Skip to main content
Top
Published in: Annals of Intensive Care 1/2017

Open Access 01-12-2017 | Research

Cost analysis of single-use (Ambu® aScope™) and reusable bronchoscopes in the ICU

Authors: S. Perbet, M. Blanquet, C. Mourgues, J. Delmas, S. Bertran, B. Longère, V. Boïko-Alaux, P. Chennell, J.-E. Bazin, J.-M. Constantin

Published in: Annals of Intensive Care | Issue 1/2017

Login to get access

Abstract

Background

Flexible optical bronchoscopes are essential for management of airways in ICU, but the conventional reusable flexible scopes have three major drawbacks: high cost of repairs, need for decontamination, and possible transmission of infectious agents. The main objective of this study was to measure the cost of bronchoalveolar lavage (BAL) and percutaneous tracheostomy (PT) using reusable bronchoscopes and single-use bronchoscopes in an ICU of an university hospital. The secondary objective was to compare the satisfaction of healthcare professionals with reusable and single-use bronchoscopes.

Methods

The study was performed between August 2009 and July 2014 in a 16-bed ICU. All BAL and PT procedures were performed by experienced healthcare professionals. Cost analysis was performed considering ICU and hospital organization. Healthcare professional satisfaction with single-use and reusable scopes was determined based on eight factors. Sensitivity analysis was performed by applying discount rates (0, 3, and 5%) and by simulation of six situations based on different assumptions.

Results

At a discount rate of 3%, the costs per BAL for the two reusable scopes were 188.86€ (scope 1) and 185.94€ (scope 2), and the costs per PT for the reusable scope 1 and scope 2 and single-use scopes were 1613.84€, 410.24€, and 204.49€, respectively. The cost per procedure for the reusable scopes depended on the number of procedures performed, maintenance costs, and decontamination costs. Healthcare professionals were more satisfied with the third-generation single-use Ambu® aScope™.

