CT accuracy of percutaneous versus open pedicle screw techniques: a series of 1609 screws

Authors: Todd M. Chapman, Daniel J. Blizzard, Christopher R. Brown

Published in: European Spine Journal | Issue 6/2016

Login to get access

Abstract

Introduction

Traditional open exposure for posterior instrumentation requires significant soft tissue mobilization and causes significant blood loss and increased recovery time. Mal-placed screws can injure nerve roots, the spinal cord, viscera, vasculature and the cardiopulmonary system. Placement of pedicle screws using a minimally invasive technique can decrease bleeding risk, damage to soft tissues, and post-operative pain. The purpose of this study is to compare the radiographic accuracy of open free-hand versus percutaneous technique for pedicle screw placement.

Methods

Consecutive patients undergoing thoracolumbar surgery from September 2006 to October 2011 with post-operative CT imaging were included in this study. Three-dimensional screw positioning within the pedicle and the vertebral body was assessed on CT. The magnitude and location of violations were measured and recorded. Facet breaches at the cephalad and caudad ends of the construct were documented and graded.

Results

Two-hundred and twenty-three patients met the inclusion criteria for a total of 1609 pedicle screws. Seven-hundred and twenty-four screws were placed using a standard open free-hand technique and 885 were placed percutaneously. There was a significant difference in overall pedicle breach rates: 7.5 % for open and 4.7 % for percutaneous techniques. The magnitude of breach was greater for the percutaneous technique compared to the open technique: 5.4 versus 3.7 mm, respectively. The difference in vertebral body breaches was also significant: 11.3 % for open and 3.6 % for percutaneous. The rates of facet breach did not significantly differ.

