Published in: Intensive Care Medicine 7/2003

01-07-2003 | Neonatal and Pediatric Intensive Care

Comparison of two different CPAP systems by tidal breathing parameters

Authors: Thomas Hückstädt, Bertram Foitzik, Roland R. Wauer, Gerd Schmalisch

Published in: Intensive Care Medicine | Issue 7/2003

Login to get access

Abstract

Objective

Comparison of tidal breathing and pressure fluctuation of the continuous positive airway pressure (CPAP) associated with the use of the valveless Infant Flow System versus the conventional constant-flow CPAP (Babylog 8000) in preterm infants.

Design

Randomized cross-over trial.

Setting

Neonatal intensive care unit level III.

Patients

Twenty infants; median (range): birth weight 1,035 g (640–4,110 g), actual weight 1,165 g (820–4,250 g), gestational age at birth 27 (26–40) weeks.

Interventions

After extubation two CPAP devices (Infant Flow System vs Babylog 8000) were applied in a random order to the same infant. Fluctuations of the applied pressure during the breathing cycle and tidal breathing parameters were measured by the flow-through technique.

Main results

Using the Infant Flow System the mean (standard deviation) inspiratory flow [1.5 (0.1) vs 1.3 (0.1) l·min-1·kg-1, P<0.05] and tidal volume were significantly increased [5.3 (1.3) vs 4.7 (1.3) ml/kg-1, P<0.05] compared to Babylog 8000. The fluctuations of the applied pressure of the Infant Flow System during the breathing cycle were significantly lower [0.1 (0.03) kPa vs 0.15 (0.08) kPa, P<0.05] compared to Babylog 8000. No differences were seen in the duration of inspiration and expiration and the time to peak tidal flow. In the Infant Flow System pressures during expiration remained stable whereas they increased during the use of Babylog 8000.

