Published in: Journal of Clinical Monitoring and Computing 4/2016

Open Access 01-08-2016 | Original Research

Diagnostic accuracy of stroke volume variation measured with uncalibrated arterial waveform analysis for the prediction of fluid responsiveness in patients with impaired left ventricular function: a prospective, observational study

Authors: L. J. Montenij, J. P. C. Sonneveld, A. P. Nierich, W. F. Buhre, E. E. C. de Waal

Published in: Journal of Clinical Monitoring and Computing | Issue 4/2016

Login to get access

Abstract

Uncalibrated arterial waveform analysis enables dynamic preload assessment in a minimally invasive fashion. Evidence about the validity of the technique in patients with impaired left ventricular function is scarce, while adequate cardiac preload assessment would be of great value in these patients. The aim of this study was to investigate the diagnostic accuracy of stroke volume variation (SVV) measured with the FloTrac/Vigileo™ system in patients with impaired left ventricular function. In this prospective, observational study, 22 patients with a left ventricular ejection fraction of 40 % or less undergoing elective coronary artery bypass grafting were included. Patients were considered fluid responsive if cardiac output increased with 15 % or more after volume loading (7 ml kg−1 ideal body weight). The following variables were calculated: area under the receiver operating characteristics (ROC) curve, ideal cut-off value for SVV, sensitivity, specificity, positive and negative predictive values, and overall accuracy. In addition, SVV cut-off points to obtain 90 % true positive and 90 % true negative predictions were determined. ROC analysis revealed an area under the curve of 0.70 [0.47; 0.92]. The ideal SVV cut-off value was 10 %, with a corresponding sensitivity and specificity of 56 and 69 % respectively. Overall accuracy was 64 %, positive and negative predictive values were 69 and 56 % respectively. SVV values to obtain more than 90 % true positive and negative predictions were 16 and 6 % respectively. The ability of uncalibrated arterial waveform analysis SVV to predict fluid responsiveness in patients with impaired LVF was low.
Literature
1.
go back to reference Bundgaard-Nielsen M, Secher NH, Khelet H. ’Liberal’ vs. ‘restricted’ perioperative fluid therapy—a critical assessment of the evidence. Acta Anaesthesiol Scand. 2009;53:843–51.CrossRefPubMed Bundgaard-Nielsen M, Secher NH, Khelet H. ’Liberal’ vs. ‘restricted’ perioperative fluid therapy—a critical assessment of the evidence. Acta Anaesthesiol Scand. 2009;53:843–51.CrossRefPubMed
2.
go back to reference Wiedemann HP, Wheeler AP, Bernard GR, Thompson BT, Hayden D, De Boisblanc B, Connors AF Jr, Hite RD, Harabin AL. Comparison of two fluid-management strategies in acute lung injury. N Engl J Med. 2006;354:2564–75.CrossRefPubMed Wiedemann HP, Wheeler AP, Bernard GR, Thompson BT, Hayden D, De Boisblanc B, Connors AF Jr, Hite RD, Harabin AL. Comparison of two fluid-management strategies in acute lung injury. N Engl J Med. 2006;354:2564–75.CrossRefPubMed
3.
go back to reference Silva JM Jr, de Oliveira AM, Noqueira FM, Vianna PM, Pereira Filho MC, Dias LF, Maia VP, Neucamp CD, Amendola CP, Carmona MJ, Malbouisson LM. The effect of excess fluid balance on the mortality rate of surgical patients: a multicenter prospective study. Crit Care. 2013;17:R288.CrossRefPubMedPubMedCentral Silva JM Jr, de Oliveira AM, Noqueira FM, Vianna PM, Pereira Filho MC, Dias LF, Maia VP, Neucamp CD, Amendola CP, Carmona MJ, Malbouisson LM. The effect of excess fluid balance on the mortality rate of surgical patients: a multicenter prospective study. Crit Care. 2013;17:R288.CrossRefPubMedPubMedCentral
