01-01-2009 | Symposium: Papers Presented at the Hip Society Meetings 2008

Computer-assisted versus Manual Alignment in THA: A Probabilistic Approach to Range of Motion

Authors: Anthony J. Petrella, PhD, Joshua Q. Stowe, MS, Darryl D. D’Lima, MD, PhD, Paul J. Rullkoetter, PhD, Peter J. Laz, PhD

Published in: Clinical Orthopaedics and Related Research® | Issue 1/2009

Login to get access

Abstract

Dislocation remains a major complication after THA, and range of motion before impingement is important in joint stability. Variability in implant alignment affects resultant range of motion. We used a probabilistic modeling approach to assess the effects of implant alignment variability based on manual and computer-assisted surgical (CAS) techniques on resultant range of motion after THA. We implemented a contact detection algorithm within a probabilistic analysis framework. The normally distributed alignment variables (mean ± 1 standard deviation) were cup abduction (manual = 45° ± 7.6°, CAS = 45° ± 5.7°), cup anteversion (manual = 20° ± 9.6°, CAS = 20° ± 4.5°), and stem anteversion (manual and CAS = 10° ± 1.5°). The outcomes of the probabilistic analysis were range of motion distributions with 1% and 99% bounds. The upper bounds of motion for manual and CAS alignment were similar because bony impingement was the limiting factor. The lower bounds of range of motion were substantially different depending on the type of surgical alignment; manual alignment produced a smaller range of motion in 3% to 5% of cases. CAS implant alignment produced range of motion values above minimum acceptable levels in all cases simulated.
Literature
1.
go back to reference Bartz RL, Nobel PC, Kadakia NR, Tullos HS. The effect of femoral component head size on posterior dislocation of the artificial hip joint. J Bone Joint Surg Am. 2000;82:1300–1307.PubMed Bartz RL, Nobel PC, Kadakia NR, Tullos HS. The effect of femoral component head size on posterior dislocation of the artificial hip joint. J Bone Joint Surg Am. 2000;82:1300–1307.PubMed
2.
go back to reference Daly PJ, Morrey BF. Operative correction of an unstable total hip arthroplasty. J Bone Joint Surg Am. 1992;74:1334–1343.PubMed Daly PJ, Morrey BF. Operative correction of an unstable total hip arthroplasty. J Bone Joint Surg Am. 1992;74:1334–1343.PubMed
3.
go back to reference D’Lima DD, Urquhart AG, Buehler KO, Walker RH, Colwell CW. The effect of the orientation of the acetabular and femoral components on the range of motion of the hip at different head-neck ratios. J Bone Joint Surg Am. 2000;82:315–321.PubMed D’Lima DD, Urquhart AG, Buehler KO, Walker RH, Colwell CW. The effect of the orientation of the acetabular and femoral components on the range of motion of the hip at different head-neck ratios. J Bone Joint Surg Am. 2000;82:315–321.PubMed
4.
go back to reference Dorr LD, Wan Z. Causes of and treatment protocol for instability of total hip replacement. Clin Orthop Relat Res. 1998;355:144–151.PubMedCrossRef Dorr LD, Wan Z. Causes of and treatment protocol for instability of total hip replacement. Clin Orthop Relat Res. 1998;355:144–151.PubMedCrossRef
5.
go back to reference Easley S, Pal S, Tomaszewski P, Petrella A, Rullkoetter P, Laz P. Finite element-based probabilistic analysis tool for orthopaedic applications. Comput Methods Programs Biomed. 2007;85:32–40.PubMedCrossRef Easley S, Pal S, Tomaszewski P, Petrella A, Rullkoetter P, Laz P. Finite element-based probabilistic analysis tool for orthopaedic applications. Comput Methods Programs Biomed. 2007;85:32–40.PubMedCrossRef
6.
go back to reference Fackler CD, Poss R. Dislocation in total hip arthroplasties. Clin Orthop Relat Res. 1980;151:169–178.PubMed Fackler CD, Poss R. Dislocation in total hip arthroplasties. Clin Orthop Relat Res. 1980;151:169–178.PubMed
7.
go back to reference Grigoris P, Grecula MJ, Amstutz HC. Tripolar hip replacement for recurrent prosthetic dislocation. Clin Orthop Relat Res. 1994;304:148–155.PubMed Grigoris P, Grecula MJ, Amstutz HC. Tripolar hip replacement for recurrent prosthetic dislocation. Clin Orthop Relat Res. 1994;304:148–155.PubMed
