Detection of Cement Leakage After Vertebroplasty with a Non-Flat-Panel Angio Unit Compared to Multidetector Computed Tomography—An Ex Vivo Study

Authors: Clemens Baumann, Heiko Fuchs, Kerstin Westphalen, Johannes Hierholzer

Published in: CardioVascular and Interventional Radiology | Issue 6/2008

Login to get access

Abstract

The purpose of this study was to investigate the detection of cement leakages after vertebroplasty using angiographic computed tomography (ACT) in a non-flat-panel angio unit compared to multidetector computed tomography (MDCT). Vertebroplasty was performed in 19 of 33 cadaver vertebrae (23 thoracic and 10 lumbar segments). In the angio suite, ACT (190°; 1.5° per image) was performed to obtain volumetric data. Another volumetric data set of the specimen was obtained by MDCT using a standard algorithm. Nine multiplanar reconstructions in standardized axial, coronal, and sagittal planes of every vertebra were generated from both data sets. Images were evaluated on the basis of a nominal scale with 18 criteria, comprising osseous properties (e.g., integrity of the end plate) and cement distribution (e.g., presence of intraspinal cement). MDCT images were regarded as gold standard and analyzed by two readers in a consensus mode. Rotational acquisitions were analyzed by six blinded readers. Results were correlated with the gold standard using Cohen’s κ-coefficient analysis. Furthermore, interobserver variability was calculated. Correlation with the gold standard ranged from no correlation (osseous margins of the neuroforamen, κ = 0.008) to intermediate (trace of vertebroplasty canula; κ = 0.615) for criteria referring to osseous morphology. However, there was an excellent correlation for those criteria referring to cement distribution, with κ values ranging from 0.948 (paravertebral cement distribution) to 0.972 (intraspinal cement distribution). With a minimum of κ = 0.768 (“good correlation”) and a maximum of κ = 0.91 (“excellent”), interobserver variability was low. In conclusion, ACT in an angio suite without a flat-panel detector depicts a cement leakage after vertebroplasty as well as MDCT. However, the method does not provide sufficient depiction of osseous morphology.
Literature
1.
go back to reference Hide IG, Gangi A (2004) Percutaneous vertebroplasty: history, technique and current perspectives. Clin Radiol 59(6):461–467PubMedCrossRef Hide IG, Gangi A (2004) Percutaneous vertebroplasty: history, technique and current perspectives. Clin Radiol 59(6):461–467PubMedCrossRef
2.
go back to reference Galibert P, Deramond H, Rosat P, Le Gars D (1987) Preliminary note on the treatment of vertebral angioma by percutaneous acrylic vertebroplasty. Neurochirurgie 33(2):166–168PubMed Galibert P, Deramond H, Rosat P, Le Gars D (1987) Preliminary note on the treatment of vertebral angioma by percutaneous acrylic vertebroplasty. Neurochirurgie 33(2):166–168PubMed
3.
go back to reference Mathis JM (2003) Percutaneous vertebroplasty: complication avoidance and technique optimization. AJNR Am J Neuroradiol 24(8):1697–1706PubMed Mathis JM (2003) Percutaneous vertebroplasty: complication avoidance and technique optimization. AJNR Am J Neuroradiol 24(8):1697–1706PubMed
4.
go back to reference Li YY, Hsu RW, Cheng CC, Huang TJ (2007) Minimally invasive vertebroplasty managed by a two C-arm fluoroscopic technique. Minim Invas Ther Allied Technol 16(6):350–354CrossRef Li YY, Hsu RW, Cheng CC, Huang TJ (2007) Minimally invasive vertebroplasty managed by a two C-arm fluoroscopic technique. Minim Invas Ther Allied Technol 16(6):350–354CrossRef
5.
go back to reference Helmberger T, Bohndorf K, Hierholzer J, Noldge G, Vorwerk D (2003) Guidelines of the German Radiological Society for percutaneous vertebroplasty. Radiologe 43(9):703–708PubMedCrossRef Helmberger T, Bohndorf K, Hierholzer J, Noldge G, Vorwerk D (2003) Guidelines of the German Radiological Society for percutaneous vertebroplasty. Radiologe 43(9):703–708PubMedCrossRef
6.