4.
go back to reference Marik PE, Baram M, Vahid B. Does central venous pressure predict fluid responsiveness? A systematic review of the literature and the tale of seven mares. Chest. 2008;134:172–8.CrossRefPubMed Marik PE, Baram M, Vahid B. Does central venous pressure predict fluid responsiveness? A systematic review of the literature and the tale of seven mares. Chest. 2008;134:172–8.CrossRefPubMed
6.
go back to reference Marik PE, Cavalazzi R, Vasu T, Hirani A. Dynamic changes in arterial waveform derived variables and fluid responsiveness in mechanically ventilated patients: a systematic review of the literature. Crit Care. 2009;37:2642–7.CrossRef Marik PE, Cavalazzi R, Vasu T, Hirani A. Dynamic changes in arterial waveform derived variables and fluid responsiveness in mechanically ventilated patients: a systematic review of the literature. Crit Care. 2009;37:2642–7.CrossRef
7.
go back to reference Montenij LJ, De Waal EE, Buhre WF. Arterial waveform analysis in anesthesia and critical care. Curr Opin Anaesthesiol. 2011;24:651–6.CrossRefPubMed Montenij LJ, De Waal EE, Buhre WF. Arterial waveform analysis in anesthesia and critical care. Curr Opin Anaesthesiol. 2011;24:651–6.CrossRefPubMed
8.
go back to reference Reuter DA, Kirchner A, Felbinger TW, Weis FC, Kilger E, Lamm P, Goetz E. Usefulness of left ventricular stroke volume variation to assess fluid responsiveness in patients with reduced cardiac function. Crit Care Med. 2003;31:1399–404.CrossRefPubMed Reuter DA, Kirchner A, Felbinger TW, Weis FC, Kilger E, Lamm P, Goetz E. Usefulness of left ventricular stroke volume variation to assess fluid responsiveness in patients with reduced cardiac function. Crit Care Med. 2003;31:1399–404.CrossRefPubMed
9.
go back to reference Yancy CW, Jessup M, Bozkurt B, et al. 2013 ACCF/AHA guideline for the management of heart failure: a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines. Circulation. 2013;128:e240–327.CrossRefPubMed Yancy CW, Jessup M, Bozkurt B, et al. 2013 ACCF/AHA guideline for the management of heart failure: a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines. Circulation. 2013;128:e240–327.CrossRefPubMed
10.
go back to reference Manecke GR. Edwards FloTrac sensor and Vigileo monitor: easy, accurate, reliable cardiac output assessment using the arterial pulse wave. Expert Rev Med Devices. 2005;2:523–7.CrossRefPubMed Manecke GR. Edwards FloTrac sensor and Vigileo monitor: easy, accurate, reliable cardiac output assessment using the arterial pulse wave. Expert Rev Med Devices. 2005;2:523–7.CrossRefPubMed
11.
go back to reference Hashim B, Lerner AB. The FloTrac™ System—measurement of stroke volume and the assessment of dynamic fluid loading. Int Anesthesiol Clin. 2010;48:45–56.CrossRefPubMed Hashim B, Lerner AB. The FloTrac™ System—measurement of stroke volume and the assessment of dynamic fluid loading. Int Anesthesiol Clin. 2010;48:45–56.CrossRefPubMed
12.
go back to reference Jansen JR, Schreuder JJ, Settels JJ, Kloek JJ, Versprille A. An adequate strategy for the thermodilution technique in patients during mechanical ventilation. Intensive Care Med. 1990;16:422–5.CrossRefPubMed Jansen JR, Schreuder JJ, Settels JJ, Kloek JJ, Versprille A. An adequate strategy for the thermodilution technique in patients during mechanical ventilation. Intensive Care Med. 1990;16:422–5.CrossRefPubMed
13.
