Published in: European Spine Journal 11/2014

01-11-2014 | Original Article

Circumferential dynamic stabilization of the lumbar spine: a biomechanical analysis

Authors: Wolfram Käfer, Balkan Cakir, Stefan Midderhoff, Heiko Reichel, Hans-Joachim Wilke

Published in: European Spine Journal | Issue 11/2014

Login to get access

Abstract

Purpose

To assess segmental angulation and mobility following implantation of the Charité artificial disc in combination with the posterior dynamic fixation device dynamic stabilization system (DSS) and the interspinous spacer Coflex at the L4–L5 segment, respectively.

Methods

Six human L4–L5 specimens were loaded with pure moments of ±7.5 Nm in flexion/extension, lateral bending, and axial rotation in a custom-made spine tester. The testing protocol was as follows: (a) intact condition, (b) destabilization by resection of the anterior longitudinal ligament (ALL), (c) implantation of the Charité with retained posterior longitudinal ligament (PLL), (d) supplemental DSS implantation, (e) removal of DSS rods and PLL resection, (f) DSS rod re-implantation, (g) enlargement of rod length, and (h) removal of DSS and implantation of Coflex. Range of motion (ROM), neutral zone, and segmental angulation were determined.

Results

ALL resection did not influence significantly ROM. TDR increased lateral bending and axial rotation only after resection of the PLL, whereas flexion/extension remained unchanged. DSS limited all degrees of freedom prior to and after PLL resection. Rod length enlargement had no significant effect. Coflex limited significantly flexion/extension compared to the intact state and TDR, whereas lateral bending and axial rotation remained unchanged. TDR increased lordosis, whereas Coflex had a substantial kyphosing effect.