go back to reference McGraw JK, Cardella J, Barr JD, Mathis JM, Sanchez O, Schwartzberg MS, Swan TL, Sacks D (2003) Society of Interventional Radiology quality improvement guidelines for percutaneous vertebroplasty. J Vasc Interv Radiol 14(9; Pt 2):S311–S315PubMed McGraw JK, Cardella J, Barr JD, Mathis JM, Sanchez O, Schwartzberg MS, Swan TL, Sacks D (2003) Society of Interventional Radiology quality improvement guidelines for percutaneous vertebroplasty. J Vasc Interv Radiol 14(9; Pt 2):S311–S315PubMed
7.
go back to reference Gailloud P, Oishi S, Murphy K (2005) Three-dimensional fusion digital subtraction angiography: new reconstruction algorithm for simultaneous three-dimensional rendering of osseous and vascular information obtained during rotational angiography. Am J Neuroradiol 26(4):908–911PubMed Gailloud P, Oishi S, Murphy K (2005) Three-dimensional fusion digital subtraction angiography: new reconstruction algorithm for simultaneous three-dimensional rendering of osseous and vascular information obtained during rotational angiography. Am J Neuroradiol 26(4):908–911PubMed
8.
go back to reference Buhk JH, Elolf E, Knauth M (2006) Angiographic computed tomography is comparable to multislice computed tomography in lumbar myelographic imaging. J Comput Assist Tomogr 30(5):739–741PubMedCrossRef Buhk JH, Elolf E, Knauth M (2006) Angiographic computed tomography is comparable to multislice computed tomography in lumbar myelographic imaging. J Comput Assist Tomogr 30(5):739–741PubMedCrossRef
9.
go back to reference Meyer BC, Frericks BB, Albrecht T, Wolf KJ, Wacker FK (2007) Contrast-enhanced abdominal angiographic CT for intra-abdominal tumor embolization: a new tool for vessel and soft tissue visualization. CardioVasc Interv Radiol 30(4):743–749 Meyer BC, Frericks BB, Albrecht T, Wolf KJ, Wacker FK (2007) Contrast-enhanced abdominal angiographic CT for intra-abdominal tumor embolization: a new tool for vessel and soft tissue visualization. CardioVasc Interv Radiol 30(4):743–749
10.
go back to reference Binkert CA, Alencar H, Singh J, Baum RA (2006) Translumbar type II endoleak repair using angiographic CT. J Vasc Interv Radiol 17(8):1349–1353PubMedCrossRef Binkert CA, Alencar H, Singh J, Baum RA (2006) Translumbar type II endoleak repair using angiographic CT. J Vasc Interv Radiol 17(8):1349–1353PubMedCrossRef
11.
go back to reference Eide KR, Odegard A, Myhre HO, Haraldseth O (2007) Initial observations of endovascular aneurysm repair using Dyna-CT. J Endovasc Ther 14(1):50–53PubMedCrossRef Eide KR, Odegard A, Myhre HO, Haraldseth O (2007) Initial observations of endovascular aneurysm repair using Dyna-CT. J Endovasc Ther 14(1):50–53PubMedCrossRef
12.
go back to reference Benndorf G, Strother CM, Claus B, Naeini R, Morsi H, Klucznik R, Mawad ME (2005) Angiographic CT in cerebrovascular stenting. Am J Neuroradiol 26(7):1813–1818PubMed Benndorf G, Strother CM, Claus B, Naeini R, Morsi H, Klucznik R, Mawad ME (2005) Angiographic CT in cerebrovascular stenting. Am J Neuroradiol 26(7):1813–1818PubMed
13.
go back to reference Rock C, Linsenmaier U, Brandl R, Kotsianos D, Wirth S, Kaltschmidt R, Euler E, Mutschler W, Pfeifer KJ (2001) Introduction of a new mobile C-arm/CT combination equipment (ISO-C-3D). Initial results of 3-D sectional imaging. Unfallchirurg 104(9):827–833PubMedCrossRef Rock C, Linsenmaier U, Brandl R, Kotsianos D, Wirth S, Kaltschmidt R, Euler E, Mutschler W, Pfeifer KJ (2001) Introduction of a new mobile C-arm/CT combination equipment (ISO-C-3D). Initial results of 3-D sectional imaging. Unfallchirurg 104(9):827–833PubMedCrossRef
14.
go back to reference Cohen J (1960) A coefficient of agreement for nominal scales. Educ Psychol Meas 2037–2046 Cohen J (1960) A coefficient of agreement for nominal scales. Educ Psychol Meas 2037–2046
15.