Conclusions

Within-subject comparisons of tidal breathing parameters of the two CPAP devices Infant Flow System and Babylog 8000 show: (1) a significant influence of the system used; and (2) that the valveless Infant Flow System increases air flow and tidal volume with less fluctuations in CPAP pressures during the breathing cycle.
Literature
1.
go back to reference Gregory GA, Kitterman JA, Phibbs RH, Tooley WH, Hamilton WK (1971) Treatment of idiopathic respiratory distress syndrome with continuous positive airway pressure. N Engl J Med 284:1333–1340 Gregory GA, Kitterman JA, Phibbs RH, Tooley WH, Hamilton WK (1971) Treatment of idiopathic respiratory distress syndrome with continuous positive airway pressure. N Engl J Med 284:1333–1340
2.
go back to reference Joris N, Sudre P, Moessinger A (2000) L'application de la CPAP précoce chez les nouveau-nés de moins de 34 semaines de gestation diminue le taux d'intubation et la durée de l'oxygénothérapie sans modifier la mortalité et les morbités. Schweiz Med Wochenschr 130:1887–1893PubMed Joris N, Sudre P, Moessinger A (2000) L'application de la CPAP précoce chez les nouveau-nés de moins de 34 semaines de gestation diminue le taux d'intubation et la durée de l'oxygénothérapie sans modifier la mortalité et les morbités. Schweiz Med Wochenschr 130:1887–1893PubMed
3.
go back to reference Morley CJ, Davis P (2001) Continuous positive airway pressure: randomized, controlled trial in Australia. Pediatrics 108:1383 Morley CJ, Davis P (2001) Continuous positive airway pressure: randomized, controlled trial in Australia. Pediatrics 108:1383
4.
go back to reference de Klerk AM, de Klerk RK (2001) Use of continuous positive airway pressure in preterm infants: comments and experience from New Zealand. Pediatrics 108:761–763PubMed de Klerk AM, de Klerk RK (2001) Use of continuous positive airway pressure in preterm infants: comments and experience from New Zealand. Pediatrics 108:761–763PubMed
5.
go back to reference da Silva WJ, Abbasi S, Pereira G, Bhutani VK (1994) Role of positive end-expiratory pressure changes on functional residual capacity in surfactant treated preterm infants. Pediatr Pulmonol 18:89–92PubMed da Silva WJ, Abbasi S, Pereira G, Bhutani VK (1994) Role of positive end-expiratory pressure changes on functional residual capacity in surfactant treated preterm infants. Pediatr Pulmonol 18:89–92PubMed
6.
go back to reference Andreasson B, Lindroth M, Svenningsen NW, Jonson B (1988) Effects on respiration of CPAP immediately after extubation in the very preterm infant. Pediatr Pulmonol 4:213–218PubMed Andreasson B, Lindroth M, Svenningsen NW, Jonson B (1988) Effects on respiration of CPAP immediately after extubation in the very preterm infant. Pediatr Pulmonol 4:213–218PubMed
7.
go back to reference Miller MJ, Carlo WA, Martin RJ (1985) Continuous positive airway pressure selectively reduces obstructive apnea in preterm infants. J Pediatr 106:91–94PubMed Miller MJ, Carlo WA, Martin RJ (1985) Continuous positive airway pressure selectively reduces obstructive apnea in preterm infants. J Pediatr 106:91–94PubMed
8.
go back to reference Davis PG, Henderson-Smart DJ (1999) Post-extubation prophylactic nasal continuous positive airway pressure in preterm infants: systematic review and meta-analysis. J Paediatr Child Health 35:367–371PubMed Davis PG, Henderson-Smart DJ (1999) Post-extubation prophylactic nasal continuous positive airway pressure in preterm infants: systematic review and meta-analysis. J Paediatr Child Health 35:367–371PubMed
9.
go back to reference Dimitriou G, Greenough A, Kavvadia V, Laubscher B, Alexiou C, Pavlou V, Mantagos S (2000) Elective use of nasal continuous positive airway pressure following extubation of preterm infants. Eur J Pediatr 159:434–439PubMed Dimitriou G, Greenough A, Kavvadia V, Laubscher B, Alexiou C, Pavlou V, Mantagos S (2000) Elective use of nasal continuous positive airway pressure following extubation of preterm infants. Eur J Pediatr 159:434–439PubMed
10.
go back to reference Lee KS, Dunn MS, Fenwick M, Shennan AT (1998) A comparison of underwater bubble continuous positive airway pressure with ventilator-derived continuous positive airway pressure in premature neonates ready for extubation. Biol Neonate 73:69–75CrossRefPubMed Lee KS, Dunn MS, Fenwick M, Shennan AT (1998) A comparison of underwater bubble continuous positive airway pressure with ventilator-derived continuous positive airway pressure in premature neonates ready for extubation. Biol Neonate 73:69–75CrossRefPubMed
11.
go back to reference Benveniste D, Poul Pedersen JE (1968) A valve substitute with no moving parts, for artificial ventilation in newborn and small infants. Brit J Anaesth 40:464–470PubMed Benveniste D, Poul Pedersen JE (1968) A valve substitute with no moving parts, for artificial ventilation in newborn and small infants. Brit J Anaesth 40:464–470PubMed
12.