Conclusions

The cost per procedure for the single-use scope was not superior to that for reusable scopes. The choice of single-use or reusable bronchoscopes in an ICU should consider the frequency of procedures and the number of bronchoscopes needed.
Appendix
Available only for authorised users
Literature
1.
go back to reference Du Rand IA, Blaikley J, Booton R, Chaudhuri N, Gupta V, Khalid S, et al. British Thoracic Society guideline for diagnostic flexible bronchoscopy in adults: accredited by NICE. Thorax. 2013;68(Suppl 1):i1–44.CrossRefPubMed Du Rand IA, Blaikley J, Booton R, Chaudhuri N, Gupta V, Khalid S, et al. British Thoracic Society guideline for diagnostic flexible bronchoscopy in adults: accredited by NICE. Thorax. 2013;68(Suppl 1):i1–44.CrossRefPubMed
2.
go back to reference Cheung NH, Napolitano LM. Tracheostomy: epidemiology, indications, timing, technique, and outcomes. Respir Care. 2014;59:895–915 (discussion 6–9).CrossRefPubMed Cheung NH, Napolitano LM. Tracheostomy: epidemiology, indications, timing, technique, and outcomes. Respir Care. 2014;59:895–915 (discussion 6–9).CrossRefPubMed
3.
go back to reference Madsen KR, Guldager H, Rewers M, Weber SO, Kobke-Jacobsen K, White J. Danish Guidelines 2015 for percutaneous. Dan Med J. 2015;2015:61. Madsen KR, Guldager H, Rewers M, Weber SO, Kobke-Jacobsen K, White J. Danish Guidelines 2015 for percutaneous. Dan Med J. 2015;2015:61.
4.
go back to reference Mehta AC, Curtis PS, Scalzitti ML, Meeker DP. The high price of bronchoscopy. Maintenance and repair of the flexible fiberoptic bronchoscope. Chest. 1990;98:448–54.CrossRefPubMed Mehta AC, Curtis PS, Scalzitti ML, Meeker DP. The high price of bronchoscopy. Maintenance and repair of the flexible fiberoptic bronchoscope. Chest. 1990;98:448–54.CrossRefPubMed
5.
go back to reference Organization WH. WHO tables on tissue infectivity distribution in transmissible spongiform encephalopathies/Expert Committee on Biological Standardization, 18 to 22 October 2010; 2010. Organization WH. WHO tables on tissue infectivity distribution in transmissible spongiform encephalopathies/Expert Committee on Biological Standardization, 18 to 22 October 2010; 2010.
6.
go back to reference Piepho T, Werner C, Noppens RR. Evaluation of the novel, single-use, flexible aScope for tracheal intubation in the simulated difficult airway and first clinical experiences. Anaesthesia. 2010;65:820–5.CrossRefPubMed Piepho T, Werner C, Noppens RR. Evaluation of the novel, single-use, flexible aScope for tracheal intubation in the simulated difficult airway and first clinical experiences. Anaesthesia. 2010;65:820–5.CrossRefPubMed
7.
go back to reference Vijayakumar M, Clarke A, Wilkes AR, Goodwin N, Hodzovic I. Comparison of the manoeuvrability and ease of use of the Ambu aScope and Olympus re-usable fibrescope in a manikin. Anaesthesia. 2011;66:689–93.CrossRefPubMed Vijayakumar M, Clarke A, Wilkes AR, Goodwin N, Hodzovic I. Comparison of the manoeuvrability and ease of use of the Ambu aScope and Olympus re-usable fibrescope in a manikin. Anaesthesia. 2011;66:689–93.CrossRefPubMed
8.
go back to reference Krugel V, Bathory I, Frascarolo P, Schoettker P. Comparison of the single-use Ambu(R) aScope 2 vs the conventional fibrescope for tracheal intubation in patients with cervical spine immobilisation by a semirigid collar*. Anaesthesia. 2013;68:21–6.CrossRefPubMed Krugel V, Bathory I, Frascarolo P, Schoettker P. Comparison of the single-use Ambu(R) aScope 2 vs the conventional fibrescope for tracheal intubation in patients with cervical spine immobilisation by a semirigid collar*. Anaesthesia. 2013;68:21–6.CrossRefPubMed
9.
go back to reference Mankikian J, Ehrmann S, Guilleminault L, Le Fol T, Barc C, Ferrandiere M, et al. An evaluation of a new single-use flexible bronchoscope with a large suction channel: reliability of bronchoalveolar lavage in ventilated piglets and initial clinical experience. Anaesthesia. 2014;69:701–6.CrossRefPubMed Mankikian J, Ehrmann S, Guilleminault L, Le Fol T, Barc C, Ferrandiere M, et al. An evaluation of a new single-use flexible bronchoscope with a large suction channel: reliability of bronchoalveolar lavage in ventilated piglets and initial clinical experience. Anaesthesia. 2014;69:701–6.CrossRefPubMed
10.
go back to reference Gupta D, Wang H. Cost-effectiveness analysis of flexible optical scopes for tracheal intubation: a descriptive comparative study of reusable and single-use scopes. J Clin Anesth. 2011;23:632–5.CrossRefPubMed Gupta D, Wang H. Cost-effectiveness analysis of flexible optical scopes for tracheal intubation: a descriptive comparative study of reusable and single-use scopes. J Clin Anesth. 2011;23:632–5.CrossRefPubMed
11.
go back to reference Aissou M, Coroir M, Debes C, Camus T, Hadri N, Gutton C. et al [Cost analysis comparing single-use (Ambu(R) aScope) and conventional reusable fiberoptic flexible scopes for difficult tracheal intubation]. Ann Fr Anesth Reanim. 2013;32:291–5.CrossRefPubMed Aissou M, Coroir M, Debes C, Camus T, Hadri N, Gutton C. et al [Cost analysis comparing single-use (Ambu(R) aScope) and conventional reusable fiberoptic flexible scopes for difficult tracheal intubation]. Ann Fr Anesth Reanim. 2013;32:291–5.CrossRefPubMed
12.
go back to reference Tvede MF, Kristensen MS, Nyhus-Andreasen M. A cost analysis of reusable and disposable flexible optical scopes for intubation. Acta Anaesthesiol Scand. 2012;56:577–84.CrossRefPubMed Tvede MF, Kristensen MS, Nyhus-Andreasen M. A cost analysis of reusable and disposable flexible optical scopes for intubation. Acta Anaesthesiol Scand. 2012;56:577–84.CrossRefPubMed
13.
go back to reference Drummond MF, Sculpher MJ, Torrance GW, O’Brien BJ, Stoddart GL. Methods for the economic evaluation of health care programme. 3rd ed. Oxford: Oxford University Press; 2005. Drummond MF, Sculpher MJ, Torrance GW, O’Brien BJ, Stoddart GL. Methods for the economic evaluation of health care programme. 3rd ed. Oxford: Oxford University Press; 2005.
14.
go back to reference Lunn W, Garland R, Gryniuk L, Smith L, Feller-Kopman D, Ernst A. Reducing maintenance and repair costs in an interventional pulmonology program. Chest. 2005;127:1382–7.PubMed Lunn W, Garland R, Gryniuk L, Smith L, Feller-Kopman D, Ernst A. Reducing maintenance and repair costs in an interventional pulmonology program. Chest. 2005;127:1382–7.PubMed
15.
go back to reference Choure A, Manali E, Krizmanich G, Gildea T, Mehta A. High price of bronchoscopy: cost of maintenance and repair of flexible bronchoscopes. J Bronchol. 2005;12:147–50.CrossRef Choure A, Manali E, Krizmanich G, Gildea T, Mehta A. High price of bronchoscopy: cost of maintenance and repair of flexible bronchoscopes. J Bronchol. 2005;12:147–50.CrossRef
16.
go back to reference Pujol E, Lopez AM, Valero R. Use of the Ambu((R)) aScope in 10 patients with predicted difficult intubation. Anaesthesia. 2010;65:1037–40.CrossRefPubMed Pujol E, Lopez AM, Valero R. Use of the Ambu((R)) aScope in 10 patients with predicted difficult intubation. Anaesthesia. 2010;65:1037–40.CrossRefPubMed
Metadata
Title
Cost analysis of single-use (Ambu® aScope™) and reusable bronchoscopes in the ICU
Authors
S. Perbet
M. Blanquet
C. Mourgues
J. Delmas
S. Bertran
B. Longère
V. Boïko-Alaux
P. Chennell
J.-E. Bazin
J.-M. Constantin
Publication date
01-12-2017
Publisher
Springer Paris
Published in
Annals of Intensive Care / Issue 1/2017
Electronic ISSN: 2110-5820
DOI
https://doi.org/10.1186/s13613-016-0228-3

Other articles of this Issue 1/2017

Annals of Intensive Care 1/2017 Go to the issue