Comparative Assessment of Energy-Mapping Approaches in CT-Based Attenuation Correction for PET

Authors: Mohammad R. Ay, Maryam Shirmohammad, Saeed Sarkar, Arman Rahmim, Habib Zaidi

Published in: Molecular Imaging and Biology | Issue 1/2011

Login to get access

Abstract

Introduction

Reliable quantification in positron emission tomography (PET) requires accurate attenuation correction of emission data, which in turn entails accurate determination of the attenuation map (µ-map) of the object under study. One of the main steps involved in CT-based attenuation correction (CTAC) is energy-mapping, or the conversion of linear attenuation coefficients (µ) calculated at the effective CT energy to those corresponding to 511 keV.

Materials and methods

The aim of this study is to compare different energy-mapping techniques including scaling, segmentation, the hybrid method, the bilinear calibration curve technique and the dual-energy approach to generate the µ-maps required for attenuation correction. In addition, our newly proposed method involving a quadratic polynomial calibration curve was also assessed. The µ-maps generated for both phantom and clinical studies were assessed qualitatively and quantitatively. A cylindrical polyethylene phantom containing different concentrations of K2HPO4 in water was scanned and the µ-maps calculated from the corresponding CT images using the above-referenced energy-mapping methods. The CT images of five whole-body data sets acquired on a GE Discovery LS PET/CT scanner were employed to generate µ-maps using different energy-mapping approaches that were compared with the µ-maps generated at 511 keV using 68Ge/68Ga rod sources. In another experiment, the evaluation was performed on PET images of a clinical study corrected for attenuation using µ-maps generated using the above described methods. The evaluation was performed for three different tissue types, namely, soft tissue, lung, and bone.