go back to reference Brendstrup A, Benveniste D, Poul Pedersen JE (1975) The magnitude of air admixture with a jet device in paediatric anaesthesia. Brit J Anaesth 47:1335–1337PubMed Brendstrup A, Benveniste D, Poul Pedersen JE (1975) The magnitude of air admixture with a jet device in paediatric anaesthesia. Brit J Anaesth 47:1335–1337PubMed
13.
go back to reference Reba I (1966) Applications of the coanda effect. Sci Am 214:84–92 Reba I (1966) Applications of the coanda effect. Sci Am 214:84–92
14.
go back to reference Ahluwalia JS, White DK, Morley CJ (1998) Infant Flow Driver or single prong nasal con-tinuous positive airway pressure: short-term physiological effects. Acta Paediatr 87:325–327CrossRefPubMed Ahluwalia JS, White DK, Morley CJ (1998) Infant Flow Driver or single prong nasal con-tinuous positive airway pressure: short-term physiological effects. Acta Paediatr 87:325–327CrossRefPubMed
15.
go back to reference Courtney SE, Pyon KH, Saslow JG, Arnold GK, Pandit PB, Habib RH (2001) Lung recruitment and breathing pattern during variable versus continuous flow nasal continuous airway pressure in premature infants: an evaluation of three devices. Pediatrics 107:304–308PubMed Courtney SE, Pyon KH, Saslow JG, Arnold GK, Pandit PB, Habib RH (2001) Lung recruitment and breathing pattern during variable versus continuous flow nasal continuous airway pressure in premature infants: an evaluation of three devices. Pediatrics 107:304–308PubMed
16.
go back to reference Pandit P, Courtney S, Pyon K, Saslow J, Habib R (2001) Work of breathing during constant- and variable-flow nasal continuous positive airway pressure in preterm neonates. Pediatrics 108:682–685PubMed Pandit P, Courtney S, Pyon K, Saslow J, Habib R (2001) Work of breathing during constant- and variable-flow nasal continuous positive airway pressure in preterm neonates. Pediatrics 108:682–685PubMed
17.
go back to reference Jackson E, Stocks J, Pilgrim L, Dundas I, Dezateux C (1995) A critical assessment of uncalibrated respiratory inductance plethysmography (Respitrace) for the measurement of tidal breathing parameters in newborns and infants. Pediatr Pulmonol 20:119–124PubMed Jackson E, Stocks J, Pilgrim L, Dundas I, Dezateux C (1995) A critical assessment of uncalibrated respiratory inductance plethysmography (Respitrace) for the measurement of tidal breathing parameters in newborns and infants. Pediatr Pulmonol 20:119–124PubMed
18.
go back to reference Foitzik B, Schmidt M, Windstetter D, Wauer RR, Schmalisch G (1998) Leak measurements in spontaneously breathing premature newborns by using the flow-through technique. J Appl Physiol 85:1187–1193PubMed Foitzik B, Schmidt M, Windstetter D, Wauer RR, Schmalisch G (1998) Leak measurements in spontaneously breathing premature newborns by using the flow-through technique. J Appl Physiol 85:1187–1193PubMed
19.
go back to reference Jouvet P, Hubert P, Jarreau PH, Lofaso F, Cloup M, Harf A (1995) Assessment of neonatal ventilator performances. Int Care Med 21:753–758 Jouvet P, Hubert P, Jarreau PH, Lofaso F, Cloup M, Harf A (1995) Assessment of neonatal ventilator performances. Int Care Med 21:753–758
20.
go back to reference Bates J, Schmalisch G, Filbrun D, Stocks J (2000) Tidal breath analysis for infant pulmonary function testing. Eur Respir J 16:1180–1192CrossRefPubMed Bates J, Schmalisch G, Filbrun D, Stocks J (2000) Tidal breath analysis for infant pulmonary function testing. Eur Respir J 16:1180–1192CrossRefPubMed
21.
go back to reference Schmalisch G, Foitzik B, Wauer RR, Stocks J (2001) In vitro assessment of equipment and software used to assess tidal breathing parameters in newborns. Eur Respir J 17:108–114CrossRefPubMed Schmalisch G, Foitzik B, Wauer RR, Stocks J (2001) In vitro assessment of equipment and software used to assess tidal breathing parameters in newborns. Eur Respir J 17:108–114CrossRefPubMed
22.
go back to reference Gaultier C, Fletcher ME, Beardsmore C, England S, Motoyama E (1995) ERS-ATS Statement. Respiratory function measurements in infants: measurement conditions. Eur Respir J 8:1057–1066PubMed Gaultier C, Fletcher ME, Beardsmore C, England S, Motoyama E (1995) ERS-ATS Statement. Respiratory function measurements in infants: measurement conditions. Eur Respir J 8:1057–1066PubMed
23.
go back to reference Schmalisch G, Foitzik B, Wauer R, Stocks J (2001) Effects of apparatus dead space on breathing parameters in newborns: ``flow-through`` versus conventional techniques. Eur Respir J 17:108–114CrossRefPubMed Schmalisch G, Foitzik B, Wauer R, Stocks J (2001) Effects of apparatus dead space on breathing parameters in newborns: ``flow-through`` versus conventional techniques. Eur Respir J 17:108–114CrossRefPubMed
24.
go back to reference Moa G, Nilsson K, Zetterstrom H, Jonsson L (1998) A new device for administration of nasal continuous airway pressure in the newborn: an experimental study. Crit Care Med 12:1238–1242 Moa G, Nilsson K, Zetterstrom H, Jonsson L (1998) A new device for administration of nasal continuous airway pressure in the newborn: an experimental study. Crit Care Med 12:1238–1242
