Skip to main content
Top
Published in: Knee Surgery, Sports Traumatology, Arthroscopy 1/2013

01-01-2013 | Knee

Differences in fixation stability between spacer plate and plate fixator following high tibial osteotomy

Authors: D. Pape, D. Kohn, N. van Giffen, A. Hoffmann, R. Seil, O. Lorbach

Published in: Knee Surgery, Sports Traumatology, Arthroscopy | Issue 1/2013

Login to get access

Abstract

Purpose

Since in vivo stability following high tibial osteotomy is unknown, surgeons customize the postoperative rehabilitation to the assumed implant stability, leaving us with numerous rehabilitation protocols. The purpose of the study is to quantify the fixation stability of different open-wedge high tibial osteotomy implants. It is hypothesized that the higher fixation stability of a plate fixator justifies early weight bearing.

Methods

In this prospective 30-subject clinical trial, fixation stability was determined over a 2-year period using radiostereometric analysis (RSA). Patients were assigned to two angle-stable osteotomy plates: a spacer plate with 6 postoperative weeks of feather-touch weight bearing versus a plate fixator with 2 postoperative weeks of feather-touch weight bearing.

Results

Postoperative RSA data showed a significant higher lateral translation of the distal tibia and a significantly increased subsidence, varus and internal rotation of the tibial head in the spacer plate compared to the plate fixator group. Weight bearing following spacer plate fixation induced significant micromotion 6 weeks after surgery. Three months after surgery, bone healing was achieved regardless of the used implant.

Conclusions

Early weight bearing is appropriate for plate fixator fixation. The 6-week period of delayed weight bearing following spacer plate fixation is inadequate and should be prolonged presumably up to 8–10 weeks to avoid pseudarthrosis and/or recurrence of varus angulation.