Conclusions

This study demonstrates that posterior dynamic stabilization in combination with TDR reduces flexion/extension ROM and segmental lordosis in a monosegmental biomechanical model.
Literature
1.
go back to reference Cakir B, Richter M, Kafer W, Puhl W, Schmidt R (2005) The impact of total lumbar disc replacement on segmental and total lumbar lordosis. Clin Biomech (Bristol, Avon) 20(4):357–364CrossRef Cakir B, Richter M, Kafer W, Puhl W, Schmidt R (2005) The impact of total lumbar disc replacement on segmental and total lumbar lordosis. Clin Biomech (Bristol, Avon) 20(4):357–364CrossRef
2.
go back to reference Rohlmann A, Zander T, Bergmann G (2005) Effect of total disc replacement with ProDisc on intersegmental rotation of the lumbar spine. Spine 30(7):738–743PubMedCrossRef Rohlmann A, Zander T, Bergmann G (2005) Effect of total disc replacement with ProDisc on intersegmental rotation of the lumbar spine. Spine 30(7):738–743PubMedCrossRef
3.
go back to reference Kafer W, Clessienne CB, Daxle M, Kocak T, Reichel H, Cakir B (2008) Posterior component impingement after lumbar total disc replacement: a radiographic analysis of 66 ProDisc-L prostheses in 56 patients. Spine 33(22):2444–2449PubMedCrossRef Kafer W, Clessienne CB, Daxle M, Kocak T, Reichel H, Cakir B (2008) Posterior component impingement after lumbar total disc replacement: a radiographic analysis of 66 ProDisc-L prostheses in 56 patients. Spine 33(22):2444–2449PubMedCrossRef
4.
go back to reference Hallab N, Link HD, McAfee PC (2003) Biomaterial optimization in total disc arthroplasty. Spine 28(20):139–152CrossRef Hallab N, Link HD, McAfee PC (2003) Biomaterial optimization in total disc arthroplasty. Spine 28(20):139–152CrossRef
5.
go back to reference Kurtz SM, Peloza J, Siskey R, Villarraga ML (2005) Analysis of a retrieved polyethylene total disc replacement component. Spine J 5(3):344–350PubMedCrossRef Kurtz SM, Peloza J, Siskey R, Villarraga ML (2005) Analysis of a retrieved polyethylene total disc replacement component. Spine J 5(3):344–350PubMedCrossRef
6.
go back to reference Kurtz SM, van Ooij A, Ross R, de Waal Malefijt J, Peloza J, Ciccarelli L, Villarraga ML (2007) Polyethylene wear and rim fracture in total disc arthroplasty. Spine J 7(1):12–21PubMedCrossRef Kurtz SM, van Ooij A, Ross R, de Waal Malefijt J, Peloza J, Ciccarelli L, Villarraga ML (2007) Polyethylene wear and rim fracture in total disc arthroplasty. Spine J 7(1):12–21PubMedCrossRef
7.
go back to reference van Ooij A, Kurtz SM, Stessels F, Noten H, van Rhijn L (2007) Polyethylene wear debris and long-term clinical failure of the Charite disc prosthesis: a study of 4 patients. Spine 32(2):223–229PubMedCrossRef van Ooij A, Kurtz SM, Stessels F, Noten H, van Rhijn L (2007) Polyethylene wear debris and long-term clinical failure of the Charite disc prosthesis: a study of 4 patients. Spine 32(2):223–229PubMedCrossRef
8.
go back to reference van Ooij A, Oner FC, Verbout AJ (2003) Complications of artificial disc replacement: a report of 27 patients with the SB Charite disc. J Spinal Disord Tech 16(4):369–383PubMedCrossRef van Ooij A, Oner FC, Verbout AJ (2003) Complications of artificial disc replacement: a report of 27 patients with the SB Charite disc. J Spinal Disord Tech 16(4):369–383PubMedCrossRef
9.
go back to reference Punt IM, Visser VM, van Rhijn LW, Kurtz SM, Antonis J, Schurink GW, van Ooij A (2008) Complications and reoperations of the SB Charite lumbar disc prosthesis: experience in 75 patients. Eur Spine J 17(1):36–43PubMedCrossRefPubMedCentral Punt IM, Visser VM, van Rhijn LW, Kurtz SM, Antonis J, Schurink GW, van Ooij A (2008) Complications and reoperations of the SB Charite lumbar disc prosthesis: experience in 75 patients. Eur Spine J 17(1):36–43PubMedCrossRefPubMedCentral
