. Med Phys 42:6572–6578CrossRef Gandhi D, Crotty D, Stevens G, Schmidt TG (2015) Technical note: phantom study to evaluate the dose and image quality effects of a computed tomography organ-based tube current modulation technique. Med Phys 42:6572–6578CrossRef
8.
go back to reference Euler A, Szucs-Farkas Z, Falkowski AL, Kawel-Böhm N, D’Errico L, Kopp S et al (2015) Organ-based tube current modulation in a clinical context: dose reduction may be largely overestimated in breast tissue. Eur Radiol. doi:10.1007/s00330-015-4085-5 CrossRefPubMed Euler A, Szucs-Farkas Z, Falkowski AL, Kawel-Böhm N, D’Errico L, Kopp S et al (2015) Organ-based tube current modulation in a clinical context: dose reduction may be largely overestimated in breast tissue. Eur Radiol. doi:10.​1007/​s00330-015-4085-5 CrossRefPubMed
9.
go back to reference Taylor S, Litmanovich DE, Shahzad M, Bankier AA, Genovis PA, Tack D (2015) Organ-based tube current modulation: are women’s breasts positioned in the reduced-dose zone? Radiology 274:260–266CrossRef Taylor S, Litmanovich DE, Shahzad M, Bankier AA, Genovis PA, Tack D (2015) Organ-based tube current modulation: are women’s breasts positioned in the reduced-dose zone? Radiology 274:260–266CrossRef
11.
go back to reference Khatonabadi M, Di Z, Mathieu K, Kim HJ, Lu P, Cody D et al (2012) A comparison of methods to estimate organ doses in CT when utilizing approximations to the tube current modulation function. Med Phys 39:5212–5228CrossRef Khatonabadi M, Di Z, Mathieu K, Kim HJ, Lu P, Cody D et al (2012) A comparison of methods to estimate organ doses in CT when utilizing approximations to the tube current modulation function. Med Phys 39:5212–5228CrossRef
12.
go back to reference American Association of Physicists in Medicine (2011) Size-specific dose estimates (SSDE) in pediatric and adult body CT examinations. Technical report 204. AAPM, College Park American Association of Physicists in Medicine (2011) Size-specific dose estimates (SSDE) in pediatric and adult body CT examinations. Technical report 204. AAPM, College Park
13.
go back to reference Zhang G, Marshall N, Bogaerts R, Jacobs R, Bosmans H (2013) Monte Carlo modeling for dose assessment in cone beam CT for oral and maxillofacial applications. Med Phys 40:072103-1-12CrossRef Zhang G, Marshall N, Bogaerts R, Jacobs R, Bosmans H (2013) Monte Carlo modeling for dose assessment in cone beam CT for oral and maxillofacial applications. Med Phys 40:072103-1-12CrossRef
14.
go back to reference Lopez-Rendon X, Zhang G, Bosmans H, Oyen R, Zanca F (2014) Implementing the complete beam hardening effect of the bowtie filter versus scaling beam intensities: effects on dosimetric applications in computed tomography. J Med Imaging (Bellingham) 1:033507CrossRef Lopez-Rendon X, Zhang G, Bosmans H, Oyen R, Zanca F (2014) Implementing the complete beam hardening effect of the bowtie filter versus scaling beam intensities: effects on dosimetric applications in computed tomography. J Med Imaging (Bellingham) 1:033507CrossRef
15.
go back to reference Gomez-Ros JM, Moraleda M, López MA, Navarro T, Navarro JF (2007) Monte Carlo based voxel phantoms for in vivo internal dosimetry. Rad Prot Dosim 125:161–165 Gomez-Ros JM, Moraleda M, López MA, Navarro T, Navarro JF (2007) Monte Carlo based voxel phantoms for in vivo internal dosimetry. Rad Prot Dosim 125:161–165
16.
go back to reference International Commission on Radiological Protection (2009) ICRP publication 110: adult reference computational phantoms. Ann ICRP 39:1–166 International Commission on Radiological Protection (2009) ICRP publication 110: adult reference computational phantoms. Ann ICRP 39:1–166
17.
go back to reference International Commission on Radiological Protection (2002) ICRP publication 89: basic anatomical and physiological data for use in radiological protection: reference values. Ann ICRP 32:1–277 International Commission on Radiological Protection (2002) ICRP publication 89: basic anatomical and physiological data for use in radiological protection: reference values. Ann ICRP 32:1–277
18.
go back to reference Ketelsen D, Buchgeister M, Fenchel M, Schmidt B, Flohr T, Syha R et al (2012) Automated computed tomography dose-saving algorithm to protect radiosensitive tissues. Invest Radiol 47:148–152CrossRef Ketelsen D, Buchgeister M, Fenchel M, Schmidt B, Flohr T, Syha R et al (2012) Automated computed tomography dose-saving algorithm to protect radiosensitive tissues. Invest Radiol 47:148–152CrossRef
19.
go back to reference Paul J, Banckwitz R, Krauss B, Vogl TJ, Maentele W, Bauer RW (2011) Estimation and comparison of effective dose in standard chest CT by organ dose measurements and dose-length-product methods and assessment of the influence of CT tube potential (energy dependency) of effective dose in a dual-source CT. Eur J Radiol 81:e507–512CrossRef Paul J, Banckwitz R, Krauss B, Vogl TJ, Maentele W, Bauer RW (2011) Estimation and comparison of effective dose in standard chest CT by organ dose measurements and dose-length-product methods and assessment of the influence of CT tube potential (energy dependency) of effective dose in a dual-source CT. Eur J Radiol 81:e507–512CrossRef
21.
go back to reference Primak AN, McCollough CH, Bruesewitz MR, Zhang J, Fletcher JG (2006) Relationship between noise, dose and pitch in cardiac multi-detector row CT. Radiographics 26:1785–1794CrossRef Primak AN, McCollough CH, Bruesewitz MR, Zhang J, Fletcher JG (2006) Relationship between noise, dose and pitch in cardiac multi-detector row CT. Radiographics 26:1785–1794CrossRef
22.
go back to reference Liu H, Gao Y, Ding A, Caracappa PF, Xu G (2015) The profound effects of patient arm positioning on organ doses from CT procedures calculated using Monte Carlo simulations and deformable phantoms. Rad Prot Dosim 164:368–375CrossRef Liu H, Gao Y, Ding A, Caracappa PF, Xu G (2015) The profound effects of patient arm positioning on organ doses from CT procedures calculated using Monte Carlo simulations and deformable phantoms. Rad Prot Dosim 164:368–375CrossRef
23.
go back to reference Brink M, de Lange F, Oostveen LJ, Dekker HM, Kool DR, Deunk J (2008) Arm raising at exposure-controlled multidetector trauma CT of thoracoabdominal region: higher image quality, lower radiation dose. Radiology 249:661–670CrossRef Brink M, de Lange F, Oostveen LJ, Dekker HM, Kool DR, Deunk J (2008) Arm raising at exposure-controlled multidetector trauma CT of thoracoabdominal region: higher image quality, lower radiation dose. Radiology 249:661–670CrossRef
Metadata
Title
Do we need 3D tube current modulation information for accurate organ dosimetry in chest CT? Protocols dose comparisons
Authors
Xochitl Lopez-Rendon
Guozhi Zhang
Walter Coudyzer
Wim Develter
Hilde Bosmans
Federica Zanca
Publication date
01-11-2017
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 11/2017
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-017-4863-3

Other articles of this Issue 11/2017

European Radiology 11/2017 Go to the issue