Commissioning and early experience with a new-generation low-energy linear accelerator with advanced delivery and imaging functionalities

Authors: Alessandro Clivio, Giorgia Nicolini, Eugenio Vanetti, Antonella Fogliata, Luca Cozzi

Published in: Radiation Oncology | Issue 1/2011

Login to get access

Abstract

Background

A new-generation low-energy linear accelerator (UNIQUE) was introduced in the clinical arena during 2009 by Varian Medical Systems. The world's first UNIQUE was installed at Oncology Institute of Southern Switzerland and put into clinical operation in June 2010. The aim of the present contribution was to report experience about its commissioning and first year results from clinical operation.

Methods

Commissioning data, beam characteristics and the modeling into the treatment planning system were summarized. Imaging system of UNIQUE included a 2D-2D matching capability and tests were performed to identify system repositioning capability. Finally, since the system is capable of delivering volumetric modulated arc therapy with RapidArc, a summary of the tests performed for such modality to assess its performance in preclinical settings and during clinical usage was included.

Results

Isocenter virtual diameter was measured as less than 0.2 mm. Observed accuracy of isocenter determination and repositioning for 2D-2D matching procedures in image guidance was <1.2 mm. Concerning reproducibility and stability over a period of 1 year, deviations from reference were found <0.3 ± 0.2% for linac output, <0.1% for homogeneity, similarly to symmetry. Rotational accuracy of the entire gantry-portal imager system showed a maximum deviation from nominal 0.0 of <1.2 mm. Pre treatment quality assurance of RapidArc plans resulted with a Gamma Agreement Index (fraction of points passing the gamma criteria) of 97.0 ± 1.6% on the first 182 arcs verified.