10.
go back to reference Siepe CJ, Mayer HM, Heinz-Leisenheimer M, Korge A (2007) Total lumbar disc replacement: different results for different levels. Spine 32(7):782–790PubMedCrossRef Siepe CJ, Mayer HM, Heinz-Leisenheimer M, Korge A (2007) Total lumbar disc replacement: different results for different levels. Spine 32(7):782–790PubMedCrossRef
11.
go back to reference Siepe CJ, Korge A, Grochulla F, Mehren C, Mayer HM (2008) Analysis of post-operative pain patterns following total lumbar disc replacement: results from fluoroscopically guided spine infiltrations. Eur Spine J 17(1):44–56PubMedCrossRefPubMedCentral Siepe CJ, Korge A, Grochulla F, Mehren C, Mayer HM (2008) Analysis of post-operative pain patterns following total lumbar disc replacement: results from fluoroscopically guided spine infiltrations. Eur Spine J 17(1):44–56PubMedCrossRefPubMedCentral
12.
go back to reference Shim CS, Lee SH, Shin HD, Kang HS, Choi WC, Jung B, Choi G, Ahn Y, Lee S, Lee HY (2007) CHARITE versus ProDisc: a comparative study of a minimum 3-year follow-up. Spine 32(9):1012–1018PubMedCrossRef Shim CS, Lee SH, Shin HD, Kang HS, Choi WC, Jung B, Choi G, Ahn Y, Lee S, Lee HY (2007) CHARITE versus ProDisc: a comparative study of a minimum 3-year follow-up. Spine 32(9):1012–1018PubMedCrossRef
13.
go back to reference Goel VK, Grauer JN, Patel T, Biyani A, Sairyo K, Vishnubhotla S, Matyas A, Cowgill I, Shaw M, Long R, Dick D, Panjabi MM, Serhan H (2005) Effects of charite artificial disc on the implanted and adjacent spinal segments mechanics using a hybrid testing protocol. Spine 30(24):2755–2764PubMedCrossRef Goel VK, Grauer JN, Patel T, Biyani A, Sairyo K, Vishnubhotla S, Matyas A, Cowgill I, Shaw M, Long R, Dick D, Panjabi MM, Serhan H (2005) Effects of charite artificial disc on the implanted and adjacent spinal segments mechanics using a hybrid testing protocol. Spine 30(24):2755–2764PubMedCrossRef
14.
go back to reference Wiseman CM, Lindsey DP, Fredrick AD, Yerby SA (2005) The effect of an interspinous process implant on facet loading during extension. Spine 30(8):903–907PubMedCrossRef Wiseman CM, Lindsey DP, Fredrick AD, Yerby SA (2005) The effect of an interspinous process implant on facet loading during extension. Spine 30(8):903–907PubMedCrossRef
15.
go back to reference Wilke HJ, Drumm J, Haussler K, Mack C, Steudel WI, Kettler A (2008) Biomechanical effect of different lumbar interspinous implants on flexibility and intradiscal pressure. Eur Spine J 17(8):1049–1056PubMedCrossRefPubMedCentral Wilke HJ, Drumm J, Haussler K, Mack C, Steudel WI, Kettler A (2008) Biomechanical effect of different lumbar interspinous implants on flexibility and intradiscal pressure. Eur Spine J 17(8):1049–1056PubMedCrossRefPubMedCentral
16.
go back to reference Kettler A, Drumm J, Heuer F, Haeussler K, Mack C, Claes L, Wilke HJ (2008) Can a modified interspinous spacer prevent instability in axial rotation and lateral bending? A biomechanical in vitro study resulting in a new idea. Clin Biomech (Bristol, Avon) 23(2):242–247CrossRef Kettler A, Drumm J, Heuer F, Haeussler K, Mack C, Claes L, Wilke HJ (2008) Can a modified interspinous spacer prevent instability in axial rotation and lateral bending? A biomechanical in vitro study resulting in a new idea. Clin Biomech (Bristol, Avon) 23(2):242–247CrossRef
17.
go back to reference Nockels RP (2005) Dynamic stabilization in the surgical management of painful lumbar spinal disorders. Spine 30(16 Suppl):S68–S72PubMedCrossRef Nockels RP (2005) Dynamic stabilization in the surgical management of painful lumbar spinal disorders. Spine 30(16 Suppl):S68–S72PubMedCrossRef
18.
