Open Access 01-12-2011 | Technical advance

Design of a factorial experiment with randomization restrictions to assess medical device performance on vascular tissue

Authors: Wiebke S Diestelkamp, Carissa M Krane, Margaret F Pinnell

Published in: BMC Medical Research Methodology | Issue 1/2011

Login to get access

Abstract

Background

Energy-based surgical scalpels are designed to efficiently transect and seal blood vessels using thermal energy to promote protein denaturation and coagulation. Assessment and design improvement of ultrasonic scalpel performance relies on both in vivo and ex vivo testing. The objective of this work was to design and implement a robust, experimental test matrix with randomization restrictions and predictive statistical power, which allowed for identification of those experimental variables that may affect the quality of the seal obtained ex vivo.

Methods

The design of the experiment included three factors: temperature (two levels); the type of solution used to perfuse the artery during transection (three types); and artery type (two types) resulting in a total of twelve possible treatment combinations. Burst pressures of porcine carotid and renal arteries sealed ex vivo were assigned as the response variable.

Results

The experimental test matrix was designed and carried out as a split-plot experiment in order to assess the contributions of several variables and their interactions while accounting for randomization restrictions present in the experimental setup. The statistical software package SAS was utilized and PROC MIXED was used to account for the randomization restrictions in the split-plot design. The combination of temperature, solution, and vessel type had a statistically significant impact on seal quality.