8.
go back to reference Haldar A, Mahadevan S. Probability, Reliability and Statistical Methods in Engineering Design. New York, NY: Wiley; 2000. Haldar A, Mahadevan S. Probability, Reliability and Statistical Methods in Engineering Design. New York, NY: Wiley; 2000.
9.
go back to reference Harris WH. Advances in surgical technique for total hip replacement. Clin Orthop Relat Res. 1980;146:188–204.PubMed Harris WH. Advances in surgical technique for total hip replacement. Clin Orthop Relat Res. 1980;146:188–204.PubMed
10.
go back to reference Hedlundh U, Ahnfelt L, Hybbinette CH. Dislocations and the femoral head size in primary total hip arthroplasty. Clin Orthop Relat Res. 1996;333:226–233.PubMedCrossRef Hedlundh U, Ahnfelt L, Hybbinette CH. Dislocations and the femoral head size in primary total hip arthroplasty. Clin Orthop Relat Res. 1996;333:226–233.PubMedCrossRef
11.
go back to reference Jaramaz B, DiGioia III AM, Blackwell M, Nikou C. Computer assisted measurement of cup placement in total hip replacement. Clin Orthop Relat Res. 1998;354:70–81.PubMedCrossRef Jaramaz B, DiGioia III AM, Blackwell M, Nikou C. Computer assisted measurement of cup placement in total hip replacement. Clin Orthop Relat Res. 1998;354:70–81.PubMedCrossRef
12.
go back to reference Jaramaz B, Nikou C, Simon D, DiGioia III A. Range of motion after total hip arthroplasty: experimental verification of the analytical simulator. Lect Notes Comput Sci. 1997;1205:573.CrossRef Jaramaz B, Nikou C, Simon D, DiGioia III A. Range of motion after total hip arthroplasty: experimental verification of the analytical simulator. Lect Notes Comput Sci. 1997;1205:573.CrossRef
13.
go back to reference Jolles BM, Genoud P, Hoffmeyer P. Accuracy of computer-assisted cup placement in total hip arthroplasty. Int Congr Ser. 2001;1230:314–318.CrossRef Jolles BM, Genoud P, Hoffmeyer P. Accuracy of computer-assisted cup placement in total hip arthroplasty. Int Congr Ser. 2001;1230:314–318.CrossRef
14.
go back to reference Jolles BM, Genoud P, Hoffmeyer P. Computer-assisted cup placement techniques in total hip arthroplasty improve accuracy of placement. Clin Orthop Relat Res. 2004;426:174–179.PubMedCrossRef Jolles BM, Genoud P, Hoffmeyer P. Computer-assisted cup placement techniques in total hip arthroplasty improve accuracy of placement. Clin Orthop Relat Res. 2004;426:174–179.PubMedCrossRef
15.
go back to reference Kalteis T, Handel M, Herold T, Perlick L, Baethis H, Grifka J. Greater accuracy in positioning of the acetabular cup by using an image-free navigation system. Int Orthop. 2005;29:272–276.PubMedCrossRef Kalteis T, Handel M, Herold T, Perlick L, Baethis H, Grifka J. Greater accuracy in positioning of the acetabular cup by using an image-free navigation system. Int Orthop. 2005;29:272–276.PubMedCrossRef
16.
go back to reference Kessler O, Patil S, Stefan W, Mayr E, Colwell CW, D’Lima DD. Bony impingement affects range of motion after total hip arthroplasty: a subject-specific approach. J Orthop Res. 2008;26:443–452.PubMedCrossRef Kessler O, Patil S, Stefan W, Mayr E, Colwell CW, D’Lima DD. Bony impingement affects range of motion after total hip arthroplasty: a subject-specific approach. J Orthop Res. 2008;26:443–452.PubMedCrossRef
17.
go back to reference Kluess D, Martin H, Mittelmeier W, Schmitz K-P, Bader R. Influence of femoral head size on impingement, dislocation and stress distribution in total hip replacement. Med Eng Phys. 2007;29:465–471.PubMedCrossRef Kluess D, Martin H, Mittelmeier W, Schmitz K-P, Bader R. Influence of femoral head size on impingement, dislocation and stress distribution in total hip replacement. Med Eng Phys. 2007;29:465–471.PubMedCrossRef
18.
go back to reference Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am. 1978;60:217–220.PubMed Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am. 1978;60:217–220.PubMed
19.