go back to reference Schmoelz W, Erhart S, Unger S, Disch AC (2012) Biomechanical evaluation of a posterior non-fusion instrumentation of the lumbar spine. Eur Spine J 21(5):939–945PubMedCrossRefPubMedCentral Schmoelz W, Erhart S, Unger S, Disch AC (2012) Biomechanical evaluation of a posterior non-fusion instrumentation of the lumbar spine. Eur Spine J 21(5):939–945PubMedCrossRefPubMedCentral
19.
go back to reference Niosi CA, Zhu QA, Wilson DC, Keynan O, Wilson DR, Oxland TR (2006) Biomechanical characterization of the three-dimensional kinematic behaviour of the Dynesys dynamic stabilization system: an in vitro study. Eur Spine J 15(6):913–922PubMedCrossRefPubMedCentral Niosi CA, Zhu QA, Wilson DC, Keynan O, Wilson DR, Oxland TR (2006) Biomechanical characterization of the three-dimensional kinematic behaviour of the Dynesys dynamic stabilization system: an in vitro study. Eur Spine J 15(6):913–922PubMedCrossRefPubMedCentral
20.
go back to reference Schmoelz W, Huber JF, Nydegger T, Dipl I, Claes L, Wilke HJ (2003) Dynamic stabilization of the lumbar spine and its effects on adjacent segments: an in vitro experiment. J Spinal Disord Tech 16(4):418–423PubMedCrossRef Schmoelz W, Huber JF, Nydegger T, Dipl I, Claes L, Wilke HJ (2003) Dynamic stabilization of the lumbar spine and its effects on adjacent segments: an in vitro experiment. J Spinal Disord Tech 16(4):418–423PubMedCrossRef
21.
go back to reference Schmoelz W, Huber JF, Nydegger T, Claes L, Wilke HJ (2006) Influence of a dynamic stabilisation system on load bearing of a bridged disc: an in vitro study of intradiscal pressure. Eur Spine J 15(8):1276–1285PubMedCrossRefPubMedCentral Schmoelz W, Huber JF, Nydegger T, Claes L, Wilke HJ (2006) Influence of a dynamic stabilisation system on load bearing of a bridged disc: an in vitro study of intradiscal pressure. Eur Spine J 15(8):1276–1285PubMedCrossRefPubMedCentral
22.
go back to reference Beastall J, Karadimas E, Siddiqui M, Nicol M, Hughes J, Smith F, Wardlaw D (2007) The Dynesys lumbar spinal stabilization system: a preliminary report on positional magnetic resonance imaging findings. Spine 32(6):685–690PubMedCrossRef Beastall J, Karadimas E, Siddiqui M, Nicol M, Hughes J, Smith F, Wardlaw D (2007) The Dynesys lumbar spinal stabilization system: a preliminary report on positional magnetic resonance imaging findings. Spine 32(6):685–690PubMedCrossRef
23.
go back to reference Link HD (2002) History, design and biomechanics of the LINK SB Charite artificial disc. Eur Spine J 11(Suppl 2):S98–S105PubMedPubMedCentral Link HD (2002) History, design and biomechanics of the LINK SB Charite artificial disc. Eur Spine J 11(Suppl 2):S98–S105PubMedPubMedCentral
24.
go back to reference Wilke HJ, Heuer F, Schmidt H (2009) Prospective design delineation and subsequent in vitro evaluation of a new posterior dynamic stabilization system. Spine (Phila Pa 1976) 34(3):255–261CrossRef Wilke HJ, Heuer F, Schmidt H (2009) Prospective design delineation and subsequent in vitro evaluation of a new posterior dynamic stabilization system. Spine (Phila Pa 1976) 34(3):255–261CrossRef
25.
go back to reference Wilke HJ, Rohlmann F, Neidlinger-Wilke C, Werner K, Claes L, Kettler A (2006) Validity and interobserver agreement of a new radiographic grading system for intervertebral disc degeneration: part I. Lumbar spine. Eur Spine J 15(6):720–730PubMedCrossRefPubMedCentral Wilke HJ, Rohlmann F, Neidlinger-Wilke C, Werner K, Claes L, Kettler A (2006) Validity and interobserver agreement of a new radiographic grading system for intervertebral disc degeneration: part I. Lumbar spine. Eur Spine J 15(6):720–730PubMedCrossRefPubMedCentral
26.