Results and Discussion

All energy-mapping methods yielded almost similar results for soft tissues. The mean relative differences between scaling, segmentation, hybrid, bilinear, and quadratic polynomial calibration curve methods and the transmission scan serving as reference were 6.60%, 6.56%, 6.60%, 5.96%, and 7.36%, respectively. However, the scaling method produced the largest difference (16%) for bone tissues. For lung tissues, the segmentation method produced the largest difference (14.9%). The results for reconstructed PET images followed a similar trend. For soft tissues, all energy-mapping methods yield results in nearly the same range. However, in bone tissues, the scaling method resulted in considerable bias in the µ-maps and the reconstructed PET images. The segmentation method also produced noticeable bias especially in regions with variable densities such as the lung, since a single µ is assigned to the lungs. Apart from the aforementioned case, despite small differences in the generated µ-maps, the use of different energy-mapping methods does not affect, to a visible or measurable extent, the reconstructed PET images.
Literature
1.
go back to reference Sanchez-Crespo A, Andreo P, Larsson SA (2004) Positron flight in human tissues and its influence on PET image spatial resolution. Eur J Nucl Med Mol Imaging 31:44–51CrossRefPubMed Sanchez-Crespo A, Andreo P, Larsson SA (2004) Positron flight in human tissues and its influence on PET image spatial resolution. Eur J Nucl Med Mol Imaging 31:44–51CrossRefPubMed
2.
go back to reference Rousset O, Rahmim A, Alavi A, Zaidi H (2007) Partial volume correction strategies in PET. PET Clinics 2:235–249CrossRef Rousset O, Rahmim A, Alavi A, Zaidi H (2007) Partial volume correction strategies in PET. PET Clinics 2:235–249CrossRef
3.
go back to reference Zaidi H, Koral KF (2004) Scatter modelling and compensation in emission tomography. Eur J Nucl Med Mol Imaging 31:761–782CrossRefPubMed Zaidi H, Koral KF (2004) Scatter modelling and compensation in emission tomography. Eur J Nucl Med Mol Imaging 31:761–782CrossRefPubMed
4.
go back to reference Zaidi H, Hasegawa BH (2003) Determination of the attenuation map in emission tomography. J Nucl Med 44:291–315PubMed Zaidi H, Hasegawa BH (2003) Determination of the attenuation map in emission tomography. J Nucl Med 44:291–315PubMed
5.
go back to reference Osman MM, Cohade C, Nakamoto Y, Wahl RL (2003) Respiratory motion artifacts on PET emission images obtained using CT attenuation correction on PET-CT. Eur J Nucl Med Mol Imaging 30:603–606CrossRefPubMed Osman MM, Cohade C, Nakamoto Y, Wahl RL (2003) Respiratory motion artifacts on PET emission images obtained using CT attenuation correction on PET-CT. Eur J Nucl Med Mol Imaging 30:603–606CrossRefPubMed
6.
go back to reference Lartizien C, Kinahan PE, Swensson R, Comtat C, Lin M, Villemagne V, et al (2003) Evaluating image reconstruction methods for tumor detection in 3-dimensional whole-body PET oncology imaging. J Nucl Med 44:276–290PubMed Lartizien C, Kinahan PE, Swensson R, Comtat C, Lin M, Villemagne V, et al (2003) Evaluating image reconstruction methods for tumor detection in 3-dimensional whole-body PET oncology imaging. J Nucl Med 44:276–290PubMed
7.
go back to reference Bai C, Kinahan PE, Brasse D, Comtat C, Townsend DW, Meltzer CC, et al (2003) An analytic study of the effects of attenuation on tumor detection in whole-body PET oncology imaging. J Nucl Med 44:1855–1861PubMed Bai C, Kinahan PE, Brasse D, Comtat C, Townsend DW, Meltzer CC, et al (2003) An analytic study of the effects of attenuation on tumor detection in whole-body PET oncology imaging. J Nucl Med 44:1855–1861PubMed
8.
go back to reference Kinahan PE, Hasegawa BH, Beyer T (2003) X-ray-based attenuation correction for positron emission tomography/computed tomography scanners. Semin Nucl Med 33:166–179CrossRefPubMed Kinahan PE, Hasegawa BH, Beyer T (2003) X-ray-based attenuation correction for positron emission tomography/computed tomography scanners. Semin Nucl Med 33:166–179CrossRefPubMed
9.
go back to reference Kinahan PE, Townsend DW, Beyer T, Sashin D (1998) Attenuation correction for a combined 3D PET/CT scanner. Med Phys 25:2046–2053CrossRefPubMed Kinahan PE, Townsend DW, Beyer T, Sashin D (1998) Attenuation correction for a combined 3D PET/CT scanner. Med Phys 25:2046–2053CrossRefPubMed
10.
