Colorectal anastomosis facilitated by the use of the LigaSure® sealing device: comparative study in an animal model

Authors: Francisco Sánchez-De Pedro, Carlos Moreno-Sanz, Antonio Morandeira-Rivas, Jose María Tenías-Burillo, Cristina Alhambra-Rodríguez De Guzmán

Published in: Surgical Endoscopy | Issue 2/2014

Login to get access

Abstract

Background

Recently, there has been speculation about the possibility of fusing intestinal tissue using bipolar sealing devices. In this study we compare in a porcine model the anastomoses created using the LigaSure® device with those created with a stapler after section and closure of a rectal stump.

Methods

Thirty pigs underwent laparoscopic colorectal anastomosis. In group A (n = 15), the division of the intestine and distal stump closure were done with a 10-mm LigaSure Atlas® device. In group B, these steps were carried out using an endoscopic stapler. Subsequently, the colorectal anastomosis was performed using circular stapling in both groups. The 4-week follow-up included health status, weight gain, blood tests, X-rays, and colonoscopy. Anastomotic tissue was processed to study the mechanical tensile strength and histopathology.

Results

There was no difference in the rate of conversion to open surgery or in average operating time between the groups. In the sealing device group, there was a significantly higher rate of failure in rectal stump closure (p = 0.042). There was one death in group B due to anastomotic leak. There was no difference in adhesion formation or stenosis. Mid-section anastomosis area was 89.7 mm2 in group A compared with 100 mm2 in group B (p = 0.52). In tensile strength studies, the maximum load resisted by the sample was 13.8 ± 4.9 N (group A) versus 15.7 ± 4.4 N (group B) (p = 0.17). There was no difference between the groups in degree of reepithelialization, number of inflammatory cells, or the presence of microabscesses.