Conclusions

The design and implementation of a split-plot experimental test-matrix provided a mechanism for addressing the existing technical randomization restrictions of ex vivo ultrasonic scalpel performance testing, while preserving the ability to examine the potential effects of independent factors or variables. This method for generating the experimental design and the statistical analyses of the resulting data are adaptable to a wide variety of experimental problems involving large-scale tissue-based studies of medical or experimental device efficacy and performance.
Literature
1.
go back to reference Foschi D, Cellerino P, Corsi F, Tiadelli T, Morandi E, Rizzi A, Trabucchi E: The mechanisms of blood vessel closure in humans by the application of ultrasonic energy. Surg Endosc. 2002, 16 (5): 814-819. 10.1007/s00464-001-9074-x.CrossRefPubMed Foschi D, Cellerino P, Corsi F, Tiadelli T, Morandi E, Rizzi A, Trabucchi E: The mechanisms of blood vessel closure in humans by the application of ultrasonic energy. Surg Endosc. 2002, 16 (5): 814-819. 10.1007/s00464-001-9074-x.CrossRefPubMed
2.
go back to reference Harold KL, Pollinger H, Matthew BD, Kercher KW, Sing RF, Heniford BT: Comparison of ultrasonic energy, bipolar thermal energy, and vascular clips for the hemostasis of small-, medium- and large-siazed arteries. Surg Endosc. 2003, 17 (8): 1228-1230. 10.1007/s00464-002-8833-7.CrossRefPubMed Harold KL, Pollinger H, Matthew BD, Kercher KW, Sing RF, Heniford BT: Comparison of ultrasonic energy, bipolar thermal energy, and vascular clips for the hemostasis of small-, medium- and large-siazed arteries. Surg Endosc. 2003, 17 (8): 1228-1230. 10.1007/s00464-002-8833-7.CrossRefPubMed
3.
go back to reference Landman J, Kerbl K, Rehman J, Andreoni C, Humphrey PA, Collyer W, Olweny E, Sundaram C, Clayman R: Evaluation of a vessel sealing system, bipolar electrosurgery, harmonic scalpel, titanium clips, endoscopic gastrointestinal anastomosis vascular staples and sutures for arterial and venous ligation in a porcine model. J Urol. 2003, 169 (2): 697-700. 10.1016/S0022-5347(05)63995-X.CrossRefPubMed Landman J, Kerbl K, Rehman J, Andreoni C, Humphrey PA, Collyer W, Olweny E, Sundaram C, Clayman R: Evaluation of a vessel sealing system, bipolar electrosurgery, harmonic scalpel, titanium clips, endoscopic gastrointestinal anastomosis vascular staples and sutures for arterial and venous ligation in a porcine model. J Urol. 2003, 169 (2): 697-700. 10.1016/S0022-5347(05)63995-X.CrossRefPubMed
4.
go back to reference Krane C, Pinnell M, Gardner C, Thompson M, Coleman J, Wilkens R: Mechanical test methods for assessing porcine carotid and uterine artery burst pressure following ex vivo ultrasonic ligature seal and transection. J Test Eval. 2011, 39 (4): 1-8. Krane C, Pinnell M, Gardner C, Thompson M, Coleman J, Wilkens R: Mechanical test methods for assessing porcine carotid and uterine artery burst pressure following ex vivo ultrasonic ligature seal and transection. J Test Eval. 2011, 39 (4): 1-8.
5.
go back to reference Bubenik LJ, Hosgood G, Vasanjee SC: Bursting tension of medium and large canine arteries sealed with ultrasonic energy or suture ligation. Vet Surg. 2005, 34 (3): 289-293. 10.1111/j.1532-950x.2005.00043.x.CrossRefPubMed Bubenik LJ, Hosgood G, Vasanjee SC: Bursting tension of medium and large canine arteries sealed with ultrasonic energy or suture ligation. Vet Surg. 2005, 34 (3): 289-293. 10.1111/j.1532-950x.2005.00043.x.CrossRefPubMed
6.
go back to reference Novitsky YW, Rosen MJ, Harrell AG, Sing RF: Evaluation of the efficacy of the electrosurgical bipolar vessel sealer (LigaSure) devices in sealing lymphatic vessels. Surg Innov. 2005, 12 (2): 155-160. 10.1177/155335060501200215.CrossRefPubMed Novitsky YW, Rosen MJ, Harrell AG, Sing RF: Evaluation of the efficacy of the electrosurgical bipolar vessel sealer (LigaSure) devices in sealing lymphatic vessels. Surg Innov. 2005, 12 (2): 155-160. 10.1177/155335060501200215.CrossRefPubMed
7.
go back to reference Webb DF, Lucas JM, Borkowski JJ: Factorial experiments when factor levels are not necessarily reset. J Quality Technology. 2004, 36 (1): 1-11. Webb DF, Lucas JM, Borkowski JJ: Factorial experiments when factor levels are not necessarily reset. J Quality Technology. 2004, 36 (1): 1-11.
8.
go back to reference Montgomery DC: Design and Analysis of Experiments. 2005, New Jersey: John Wiley & Sons, Inc, 7 Montgomery DC: Design and Analysis of Experiments. 2005, New Jersey: John Wiley & Sons, Inc, 7
9.