go back to reference Nussbaum DA, Gailloud P, Murphy K (2004) A review of complications associated with vertebroplasty and kyphoplasty as reported to the Food and Drug Administration medical device related web site. J Vasc Interv Radiol 15(11):1185–1192PubMed Nussbaum DA, Gailloud P, Murphy K (2004) A review of complications associated with vertebroplasty and kyphoplasty as reported to the Food and Drug Administration medical device related web site. J Vasc Interv Radiol 15(11):1185–1192PubMed
16.
go back to reference Hulme PA, Krebs J, Ferguson SJ, Berlemann U (2006) Vertebroplasty and kyphoplasty: a systematic review of 69 clinical studies. Spine 31(17):1983–2001PubMedCrossRef Hulme PA, Krebs J, Ferguson SJ, Berlemann U (2006) Vertebroplasty and kyphoplasty: a systematic review of 69 clinical studies. Spine 31(17):1983–2001PubMedCrossRef
17.
go back to reference Baumann C, Fuchs H, Kiwit J, Westphalen K, Hierholzer J (2007) Complications in percutaneous vertebroplasty associated with puncture or cement leakage. CardioVasc Interv Radiol 30(2):161–168CrossRef Baumann C, Fuchs H, Kiwit J, Westphalen K, Hierholzer J (2007) Complications in percutaneous vertebroplasty associated with puncture or cement leakage. CardioVasc Interv Radiol 30(2):161–168CrossRef
18.
go back to reference Mathis JM, Wong W (2003) Percutaneous vertebroplasty: technical considerations. J Vasc Interv Radiol 14(8):953–960PubMed Mathis JM, Wong W (2003) Percutaneous vertebroplasty: technical considerations. J Vasc Interv Radiol 14(8):953–960PubMed
19.
go back to reference Hodek-Wuerz R, Martin JB, Wilhelm K, Lovblad KO, Babic D, Rufenacht DA, Wetzel SG (2006) Percutaneous vertebroplasty: preliminary experiences with rotational acquisitions and 3D reconstructions for therapy control. CardioVasc Interv Radiol 29(5):862–865CrossRef Hodek-Wuerz R, Martin JB, Wilhelm K, Lovblad KO, Babic D, Rufenacht DA, Wetzel SG (2006) Percutaneous vertebroplasty: preliminary experiences with rotational acquisitions and 3D reconstructions for therapy control. CardioVasc Interv Radiol 29(5):862–865CrossRef
20.
go back to reference Singh J, Carrino JA, Alencar H, Binkert CA (2007) Comparison of angiographic CT and spiral CT to assess cement distribution after vertebral augmentation. J Vasc Interv Radiol 18(12):1547–1551PubMedCrossRef Singh J, Carrino JA, Alencar H, Binkert CA (2007) Comparison of angiographic CT and spiral CT to assess cement distribution after vertebral augmentation. J Vasc Interv Radiol 18(12):1547–1551PubMedCrossRef
21.
go back to reference Wieners G, Pech M, Beck A, Konig B, Erdmenger U, Stockle U, Wust P, Felix R, Schroder RJ (2005) Comparison of radiation dose and image quality of Siremobil-IsoC(3D) with a 16-slice spiral CT for diagnosis and intervention in the human pelvic bone. Rofo 177(2):258–264PubMed Wieners G, Pech M, Beck A, Konig B, Erdmenger U, Stockle U, Wust P, Felix R, Schroder RJ (2005) Comparison of radiation dose and image quality of Siremobil-IsoC(3D) with a 16-slice spiral CT for diagnosis and intervention in the human pelvic bone. Rofo 177(2):258–264PubMed
22.
go back to reference Rock C, Kotsianos D, Linsenmaier U, Fischer T, Brandl R, Vill F, Wirth S, Kaltschmidt R, Euler E, Pfeifer KJ, Reiser M (2002) Studies on image quality, high contrast resolution and dose for the axial skeleton and limbs with a new, dedicated CT system (ISO-C-3 D). Rofo 174(2):170–176PubMed Rock C, Kotsianos D, Linsenmaier U, Fischer T, Brandl R, Vill F, Wirth S, Kaltschmidt R, Euler E, Pfeifer KJ, Reiser M (2002) Studies on image quality, high contrast resolution and dose for the axial skeleton and limbs with a new, dedicated CT system (ISO-C-3 D). Rofo 174(2):170–176PubMed
Metadata
Title
Detection of Cement Leakage After Vertebroplasty with a Non-Flat-Panel Angio Unit Compared to Multidetector Computed Tomography—An Ex Vivo Study
Authors
Clemens Baumann
Heiko Fuchs
Kerstin Westphalen
Johannes Hierholzer
Publication date
01-11-2008
Publisher
Springer-Verlag
Published in
CardioVascular and Interventional Radiology / Issue 6/2008
Print ISSN: 0174-1551
Electronic ISSN: 1432-086X
DOI
https://doi.org/10.1007/s00270-008-9385-6

Other articles of this Issue 6/2008

CardioVascular and Interventional Radiology 6/2008 Go to the issue