Conclusions

The results of the commissioning tests and of the first period of clinical operation, resulted meeting specifications and having good margins respect to tolerances. UNIQUE was put into operation for all delivery techniques; in particular, as shown by the pre-treatment quality assurance results, it enabled accurate and safe delivery of RapidArc plans.
Appendix
Available only for authorised users
Literature
1.
go back to reference Das IJ, Cheng CW, Watts RJ, Ahnesjö A, Gibbons J, Li XA, Lowenstein J, Mitra RK, Simon WE, Zhu TC: TG-106 of the Therapy Physics Committee of the AAPM. Accelerator beam data commissioning equipment and procedures: report of the TG-106 of the Therapy Physics Committee of the AAPM. Med Phys 2008, 35: 4186-4215. 10.1118/1.2969070CrossRefPubMed Das IJ, Cheng CW, Watts RJ, Ahnesjö A, Gibbons J, Li XA, Lowenstein J, Mitra RK, Simon WE, Zhu TC: TG-106 of the Therapy Physics Committee of the AAPM. Accelerator beam data commissioning equipment and procedures: report of the TG-106 of the Therapy Physics Committee of the AAPM. Med Phys 2008, 35: 4186-4215. 10.1118/1.2969070CrossRefPubMed
2.
go back to reference Aletti P, Bey P, Chauvel P, Chavaudra J, Costa A, Donnareix D, Gaboriaud G, Lagrange JL, Manny C, Ponvert D, Rozan R, Valinta D, Van Dam J: Recommendations for a quality assurance programme in external radiotherapy. 1995. Aletti P, Bey P, Chauvel P, Chavaudra J, Costa A, Donnareix D, Gaboriaud G, Lagrange JL, Manny C, Ponvert D, Rozan R, Valinta D, Van Dam J: Recommendations for a quality assurance programme in external radiotherapy. 1995.
3.
go back to reference Kapanen M, Tenhunen M, Hämäläinen T, Sipilä P, Parkkinen R, Järvinen H: Analysis of quality control data of eight modern radiotherapy linear accelerators: the short- and long-term behaviours of the outputs and the reproducibility of quality control measurements. Phys Med Biol 2006, 51: 3581-3592. 10.1088/0031-9155/51/14/020CrossRefPubMed Kapanen M, Tenhunen M, Hämäläinen T, Sipilä P, Parkkinen R, Järvinen H: Analysis of quality control data of eight modern radiotherapy linear accelerators: the short- and long-term behaviours of the outputs and the reproducibility of quality control measurements. Phys Med Biol 2006, 51: 3581-3592. 10.1088/0031-9155/51/14/020CrossRefPubMed
4.
go back to reference Brahme A: Dosimetric precision requirements in radiation therapy. Acta Radiol Oncol 1984, 23: 379-391. 10.3109/02841868409136037CrossRefPubMed Brahme A: Dosimetric precision requirements in radiation therapy. Acta Radiol Oncol 1984, 23: 379-391. 10.3109/02841868409136037CrossRefPubMed
5.
go back to reference Swiss Society of Radiobiology and Medical Physics: Report number 11. Quality control of medical electron accelerators. 2003. Swiss Society of Radiobiology and Medical Physics: Report number 11. Quality control of medical electron accelerators. 2003.
6.
go back to reference Fogliata A, Nicolini G, Vanetti E, Clivio A, Cozzi L: Dosimetric validation of the Anisotropic Analytical Algorithm for photon dose calculation: fundamental characterisation in water. Phys Med Biol 2006, 51: 1421-1438. 10.1088/0031-9155/51/6/004CrossRefPubMed Fogliata A, Nicolini G, Vanetti E, Clivio A, Cozzi L: Dosimetric validation of the Anisotropic Analytical Algorithm for photon dose calculation: fundamental characterisation in water. Phys Med Biol 2006, 51: 1421-1438. 10.1088/0031-9155/51/6/004CrossRefPubMed
7.
go back to reference Ulmer W, Pyyry J, Kaissl WA: 3D photon superposition/convolution algorithm and its foundation on results of Monte Carlo calculations. Phys Med Biol 2005, 50: 1767-1790. 10.1088/0031-9155/50/8/010CrossRefPubMed Ulmer W, Pyyry J, Kaissl WA: 3D photon superposition/convolution algorithm and its foundation on results of Monte Carlo calculations. Phys Med Biol 2005, 50: 1767-1790. 10.1088/0031-9155/50/8/010CrossRefPubMed
8.
go back to reference Nicolini G, Vanetti E, Clivio A, Fogliata A, Boka G, Cozzi L: Testing the portal imager GLAaS algorithm for machine quality assurance. Radiat Oncol 2008, 3: 14. 10.1186/1748-717X-3-14PubMedCentralCrossRefPubMed Nicolini G, Vanetti E, Clivio A, Fogliata A, Boka G, Cozzi L: Testing the portal imager GLAaS algorithm for machine quality assurance. Radiat Oncol 2008, 3: 14. 10.1186/1748-717X-3-14PubMedCentralCrossRefPubMed
9.
go back to reference Low DA, Harms WB, Mutic S, Purdy JA: A technique for quantitative evaluation of dose distributions. Med Phys 1998, 25: 656-661. 10.1118/1.598248CrossRefPubMed Low DA, Harms WB, Mutic S, Purdy JA: A technique for quantitative evaluation of dose distributions. Med Phys 1998, 25: 656-661. 10.1118/1.598248CrossRefPubMed
10.
go back to reference Yoo S, Kim G, Hammoud R, Elder E, Pawlicki T, Guan H, Fox T, Luxton G, Yn FF, Munro P: A quality assurance program for the on board imager. Med Phys 2006, 33: 4431-4447. 10.1118/1.2362872CrossRefPubMed Yoo S, Kim G, Hammoud R, Elder E, Pawlicki T, Guan H, Fox T, Luxton G, Yn FF, Munro P: A quality assurance program for the on board imager. Med Phys 2006, 33: 4431-4447. 10.1118/1.2362872CrossRefPubMed
11.
go back to reference Nicolini G, Vanetti E, Clivio A, Fogliata A, Korreman S, Bocanek J, Cozzi L: The GLAaS algorithm for portal dosimetry and quality assurance of RapidArc, an intensity modulated rotational therapy. Radiat Oncol 2008, 3: 24. 10.1186/1748-717X-3-24PubMedCentralCrossRefPubMed Nicolini G, Vanetti E, Clivio A, Fogliata A, Korreman S, Bocanek J, Cozzi L: The GLAaS algorithm for portal dosimetry and quality assurance of RapidArc, an intensity modulated rotational therapy. Radiat Oncol 2008, 3: 24. 10.1186/1748-717X-3-24PubMedCentralCrossRefPubMed
12.
go back to reference Cozzi L, Dinshaw KA, Shrivastava SK, Mahantshetty U, Engineer R, Deshpande DD, Jamema SV, Vanetti E, Clivio A, Nicolini G, Fogliata A: A treatment planning study comparing volumetric arc modulation with RapidArc and fixed field IMRT for cervix uteri radiotherapy. Radiother Oncol 2008, 89: 180-191. 10.1016/j.radonc.2008.06.013CrossRefPubMed Cozzi L, Dinshaw KA, Shrivastava SK, Mahantshetty U, Engineer R, Deshpande DD, Jamema SV, Vanetti E, Clivio A, Nicolini G, Fogliata A: A treatment planning study comparing volumetric arc modulation with RapidArc and fixed field IMRT for cervix uteri radiotherapy. Radiother Oncol 2008, 89: 180-191. 10.1016/j.radonc.2008.06.013CrossRefPubMed
13.
go back to reference Ling C, Zhang P, Archambault Y, Bocanek J, Tank G, LoSasso T: Commissioning and quality assurance of RapidArc radiotherapy delivery system. Int J Radiat Oncol Biol Phys 2008, 72: 575-581. 10.1016/j.ijrobp.2008.05.060CrossRefPubMed Ling C, Zhang P, Archambault Y, Bocanek J, Tank G, LoSasso T: Commissioning and quality assurance of RapidArc radiotherapy delivery system. Int J Radiat Oncol Biol Phys 2008, 72: 575-581. 10.1016/j.ijrobp.2008.05.060CrossRefPubMed
Metadata
Title
Commissioning and early experience with a new-generation low-energy linear accelerator with advanced delivery and imaging functionalities
Authors
Alessandro Clivio
Giorgia Nicolini
Eugenio Vanetti
Antonella Fogliata
Luca Cozzi
Publication date
01-12-2011
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2011
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/1748-717X-6-129

Other articles of this Issue 1/2011

Radiation Oncology 1/2011 Go to the issue