go back to reference Wilke HJ, Claes L, Schmitt H, Wolf S (1994) A universal spine tester for in vitro experiments with muscle force simulation. Eur Spine J 3(2):91–97PubMedCrossRef Wilke HJ, Claes L, Schmitt H, Wolf S (1994) A universal spine tester for in vitro experiments with muscle force simulation. Eur Spine J 3(2):91–97PubMedCrossRef
27.
go back to reference Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476):307–310PubMedCrossRef Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476):307–310PubMedCrossRef
28.
go back to reference Wilke HJ, Wenger K, Claes L (1998) Testing criteria for spinal implants: recommendations for the standardization of in vitro stability testing of spinal implants. Eur Spine J 7(2):148–154PubMedCrossRefPubMedCentral Wilke HJ, Wenger K, Claes L (1998) Testing criteria for spinal implants: recommendations for the standardization of in vitro stability testing of spinal implants. Eur Spine J 7(2):148–154PubMedCrossRefPubMedCentral
29.
go back to reference Wilke HJ, Rohlmann A, Neller S, Schultheiss M, Bergmann G, Graichen F, Claes LE (2001) Is it possible to simulate physiologic loading conditions by applying pure moments? A comparison of in vivo and in vitro load components in an internal fixator. Spine 26(6):636–642PubMedCrossRef Wilke HJ, Rohlmann A, Neller S, Schultheiss M, Bergmann G, Graichen F, Claes LE (2001) Is it possible to simulate physiologic loading conditions by applying pure moments? A comparison of in vivo and in vitro load components in an internal fixator. Spine 26(6):636–642PubMedCrossRef
30.
go back to reference Cakir B, Richter M, Puhl W, Schmidt R (2006) Reliability of motion measurements after total disc replacement: the spike and the fin method. Eur Spine J 15(2):165–173PubMedCrossRefPubMedCentral Cakir B, Richter M, Puhl W, Schmidt R (2006) Reliability of motion measurements after total disc replacement: the spike and the fin method. Eur Spine J 15(2):165–173PubMedCrossRefPubMedCentral
32.
go back to reference Cunningham BW, Gordon JD, Dmitriev AE, Hu N, McAfee PC (2003) Biomechanical evaluation of total disc replacement arthroplasty: an in vitro human cadaveric model. Spine 28(20):S110–S117PubMedCrossRef Cunningham BW, Gordon JD, Dmitriev AE, Hu N, McAfee PC (2003) Biomechanical evaluation of total disc replacement arthroplasty: an in vitro human cadaveric model. Spine 28(20):S110–S117PubMedCrossRef
33.
go back to reference Panjabi M, Henderson G, Abjornson C, Yue J (2007) Multidirectional testing of one- and two-level ProDisc-L versus simulated fusions. Spine 32(12):1311–1319PubMedCrossRef Panjabi M, Henderson G, Abjornson C, Yue J (2007) Multidirectional testing of one- and two-level ProDisc-L versus simulated fusions. Spine 32(12):1311–1319PubMedCrossRef
34.
go back to reference Panjabi M, Malcolmson G, Teng E, Tominaga Y, Henderson G, Serhan H (2007) Hybrid testing of lumbar CHARITE discs versus fusions. Spine 32(9):959–966 (discussion 967)PubMedCrossRef Panjabi M, Malcolmson G, Teng E, Tominaga Y, Henderson G, Serhan H (2007) Hybrid testing of lumbar CHARITE discs versus fusions. Spine 32(9):959–966 (discussion 967)PubMedCrossRef
35.
go back to reference Wilke HJ, Schmidt H, Werner K, Schmolz W, Drumm J (2006) Biomechanical evaluation of a new total posterior-element replacement system. Spine 31(24):2790–2796 (discussion 2797)PubMedCrossRef Wilke HJ, Schmidt H, Werner K, Schmolz W, Drumm J (2006) Biomechanical evaluation of a new total posterior-element replacement system. Spine 31(24):2790–2796 (discussion 2797)PubMedCrossRef
36.
go back to reference Leivseth G, Braaten S, Frobin W, Brinckmann P (2006) Mobility of lumbar segments instrumented with a ProDisc II prosthesis: a two-year follow-up study. Spine 31(15):1726–1733PubMedCrossRef Leivseth G, Braaten S, Frobin W, Brinckmann P (2006) Mobility of lumbar segments instrumented with a ProDisc II prosthesis: a two-year follow-up study. Spine 31(15):1726–1733PubMedCrossRef