go back to reference Federer WT, King F: Variations on Split Plot and Split Block Experiment Designs. 2007, New Jersey: John Wiley & Sons, IncCrossRef Federer WT, King F: Variations on Split Plot and Split Block Experiment Designs. 2007, New Jersey: John Wiley & Sons, IncCrossRef
10.
go back to reference Ganju J, Lucas JM: Detecting randomization restrictions caused by factors. J Stat Plann Inference. 1999, 81: 129-140. 10.1016/S0378-3758(99)00002-6.CrossRef Ganju J, Lucas JM: Detecting randomization restrictions caused by factors. J Stat Plann Inference. 1999, 81: 129-140. 10.1016/S0378-3758(99)00002-6.CrossRef
11.
go back to reference Ganju J, Lucas JM: Bias in test statistics when restrictions in randomization are caused by factors. Comm Statist Theory Methods. 1997, 26 (1): 47-63. 10.1080/03610929708831901.CrossRef Ganju J, Lucas JM: Bias in test statistics when restrictions in randomization are caused by factors. Comm Statist Theory Methods. 1997, 26 (1): 47-63. 10.1080/03610929708831901.CrossRef
12.
go back to reference Kovalski SM, Vining GG, Montgomery DC, Borror CM: Modifying a central composite design to model the process mean and variance when there are hard-to-change factors. J Roy Statist Soc Ser C. 2006, 55 (5): 615-630. 10.1111/j.1467-9876.2006.00556.x.CrossRef Kovalski SM, Vining GG, Montgomery DC, Borror CM: Modifying a central composite design to model the process mean and variance when there are hard-to-change factors. J Roy Statist Soc Ser C. 2006, 55 (5): 615-630. 10.1111/j.1467-9876.2006.00556.x.CrossRef
13.
go back to reference Webb DF, Lucas JM: Blocking strategies for factorial experiments with hard-to-change factors. Proceedings of the Section on Quality and Productivity of the American Statistical Association. 2004, 2181-2188. Webb DF, Lucas JM: Blocking strategies for factorial experiments with hard-to-change factors. Proceedings of the Section on Quality and Productivity of the American Statistical Association. 2004, 2181-2188.
14.
go back to reference Dodde R, Shih A, Advincula AP: A novel technique for demonstrating the real-time subsurface tissue thermal profile of two energized surgical instruments. Journal of Minimally Invasive Gynecology. 2009, 16: 599-603. 10.1016/j.jmig.2009.05.018.CrossRefPubMed Dodde R, Shih A, Advincula AP: A novel technique for demonstrating the real-time subsurface tissue thermal profile of two energized surgical instruments. Journal of Minimally Invasive Gynecology. 2009, 16: 599-603. 10.1016/j.jmig.2009.05.018.CrossRefPubMed
15.
go back to reference Lim CBB, Goldin RD, Elson DS, Darzi A, Hanna GB: In vivo thermography during small bowel fusion using radiofrequency energy. Surg Endosc. 2010, 24: 2465-2474. 10.1007/s00464-010-0987-0.CrossRefPubMed Lim CBB, Goldin RD, Elson DS, Darzi A, Hanna GB: In vivo thermography during small bowel fusion using radiofrequency energy. Surg Endosc. 2010, 24: 2465-2474. 10.1007/s00464-010-0987-0.CrossRefPubMed
16.
go back to reference Wallwiener CW, Rajab TK, Krämer B, Isaacson KB, Bruckner S, Wallwiener M: Quantifying electrosurgery-induced thermal effects and damage to human tissue: An exploratory study with the fallopian tube as a novel in-vivo in-situ model. Journal Minimally Invasive Gynecology. 2010, 17: 70-77. 10.1016/j.jmig.2009.09.007.CrossRef Wallwiener CW, Rajab TK, Krämer B, Isaacson KB, Bruckner S, Wallwiener M: Quantifying electrosurgery-induced thermal effects and damage to human tissue: An exploratory study with the fallopian tube as a novel in-vivo in-situ model. Journal Minimally Invasive Gynecology. 2010, 17: 70-77. 10.1016/j.jmig.2009.09.007.CrossRef
17.
go back to reference Koch C, Friedrich T, Matternich F, Tannapfel A, Reimann HP, Eichfeld U: Determination of temperature elevation in tissue during the application of the harmonic scalpel. Ultrasound in Med & Biol. 2003, 29 (2): 301-309. 10.1016/S0301-5629(02)00727-5.CrossRef Koch C, Friedrich T, Matternich F, Tannapfel A, Reimann HP, Eichfeld U: Determination of temperature elevation in tissue during the application of the harmonic scalpel. Ultrasound in Med & Biol. 2003, 29 (2): 301-309. 10.1016/S0301-5629(02)00727-5.CrossRef
Metadata
Title
Design of a factorial experiment with randomization restrictions to assess medical device performance on vascular tissue
Authors
Wiebke S Diestelkamp
Carissa M Krane
Margaret F Pinnell
Publication date
01-12-2011
Publisher
BioMed Central
Published in
BMC Medical Research Methodology / Issue 1/2011
Electronic ISSN: 1471-2288
DOI
https://doi.org/10.1186/1471-2288-11-75

Other articles of this Issue 1/2011

BMC Medical Research Methodology 1/2011 Go to the issue