Discussion and conclusion

This is the largest series comparing the accuracy of percutaneous to open pedicle screw placement. The rates of pedicle, vertebral body, and facets breaches in the percutaneous group were similar to the rates in the open technique group as well as rates reported in the literature. This demonstrates that the percutaneous technique described here is an accurate alternative to standard open free-hand technique.
Literature
1.
go back to reference France JC et al (1999) A randomized prospective study of posterolateral lumbar fusion. Outcomes with and without pedicle screw instrumentation. Spine (Phila Pa 1976) 24(6):553–560CrossRef France JC et al (1999) A randomized prospective study of posterolateral lumbar fusion. Outcomes with and without pedicle screw instrumentation. Spine (Phila Pa 1976) 24(6):553–560CrossRef
2.
go back to reference Belmont PJ Jr et al (2001) In vivo accuracy of thoracic pedicle screws. Spine (Phila Pa 1976) 26(21):2340–2346CrossRef Belmont PJ Jr et al (2001) In vivo accuracy of thoracic pedicle screws. Spine (Phila Pa 1976) 26(21):2340–2346CrossRef
3.
go back to reference Belmont PJ Jr et al (2002) Accuracy of thoracic pedicle screws in patients with and without coronal plane spinal deformities. Spine (Phila Pa 1976) 27(14):1558–1566CrossRef Belmont PJ Jr et al (2002) Accuracy of thoracic pedicle screws in patients with and without coronal plane spinal deformities. Spine (Phila Pa 1976) 27(14):1558–1566CrossRef
4.
go back to reference Castro WH et al (1996) Accuracy of pedicle screw placement in lumbar vertebrae. Spine (Phila Pa 1976) 21(11):1320–1324CrossRef Castro WH et al (1996) Accuracy of pedicle screw placement in lumbar vertebrae. Spine (Phila Pa 1976) 21(11):1320–1324CrossRef
5.
go back to reference Choi WW, Green BA, Levi AD (2000) Computer-assisted fluoroscopic targeting system for pedicle screw insertion. Neurosurgery 47(4):872–878CrossRefPubMed Choi WW, Green BA, Levi AD (2000) Computer-assisted fluoroscopic targeting system for pedicle screw insertion. Neurosurgery 47(4):872–878CrossRefPubMed
7.
go back to reference Kim HJ et al (2011) The biomechanical effect of pedicle screws’ insertion angle and position on the superior adjacent segment in one segment lumbar fusion. Spine (Phila Pa 1976). 37(19):1637–1644 Kim HJ et al (2011) The biomechanical effect of pedicle screws’ insertion angle and position on the superior adjacent segment in one segment lumbar fusion. Spine (Phila Pa 1976). 37(19):1637–1644
8.
go back to reference Mirza SK et al (2003) Accuracy of thoracic vertebral body screw placement using standard fluoroscopy, fluoroscopic image guidance, and computed tomographic image guidance: a cadaver study. Spine (Phila Pa 1976) 28(4):402–413 Mirza SK et al (2003) Accuracy of thoracic vertebral body screw placement using standard fluoroscopy, fluoroscopic image guidance, and computed tomographic image guidance: a cadaver study. Spine (Phila Pa 1976) 28(4):402–413
9.
go back to reference Raley DA, Mobbs RJ (2012) Retrospective computed tomography scan analysis of percutaneously inserted pedicle screws for posterior transpedicular stabilization of the thoracic and lumbar spine: accuracy and complication rates. Spine (Phila Pa 1976) 37(12):1092–1100CrossRef Raley DA, Mobbs RJ (2012) Retrospective computed tomography scan analysis of percutaneously inserted pedicle screws for posterior transpedicular stabilization of the thoracic and lumbar spine: accuracy and complication rates. Spine (Phila Pa 1976) 37(12):1092–1100CrossRef
10.
go back to reference Schizas C et al (2007) Computer tomography assessment of pedicle screw insertion in percutaneous posterior transpedicular stabilization. Eur Spine J 16(5):613–617CrossRefPubMedPubMedCentral Schizas C et al (2007) Computer tomography assessment of pedicle screw insertion in percutaneous posterior transpedicular stabilization. Eur Spine J 16(5):613–617CrossRefPubMedPubMedCentral
11.
go back to reference Chen Z et al (2008) Technical factors related to the incidence of adjacent superior segment facet joint violation after transpedicular instrumentation in the lumbar spine. Eur Spine J 17(11):1476–1480CrossRefPubMedPubMedCentral Chen Z et al (2008) Technical factors related to the incidence of adjacent superior segment facet joint violation after transpedicular instrumentation in the lumbar spine. Eur Spine J 17(11):1476–1480CrossRefPubMedPubMedCentral
12.
go back to reference Moshirfar A et al (2006) Computed tomography evaluation of superior-segment facet-joint violation after pedicle instrumentation of the lumbar spine with a midline surgical approach. Spine (Phila Pa 1976) 31(22):2624–2629CrossRef Moshirfar A et al (2006) Computed tomography evaluation of superior-segment facet-joint violation after pedicle instrumentation of the lumbar spine with a midline surgical approach. Spine (Phila Pa 1976) 31(22):2624–2629CrossRef
13.
go back to reference Patel RD et al (2011) Facet violation with the placement of percutaneous pedicle screws. Spine (Phila Pa 1976) 36(26):E1749–E1752CrossRef Patel RD et al (2011) Facet violation with the placement of percutaneous pedicle screws. Spine (Phila Pa 1976) 36(26):E1749–E1752CrossRef
14.
go back to reference Shah RR et al (2003) Radiologic evaluation of adjacent superior segment facet joint violation following transpedicular instrumentation of the lumbar spine. Spine (Phila Pa 1976) 28(3):272–275 Shah RR et al (2003) Radiologic evaluation of adjacent superior segment facet joint violation following transpedicular instrumentation of the lumbar spine. Spine (Phila Pa 1976) 28(3):272–275
15.
go back to reference Foley KT, Gupta SK (2002) Percutaneous pedicle screw fixation of the lumbar spine: preliminary clinical results. J Neurosurg 97(1 Suppl):7–12PubMed Foley KT, Gupta SK (2002) Percutaneous pedicle screw fixation of the lumbar spine: preliminary clinical results. J Neurosurg 97(1 Suppl):7–12PubMed
16.
go back to reference Foley KT, Holly LT, Schwender JD (2003) Minimally invasive lumbar fusion. Spine (Phila Pa 1976) 28(15 Suppl):S26–S35 Foley KT, Holly LT, Schwender JD (2003) Minimally invasive lumbar fusion. Spine (Phila Pa 1976) 28(15 Suppl):S26–S35
17.
go back to reference Foley KT, Lefkowitz MA (2002) Advances in minimally invasive spine surgery. Clin Neurosurg 49:499–517PubMed Foley KT, Lefkowitz MA (2002) Advances in minimally invasive spine surgery. Clin Neurosurg 49:499–517PubMed
18.
go back to reference Khoo LT et al (2002) Minimally invasive percutaneous posterior lumbar interbody fusion. Neurosurgery 51(5 Suppl):S166–1PubMed Khoo LT et al (2002) Minimally invasive percutaneous posterior lumbar interbody fusion. Neurosurgery 51(5 Suppl):S166–1PubMed
19.
go back to reference Perez-Cruet MJ et al (2002) Microendoscopic lumbar discectomy: technical note. Neurosurgery 51(5 Suppl):S129–S136PubMed Perez-Cruet MJ et al (2002) Microendoscopic lumbar discectomy: technical note. Neurosurgery 51(5 Suppl):S129–S136PubMed
20.
go back to reference Babu R et al (2012) Comparison of superior-level facet joint violations during open and percutaneous pedicle screw placement. Neurosurgery 71(5):962–970CrossRefPubMedPubMedCentral Babu R et al (2012) Comparison of superior-level facet joint violations during open and percutaneous pedicle screw placement. Neurosurgery 71(5):962–970CrossRefPubMedPubMedCentral
21.
go back to reference Team RDC (2011) R: a language and environment for statistical computing. R Foundation for Statistical Computing, Vienna Team RDC (2011) R: a language and environment for statistical computing. R Foundation for Statistical Computing, Vienna
Metadata
Title
CT accuracy of percutaneous versus open pedicle screw techniques: a series of 1609 screws
Authors
Todd M. Chapman
Daniel J. Blizzard
Christopher R. Brown
Publication date
01-06-2016
Publisher
Springer Berlin Heidelberg
Published in
European Spine Journal / Issue 6/2016
Print ISSN: 0940-6719
Electronic ISSN: 1432-0932
DOI
https://doi.org/10.1007/s00586-015-4163-z

Other articles of this Issue 6/2016

European Spine Journal 6/2016 Go to the issue