37.
go back to reference Jackson RP, McManus AC (1994) Radiographic analysis of sagittal plane alignment and balance in standing volunteers and patients with low back pain matched for age, sex, and size. A prospective controlled clinical study. Spine 19(14):1611–1618PubMedCrossRef Jackson RP, McManus AC (1994) Radiographic analysis of sagittal plane alignment and balance in standing volunteers and patients with low back pain matched for age, sex, and size. A prospective controlled clinical study. Spine 19(14):1611–1618PubMedCrossRef
38.
go back to reference Vialle R, Levassor N, Rillardon L, Templier A, Skalli W, Guigui P (2005) Radiographic analysis of the sagittal alignment and balance of the spine in asymptomatic subjects. J Bone Joint Surg Am 87(2):260–267PubMedCrossRef Vialle R, Levassor N, Rillardon L, Templier A, Skalli W, Guigui P (2005) Radiographic analysis of the sagittal alignment and balance of the spine in asymptomatic subjects. J Bone Joint Surg Am 87(2):260–267PubMedCrossRef
39.
go back to reference Dooris AP, Goel VK, Grosland NM, Gilbertson LG, Wilder DG (2001) Load-sharing between anterior and posterior elements in a lumbar motion segment implanted with an artificial disc. Spine 26(6):122–129CrossRef Dooris AP, Goel VK, Grosland NM, Gilbertson LG, Wilder DG (2001) Load-sharing between anterior and posterior elements in a lumbar motion segment implanted with an artificial disc. Spine 26(6):122–129CrossRef
40.
go back to reference Zander T, Rohlmann A, Bergmann G (2004) Analysis of simulated single ligament transection on the mechanical behaviour of a lumbar functional spinal unit. Biomed Tech (Berl) 49(1–2):27–32CrossRef Zander T, Rohlmann A, Bergmann G (2004) Analysis of simulated single ligament transection on the mechanical behaviour of a lumbar functional spinal unit. Biomed Tech (Berl) 49(1–2):27–32CrossRef
41.
go back to reference Hopf C, Heeckt H, Beske C (2004) Indikation, Biomechanik und Frühergebnisse des künstlichen Bandscheibenersatzes. Z Orthop Ihre Grenzgeb 142(2):153–158PubMedCrossRef Hopf C, Heeckt H, Beske C (2004) Indikation, Biomechanik und Frühergebnisse des künstlichen Bandscheibenersatzes. Z Orthop Ihre Grenzgeb 142(2):153–158PubMedCrossRef
42.
go back to reference McAfee PC, Cunningham BW, Hayes V, Sidiqi F, Dabbah M, Sefter JC, Hu N, Beatson H (2006) Biomechanical analysis of rotational motions after disc arthroplasty: implications for patients with adult deformities. Spine 31(19 Suppl):152–160CrossRef McAfee PC, Cunningham BW, Hayes V, Sidiqi F, Dabbah M, Sefter JC, Hu N, Beatson H (2006) Biomechanical analysis of rotational motions after disc arthroplasty: implications for patients with adult deformities. Spine 31(19 Suppl):152–160CrossRef
43.
go back to reference Cakir B, Richter M, Schmoelz W, Schmidt R, Reichel H, Wilke HJ (2012) Resect or not to resect: the role of posterior longitudinal ligament in lumbar total disc replacement. Eur Spine J 21(Suppl 5):S592–S598PubMedCrossRef Cakir B, Richter M, Schmoelz W, Schmidt R, Reichel H, Wilke HJ (2012) Resect or not to resect: the role of posterior longitudinal ligament in lumbar total disc replacement. Eur Spine J 21(Suppl 5):S592–S598PubMedCrossRef
44.
go back to reference Geisler FH (2006) The CHARITE artificial disc: design history, FDA IDE study results, and surgical technique. Clin Neurosurg 53:223–228PubMed Geisler FH (2006) The CHARITE artificial disc: design history, FDA IDE study results, and surgical technique. Clin Neurosurg 53:223–228PubMed
45.