go back to reference Ay M, Zaidi H (2006) Computed Tomography-based attenuation correction in neurological positron emission tomography: evaluation of the effect of x-ray tube voltage on quantitative analysis. Nucl Med Commun 27:339–346CrossRef Ay M, Zaidi H (2006) Computed Tomography-based attenuation correction in neurological positron emission tomography: evaluation of the effect of x-ray tube voltage on quantitative analysis. Nucl Med Commun 27:339–346CrossRef
11.
go back to reference Mawlawi O, Erasmus JJ, Munden RF, Pan T, Knight AE, Macapinlac HA, et al (2006) Quantifying the effect of IV contrast media on integrated PET/CT: clinical evaluation. AJR Am J Roentgenol 186:308–319CrossRefPubMed Mawlawi O, Erasmus JJ, Munden RF, Pan T, Knight AE, Macapinlac HA, et al (2006) Quantifying the effect of IV contrast media on integrated PET/CT: clinical evaluation. AJR Am J Roentgenol 186:308–319CrossRefPubMed
12.
go back to reference Ahmadian A, Ay MR, Bidgoli JH, Sarkar S, Zaidi H (2008) Correction of oral contrast artifacts in CT-based attenuation correction of PET images using an automated segmentation algorithm. Eur J Nucl Med Mol Imaging 35:1812–1823CrossRefPubMed Ahmadian A, Ay MR, Bidgoli JH, Sarkar S, Zaidi H (2008) Correction of oral contrast artifacts in CT-based attenuation correction of PET images using an automated segmentation algorithm. Eur J Nucl Med Mol Imaging 35:1812–1823CrossRefPubMed
13.
go back to reference Ay M, Zaidi H (2006) Assessment of errors caused by x-ray scatter and use of contrast medium when using CT-based attenuation correction in PET. Eur J Nucl Med Mol Imaging 33:1301–1313CrossRefPubMed Ay M, Zaidi H (2006) Assessment of errors caused by x-ray scatter and use of contrast medium when using CT-based attenuation correction in PET. Eur J Nucl Med Mol Imaging 33:1301–1313CrossRefPubMed
14.
go back to reference DiFilippo FP, Brunken RC (2005) Do implanted pacemaker leads and ICD leads cause metal-related artifact in cardiac PET/CT? J Nucl Med 46:436–443PubMed DiFilippo FP, Brunken RC (2005) Do implanted pacemaker leads and ICD leads cause metal-related artifact in cardiac PET/CT? J Nucl Med 46:436–443PubMed
15.
go back to reference Lemmens C, Montandon M-L, Nuyts J, Ratib O, Dupont P, Zaidi H (2008) Impact of metal artefacts due to EEG electrodes in brain PET/CT imaging. Phys Med Biol 53:4417–4429CrossRefPubMed Lemmens C, Montandon M-L, Nuyts J, Ratib O, Dupont P, Zaidi H (2008) Impact of metal artefacts due to EEG electrodes in brain PET/CT imaging. Phys Med Biol 53:4417–4429CrossRefPubMed
16.
go back to reference Bai C, Shao L, Da Silva A, Zhao Z (2003) A generalized model for the conversion from CT numbers to linear attenuation coefficients. IEEE Trans Nucl Sci 50:1510–1515CrossRef Bai C, Shao L, Da Silva A, Zhao Z (2003) A generalized model for the conversion from CT numbers to linear attenuation coefficients. IEEE Trans Nucl Sci 50:1510–1515CrossRef
17.
go back to reference Beyer T, Kinahan PE, Townsend DW, Sashin D (1994) The use of X-ray CT for attenuation correction of PET data. Proc IEEE Nuclear Science Symposium and Medical Imaging Conference. 30 Oct.-5 Nov., Norfolk, VA, USA, 1573–1577 Beyer T, Kinahan PE, Townsend DW, Sashin D (1994) The use of X-ray CT for attenuation correction of PET data. Proc IEEE Nuclear Science Symposium and Medical Imaging Conference. 30 Oct.-5 Nov., Norfolk, VA, USA, 1573–1577
18.
go back to reference Burger C, Goerres G, Schoenes S, Buck A, Lonn AHR, Von Schulthess GK (2002) PET attenuation coefficients from CT images: experimental evaluation of the transformation of CT into PET 511-keV attenuation coefficients. Eur J Nucl Med 29:922–927CrossRef Burger C, Goerres G, Schoenes S, Buck A, Lonn AHR, Von Schulthess GK (2002) PET attenuation coefficients from CT images: experimental evaluation of the transformation of CT into PET 511-keV attenuation coefficients. Eur J Nucl Med 29:922–927CrossRef
19.
go back to reference Guy MJ, Castellano-Smith IA, Flower MA, Flux GD, Ott RJ, Visvikis D (1998) DETECT-dual energy transmission estimation CT-for improved attenuation correction in SPECT and PET. IEEE Trans Nucl Sci 45:1261–1267CrossRef Guy MJ, Castellano-Smith IA, Flower MA, Flux GD, Ott RJ, Visvikis D (1998) DETECT-dual energy transmission estimation CT-for improved attenuation correction in SPECT and PET. IEEE Trans Nucl Sci 45:1261–1267CrossRef
20.
go back to reference Kinahan PE, Alessio AM, Fessler JA (2006) Dual energy CT attenuation correction methods for quantitative assessment of response to cancer therapy with PET/CT imaging. Technol Cancer Res Treat 5:319–327PubMed Kinahan PE, Alessio AM, Fessler JA (2006) Dual energy CT attenuation correction methods for quantitative assessment of response to cancer therapy with PET/CT imaging. Technol Cancer Res Treat 5:319–327PubMed