go back to reference McCollum DE, Gray WJ. Dislocation after total hip arthroplasty. Causes and prevention. Clin Orthop Relat Res. 1990;261:159–170.PubMed McCollum DE, Gray WJ. Dislocation after total hip arthroplasty. Causes and prevention. Clin Orthop Relat Res. 1990;261:159–170.PubMed
20.
go back to reference Minoda Y, Kadowaki T, Kim M. Acetabular component orientation in 834 total hip arthroplasties using a manual technique. Clin Orthop Relat Res. 2006;445:186–191.PubMed Minoda Y, Kadowaki T, Kim M. Acetabular component orientation in 834 total hip arthroplasties using a manual technique. Clin Orthop Relat Res. 2006;445:186–191.PubMed
21.
go back to reference Morrey BF. Instability after total hip arthroplasty. Orthop Clin North Am. 1992;23:237–248.PubMed Morrey BF. Instability after total hip arthroplasty. Orthop Clin North Am. 1992;23:237–248.PubMed
22.
go back to reference Müller ME. Total hip prostheses. Clin Orthop Relat Res. 1970;72:46–68.PubMed Müller ME. Total hip prostheses. Clin Orthop Relat Res. 1970;72:46–68.PubMed
23.
go back to reference Nadzadi ME, Pedersen DR, Yack HJ, Callaghan JJ, Brown TD. Kinematics, kinetics, and finite element analysis of commonplace maneuvers at risk for total hip dislocation. J Biomech. 2003;36:577–591.PubMedCrossRef Nadzadi ME, Pedersen DR, Yack HJ, Callaghan JJ, Brown TD. Kinematics, kinetics, and finite element analysis of commonplace maneuvers at risk for total hip dislocation. J Biomech. 2003;36:577–591.PubMedCrossRef
24.
go back to reference Parratte S, Argenson JA. Validation and usefulness of a computer-assisted cup-positioning system in total hip arthroplasty. A prospective, randomized, controlled study. J Bone Joint Surg Am. 2007;89:494–499.PubMedCrossRef Parratte S, Argenson JA. Validation and usefulness of a computer-assisted cup-positioning system in total hip arthroplasty. A prospective, randomized, controlled study. J Bone Joint Surg Am. 2007;89:494–499.PubMedCrossRef
25.
go back to reference Pierchon F, Pasquier G, Cotten A, Fontaine C, Clarisse J, Duquennoy A. Causes of dislocation of total hip arthroplasty. CT study of component alignment. J Bone Joint Surg Br. 1994;76:45–48.PubMed Pierchon F, Pasquier G, Cotten A, Fontaine C, Clarisse J, Duquennoy A. Causes of dislocation of total hip arthroplasty. CT study of component alignment. J Bone Joint Surg Br. 1994;76:45–48.PubMed
26.
go back to reference Seel MJ, Hafez MA, Eckman K, Jaramaz B, Davidson D, DiGioia III AM. Three-dimensional planning and virtual radiographs in revision total hip arthroplasty for instability. Clin Orthop Relat Res. 2006;442:35–38.PubMed Seel MJ, Hafez MA, Eckman K, Jaramaz B, Davidson D, DiGioia III AM. Three-dimensional planning and virtual radiographs in revision total hip arthroplasty for instability. Clin Orthop Relat Res. 2006;442:35–38.PubMed
27.
go back to reference Turner RS. Postoperative total hip prosthetic femoral head dislocations Incidence, etiologic factors, and management. Clin Orthop Relat Res. 1994;301:196–204.PubMed Turner RS. Postoperative total hip prosthetic femoral head dislocations Incidence, etiologic factors, and management. Clin Orthop Relat Res. 1994;301:196–204.PubMed
29.
go back to reference Van Sint Jan S, Salvia P, Hilal I, Sholukha V, Rooze M, Clapworthy G. Registration of 6-DOFs electrogoniometry and CT medical imaging for 3D joint modeling. J Biomech. 2002;35:1475–1484.PubMedCrossRef Van Sint Jan S, Salvia P, Hilal I, Sholukha V, Rooze M, Clapworthy G. Registration of 6-DOFs electrogoniometry and CT medical imaging for 3D joint modeling. J Biomech. 2002;35:1475–1484.PubMedCrossRef
30.
go back to reference White Jr RE, Forness TJ, Allman JK, Junick DW. Effect of posterior capsular repair on early dislocation in primary total hip replacement. Clin Orthop Relat Res. 2001;393:163–167.PubMedCrossRef White Jr RE, Forness TJ, Allman JK, Junick DW. Effect of posterior capsular repair on early dislocation in primary total hip replacement. Clin Orthop Relat Res. 2001;393:163–167.PubMedCrossRef
31.
