Published in: European Archives of Oto-Rhino-Laryngology 9/2008

Open Access 01-09-2008 | Miscellaneous

Comparison of speech intelligibility in quiet and in noise after hearing aid fitting according to a purely prescriptive and a comparative fitting procedure

Authors: Mick Metselaar, Bert Maat, Pieta Krijnen, Hans Verschuure, Wouter Dreschler, Louw Feenstra

Published in: European Archives of Oto-Rhino-Laryngology | Issue 9/2008

Login to get access

Abstract

We compared two different types of hearing-aid fitting procedures in a double-blind randomized clinical study. Hearing aid fittings based on a purely prescriptive procedure (the NAL-RP formula) were compared to a comparative fitting procedure based on optimizing speech intelligibility scores. Main outcome measures were improvement of speech intelligibility scores in quiet and in noise. Data were related to the real-ear insertion responses that were measured after fitting. For analysis purposes subgroups were composed according to degree of hearing loss, characterized by unaided speech intelligibility in quiet, previous experience with hearing aids, unilateral or bilateral fittings and type of hearing aid. We found equal improvement of speech intelligibility in quiet, while fitting according to the prescriptive formula resulted in a somewhat better performance as expressed by the speech-to-noise ratio in comparison to the comparative procedure. Both procedures resulted in comparable real-ear insertion responses.
Literature
1.
go back to reference Gatehouse S (1993) Hearing aid evaluation: limitations of present procedures and future requirements. JSLPA Monogr Suppl 1 Gatehouse S (1993) Hearing aid evaluation: limitations of present procedures and future requirements. JSLPA Monogr Suppl 1
2.
go back to reference Byrne D, Cotton S (1988) Evaluation of the National Acoustic Laboratories’ new hearing aid selection procedure. J Speech Hear Res 31:178–186PubMed Byrne D, Cotton S (1988) Evaluation of the National Acoustic Laboratories’ new hearing aid selection procedure. J Speech Hear Res 31:178–186PubMed
3.
go back to reference Byrne D, Dillon H (1986) The National Acoustic Laboratories’ (NAL) new procedure for selecting the gain and frequency response of a hearing aid. Ear Hear 7:257–265PubMedCrossRef Byrne D, Dillon H (1986) The National Acoustic Laboratories’ (NAL) new procedure for selecting the gain and frequency response of a hearing aid. Ear Hear 7:257–265PubMedCrossRef
4.
go back to reference Byrne D, Parkinson A, Newall PH (1990) Hearing aid gain and frequency response requirements for the severely/profoundly hearing impaired. Ear Hear 11:40–49PubMedCrossRef Byrne D, Parkinson A, Newall PH (1990) Hearing aid gain and frequency response requirements for the severely/profoundly hearing impaired. Ear Hear 11:40–49PubMedCrossRef
5.
go back to reference Byrne D, Tonisson W (1976) Selecting the gain of hearing aids for persons with sensorineural hearing impairments. Scand Audiol 5:51–59 Byrne D, Tonisson W (1976) Selecting the gain of hearing aids for persons with sensorineural hearing impairments. Scand Audiol 5:51–59
6.
go back to reference Carhart R (1946) Tests for selection of hearing aids. Laryngoscope 56:780–794PubMed Carhart R (1946) Tests for selection of hearing aids. Laryngoscope 56:780–794PubMed
7.
go back to reference Bosman A (1989) Speech perception by the hearing impaired. Thesis, Utrecht University, The Netherlands Bosman A (1989) Speech perception by the hearing impaired. Thesis, Utrecht University, The Netherlands
8.
go back to reference Smoorenburg GF (1985) Het spraakaudiogram in relatie tot de gebruikte methode. In: Spraakaudiometrie. Uitgave Nederlandse Vereniging voor Audiologie Smoorenburg GF (1985) Het spraakaudiogram in relatie tot de gebruikte methode. In: Spraakaudiometrie. Uitgave Nederlandse Vereniging voor Audiologie
9.