21.
go back to reference Shirmohammad M, Ay M, Sarkar S, Ghadiri H, Rahmim A (2008) Comparative assessment of different energy mapping methods for generation of 511 KeV attenuation map from CT images in PET/CT systems: a phantom study. The Fifth IEEE International Symposium on Biomedical Imaging: From Nano to Macro. Paris, France, pp 644–647 Shirmohammad M, Ay M, Sarkar S, Ghadiri H, Rahmim A (2008) Comparative assessment of different energy mapping methods for generation of 511 KeV attenuation map from CT images in PET/CT systems: a phantom study. The Fifth IEEE International Symposium on Biomedical Imaging: From Nano to Macro. Paris, France, pp 644–647
22.
go back to reference Shirmohammad M, Ay M, Rahmim A, Sarkar S, Zaidi H (2008) A novel energy mapping method for attenuation map generation at 511 keV in computed tomography based attenuation correction [abstract]. Eur J Nucl Med Mol Imaging 35:S146CrossRef Shirmohammad M, Ay M, Rahmim A, Sarkar S, Zaidi H (2008) A novel energy mapping method for attenuation map generation at 511 keV in computed tomography based attenuation correction [abstract]. Eur J Nucl Med Mol Imaging 35:S146CrossRef
23.
go back to reference Flohr TG, McCollough CH, Bruder H, Petersilka M, Gruber K, Suss C, et al (2006) First performance evaluation of a dual-source CT (DSCT) system. Eur Radiol 16:256–268CrossRefPubMed Flohr TG, McCollough CH, Bruder H, Petersilka M, Gruber K, Suss C, et al (2006) First performance evaluation of a dual-source CT (DSCT) system. Eur Radiol 16:256–268CrossRefPubMed
24.
go back to reference Greess H, Lutze J, Nomayr A, Wolf H, Hothorn T, Kalender WA, et al (2004) Dose reduction in subsecond multislice spiral CT examination of children by online tube current modulation. Eur Radiol 14:995–999CrossRefPubMed Greess H, Lutze J, Nomayr A, Wolf H, Hothorn T, Kalender WA, et al (2004) Dose reduction in subsecond multislice spiral CT examination of children by online tube current modulation. Eur Radiol 14:995–999CrossRefPubMed
25.
go back to reference Wang J, Li T, Lu H, Liang Z (2006) Noise reduction for low-dose single-slice helical CT sinograms. IEEE Trans Nucl Sci 53:1230–1237CrossRefPubMed Wang J, Li T, Lu H, Liang Z (2006) Noise reduction for low-dose single-slice helical CT sinograms. IEEE Trans Nucl Sci 53:1230–1237CrossRefPubMed
27.
go back to reference Rahmim A, Tang J, Lodge MA, Lashkari S, Ay MR, Lautamaki R, et al (2008) Analytic system matrix resolution modeling in PET: an application to Rb-82 cardiac imaging. Phys Med Biol 53:5947–5965CrossRefPubMed Rahmim A, Tang J, Lodge MA, Lashkari S, Ay MR, Lautamaki R, et al (2008) Analytic system matrix resolution modeling in PET: an application to Rb-82 cardiac imaging. Phys Med Biol 53:5947–5965CrossRefPubMed
28.
go back to reference Robinson PJ, Kreel L (1979) Pulmonary tissue attenuation with computed tomography: comparison of inspiration and expiration scans. J Comput Assist Tomogr 3:740–748PubMed Robinson PJ, Kreel L (1979) Pulmonary tissue attenuation with computed tomography: comparison of inspiration and expiration scans. J Comput Assist Tomogr 3:740–748PubMed
29.
go back to reference Zaidi H (2007) Is MRI-guided attenuation correction a viable option for dual-modality PET/MR imaging? Radiology 244:639–642CrossRefPubMed Zaidi H (2007) Is MRI-guided attenuation correction a viable option for dual-modality PET/MR imaging? Radiology 244:639–642CrossRefPubMed
30.
go back to reference Nahmias C, Lemmens C, Faul D, Carlson E, Long M, Blodgett T, et al (2008) Does reducing CT artifacts from dental implants influence the PET interpretation in PET/CT studies of oral cancer and head and neck cancer? J Nucl Med 49:1047–1052CrossRefPubMed Nahmias C, Lemmens C, Faul D, Carlson E, Long M, Blodgett T, et al (2008) Does reducing CT artifacts from dental implants influence the PET interpretation in PET/CT studies of oral cancer and head and neck cancer? J Nucl Med 49:1047–1052CrossRefPubMed
Metadata
Title
Comparative Assessment of Energy-Mapping Approaches in CT-Based Attenuation Correction for PET
Authors
Mohammad R. Ay
Maryam Shirmohammad
Saeed Sarkar
Arman Rahmim
Habib Zaidi
Publication date
01-02-2011
Publisher
Springer-Verlag
Published in
Molecular Imaging and Biology / Issue 1/2011
Print ISSN: 1536-1632
Electronic ISSN: 1860-2002
DOI
https://doi.org/10.1007/s11307-010-0303-3

Other articles of this Issue 1/2011

Molecular Imaging and Biology 1/2011 Go to the issue