Abstract

Objectives

To compare the cleansing efficacy of a representative “ten seconds” auto-cleaning device with that of uninstructed manual toothbrushing in a pilot study.

Materials and methods

Twenty periodontally healthy probands refrained from oral hygiene for 3 days. Baseline full-mouth plaque scores (Rustogi Modified Navy Plaque Index, RMNPI) were assessed. After randomization, probands cleaned their teeth either with the auto-cleaning test device according to the manufacturer’s protocol or with a manual toothbrush. Plaque reduction was assessed by two aligned blinded investigators. After a 2-week recovery, the clinical investigation was repeated in a crossover design. The brushing pattern of the auto-cleaning device was analyzed in probands’ casts.

Results

Full-mouth plaque reduction was 11.37 ± 3.70% for the auto-cleaning device and 31.39 ± 5.27% for manual toothbrushing (p < 0.0001). The investigation of the auto-cleaning device’s brushing pattern in dental casts revealed a positive relationship of bristle rows in contact with tooth surfaces and the cleansing efficacy in the respective areas. A maximum of 2/4 bristle rows were in contact with the tooth surfaces; in some areas, the bristles had no contact to the teeth.

Conclusions

Uninstructed manual toothbrushing is superior to auto-cleaning. The alignment and density of the auto-cleaning device’s bristle rows need to be improved, and assorted sizes would be necessary to cover different jaw shapes.

Clinical relevance

The auto-cleaning device has been developed to accommodate individuals with poor dexterity or compliance. To date, it is unable to provide sufficient plaque reduction due to an inappropriate bristle alignment and poor fit with diverse dental arches.
Literature
6.
go back to reference Deinzer R, Harnacke D, Mengel R, Telzer M, Lotzmann U, Wöstmann B (2016) Effectiveness of computer-based training on toothbrush skills of patients treated with crowns: a randomized controlled trial. J Periodontol 87:1333–1342CrossRef Deinzer R, Harnacke D, Mengel R, Telzer M, Lotzmann U, Wöstmann B (2016) Effectiveness of computer-based training on toothbrush skills of patients treated with crowns: a randomized controlled trial. J Periodontol 87:1333–1342CrossRef
13.
go back to reference Beal D, Ngo T, Feng Y, Cook D, Grau DG, Weber DA (2000) Development and laboratory evaluation of a new toothbrush with a novel brush head design. Am J Dent 13:5A–14A Beal D, Ngo T, Feng Y, Cook D, Grau DG, Weber DA (2000) Development and laboratory evaluation of a new toothbrush with a novel brush head design. Am J Dent 13:5A–14A
15.
go back to reference Cugini M, Thompson M, Warren PR (2006) Correlations between two plaque indices in the assessment of toothbrush effectiveness. J Contemp Dent Pract 7:1–9CrossRef Cugini M, Thompson M, Warren PR (2006) Correlations between two plaque indices in the assessment of toothbrush effectiveness. J Contemp Dent Pract 7:1–9CrossRef
Metadata
Title
Cleansing efficacy of an auto-cleaning electronic toothbrushing device: a randomized-controlled crossover pilot study
Authors
Dagmar Schnabl
Vera Wiesmüller
Vera Hönlinger
Simon Wimmer
Emanuel Bruckmoser
Ines Kapferer-Seebacher
Publication date
01-01-2021
Publisher
Springer Berlin Heidelberg
Published in
Clinical Oral Investigations / Issue 1/2021
Print ISSN: 1432-6981
Electronic ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-020-03359-5