Open Access 01-12-2022 | Research

Development of a "Scissor-Tip-Separator" for adjustment of scissor blade separation and prevention of scissor blade damage during steam sterilization

Authors: Natthacha Chiannilkulchai, Peinjit Bhumisirikul

Published in: Patient Safety in Surgery | Issue 1/2022

Login to get access

Abstract

Background

Reprocess reusable surgical instruments during steam sterilization; damage occurs to sharp scissor blades in close position, so steam cannot reach the blades. Surgical instruments' management requires standards to ensure patient safety and prevent harmful pathogens, especially in the COVID-19 pandemic. Although various devices can separate scissor blades, they do not prevent damage to cutting edges. To address the above problem, we developed a new scissor protector, the "Scissor-Tip-Separator," and evaluated its efficacy.

Methods

The "Scissor-Tip-Separator" design follows the steam sterilization guideline that instrument tips must be separated. The locking handles and V groove mechanism keep the scissor blades separated while preventing damage to the cutting edges. For efficacy assessment, purposive sampling was performed to select 44 Thai perioperative nurses at Ramathibodi Hospital, Bangkok, Thailand, to evaluate the "Scissor-Tip-Separators" in 450 sterile instrument containers. All participants evaluated surgical scissors placed in the "Scissor-Tip-Separators" during instrument setup, following a problem record checklist. At the end of the fifth use, participants were asked to complete the "Scissor-Tip-Separator" Effectiveness Scale, which was used to test the structural design of the "Scissor-Tip-Separator" in terms of function, usability, and safety. The Adenosine Triphosphate surface test was also used to validate the "Scissor-Tip-Separator" cleanliness. Data were collected from August 2020 to November 2020, then analyzed via descriptive statistics.

Results

The "Scissor-Tip-Separator" met the cleaning validation criteria, and in 44 uses, the physical property remained the same. The scissor shank was discovered loose from the handle before it had been unlocked (0.2–0.4%) at the 45th use. Based on participants' opinions, the overall instrument effectiveness was high in terms of function, usability, and safety.

Conclusion

The "Scissor-Tip-Separator" regulates scissor blade separation under sterilization guidelines; it prevents damage to cutting edges, thus ensuring patient safety. It protects against losses in a sterile field and can prevent hand injuries.
Literature
5.
go back to reference van Doremalen N, Bushmaker T, Morris DH, Holbrook MG, Gamble A, Williamson BN, Tamin A, Harcourt JL, Thornburg NJ, Gerber SI, Lloyd-Smith JO, de Wit E, Munster VJ. Aerosol and surface stability of SARS-CoV-2 as compared with SARS-CoV-1. N Engl J Med. 2020;382(16):1564–7. https://doi.org/10.1056/nejmc2004973.CrossRefPubMed van Doremalen N, Bushmaker T, Morris DH, Holbrook MG, Gamble A, Williamson BN, Tamin A, Harcourt JL, Thornburg NJ, Gerber SI, Lloyd-Smith JO, de Wit E, Munster VJ. Aerosol and surface stability of SARS-CoV-2 as compared with SARS-CoV-1. N Engl J Med. 2020;382(16):1564–7. https://​doi.​org/​10.​1056/​nejmc2004973.CrossRefPubMed
12.
go back to reference Committee AftAoMISS, Instrumentation AftAoM, Institute ANS. ANSI/AAMI ST79: 2017 Comprehensive Guide to Steam Sterilization and Sterility Assurance in Health: Association for the Advancement of Medical Instrumentation; 2017. Committee AftAoMISS, Instrumentation AftAoM, Institute ANS. ANSI/AAMI ST79: 2017 Comprehensive Guide to Steam Sterilization and Sterility Assurance in Health: Association for the Advancement of Medical Instrumentation; 2017.
14.
go back to reference AORN. Guidelines for Perioperative Practice 2018 Edition. Denver, Canada: AORN; 2018. AORN. Guidelines for Perioperative Practice 2018 Edition. Denver, Canada: AORN; 2018.
Metadata
Title
Development of a "Scissor-Tip-Separator" for adjustment of scissor blade separation and prevention of scissor blade damage during steam sterilization
Authors
Natthacha Chiannilkulchai
Peinjit Bhumisirikul
Publication date
01-12-2022
Publisher
BioMed Central
Published in
Patient Safety in Surgery / Issue 1/2022
Electronic ISSN: 1754-9493
DOI
https://doi.org/10.1186/s13037-022-00338-5

Other articles of this Issue 1/2022

Patient Safety in Surgery 1/2022 Go to the issue