Open Access 01-12-2024 | Case report

Complement-mediated dialysis reaction during regular hemodialysis treatment: a case report

Authors: Ákos Pethő, Karolina Schnabel, László Dézsi

Published in: Journal of Medical Case Reports | Issue 1/2024

Login to get access

Abstract

Background

Hemodialysis reactions (HDRs) are similar to complement activation-related pseudo allergy (CARPA), a hypersensitivity reaction that occurs when administering certain (nano)drugs intravenously. The pathomechanism of CARPA was described based on animal experiments. Typical CARPA-like dialysis reactions, which occur at the start of hemodialysis, have been reported using polysulfone dialyzers. However, to our knowledge, this is the first dialysis reaction that occurred towards the end of hemodialysis treatment.

Case presentation

This report describes a 52-year-old Caucasian male patient who had been receiving chronic hemodialysis for 3 years and exhibited a CARPA reaction during his third hour of treatment. Upon activation of the microbubble alarm, the extracorporeal system recirculated for five minutes. Following reconnection, the patient exhibited a drop in systemic blood pressure, chest pain, and dyspnea after five minutes. Symptoms disappeared spontaneously after reducing the speed of the blood pump, placing the patient in a Trendelenburg position, and administering a bolus infusion from the dialysis machine. The remaining dialysis treatment was uneventful.

Conclusion

Numerous case reports about reactions occurring with modern high-efficiency polysulfone dialyzers have been published. However, due to changes in the material structure by the manufacturers, we have not encountered such cases lately. The recently reported increase in thromboxane-B2 and pulmonary arterial pressure and complement activation upon re-infusion of extracorporeal blood following dialysis may explain the reaction observed here.
Literature
1.
go back to reference Ma S, Pu N, Ma J. Effects of high-flux hemodialysis and hemodiafiltration on the mortality of patients with end-stage kidney disease: a meta-analysis. Ren Fail. 2023;45(1):2147436.PubMedPubMedCentralCrossRef Ma S, Pu N, Ma J. Effects of high-flux hemodialysis and hemodiafiltration on the mortality of patients with end-stage kidney disease: a meta-analysis. Ren Fail. 2023;45(1):2147436.PubMedPubMedCentralCrossRef
2.
go back to reference Delgado Córdova M, Blanco N, Azaña C. Hypotension in hemodialysis secondary to a reaction to synthetic membranes. Nefrologia. 2018;38(3):329–30.PubMedCrossRef Delgado Córdova M, Blanco N, Azaña C. Hypotension in hemodialysis secondary to a reaction to synthetic membranes. Nefrologia. 2018;38(3):329–30.PubMedCrossRef
3.
go back to reference Butani L, Calogiuri G. Hypersensitivity reactions in hemodialysis patients. Ann Allergy Asthma Immunol. 2017;118(6):680–4.PubMedCrossRef Butani L, Calogiuri G. Hypersensitivity reactions in hemodialysis patients. Ann Allergy Asthma Immunol. 2017;118(6):680–4.PubMedCrossRef
4.
go back to reference Rodriguez-Sanz A, Sanchez-Villanueva R, Dominguez-Ortega J, Fiandor AM, Ruiz MP, Trocoli F, Diaz-Tejeiro R, Cadenillas C, Gonzalez E, Martinez V, Lopez-Trascasa M, Quirce S, Selgas R, Bellon T. Mechanisms involved in hypersensitivity reactions to polysulfone hemodialysis membranes. Artif Organs. 2017;41(11):E285–95.PubMedCrossRef Rodriguez-Sanz A, Sanchez-Villanueva R, Dominguez-Ortega J, Fiandor AM, Ruiz MP, Trocoli F, Diaz-Tejeiro R, Cadenillas C, Gonzalez E, Martinez V, Lopez-Trascasa M, Quirce S, Selgas R, Bellon T. Mechanisms involved in hypersensitivity reactions to polysulfone hemodialysis membranes. Artif Organs. 2017;41(11):E285–95.PubMedCrossRef
5.
go back to reference Sayeed K, et al. Anaphylactic shock at the beginning of hemodialysis. Semin Dial. 2016;29(1):81–4.PubMedCrossRef Sayeed K, et al. Anaphylactic shock at the beginning of hemodialysis. Semin Dial. 2016;29(1):81–4.PubMedCrossRef
6.
go back to reference Szebeni J. Complement activation-related pseudoallergy: a stress reaction in blood triggered by nanomedicines and biologicals. Mol Immunol. 2014;61(2):163–73.PubMedCrossRef Szebeni J. Complement activation-related pseudoallergy: a stress reaction in blood triggered by nanomedicines and biologicals. Mol Immunol. 2014;61(2):163–73.PubMedCrossRef
7.
go back to reference Dezsi L, et al. Features of complement activation-related pseudoallergy to liposomes with different surface charge and PEGylation: comparison of the porcine and rat responses. J Control Release. 2014;195:2–10.PubMedCrossRef Dezsi L, et al. Features of complement activation-related pseudoallergy to liposomes with different surface charge and PEGylation: comparison of the porcine and rat responses. J Control Release. 2014;195:2–10.PubMedCrossRef
8.
go back to reference Boer WH, et al. Acute reactions to polysulfone/polyethersulfone dialysers: literature review and management. Neth J Med. 2017;75(1):4–13.PubMed Boer WH, et al. Acute reactions to polysulfone/polyethersulfone dialysers: literature review and management. Neth J Med. 2017;75(1):4–13.PubMed
9.
go back to reference Hoseinpour V, et al. Surface modification of PES membrane via aminolysis and immobilization of carboxymethylcellulose and sulphated carboxymethylcellulose for hemodialysis. Carbohyd Polym. 2018;188:37–47.CrossRef Hoseinpour V, et al. Surface modification of PES membrane via aminolysis and immobilization of carboxymethylcellulose and sulphated carboxymethylcellulose for hemodialysis. Carbohyd Polym. 2018;188:37–47.CrossRef
10.
go back to reference Melchior P, et al. Complement activation by dialysis membranes and its association with secondary membrane formation and surface charge. Artif Organs. 2021;45(7):770–8.PubMedCrossRef Melchior P, et al. Complement activation by dialysis membranes and its association with secondary membrane formation and surface charge. Artif Organs. 2021;45(7):770–8.PubMedCrossRef
11.
go back to reference Petho A, et al. A porcine model of hemodialyzer reactions: roles of complement activation and rinsing back of extracorporeal blood. Ren Fail. 2021;43(1):1609–20.PubMedPubMedCentralCrossRef Petho A, et al. A porcine model of hemodialyzer reactions: roles of complement activation and rinsing back of extracorporeal blood. Ren Fail. 2021;43(1):1609–20.PubMedPubMedCentralCrossRef
13.
go back to reference Arenas Jiménez MD, et al. Adverse reactions to polisulphone membrane dialyzers durind hemodialysis abstract. Nefrologia. 2007;27(5):638–42. Arenas Jiménez MD, et al. Adverse reactions to polisulphone membrane dialyzers durind hemodialysis abstract. Nefrologia. 2007;27(5):638–42.
Metadata
Title
Complement-mediated dialysis reaction during regular hemodialysis treatment: a case report
Authors
Ákos Pethő
Karolina Schnabel
László Dézsi
Publication date
01-12-2024
Publisher
BioMed Central
Published in
Journal of Medical Case Reports / Issue 1/2024
Electronic ISSN: 1752-1947
DOI
https://doi.org/10.1186/s13256-024-04365-x

Other articles of this Issue 1/2024

Journal of Medical Case Reports 1/2024 Go to the issue