Conclusions

Division and sealing of the rectal stump with the LigaSure® device is feasible in the proposed experimental model, but it is less reliable than conventional closure with a stapler, since it has a significantly greater failure rate. Therefore, The LigaSure® device should not be used for this purpose in the clinical setting as this could lead to serious and dramatic complications.
Literature
2.
go back to reference Shields CA, Schechter DA, Tetzlaff P, Baily AL, Dycus S, Cosgriff N (2004) Method for creating ideal tissue fusion in soft-tissue structures using radio frequency (RF) energy. Surg Technol Int 13:49–55PubMed Shields CA, Schechter DA, Tetzlaff P, Baily AL, Dycus S, Cosgriff N (2004) Method for creating ideal tissue fusion in soft-tissue structures using radio frequency (RF) energy. Surg Technol Int 13:49–55PubMed
3.
go back to reference Miquilarena R, Coronel P, Arocha R, Troconis E, Navas H (2006) Cierre del muñón apendicular con Ligasure™ en conejos. Un reporte preliminar. Rev Venez Cir 59(1):8–11 Miquilarena R, Coronel P, Arocha R, Troconis E, Navas H (2006) Cierre del muñón apendicular con Ligasure™ en conejos. Un reporte preliminar. Rev Venez Cir 59(1):8–11
4.
go back to reference Aslan A, Karaveli C, Elpek O (2008) Laparoscopic appendectomy without clip or ligature. An experimental study. Surg Endosc 22(9):2084–2087PubMedCrossRef Aslan A, Karaveli C, Elpek O (2008) Laparoscopic appendectomy without clip or ligature. An experimental study. Surg Endosc 22(9):2084–2087PubMedCrossRef
5.
go back to reference Elemen L, Yazir Y, Tugay M, Akay A, Aydin S, Yanar K, Ceylan S (2010) LigaSure compared with ligatures and endoclips in experimental appendectomy: how safe is it? Pediatr Surg Int 26(5):539–545PubMedCrossRef Elemen L, Yazir Y, Tugay M, Akay A, Aydin S, Yanar K, Ceylan S (2010) LigaSure compared with ligatures and endoclips in experimental appendectomy: how safe is it? Pediatr Surg Int 26(5):539–545PubMedCrossRef
6.
go back to reference Khanna S, Khurana S, Vij S (2004) No clip, no ligature laparoscopic appendectomy. Surg Laparosc Endosc Percutan Tech 14(4):201–203PubMedCrossRef Khanna S, Khurana S, Vij S (2004) No clip, no ligature laparoscopic appendectomy. Surg Laparosc Endosc Percutan Tech 14(4):201–203PubMedCrossRef
7.
go back to reference Salameh JR, Schwartz JH, Hildebrandt DA (2006) Can LigaSure seal and divide the small bowel? Am J Surg 191:791–793PubMedCrossRef Salameh JR, Schwartz JH, Hildebrandt DA (2006) Can LigaSure seal and divide the small bowel? Am J Surg 191:791–793PubMedCrossRef
8.
go back to reference Smulders JF, de Hingh IH, Stavast J, Jackimowicz JJ (2007) Exploring new technologies to facilitate laparoscopic surgery: creating intestinal anastomoses without sutures or staples, using a radio-frequency-energy-driven bipolar fusion device. Surg Endosc 21(11):2105–2109PubMedCrossRef Smulders JF, de Hingh IH, Stavast J, Jackimowicz JJ (2007) Exploring new technologies to facilitate laparoscopic surgery: creating intestinal anastomoses without sutures or staples, using a radio-frequency-energy-driven bipolar fusion device. Surg Endosc 21(11):2105–2109PubMedCrossRef
9.
go back to reference Aslan A, Elpek O (2009) A novel technique of sutureless colorectal anastomosis by bipolar coagulation. Eur J Pediatr Surg 19(6):384–387PubMedCrossRef Aslan A, Elpek O (2009) A novel technique of sutureless colorectal anastomosis by bipolar coagulation. Eur J Pediatr Surg 19(6):384–387PubMedCrossRef
10.
go back to reference Winter H, Holmer C, Buhr HJ, Lindner G, Lauster R, Kraft M, Ritz JP (2010) Pilot study of bipolar radiofrequency-induced anastomotic thermofusion: exploration of therapy parameters ex vivo. Int J Colorectal Dis 25(1):129–133PubMedCrossRef Winter H, Holmer C, Buhr HJ, Lindner G, Lauster R, Kraft M, Ritz JP (2010) Pilot study of bipolar radiofrequency-induced anastomotic thermofusion: exploration of therapy parameters ex vivo. Int J Colorectal Dis 25(1):129–133PubMedCrossRef
11.
go back to reference Holmer C, Winter H, Kröger M, Nagel A, Jaenicke A, Lauster R, Kraft M, Buhr HJ, Ritz JP (2011) Bipolar radiofrequency-induced thermofusion of intestinal anastomoses—feasibility of a new anastomosis technique in porcine and rat colon. Langenbecks Arch Surg 396(4):529–533PubMedCrossRef Holmer C, Winter H, Kröger M, Nagel A, Jaenicke A, Lauster R, Kraft M, Buhr HJ, Ritz JP (2011) Bipolar radiofrequency-induced thermofusion of intestinal anastomoses—feasibility of a new anastomosis technique in porcine and rat colon. Langenbecks Arch Surg 396(4):529–533PubMedCrossRef
12.
go back to reference Moreno-Sanz C, Picazo-Yeste J, Seoane-Gonzáles J, Manzanera-Díaz M, Tadeo-Ruiz G (2008) Division of the small bowel with the LigaSure Atlas device during the right laparoscopic colectomy. J Laparoendosc Adv Surg Tech A 18(1):99–101PubMedCrossRef Moreno-Sanz C, Picazo-Yeste J, Seoane-Gonzáles J, Manzanera-Díaz M, Tadeo-Ruiz G (2008) Division of the small bowel with the LigaSure Atlas device during the right laparoscopic colectomy. J Laparoendosc Adv Surg Tech A 18(1):99–101PubMedCrossRef
13.
go back to reference Himpens J, Leman G, Sonneville T (2005) Laparoscopic Roux-en-Y gastric bypass performed without staples. Surg Endosc 19:1003PubMedCrossRef Himpens J, Leman G, Sonneville T (2005) Laparoscopic Roux-en-Y gastric bypass performed without staples. Surg Endosc 19:1003PubMedCrossRef
14.
go back to reference Ettinger JE, Ramos AC, Azaro E, Galvão-Neto MP, Mello CA, Galvão MS, Amaral PC, Carlo A, Fahel E (2006) Staplerless laparoscopic gastric bypass: a new option in bariatric surgery. Obes Surg 16(5):638–645PubMedCrossRef Ettinger JE, Ramos AC, Azaro E, Galvão-Neto MP, Mello CA, Galvão MS, Amaral PC, Carlo A, Fahel E (2006) Staplerless laparoscopic gastric bypass: a new option in bariatric surgery. Obes Surg 16(5):638–645PubMedCrossRef
15.
go back to reference Hoebeke PB, Zubair S, Oosterlinck W, Verbeeck RM (1992) Low-voltage coagulation for welding sutures watertight: an experimental study. Urol Res 20(6):25–27CrossRef Hoebeke PB, Zubair S, Oosterlinck W, Verbeeck RM (1992) Low-voltage coagulation for welding sutures watertight: an experimental study. Urol Res 20(6):25–27CrossRef
16.
go back to reference Wallwiener CW, Rajab TK, Zubke W, Isaacson KB, Enderle M, Schäller D, Wallwiener M (2008) Thermal conduction, compression, and electrical current: an evaluation of major parameters of electrosurgical vessel sealing in a porcine in vitro model. J Minim Invasive Gynecol 15(5):605–610PubMedCrossRef Wallwiener CW, Rajab TK, Zubke W, Isaacson KB, Enderle M, Schäller D, Wallwiener M (2008) Thermal conduction, compression, and electrical current: an evaluation of major parameters of electrosurgical vessel sealing in a porcine in vitro model. J Minim Invasive Gynecol 15(5):605–610PubMedCrossRef
17.
go back to reference Rahbari NN, Weitz J, Hohenberger W, Heald RJ, Moran B, Ulrich A, Holm T, Wong WD, Tiret E, Moriya Y, Laurberg S, den Dulk M, van de Velde C, Büchler MW (2010) Definition and grading of anastomotic leakage following anterior resection of the rectum: a proposal by the International Study Group of Rectal Cancer. Surgery 147(3):339–351PubMedCrossRef Rahbari NN, Weitz J, Hohenberger W, Heald RJ, Moran B, Ulrich A, Holm T, Wong WD, Tiret E, Moriya Y, Laurberg S, den Dulk M, van de Velde C, Büchler MW (2010) Definition and grading of anastomotic leakage following anterior resection of the rectum: a proposal by the International Study Group of Rectal Cancer. Surgery 147(3):339–351PubMedCrossRef
Metadata
Title
Colorectal anastomosis facilitated by the use of the LigaSure® sealing device: comparative study in an animal model
Authors
Francisco Sánchez-De Pedro
Carlos Moreno-Sanz
Antonio Morandeira-Rivas
Jose María Tenías-Burillo
Cristina Alhambra-Rodríguez De Guzmán
Publication date
01-02-2014
Publisher
Springer US
Published in
Surgical Endoscopy / Issue 2/2014
Print ISSN: 0930-2794
Electronic ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-013-3194-y

Other articles of this Issue 2/2014

Surgical Endoscopy 2/2014 Go to the issue