go back to reference De Waal EE, Rex S, Kruitwagen CL, Kalkman CJ, Buhre WF. Dynamic preload indicators fail to predict fluid responsiveness in open chest conditions. Crit Care Med. 2009;37:510–5.CrossRefPubMed De Waal EE, Rex S, Kruitwagen CL, Kalkman CJ, Buhre WF. Dynamic preload indicators fail to predict fluid responsiveness in open chest conditions. Crit Care Med. 2009;37:510–5.CrossRefPubMed
14.
go back to reference Cecconi M, Rhodes A, Poloniecki J, Della Rocca G, Grounds RM. Bench-to-bedside review: the importance of the precision of the reference technique in method comparison studies—with specific reference to the measurement of cardiac output. Crit Care. 2009;13:201–6.CrossRefPubMedPubMedCentral Cecconi M, Rhodes A, Poloniecki J, Della Rocca G, Grounds RM. Bench-to-bedside review: the importance of the precision of the reference technique in method comparison studies—with specific reference to the measurement of cardiac output. Crit Care. 2009;13:201–6.CrossRefPubMedPubMedCentral
15.
go back to reference Linnet K, Bossuyt PM, Moons KG, Reitsma JB. Quantifying the accuracy of a diagnostic test or marker. Clin Chem. 2012;58:1292–301.CrossRefPubMed Linnet K, Bossuyt PM, Moons KG, Reitsma JB. Quantifying the accuracy of a diagnostic test or marker. Clin Chem. 2012;58:1292–301.CrossRefPubMed
17.
go back to reference Cannesson M, Pestel G, Ricks C, Hoeft A, Perel A. Hemodynamic monitoring and management in patients undergoing high risk surgery: a survey among North American and European anesthesiologists. Crit Care. 2011;15:R197.CrossRefPubMedPubMedCentral Cannesson M, Pestel G, Ricks C, Hoeft A, Perel A. Hemodynamic monitoring and management in patients undergoing high risk surgery: a survey among North American and European anesthesiologists. Crit Care. 2011;15:R197.CrossRefPubMedPubMedCentral
18.
go back to reference Heenen S, De Backer D, Vincent JL. How can the response to volume expansion in patients with spontaneous respiratory movements be predicted? Crit Care. 2006;10:R102.CrossRefPubMedPubMedCentral Heenen S, De Backer D, Vincent JL. How can the response to volume expansion in patients with spontaneous respiratory movements be predicted? Crit Care. 2006;10:R102.CrossRefPubMedPubMedCentral
19.
go back to reference Reuter DA, Bayerlein J, Goepfert MS, Weis FC, Kilger E. Influence of tidal volume on left ventricular stroke volume variation measured by pulse contour analysis in mechanically ventilated patients. Intensive Care Med. 2003;29:476–80.CrossRefPubMed Reuter DA, Bayerlein J, Goepfert MS, Weis FC, Kilger E. Influence of tidal volume on left ventricular stroke volume variation measured by pulse contour analysis in mechanically ventilated patients. Intensive Care Med. 2003;29:476–80.CrossRefPubMed
20.
go back to reference Feinstein AR. The inadequacy of binary models for the clinical reality of three-zone diagnostic decisions. J Clin Epidemiol. 1990;43:109–13.CrossRefPubMed Feinstein AR. The inadequacy of binary models for the clinical reality of three-zone diagnostic decisions. J Clin Epidemiol. 1990;43:109–13.CrossRefPubMed
Metadata
Title
Diagnostic accuracy of stroke volume variation measured with uncalibrated arterial waveform analysis for the prediction of fluid responsiveness in patients with impaired left ventricular function: a prospective, observational study
Authors
L. J. Montenij
J. P. C. Sonneveld
A. P. Nierich
W. F. Buhre
E. E. C. de Waal
Publication date
01-08-2016
Publisher
Springer Netherlands
Published in
Journal of Clinical Monitoring and Computing / Issue 4/2016
Print ISSN: 1387-1307
Electronic ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-015-9743-2

Other articles of this Issue 4/2016

Journal of Clinical Monitoring and Computing 4/2016 Go to the issue