Abstract

Purpose

This study was decided to evaluate the impact of diameter and the existences of multiple side holes along the straight portion of double-J ureteral stents (DJUS) on early dysfunction of stents placed for malignant ureteral strictures.

Methods

Between April 2007 and December 2011, 141 DJUSs were placed via a percutaneous nephrostomy (PCN) tract in 110 consecutive patients with malignant ureteral strictures. 7F DJUSs with multiple side holes in the straight portion were placed in 58 ureters of 43 patients (Group 1). 8F DJUSs with three side holes in the proximal 2-cm of the straight portion were placed in 83 ureters of 67 patients (Group 2). The incidence of early DJUS dysfunction was compared between the two groups, and nephrostographic findings were evaluated in the cases of early dysfunction.

Results

Early dysfunction of the DJUS was noted in 14 of 58 patients (24.1 %) in Group 1, which was significantly higher (p = 0.001) than in Group 2 in which only 1 of 83 patients (1.2 %) had early dysfunction of the DJUS. Nephrostographic findings of early dysfunction included dilatation of the pelvicalyceal system, filling defects in the ureteral stent, and no passage of contrast media into the urinary bladder.

Conclusions

In malignant ureteral strictures, multiple side holes in the straight portion of the 7-F DJUS seem to cause early dysfunction. The 8F DJUSs with three side holes in the proximal 2-cm of the straight portion may be superior at preventing early dysfunction.
Literature
1.
go back to reference Zimskind PD, Fetter TR, Wilkerson JL (1967) Clinical use of long-term indwelling silicone rubber ureteral splints inserted cystoscopically. J Urol 97(5):840–844PubMed Zimskind PD, Fetter TR, Wilkerson JL (1967) Clinical use of long-term indwelling silicone rubber ureteral splints inserted cystoscopically. J Urol 97(5):840–844PubMed
2.
go back to reference McCullough DL (1974) ‘Shepherds crook’ self-retaining ureteral catheter. Urol Lett Club 32:54–55 McCullough DL (1974) ‘Shepherds crook’ self-retaining ureteral catheter. Urol Lett Club 32:54–55
3.
go back to reference Druy EM (1985) A dilating introducer-sheath for the antegrade insertion of ureteral stents. AJR Am J Roentgenol 145(6):1274–1276CrossRefPubMed Druy EM (1985) A dilating introducer-sheath for the antegrade insertion of ureteral stents. AJR Am J Roentgenol 145(6):1274–1276CrossRefPubMed
4.
go back to reference Finney RP (1978) Experience with new double J ureteral catheter stent. J Urol 120(6):678–681PubMed Finney RP (1978) Experience with new double J ureteral catheter stent. J Urol 120(6):678–681PubMed
5.
go back to reference Hepperlen TW, Mardis HK, Kammandel H (1978) Self-retained internal ureteral stents: a new approach. J Urol 119(6):731–734PubMed Hepperlen TW, Mardis HK, Kammandel H (1978) Self-retained internal ureteral stents: a new approach. J Urol 119(6):731–734PubMed
6.
go back to reference Mitty HA et al (1988) Experience with a new ureteral stent made of a biocompatible copolymer. Radiology 168(2):557–559CrossRefPubMed Mitty HA et al (1988) Experience with a new ureteral stent made of a biocompatible copolymer. Radiology 168(2):557–559CrossRefPubMed
7.
go back to reference Tschada RK et al (1994) Spiral-reinforced ureteral stent: an alternative for internal urinary diversion. J Endourol 8(2):119–123CrossRefPubMed Tschada RK et al (1994) Spiral-reinforced ureteral stent: an alternative for internal urinary diversion. J Endourol 8(2):119–123CrossRefPubMed
8.
go back to reference Soh KC et al (2008) Is the routine check nephrostogram following percutaneous antegrade ureteric stent placement necessary? Cardiovasc Interv Radiol 31(3):604–609CrossRef Soh KC et al (2008) Is the routine check nephrostogram following percutaneous antegrade ureteric stent placement necessary? Cardiovasc Interv Radiol 31(3):604–609CrossRef
9.
go back to reference Seymour H, Patel U (2000) Ureteric stenting: current status. Semin Interv Radiol 17:351–366CrossRef Seymour H, Patel U (2000) Ureteric stenting: current status. Semin Interv Radiol 17:351–366CrossRef
10.
go back to reference Smedley FH et al (1988) 168 double J (pigtail) ureteric catheter insertions: a retrospective review. Ann R Coll Surg Engl 70(6):377–379PubMedCentralPubMed Smedley FH et al (1988) 168 double J (pigtail) ureteric catheter insertions: a retrospective review. Ann R Coll Surg Engl 70(6):377–379PubMedCentralPubMed
11.
go back to reference Tong JC, Sparrow EM, Abraham JP (2007) Numerical simulation of the urine flow in a stented ureter. J Biomech Eng 129(2):187–192CrossRefPubMed Tong JC, Sparrow EM, Abraham JP (2007) Numerical simulation of the urine flow in a stented ureter. J Biomech Eng 129(2):187–192CrossRefPubMed
12.
go back to reference Christman MS et al (2009) Analysis of ureteral stent compression force and its role in malignant obstruction. J Urol 181(1):392–396CrossRefPubMed Christman MS et al (2009) Analysis of ureteral stent compression force and its role in malignant obstruction. J Urol 181(1):392–396CrossRefPubMed
13.
go back to reference Kunamneni A et al (2008) Urokinase-a very popular cardiovascular agent. Recent Pat Cardiovasc Drug Discov 3(1):45–58CrossRefPubMed Kunamneni A et al (2008) Urokinase-a very popular cardiovascular agent. Recent Pat Cardiovasc Drug Discov 3(1):45–58CrossRefPubMed
14.
go back to reference Docimo SG, Dewolf WC (1989) High failure rate of indwelling ureteral stents in patients with extrinsic obstruction: experience at 2 institutions. J Urol 142(2 Pt 1):277–279PubMed Docimo SG, Dewolf WC (1989) High failure rate of indwelling ureteral stents in patients with extrinsic obstruction: experience at 2 institutions. J Urol 142(2 Pt 1):277–279PubMed
15.
Metadata
Title
Comparison of Two Types of Double-J Ureteral Stents that Differ in Diameter and the Existence of Multiple Side Holes along the Straight Portion in Malignant Ureteral Strictures
Authors
Myung Gyu Song
Tae-Seok Seo
Chang Hee Lee
Kyeong Ah Kim
Jun Suk Kim
Sang Cheul Oh
Jae-Kwan Lee
Publication date
01-06-2015
Publisher
Springer US
Published in
CardioVascular and Interventional Radiology / Issue 3/2015
Print ISSN: 0174-1551
Electronic ISSN: 1432-086X
DOI
https://doi.org/10.1007/s00270-014-0963-5

Other articles of this Issue 3/2015

CardioVascular and Interventional Radiology 3/2015 Go to the issue