25.
go back to reference Klausener JF, Lee AY, Hutchinson AA (1996) Decreased imposed work with a new nasal continuous airway device. Pediatr Pulmonol 22:188–194PubMed Klausener JF, Lee AY, Hutchinson AA (1996) Decreased imposed work with a new nasal continuous airway device. Pediatr Pulmonol 22:188–194PubMed
26.
go back to reference Stocks J, Dezateux CA, Jackson EA, Hoo AF, Costeloe KL, Wade AM (1994) Analysis of tidal breathing parameters in infancy: how variable is TPTEF:TE? Am J Respir Crit Care Med 150:1347–1354 Stocks J, Dezateux CA, Jackson EA, Hoo AF, Costeloe KL, Wade AM (1994) Analysis of tidal breathing parameters in infancy: how variable is TPTEF:TE? Am J Respir Crit Care Med 150:1347–1354
27.
go back to reference Gherini S, Peters RM, Virgilio RW (1979) Mechanical work on the lungs and work of breathing with positive end-expiratory pressure and continuous positive airway pressure. Chest 76:251–256PubMed Gherini S, Peters RM, Virgilio RW (1979) Mechanical work on the lungs and work of breathing with positive end-expiratory pressure and continuous positive airway pressure. Chest 76:251–256PubMed
28.
29.
go back to reference Storme L, Truffert P, Rakza T, Lequien P (1999) Pour ou contre une utilisation precoce de la pression positive continue nasale et du surfactant exogene au cours de la maladie des membranes hyalines. Arguments physiopathologiques. Arch Pediatr 6:549–555CrossRefPubMed Storme L, Truffert P, Rakza T, Lequien P (1999) Pour ou contre une utilisation precoce de la pression positive continue nasale et du surfactant exogene au cours de la maladie des membranes hyalines. Arguments physiopathologiques. Arch Pediatr 6:549–555CrossRefPubMed
30.
go back to reference Claris O, Salle BL, Lapillonne A, Ronin E, Picaud JC, Besnier S (1996) Nouvelle technique de pression positive continue par voie nasale en neonatologie. Arch Pediatr 3:452–456CrossRefPubMed Claris O, Salle BL, Lapillonne A, Ronin E, Picaud JC, Besnier S (1996) Nouvelle technique de pression positive continue par voie nasale en neonatologie. Arch Pediatr 3:452–456CrossRefPubMed
31.
go back to reference Kavvadia V, Greenough A, Dimitriou G (2000) Prediction of extubation failure in preterm neonates. Eur J Pediatr 159:227–231CrossRefPubMed Kavvadia V, Greenough A, Dimitriou G (2000) Prediction of extubation failure in preterm neonates. Eur J Pediatr 159:227–231CrossRefPubMed
32.
go back to reference Kamper J (1999) Early nasal continuous positive airway pressure and minimal handling in the treatment of very-low-birth-weight infants. Biol Neonate 76:22–28CrossRefPubMed Kamper J (1999) Early nasal continuous positive airway pressure and minimal handling in the treatment of very-low-birth-weight infants. Biol Neonate 76:22–28CrossRefPubMed
33.
go back to reference Robertson NJ, Hamilton PA (1998) Randomised trial of elective continuous positive airway pressure (CPAP) compared with rescue CPAP after extubation. Arch Dis Child Neonatal Ed 79:F58–F60 Robertson NJ, Hamilton PA (1998) Randomised trial of elective continuous positive airway pressure (CPAP) compared with rescue CPAP after extubation. Arch Dis Child Neonatal Ed 79:F58–F60
34.
go back to reference Verder H, Albertsen P, Ebbesen F, Greisen G, Robertson B, Bertelsen A, Agertoft L, Djernes B, Nathan E, Reinholdt J (1999) Nasal continuous positive airway pressure and early surfactant therapy for respiratory distress syndrome in newborns of less than 30 weeks' gestation. Pediatrics 103:24 Verder H, Albertsen P, Ebbesen F, Greisen G, Robertson B, Bertelsen A, Agertoft L, Djernes B, Nathan E, Reinholdt J (1999) Nasal continuous positive airway pressure and early surfactant therapy for respiratory distress syndrome in newborns of less than 30 weeks' gestation. Pediatrics 103:24
35.
go back to reference Hoch B, Bernhard M, Hinsch A (1998) Different patterns of sighs in neonates and young infants. Biol Neonate 74:16–21CrossRefPubMed Hoch B, Bernhard M, Hinsch A (1998) Different patterns of sighs in neonates and young infants. Biol Neonate 74:16–21CrossRefPubMed
36.
go back to reference Wall MA (1980) Infant endotracheal tube resistance: effects of changing length, diameter, and gas density. Crit Care Med 8:38–40PubMed Wall MA (1980) Infant endotracheal tube resistance: effects of changing length, diameter, and gas density. Crit Care Med 8:38–40PubMed
Metadata
Title
Comparison of two different CPAP systems by tidal breathing parameters
Authors
Thomas Hückstädt
Bertram Foitzik
Roland R. Wauer
Gerd Schmalisch
Publication date
01-07-2003
Publisher
Springer-Verlag
Published in
Intensive Care Medicine / Issue 7/2003
Print ISSN: 0342-4642
Electronic ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-003-1785-0

Other articles of this Issue 7/2003

Intensive Care Medicine 7/2003 Go to the issue