go back to reference Lybarger SF (1963) Simplified fitting system for hearing aids. Canonsburg, Pennsylvania: Radioear Corp Lybarger SF (1963) Simplified fitting system for hearing aids. Canonsburg, Pennsylvania: Radioear Corp
10.
go back to reference International Electrotechnical Commission (IEC) (1973) IEC reference coupler for the measurements of hearing aids using earphones coupled to the ear by means of ear inserts. Publication 126, 2nd edn. Geneva, Switzerland International Electrotechnical Commission (IEC) (1973) IEC reference coupler for the measurements of hearing aids using earphones coupled to the ear by means of ear inserts. Publication 126, 2nd edn. Geneva, Switzerland
11.
go back to reference Cornelisse LE, Seewald RC, Jamieson DG (1995) The input/output formula: a theoretical approach to the fitting of personal amplification devices. J Acoust Soc Am 97:1854–1864PubMedCrossRef Cornelisse LE, Seewald RC, Jamieson DG (1995) The input/output formula: a theoretical approach to the fitting of personal amplification devices. J Acoust Soc Am 97:1854–1864PubMedCrossRef
12.
go back to reference International Electrotechnical Commission (IEC) (1983) Hearing aids. Part 7. Measurement of the performance of hearing aids for quality inspection for delivery purposes. Publication 118-7, 1st edn. Geneva, Switzerland International Electrotechnical Commission (IEC) (1983) Hearing aids. Part 7. Measurement of the performance of hearing aids for quality inspection for delivery purposes. Publication 118-7, 1st edn. Geneva, Switzerland
13.
go back to reference Verschuure J (1994) Selektie van het hoortoestel. In: Kapteyn TS, Clemens A, Glazenburg BE, Joustra J (eds) Slechthorende en hoortoestel. De Leeuw Press, Rijnsburg Verschuure J (1994) Selektie van het hoortoestel. In: Kapteyn TS, Clemens A, Glazenburg BE, Joustra J (eds) Slechthorende en hoortoestel. De Leeuw Press, Rijnsburg
14.
go back to reference Plomp R, Mimpen AM (1979) Improving the reliability of testing the speech reception threshold for sentences. Audiology 18:43–52PubMedCrossRef Plomp R, Mimpen AM (1979) Improving the reliability of testing the speech reception threshold for sentences. Audiology 18:43–52PubMedCrossRef
15.
go back to reference van Buuren RA, Festen JM, Plomp R (1995) Evaluation of a wide range of amplitude-frequency responses for the hearing impaired. J Speech Hear Res 38:211–221PubMed van Buuren RA, Festen JM, Plomp R (1995) Evaluation of a wide range of amplitude-frequency responses for the hearing impaired. J Speech Hear Res 38:211–221PubMed
16.
go back to reference Verschuure J, van Benthem PPG (1992) Effect of hearing aids on speech perception in noisy situations. Audiology 31:205–221PubMedCrossRef Verschuure J, van Benthem PPG (1992) Effect of hearing aids on speech perception in noisy situations. Audiology 31:205–221PubMedCrossRef
17.
go back to reference van den Heuvel J, Goverts STh, Kapteyn ThS (1997) Evaluation of fitting rules with a programmable hearing aid. Audiology 36:261–278PubMedCrossRef van den Heuvel J, Goverts STh, Kapteyn ThS (1997) Evaluation of fitting rules with a programmable hearing aid. Audiology 36:261–278PubMedCrossRef
18.
go back to reference Lee LW, Humes LF, Wilde G (1993) Evaluating performance with high-frequency emphasis amplification. J Am Acad Audiol 4:91–97PubMed Lee LW, Humes LF, Wilde G (1993) Evaluating performance with high-frequency emphasis amplification. J Am Acad Audiol 4:91–97PubMed
Metadata
Title
Comparison of speech intelligibility in quiet and in noise after hearing aid fitting according to a purely prescriptive and a comparative fitting procedure
Authors
Mick Metselaar
Bert Maat
Pieta Krijnen
Hans Verschuure
Wouter Dreschler
Louw Feenstra
Publication date
01-09-2008
Publisher
Springer-Verlag
Published in
European Archives of Oto-Rhino-Laryngology / Issue 9/2008
Print ISSN: 0937-4477
Electronic ISSN: 1434-4726
DOI
https://doi.org/10.1007/s00405-008-0596-x

Other articles of this Issue 9/2008

European Archives of Oto-Rhino-Laryngology 9/2008 Go to the issue