Level of evidence

Level II, diagnostic study—investigating a diagnostic test.
Literature
1.
go back to reference Agneskirchner JD, Freiling D, Hurschler C, Lobenhoffer P (2006) Primary stability of four different implants for opening wedge high tibial osteotomy. Knee Surg Sports Traumatol Arthrosc 14:291–300PubMedCrossRef Agneskirchner JD, Freiling D, Hurschler C, Lobenhoffer P (2006) Primary stability of four different implants for opening wedge high tibial osteotomy. Knee Surg Sports Traumatol Arthrosc 14:291–300PubMedCrossRef
2.
go back to reference Brinkman JM, Lobenhoffer P, Agneskirchner JD, Staubli AE, Wymenga AB, van Heerwaarden RJ (2008) Osteotomies around the knee: patient selection, stability of fixation and bone healing in high tibial osteotomies. J Bone Joint Surg Br 90:1548–1557PubMedCrossRef Brinkman JM, Lobenhoffer P, Agneskirchner JD, Staubli AE, Wymenga AB, van Heerwaarden RJ (2008) Osteotomies around the knee: patient selection, stability of fixation and bone healing in high tibial osteotomies. J Bone Joint Surg Br 90:1548–1557PubMedCrossRef
3.
go back to reference Brinkman JM, Luites JW, Wymenga AB, van Heerwaarden RJ (2010) Early full weight bearing is safe in open-wedge high tibial osteotomy. Acta Orthop 81:193–198PubMedCrossRef Brinkman JM, Luites JW, Wymenga AB, van Heerwaarden RJ (2010) Early full weight bearing is safe in open-wedge high tibial osteotomy. Acta Orthop 81:193–198PubMedCrossRef
4.
go back to reference Fowler P, Tan L, Brown G (2000) Medial opening wedge high tibial osteotomy: how I do it. Oper Tech Sports Med 8:32–38CrossRef Fowler P, Tan L, Brown G (2000) Medial opening wedge high tibial osteotomy: how I do it. Oper Tech Sports Med 8:32–38CrossRef
5.
go back to reference Insall JN, Dorr LD, Scott RD, Scott WN (1989) Rationale of the Knee Society clinical rating system. Clin Orthop Relat Res 248:13–14PubMed Insall JN, Dorr LD, Scott RD, Scott WN (1989) Rationale of the Knee Society clinical rating system. Clin Orthop Relat Res 248:13–14PubMed
6.
go back to reference Karrholm J (1989) Roentgen stereophotogrammetry: review of orthopedic applications. Acta Orthop Scand 60:491–503PubMedCrossRef Karrholm J (1989) Roentgen stereophotogrammetry: review of orthopedic applications. Acta Orthop Scand 60:491–503PubMedCrossRef
7.
go back to reference Lobenhoffer P, Agneskirchner JD (2003) Improvements in surgical technique of valgus high tibial osteotomy. Knee Surg Sports Traumatol Arthrosc 11:132–138PubMed Lobenhoffer P, Agneskirchner JD (2003) Improvements in surgical technique of valgus high tibial osteotomy. Knee Surg Sports Traumatol Arthrosc 11:132–138PubMed
8.
go back to reference Luites JW, Brinkman JM, Wymenga AB, van Heerwaarden RJ (2009) Fixation stability of opening- versus closing-wedge high tibial osteotomy: a randomised clinical trial using radiostereometry. J Bone Joint Surg Br 91:1459–1465PubMed Luites JW, Brinkman JM, Wymenga AB, van Heerwaarden RJ (2009) Fixation stability of opening- versus closing-wedge high tibial osteotomy: a randomised clinical trial using radiostereometry. J Bone Joint Surg Br 91:1459–1465PubMed
9.
go back to reference Miniaci A, Ballmer FT, Ballmer PM, Jakob RP (1989) Proximal tibial osteotomy. A new fixation device. Clin Orthop Relat Res 246:250–259PubMed Miniaci A, Ballmer FT, Ballmer PM, Jakob RP (1989) Proximal tibial osteotomy. A new fixation device. Clin Orthop Relat Res 246:250–259PubMed
10.
go back to reference Pape D, Adam F, Seil R, Georg T, Kohn D (2005) Fixation stability following high tibial osteotomy: a radiostereometric analysis. J Knee Surg 18:108–115PubMedCrossRef Pape D, Adam F, Seil R, Georg T, Kohn D (2005) Fixation stability following high tibial osteotomy: a radiostereometric analysis. J Knee Surg 18:108–115PubMedCrossRef
11.
go back to reference Pape D, Lorbach O, Schmitz C, Busch LC, Van GN, Seil R, Kohn DM (2010) Effect of a biplanar osteotomy on primary stability following high tibial osteotomy: a biomechanical cadaver study. Knee Surg Sports Traumatol Arthrosc 18:204–211PubMedCrossRef Pape D, Lorbach O, Schmitz C, Busch LC, Van GN, Seil R, Kohn DM (2010) Effect of a biplanar osteotomy on primary stability following high tibial osteotomy: a biomechanical cadaver study. Knee Surg Sports Traumatol Arthrosc 18:204–211PubMedCrossRef
12.
go back to reference Roos EM, Roos HP, Lohmander LS, Ekdahl C, Beynnon BD (1998) Knee injury and osteoarthritis outcome score (KOOS)—development of a self-administered outcome measure. J Orthop Sports Phys Ther 28:88–96PubMed Roos EM, Roos HP, Lohmander LS, Ekdahl C, Beynnon BD (1998) Knee injury and osteoarthritis outcome score (KOOS)—development of a self-administered outcome measure. J Orthop Sports Phys Ther 28:88–96PubMed
13.
go back to reference Selvik G, Alberius P, Aronson AS (1983) A roentgen stereophotogrammetric system. Construction, calibration and technical accuracy. Acta Radiol Diagn 24:343–352 Selvik G, Alberius P, Aronson AS (1983) A roentgen stereophotogrammetric system. Construction, calibration and technical accuracy. Acta Radiol Diagn 24:343–352
14.
go back to reference Sharma L, Song J, Felson DT, Cahue S, Shamiyeh E, Dunlop DD (2001) The role of knee alignment in disease progression and functional decline in knee osteoarthritis. JAMA 286:188–195PubMedCrossRef Sharma L, Song J, Felson DT, Cahue S, Shamiyeh E, Dunlop DD (2001) The role of knee alignment in disease progression and functional decline in knee osteoarthritis. JAMA 286:188–195PubMedCrossRef
15.
go back to reference Staubli AE, De SC, Babst R, Lobenhoffer P (2003) TomoFix: a new LCP-concept for open wedge osteotomy of the medial proximal tibia—early results in 92 cases. Injury 34:55–62CrossRef Staubli AE, De SC, Babst R, Lobenhoffer P (2003) TomoFix: a new LCP-concept for open wedge osteotomy of the medial proximal tibia—early results in 92 cases. Injury 34:55–62CrossRef
16.
go back to reference Staubli AE, Jacob HA (2010) Evolution of open-wedge high-tibial osteotomy: experience with a special angular stable device for internal fixation without interposition material. Int Orthop 34:167–172PubMedCrossRef Staubli AE, Jacob HA (2010) Evolution of open-wedge high-tibial osteotomy: experience with a special angular stable device for internal fixation without interposition material. Int Orthop 34:167–172PubMedCrossRef
17.
go back to reference Stoffel K, Stachowiak G, Kuster M (2004) Open wedge high tibial osteotomy: biomechanical investigation of the modified Arthrex Osteotomy Plate (Puddu Plate) and the TomoFix Plate. Clin Biomech 19:944–950CrossRef Stoffel K, Stachowiak G, Kuster M (2004) Open wedge high tibial osteotomy: biomechanical investigation of the modified Arthrex Osteotomy Plate (Puddu Plate) and the TomoFix Plate. Clin Biomech 19:944–950CrossRef
18.
go back to reference Valstar ER, Gill R, Ryd L, Flivik G, Borlin N, Karrholm J (2005) Guidelines for standardization of radiostereometry (RSA) of implants. Acta Orthop 76:563–572PubMedCrossRef Valstar ER, Gill R, Ryd L, Flivik G, Borlin N, Karrholm J (2005) Guidelines for standardization of radiostereometry (RSA) of implants. Acta Orthop 76:563–572PubMedCrossRef
19.
go back to reference van den Bekerom MP, Patt TW, Kleinhout MY, Albers GH (2008) Early complications after high tibial osteotomy: a comparison of two techniques. J Knee Surg 21:68–74PubMed van den Bekerom MP, Patt TW, Kleinhout MY, Albers GH (2008) Early complications after high tibial osteotomy: a comparison of two techniques. J Knee Surg 21:68–74PubMed
Metadata
Title
Differences in fixation stability between spacer plate and plate fixator following high tibial osteotomy
Authors
D. Pape
D. Kohn
N. van Giffen
A. Hoffmann
R. Seil
O. Lorbach
Publication date
01-01-2013
Publisher
Springer-Verlag
Published in
Knee Surgery, Sports Traumatology, Arthroscopy / Issue 1/2013
Print ISSN: 0942-2056
Electronic ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-011-1693-8

Other articles of this Issue 1/2013

Knee Surgery, Sports Traumatology, Arthroscopy 1/2013 Go to the issue