go back to reference Widmer KH, Majewski M. The impact of the CCD-angle on range of motion and cup positioning in total hip arthroplasty. Clin Biomech (Bristol Avon). 2005;20:723–728.CrossRef Widmer KH, Majewski M. The impact of the CCD-angle on range of motion and cup positioning in total hip arthroplasty. Clin Biomech (Bristol Avon). 2005;20:723–728.CrossRef
32.
go back to reference Widmer KH, Zurfluh B. Compliant positioning of total hip components for optimal range of motion. J Orthop Res. 2004;22:815–821.PubMedCrossRef Widmer KH, Zurfluh B. Compliant positioning of total hip components for optimal range of motion. J Orthop Res. 2004;22:815–821.PubMedCrossRef
33.
go back to reference Wines AP, McNicol D. Computed tomography measurement of the accuracy of component version in total hip arthroplasty. J Arthroplasty. 2006;21:696–701.PubMedCrossRef Wines AP, McNicol D. Computed tomography measurement of the accuracy of component version in total hip arthroplasty. J Arthroplasty. 2006;21:696–701.PubMedCrossRef
34.
go back to reference Woo RY, Morrey BF. Dislocations after total hip arthroplasty. J Bone Joint Surg Am. 1982;64:1295–1306.PubMed Woo RY, Morrey BF. Dislocations after total hip arthroplasty. J Bone Joint Surg Am. 1982;64:1295–1306.PubMed
35.
go back to reference Wu G, Siegler S, Allard P, Kirtley C, Leardini A, Rosenbaum D, Whittle M, D’Lima DD, Cristofolini L, Witte H, Schmid O, Stokes I. ISB recommendation on definitions of joint coordinate system of various joints for the reporting of human joint motion—part I: ankle, hip, and spine. J Biomech. 2002;35:543–548.PubMedCrossRef Wu G, Siegler S, Allard P, Kirtley C, Leardini A, Rosenbaum D, Whittle M, D’Lima DD, Cristofolini L, Witte H, Schmid O, Stokes I. ISB recommendation on definitions of joint coordinate system of various joints for the reporting of human joint motion—part I: ankle, hip, and spine. J Biomech. 2002;35:543–548.PubMedCrossRef
36.
go back to reference Wu Y-T, Millwater HR, Cruse TA. Advanced probabilistic structural analysis method for implicit performance functions. J AIAA. 1990;28:1663–1669.CrossRef Wu Y-T, Millwater HR, Cruse TA. Advanced probabilistic structural analysis method for implicit performance functions. J AIAA. 1990;28:1663–1669.CrossRef
37.
go back to reference Yoshimine F. The safe-zones for combined cup and neck anteversions that fulfill the essential range of motion and their optimum combination in total hip replacements. J Biomech. 2006;39:1315–1323.PubMedCrossRef Yoshimine F. The safe-zones for combined cup and neck anteversions that fulfill the essential range of motion and their optimum combination in total hip replacements. J Biomech. 2006;39:1315–1323.PubMedCrossRef
38.
go back to reference Yoshimine F, Ginbayashi K. A mathematical formula to calculate the theoretical range of motion for total hip replacement. J Biomech. 2002;35:989–993.PubMedCrossRef Yoshimine F, Ginbayashi K. A mathematical formula to calculate the theoretical range of motion for total hip replacement. J Biomech. 2002;35:989–993.PubMedCrossRef
39.
go back to reference Yuan L, Shih C. Dislocation after total hip arthroplasty. Arch Orthop Trauma Surg. 1999;119:263–266.PubMedCrossRef Yuan L, Shih C. Dislocation after total hip arthroplasty. Arch Orthop Trauma Surg. 1999;119:263–266.PubMedCrossRef
Metadata
Title
Computer-assisted versus Manual Alignment in THA: A Probabilistic Approach to Range of Motion
Authors
Anthony J. Petrella, PhD
Joshua Q. Stowe, MS
Darryl D. D’Lima, MD, PhD
Paul J. Rullkoetter, PhD
Peter J. Laz, PhD
Publication date
01-01-2009
Publisher
Springer-Verlag
Published in
Clinical Orthopaedics and Related Research® / Issue 1/2009
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-008-0561-4

Other articles of this Issue 1/2009

Clinical Orthopaedics and Related Research® 1/2009 Go to the issue

Symposium: Papers Presented at the Hip Society Meetings 2008

Late Remodeling Around a Proximally HA-coated Tapered Titanium Femoral Component

Symposium: Papers Presented at the Hip Society Meetings 2008

Isolated Polyethylene Exchange versus Acetabular Revision for Polyethylene Wear