go back to reference Gedet P, Haschtmann D, Thistlethwaite PA, Ferguson SJ (2009) Comparative biomechanical investigation of a modular dynamic lumbar stabilization system and the Dynesys system. Eur Spine J 18(10):1504–1511PubMedCrossRefPubMedCentral Gedet P, Haschtmann D, Thistlethwaite PA, Ferguson SJ (2009) Comparative biomechanical investigation of a modular dynamic lumbar stabilization system and the Dynesys system. Eur Spine J 18(10):1504–1511PubMedCrossRefPubMedCentral
46.
go back to reference Schulte TL, Hurschler C, Haversath M, Liljenqvist U, Bullmann V, Filler TJ, Osada N, Fallenberg EM, Hackenberg L (2008) The effect of dynamic, semi-rigid implants on the range of motion of lumbar motion segments after decompression. Eur Spine J 17(8):1057–1065PubMedCrossRefPubMedCentral Schulte TL, Hurschler C, Haversath M, Liljenqvist U, Bullmann V, Filler TJ, Osada N, Fallenberg EM, Hackenberg L (2008) The effect of dynamic, semi-rigid implants on the range of motion of lumbar motion segments after decompression. Eur Spine J 17(8):1057–1065PubMedCrossRefPubMedCentral
47.
go back to reference Legaye J (2005) Consequences defavorables sur l’equilibre sagittal du rachis du systeme de neutralisation dynamique. Rev Chir Orthop Reparatrice Appar Mot 91(6):542–550PubMedCrossRef Legaye J (2005) Consequences defavorables sur l’equilibre sagittal du rachis du systeme de neutralisation dynamique. Rev Chir Orthop Reparatrice Appar Mot 91(6):542–550PubMedCrossRef
48.
go back to reference Siddiqui M, Karadimas E, Nicol M, Smith FW, Wardlaw D (2006) Effects of X-STOP device on sagittal lumbar spine kinematics in spinal stenosis. J Spinal Disord Tech 19(5):328–333PubMedCrossRef Siddiqui M, Karadimas E, Nicol M, Smith FW, Wardlaw D (2006) Effects of X-STOP device on sagittal lumbar spine kinematics in spinal stenosis. J Spinal Disord Tech 19(5):328–333PubMedCrossRef
49.
go back to reference Siddiqui M, Nicol M, Karadimas E, Smith F, Wardlaw D (2005) The positional magnetic resonance imaging changes in the lumbar spine following insertion of a novel interspinous process distraction device. Spine 30(23):2677–2682PubMedCrossRef Siddiqui M, Nicol M, Karadimas E, Smith F, Wardlaw D (2005) The positional magnetic resonance imaging changes in the lumbar spine following insertion of a novel interspinous process distraction device. Spine 30(23):2677–2682PubMedCrossRef
50.
go back to reference Cakir B, Schmidt R, Huch K, Puhl W, Richter M (2004) Sagittales Alignement und segmentale Beweglichkeit nach endoprothetischer Versorgung lumbaler Bewegungssegmente. Z Orthop Ihre Grenzgeb 142(2):159–165PubMedCrossRef Cakir B, Schmidt R, Huch K, Puhl W, Richter M (2004) Sagittales Alignement und segmentale Beweglichkeit nach endoprothetischer Versorgung lumbaler Bewegungssegmente. Z Orthop Ihre Grenzgeb 142(2):159–165PubMedCrossRef
51.
go back to reference Wilke HJ, Schmidt R, Richter M, Schmoelz W, Reichel H, Cakir B (2012) The role of prosthesis design on segmental biomechanics: semi-constrained versus unconstrained prostheses and anterior versus posterior centre of rotation. Eur Spine J 21(Suppl 5):577–584CrossRefPubMedCentral Wilke HJ, Schmidt R, Richter M, Schmoelz W, Reichel H, Cakir B (2012) The role of prosthesis design on segmental biomechanics: semi-constrained versus unconstrained prostheses and anterior versus posterior centre of rotation. Eur Spine J 21(Suppl 5):577–584CrossRefPubMedCentral
Metadata
Title
Circumferential dynamic stabilization of the lumbar spine: a biomechanical analysis
Authors
Wolfram Käfer
Balkan Cakir
Stefan Midderhoff
Heiko Reichel
Hans-Joachim Wilke
Publication date
01-11-2014
Publisher
Springer Berlin Heidelberg
Published in
European Spine Journal / Issue 11/2014
Print ISSN: 0940-6719
Electronic ISSN: 1432-0932
DOI
https://doi.org/10.1007/s00586-014-3286-y

Other articles of this Issue 11/2014

European Spine Journal 11/2014 Go to the issue